2008T BlueStar Hemodialysis Machine recallZ-1868-2024 · Fresenius Medical Care Holdings, Inc. · initiated Apr 17, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Fresenius Medical Care Holdings, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
An increased number of complaints were reported on the Blood Pump Rotor (P/N - F40015481 REV A) after a period of use by clinics noting loose/dislodged guide sheaves.
FDA root cause category: Device Design
What the company is doing
On April 17, 2024 IMPORTANT MEDICAL DEVICE CORRECTION letters were sent to customers. Customer actions: As per the 2008T Operators Manual P/N 490122 Rev AA: Blood Pump Rotors should be inspected for proper operation, as stated in Chapter 2: Daily preparation for treatment. If you experience a dislodged guide post sleeve, bent or loose tubing guide posts, or roller cannot move freely on your current F40015481 Rev A blood pump rotor, please contact FMCRTG Technical Support Services at 800-405-1321 for a replacement Blood Pump Rotor (Part Number, F40015481, Rev C) . Have Rotor Lot Number available when requesting a replacement. Please provide this letter to your Biomedical Technician or trained personnel. Report any complaints or adverse events to [email protected] or FDA MedWatch at https://www.fda.gov/safety/medical-product-safety-information/medwatch-forms-fda-safety-reporting. Provide this Customer Notice to all those who need to be aware within your organization. Complete and return the attached reply form. Biomedical Technician/ actions: A Field Service Bulletin will be posted to fmcna.com website regarding Preventative Maintenance activities for Blood Pump Rotor Tubing Guide Inspection of the need to inspect the blood pump tubing guides during the 6-month Preventive Maintenance on all 2008 Series Hemodialysis Machines. URL: https://fmcna.com/support/product-support-documents/field-service-bulletins/ For any clinical or medical questions, please contact the Global Medical Information and Education Office at 1-855-616-2309.
Recall details
- Recall number
- Z-1868-2024
- Status
- Open, Classified
- Initiated by the company
- Apr 17, 2024
- Posted by the FDA
- May 17, 2024
- Recalling firm
- Fresenius Medical Care Holdings, Inc.
- Quantity
- 12,292 units
- Product code
- KDI
- FDA submissions
- K173972, K222952, K231125
- Distribution
- US Nationwide distribution.
Product description
2008T BlueStar Hemodialysis Machine, Model 191126 with Component: Blood Pump Rotor, Model Number: F40015481 Rev A.
Code information (models, lots, serial numbers)
Show the full code information (148k characters)
Component: Model Number: F40015481 Rev A, UDI/DI: N/A, All Rev A lots. Finished Devices: Model Number: 191126, UDI/DI: 00840861102099; Serial Numbers: 3T0S270687, 3T0S270700, 3T0S274007, 3T0S274009, 3T0S274011, 3T0S274012, 3T0S274016, 3T0S274057, 3T0S274859, 3T0S274920, 3T0S274922, 3T0S274923, 3T0S274938, 3T0S274957, 3T0S274969, 3T0S274970, 3T0S274980, 3T0S274985, 3T0S274996, 3T0S274997, 3T0S275004, 3T0S275009, 3T0S275011, 1T0S262226, 1T0S262249, 1T0S262254, 1T0S262257, 1T0S262265, 1T0S262266, 1T0S262267, 1T0S262268, 1T0S262269, 1T0S262272, 1T0S262273, 1T0S262274, 1T0S262275, 1T0S262276, 1T0S262278, 1T0S262279, 1T0S262280, 1T0S262282, 1T0S262283, 1T0S262284, 1T0S262285, 1T0S262286, 1T0S262289, 1T0S262293, 1T0S262294, 1T0S262296, 1T0S262297, 1T0S262299, 1T0S262300, 1T0S262305, 1T0S262313, 1T0S262315, 1T0S262319, 1T0S262321, 4T0S275623, 4T0S275627, 4T0S275628, 4T0S275629, 4T0S275639, 4T0S275649, 4T0S275654, 4T0S275662, 4T0S275664, 4T0S275666, 4T0S275677, 4T0S275682, 4T0S275685, 4T0S275688, 4T0S275692, 4T0S275703, 4T0S275729, 4T0S275738, 4T0S275742, 4T0S275745, 4T0S275760, 4T0S275768, 4T0S275769, 4T0S275778, 4T0S275787, 4T0S275800, 4T0S275804, 4T0S276804, 4T0S276814, 4T0S276818, 4T0S276829, 4T0S276833, 4T0S276838, 4T0S276858, 4T0S276867, 4T0S276870, 3T0S269793, 3T0S269826, 3T0S269827, 3T0S269828, 3T0S269836, 3T0S269838, 3T0S269841, 3T0S269842, 3T0S269954, 3T0S269955, 3T0S269956, 3T0S269957, 3T0S269959, 3T0S269960, 3T0S269961, 3T0S269963, 3T0S269964, 3T0S269965, 3T0S269966, 3T0S269968, 3T0S269972, 3T0S269976, 3T0S269978, 3T0S269979, 0T0S253172, 0T0S253178, 0T0S253179, 0T0S253184, 0T0S253188, 0T0S253190, 0T0S253203, 0T0S253208, 0T0S253215, 0T0S253217, 0T0S253222, 0T0S253223, 0T0S253227, 0T0S253230, 0T0S253231, 0T0S253232, 0T0S253240, 0T0S253245, 0T0S253257, 0T0S253268, 0T0S253271, 0T0S253272, 0T0S253286, 1T0S260028, 1T0S260091, 1T0S261521, 1T0S261526, 1T0S261527, 1T0S261536, 1T0S261556, 1T0S261557, 1T0S261558, 1T0S261559, 1T0S261572, 1T0S261573, 1T0S261578, 1T0S261581, 1T0S261588, 1T0S261747, 1T0S261771, 1T0S261776, 1T0S261787, 1T0S261788, 1T0S261790, 1T0S261792, 1T0S261794, 1T0S261810, 1T0S261812, 1T0S261821, 1T0S261836, 4T0S276807, 4T0S276886, 4T0S276894, 4T0S276895, 4T0S276897, 4T0S276909, 4T0S276911, 4T0S276912, 4T0S276916, 4T0S276921, 4T0S276926, 4T0S276928, 4T0S276934, 4T0S276952, 4T0S276960, 4T0S276961, 4T0S276963, 4T0S276964, 4T0S276965, 4T0S276969, 4T0S276973, 4T0S276976, 4T0S276979, 4T0S276991, 4T0S276998, 4T0S276999, 4T0S277004, 4T0S277005, 1T0S257039, 1T0S257061, 1T0S257078, 1T0S257083, 1T0S257089, 1T0S257097, 1T0S257107, 1T0S257115, 1T0S257116, 1T0S257129, 1T0S257138, 1T0S257141, 1T0S257153, 1T0S257156, 1T0S257157, 1T0S257161, 1T0S257163, 1T0S257166, 1T0S257171, 1T0S257173, 1T0S257178, 1T0S257194, 1T0S262433, 1T0S261874, 1T0S261934, 1T0S261946, 1T0S262256, 1T0S262271, 1T0S262301, 3T0S268718, 3T0S268750, 3T0S269695, 3T0S269702, 3T0S269709, 3T0S269710, 3T0S269712, 3T0S269713, 0T0S252683, 0T0S252712, 0T0S252742, 0T0S253194, 0T0S253220, 0T0S253356, 0T0S253372, 0T0S253393, 0T0S253400, 0T0S253408, 0T0S253416, 0T0S253423, 0T0S253427, 0T0S253616, 0T0S253632, 0T0S253636, 0T0S253638, 0T0S253646, 0T0S253647, 0T0S253649, 0T0S253678, 0T0S253962, 1T0S258819, 1T0S258845, 1T0S258873, 1T0S258891, 1T0S259185, 1T0S259188, 1T0S259360, 1T0S259362, 1T0S259369, 1T0S259373, 1T0S259377, 1T0S259400, 1T0S259403, 1T0S260106, 1T0S260120, 1T0S260122, 1T0S260125, 1T0S260128, 1T0S260131, 1T0S260135, 1T0S260136, 1T0S260141, 1T0S257132, 1T0S257162, 1T0S257183, 0T0S253221, 0T0S253235, 0T0S253274, 0T0S253296, 0T0S253298, 0T0S253299, 0T0S253378, 0T0S253384, 0T0S253394, 0T0S253411, 0T0S253431, 0T0S253435, 0T0S253441, 0T0S253443, 0T0S253445, 0T0S253449, 0T0S253453, 0T0S253454, 0T0S253455, 0T0S253463, 0T0S253466, 0T0S253467, 0T0S253469, 0T0S253470, 0T0S253471, 0T0S253472, 0T0S253474, 0T0S253475, 0T0S253483, 0T0S253484, 0T0S253485, 0T0S253486, 0T0S253488, 0T0S253491, 0T0S253492, 0T0S253494, 0T0S253501, 0T0S253503, 0T0S255162, 0T0S255178, 1T0S261828, 1T0S261852, 1T0S261870, 1T0S261897, 1T0S261965, 1T0S261970, 1T0S261990, 1T0S261993, 1T0S262010, 1T0S262014, 1T0S262019, 1T0S262023, 1T0S262024, 1T0S262025, 1T0S262026, 1T0S262029, 1T0S262037, 1T0S262039, 1T0S262046, 1T0S262048, 1T0S262051, 1T0S262059, 1T0S262067, 0T0S254123, 0T0S254131, 0T0S255109, 0T0S255117, 0T0S255128, 0T0S255142, 0T0S255156, 1T0S256919, 1T0S257000, 1T0S257309, 1T0S259629, 1T0S259641, 1T0S259647, 1T0S259657, 1T0S259658, 1T0S259660, 1T0S259663, 1T0S259666, 1T0S259667, 1T0S259669, 1T0S259680, 1T0S259681, 1T0S259683, 1T0S259685, 1T0S259698, 1T0S259700, 1T0S259705, 1T0S259706, 1T0S259712, 1T0S259715, 1T0S259726, 1T0S259734, 1T0S259735, 1T0S259736, 1T0S259742, 1T0S259748, 1T0S259753, 1T0S259754, 1T0S260027, 1T0S260030, 1T0S260035, 1T0S260036, 1T0S260039, 1T0S260044, 1T0S260080, 1T0S261131, 1T0S261202, 1T0S261211, 1T0S261213, 1T0S261215, 1T0S261221, 1T0S261225, 1T0S261226, 1T0S261240, 1T0S261243, 1T0S261244, 1T0S261247, 1T0S261253, 1T0S261257, 1T0S261259, 1T0S261260, 1T0S261261, 1T0S261264, 1T0S261265, 1T0S261267, 1T0S261268, 1T0S261272, 1T0S261274, 1T0S261275, 1T0S261276, 1T0S261279, 1T0S261284, 1T0S261285, 1T0S261286, 1T0S261288, 1T0S261290, 1T0S261293, 1T0S261294, 1T0S261296, 1T0S261298, 1T0S261299, 1T0S261302, 1T0S261307, 1T0S261311, 1T0S261312, 1T0S261313, 1T0S261315, 1T0S261319, 1T0S261320, 1T0S261338, 1T0S261339, 1T0S261340, 1T0S261359, 1T0S261363, 1T0S261420, 1T0S261437, 1T0S261450, 1T0S261456, 1T0S261461, 1T0S261470, 1T0S261471, 1T0S261479, 1T0S261485, 1T0S261486, 1T0S261487, 1T0S261491, 1T0S261492, 1T0S261493, 1T0S261494, 1T0S261495, 1T0S261501, 1T0S261504, 1T0S261505, 1T0S261508, 1T0S261511, 1T0S261515, 1T0S261517, 1T0S261519, 1T0S261520, 1T0S261522, 1T0S261525, 1T0S261529, 1T0S261532, 1T0S261537, 1T0S261543, 1T0S261548, 1T0S261554, 1T0S261560, 1T0S261566, 1T0S261574, 1T0S261575, 1T0S261577, 1T0S261579, 1T0S261967, 1T0S261390, 0T0S253944, 0T0S253948, 2T0S266546, 2T0S266553, 2T0S266998, 2T0S267002, 2T0S267003, 2T0S267015, 2T0S267072, 2T0S267088, 3T0S269989, 3T0S269990, 3T0S270003, 3T0S270006, 3T0S270011, 3T0S270015, 3T0S270017, 3T0S271622, 3T0S271630, 3T0S271641, 3T0S271955, 3T0S274097, 3T0S269958, 3T0S269985, 3T0S269987, 3T0S269991, 3T0S269993, 3T0S269995, 3T0S269996, 3T0S269998, 3T0S269999, 3T0S270000, 3T0S270001, 3T0S270007, 3T0S270010, 3T0S270013, 4T0S277002, 4T0S277003, 4T0S277025, 4T0S277048, 4T0S277054, 4T0S277077, 4T0S277090, 4T0S277094, 4T0S277095, 4T0S277099, 4T0S277115, 4T0S277345, 4T0S277355, 4T0S277367, 4T0S277371, 4T0S277374, 4T0S277380, 4T0S277381, 4T0S277392, 4T0S277398, 4T0S277419, 4T0S277420, 4T0S277424, 4T0S277431, 4T0S277434, 4T0S277452, 4T0S277459, 4T0S277475, 1T0S257397, 1T0S257412, 1T0S257802, 1T0S257818, 1T0S258217, 1T0S258218, 1T0S258220, 1T0S258223, 1T0S258228, 1T0S258231, 1T0S258232, 1T0S258233, 1T0S258242, 1T0S258243, 1T0S258249, 1T0S258250, 1T0S258251, 1T0S258257, 1T0S258262, 1T0S258263, 1T0S258549, 1T0S258553, 1T0S258583, 1T0S258588, 1T0S259445, 1T0S259457, 1T0S259488, 1T0S259491, 1T0S259505, 1T0S259520, 1T0S259521, 1T0S259525, 1T0S259527, 1T0S259531, 1T0S259532, 1T0S259533, 1T0S259534, 1T0S259536, 1T0S259539, 1T0S259540, 1T0S259545, 1T0S259547, 1T0S259770, 1T0S259801, 1T0S259810, 1T0S259813, 1T0S259815, 1T0S259821, 1T0S259833, 1T0S259835, 1T0S259836, 1T0S259838, 1T0S259839, 1T0S259840, 1T0S259841, 1T0S259842, 1T0S259844, 1T0S259845, 1T0S259846, 1T0S259847, 1T0S259848, 1T0S259849, 1T0S259850, 1T0S259851, 1T0S259852, 1T0S259853, 1T0S259854, 1T0S259855, 1T0S259856, 1T0S259858, 1T0S259860, 1T0S259861, 1T0S259862, 1T0S259863, 1T0S259864, 1T0S259865, 1T0S259866, 1T0S259867, 1T0S259868, 1T0S259869, 1T0S259870, 1T0S260022, 1T0S260026, 1T0S260067, 1T0S260074, 1T0S260078, 3T0S271584, 3T0S271612, 3T0S271615, 3T0S271618, 3T0S271620, 3T0S271624, 3T0S271628, 3T0S271654, 3T0S271659, 3T0S271662, 3T0S271665, 3T0S271669, 3T0S271672, 3T0S271680, 3T0S271701, 3T0S271704, 3T0S271714, 3T0S271727, 3T0S271740, 3T0S271781, 3T0S271786, 3T0S271787, 3T0S271788, 3T0S271795, 3T0S271801, 3T0S271814, 3T0S271821, 3T0S271827, 3T0S271828, 3T0S271839, 3T0S271845, 3T0S271856, 3T0S271882, 3T0S271901, 1T0S261666, 1T0S261680, 1T0S261681, 1T0S261685, 1T0S261696, 1T0S261705, 1T0S261710, 1T0S261713, 1T0S261714, 1T0S261715, 1T0S261716, 1T0S261717, 1T0S261719, 1T0S261720, 1T0S261721, 1T0S261722, 1T0S261723, 1T0S261724, 1T0S261726, 1T0S261727, 1T0S261731, 1T0S261736, 1T0S261738, 1T0S261739, 1T0S261741, 1T0S261744, 1T0S261745, 1T0S261750, 1T0S261718, 1T0S261749, 1T0S261751, 1T0S261753, 1T0S261754, 1T0S261755, 1T0S261756, 1T0S261757, 1T0S261758, 1T0S261759, 1T0S261760, 1T0S261761, 1T0S261762, 1T0S261763, 1T0S261764, 1T0S261765, 1T0S261766, 1T0S261767, 1T0S261768, 1T0S261769, 1T0S261770, 1T0S261772, 1T0S261773, 1T0S261774, 1T0S261775, 1T0S261777, 1T0S261778, 1T0S261780, 1T0S261781, 1T0S261784, 0T0S252775, 1T0S255773, 1T0S255775, 1T0S255800, 1T0S262287, 1T0S262288, 1T0S262290, 1T0S262304, 1T0S262306, 1T0S262307, 1T0S262308, 1T0S262309, 1T0S262310, 1T0S262316, 1T0S262317, 1T0S262318, 1T0S262322, 0T0S253237, 0T0S253403, 0T0S253418, 0T0S253440, 1T0S256402, 1T0S256420, 1T0S256432, 1T0S256462, 1T0S256464, 1T0S256470, 1T0S256474, 1T0S256476, 1T0S256482, 1T0S261789, 1T0S261791, 1T0S261793, 1T0S261795, 1T0S261796, 1T0S261797, 1T0S261798, 1T0S261799, 1T0S261802, 1T0S261803, 1T0S261805, 1T0S261806, 1T0S261811, 1T0S261813, 1T0S261814, 1T0S261815, 1T0S261816, 1T0S261819, 1T0S261822, 1T0S261823, 1T0S261829, 1T0S261831, 1T0S261832, 1T0S261833, 1T0S261834, 1T0S261835, 1T0S261844, 1T0S261851, 1T0S262002, 1T0S262006, 1T0S262027, 1T0S262030, 1T0S262033, 1T0S262035, 1T0S262102, 1T0S262107, 1T0S262115, 1T0S262122, 1T0S262124, 1T0S262125, 1T0S262127, 1T0S262141, 1T0S262168, 1T0S262189, 1T0S262193, 1T0S262196, 1T0S262201, 1T0S262203, 1T0S262211, 1T0S262213, 1T0S262217, 1T0S262221, 1T0S262222, 1T0S262225, 1T0S262232, 4T0S277582, 4T0S277587, 4T0S277606, 4T0S277616, 4T0S277635, 4T0S277649, 4T0S277655, 4T0S277656, 4T0S277682, 4T0S277707, 4T0S277711, 4T0S277716, 4T0S277723, 4T0S277741, 4T0S277746, 4T0S277753, 4T0S277755, 4T0S277756, 4T0S277757, 4T0S277758, 0T0S253592, 0T0S253670, 0T0S253954, 0T0S253968, 0T0S253969, 0T0S253978, 0T0S253987, 0T0S255078, 0T0S255104, 0T0S255125, 0T0S255126, 0T0S255131, 0T0S255134, 0T0S255138, 0T0S255145, 0T0S255147, 0T0S255148, 0T0S255150, 0T0S255152, 0T0S255161, 0T0S255164, 0T0S255165, 0T0S255168, 0T0S255169, 0T0S255170, 0T0S255171, 0T0S255173, 0T0S255177, 0T0S255179, 0T0S255185, 0T0S255186, 0T0S255189, 0T0S255190, 0T0S255191, 0T0S255194, 0T0S255206, 0T0S255209, 0T0S255219, 0T0S255227, 0T0S255351, 1T0S256675, 4T0S277360, 4T0S277401, 4T0S277545, 4T0S277547, 4T0S277565, 4T0S277576, 4T0S277609, 4T0S277629, 4T0S277633, 4T0S277637, 4T0S277643, 4T0S277650, 4T0S277653, 4T0S277669, 4T0S277671, 4T0S277693, 4T0S277706, 4T0S277713, 4T0S277714, 4T0S277721, 4T0S277724, 4T0S277732, 4T0S277736, 4T0S277744, 4T0S277749, 4T0S277752, 4T0S277754, 0T0S253169, 0T0S253182, 0T0S253202, 0T0S253204, 0T0S253205, 0T0S253206, 0T0S253211, 1T0S262113, 4T0S276854, 4T0S276884, 4T0S276920, 4T0S276924, 4T0S276927, 4T0S276935, 4T0S276937, 4T0S276938, 4T0S276942, 4T0S276953, 4T0S276958, 4T0S276959, 4T0S276971, 4T0S276974, 4T0S276975, 4T0S276977, 4T0S276982, 4T0S276984, 4T0S276987, 4T0S276989, 4T0S276993, 4T0S276994, 4T0S276995, 4T0S276996, 4T0S276997, 4T0S277001, 4T0S277006, 4T0S277016, 4T0S277036, 0T0S255674, 1T0S255712, 1T0S255721, 1T0S255722, 1T0S255723, 1T0S255729, 1T0S255730, 1T0S255731, 1T0S255732, 1T0S255748, 1T0S255787, 1T0S255799, 1T0S255802, 1T0S255804, 1T0S258762, 1T0S258794, 1T0S258796, 1T0S258801, 4T0S277102, 4T0S277343, 4T0S277393, 4T0S277397, 4T0S277399, 4T0S277422, 4T0S277429, 4T0S277435, 4T0S277445, 4T0S277446, 4T0S277451, 4T0S277458, 4T0S277461, 4T0S277465, 4T0S277481, 4T0S277483, 4T0S277485, 4T0S277489, 4T0S276375, 4T0S276492, 4T0S276534, 4T0S276543, 4T0S276553, 4T0S276601, 4T0S276602, 4T0S276604, 4T0S276605, 4T0S276612, 4T0S276618, 4T0S276627, 4T0S276647, 4T0S276650, 4T0S276652, 4T0S276660, 4T0S276664, 0T0S255132, 3T0S274823, 3T0S274824, 3T0S274825, 3T0S274827, 3T0S274829, 3T0S274831, 3T0S274847, 3T0S274868, 3T0S274870, 3T0S274873, 3T0S274878, 3T0S274880, 3T0S274885, 3T0S274890, 3T0S274892, 3T0S274898, 3T0S274900, 3T0S274903, 3T0S274908, 3T0S274911, 3T0S274912, 3T0S274914, 3T0S274928, 3T0S274954, 3T0S274989, 3T0S274990, 3T0S275006, 3T0S275008, 3T0S269700, 3T0S269701, 3T0S269703, 3T0S269722, 3T0S269723, 3T0S269724, 3T0S269728, 3T0S269732, 3T0S269739, 3T0S269742, 3T0S269743, 3T0S269747, 3T0S269754, 3T0S269756, 3T0S269759, 3T0S269760, 3T0S269761, 3T0S269762, 3T0S269765, 3T0S269767, 3T0S269770, 3T0S269774, 3T0S269775, 3T0S269776, 3T0S269777, 3T0S269778, 3T0S269780, 3T0S269782, 1T0S256441, 1T0S257212, 1T0S257221, 1T0S260031, 1T0S261137, 1T0S261142, 1T0S261146, 1T0S261164, 1T0S261173, 1T0S261174, 1T0S261175, 1T0S261182, 1T0S261190, 1T0S261192, 1T0S261195, 1T0S261196, 1T0S261198, 1T0S261204, 1T0S261206, 1T0S261209, 1T0S261210, 1T0S261214, 1T0S261217, 1T0S261218, 1T0S261220, 1T0S261222, 1T0S261223, 1T0S261224, 1T0S261227, 1T0S261229, 1T0S261231, 1T0S261233, 1T0S261239, 1T0S261241, 1T0S261251, 1T0S261254, 1T0S261255, 1T0S261256, 1T0S261270, 1T0S261271, 1T0S261273, 1T0S261326, 2T0S264214, 0T0S251949, 0T0S251950, 0T0S251956, 0T0S251958, 0T0S251973, 0T0S252587, 0T0S252588, 0T0S252620, 0T0S252629, 3T0S273953, 3T0S274066, 3T0S274093, 3T0S274158, 0T0S251939, 0T0S253990, 0T0S253991, 0T0S254591, 0T0S254600, 0T0S254606, 1T0S257217, 1T0S257219, 1T0S257223, 1T0S257231, 2T0S267238, 4T0S276815, 4T0S276823, 4T0S276826, 4T0S276832, 4T0S276842, 4T0S276843, 4T0S276845, 4T0S276849, 4T0S276853, 4T0S276861, 4T0S276863, 4T0S276872, 4T0S276873, 4T0S276874, 4T0S276876, 4T0S276877, 4T0S276881, 4T0S276883, 4T0S276901, 4T0S276908, 0T0S248610, 1T0S256093, 0T0S253975, 0T0S253980, 1T0S259251, 1T0S259274, 2T0S264321, 2T0S264965, 3T0S270092, 0T0S253558, 0T0S253578, 0T0S253611, 0T0S253625, 0T0S253643, 0T0S253645, 0T0S254515, 0T0S254538, 3T0S271806, 1T0S257800, 1T0S257824, 1T0S258751, 1T0S258770, 1T0S258786, 1T0S258789, 1T0S258793, 1T0S258797, 1T0S258980, 1T0S258997, 1T0S259001, 1T0S259010, 2T0S265874, 2T0S264422, 2T0S265431, 2T0S266852, 2T0S266878, 2T0S266967, 2T0S266978, 0T0S253588, 0T0S254133, 0T0S255246, 0T0S255270, 0T0S255283, 0T0S255285, 0T0S255291, 0T0S255308, 0T0S255319, 0T0S255321, 0T0S255328, 0T0S255329, 0T0S255332, 0T0S255339, 0T0S255341, 0T0S255343, 0T0S255344, 0T0S255346, 0T0S255350, 0T0S255354, 0T0S255358, 0T0S255363, 0T0S255382, 0T0S255386, 0T0S255659, 0T0S255669, 0T0S255684, 0T0S255695, 0T0S255702, 1T0S255714, 1T0S255717, 1T0S255728, 1T0S255750, 1T0S255755, 1T0S255759, 1T0S255777, 1T0S256121, 1T0S256128, 1T0S262110, 1T0S262114, 1T0S262117, 1T0S262131, 1T0S262133, 1T0S262135, 1T0S262140, 1T0S262154, 1T0S262155, 1T0S262157, 1T0S262170, 1T0S262194, 1T0S262209, 1T0S262210, 1T0S262218, 3T0S271637, 3T0S271649, 3T0S271762, 3T0S271794, 1T0S256977, 1T0S257369, 1T0S257897, 1T0S258573, 1T0S258575, 1T0S258590, 3T0S270580, 3T0S270627, 3T0S270632, 3T0S270652, 3T0S270726, 3T0S270729, 3T0S270745, 3T0S270767, 3T0S270780, 3T0S270788, 3T0S271170, 3T0S271175, 3T0S271822, 3T0S271829, 3T0S271850, 3T0S271858, 3T0S271897, 3T0S271907, 3T0S271916, 3T0S271917, 3T0S274040, 3T0S274042, 3T0S274059, 3T0S274077, 3T0S274082, 3T0S274088, 3T0S274120, 3T0S274125, 3T0S274174, 2T0S264169, 2T0S264198, 2T0S264207, 2T0S264208, 1T0S256823, 1T0S256827, 1T0S256830, 1T0S256831, 1T0S256834, 1T0S256837, 1T0S256847, 1T0S257281, 1T0S257322, 1T0S257343, 1T0S257381, 1T0S257395, 1T0S257400, 1T0S257405, 2T0S266573, 2T0S266590, 2T0S266597, 2T0S266599, 2T0S266600, 2T0S266602, 2T0S266605, 2T0S266611, 2T0S266613, 2T0S266615, 2T0S266622, 2T0S266623, 2T0S266626, 2T0S266627, 2T0S266628, 2T0S266860, 2T0S267639, 2T0S266857, 2T0S266861, 2T0S266887, 2T0S266890, 3T0S273853, 3T0S273856, 3T0S273875, 3T0S273892, 3T0S274058, 3T0S274060, 3T0S274083, 3T0S274089, 3T0S274090, 3T0S274092, 3T0S274096, 3T0S274100, 3T0S274108, 3T0S274129, 3T0S274130, 3T0S274138, 3T0S274143, 3T0S274146, 3T0S274147, 3T0S274153, 3T0S274155, 3T0S274161, 3T0S274166, 3T0S274167, 3T0S274169, 3T0S274170, 1T0S255742, 1T0S255743, 1T0S255746, 1T0S255768, 1T0S255771, 1T0S255772, 1T0S255774, 1T0S255778, 1T0S255780, 1T0S255782, 1T0S255783, 1T0S255789, 1T0S255790, 1T0S255795, 1T0S255803, 1T0S256154, 1T0S256157, 4T0S275857, 4T0S275864, 4T0S275869, 4T0S275870, 4T0S275882, 4T0S275887, 4T0S275888, 4T0S275889, 4T0S275894, 4T0S275895, 4T0S275908, 4T0S275909, 4T0S275913, 4T0S275914, 4T0S275925, 4T0S275926, 4T0S275939, 4T0S275961, 0T0S255036, 0T0S255053, 0T0S255064, 0T0S255071, 0T0S255077, 3T0S271166, 3T0S271173, 3T0S271197, 3T0S271205, 3T0S271709, 3T0S271747, 3T0S271765, 3T0S271818, 3T0S271842, 3T0S271848, 3T0S271854, 3T0S271861, 3T0S271864, 3T0S271866, 3T0S271867, 3T0S271870, 3T0S271875, 3T0S271884, 3T0S271886, 3T0S271898, 3T0S271913, 3T0S271919, 3T0S271928, 2T0S267142, 2T0S267166, 2T0S267170, 2T0S267171, 2T0S267172, 2T0S267176, 2T0S267188, 2T0S267194, 2T0S267204, 3T0S270658, 3T0S270673, 3T0S270758, 3T0S270762, 3T0S270766, 3T0S270793, 3T0S270795, 3T0S270796, 3T0S270802, 3T0S270805, 3T0S270806, 3T0S270808, 3T0S270811, 3T0S270814, 3T0S271199, 0T0S254027, 0T0S254047, 0T0S254062, 0T0S254064, 0T0S254069, 0T0S254070, 0T0S254072, 0T0S254086, 1T0S257022, 1T0S257082, 1T0S257103, 1T0S257110, 1T0S257113, 1T0S257114, 1T0S257125, 1T0S257127, 1T0S257842, 1T0S257859, 1T0S257878, 1T0S259668, 1T0S260072, 1T0S260077, 1T0S260109, 1T0S261235, 1T0S261371, 1T0S261380, 1T0S261393, 1T0S261395, 1T0S261399, 1T0S261413, 1T0S261415, 1T0S261417, 1T0S261421, 1T0S261424, 1T0S261455, 1T0S261465, 1T0S261472, 1T0S261477, 2T0S266947, 2T0S267058, 2T0S267071, 2T0S267137, 2T0S267167, 2T0S267207, 2T0S267222, 2T0S267223, 2T0S267225, 2T0S267230, 2T0S267235, 2T0S267240, 2T0S267257, 2T0S267274, 2T0S267275, 2T0S267279, 2T0S267281, 2T0S267282, 2T0S267283, 2T0S267302, 2T0S267305, 2T0S267306, 1T0S261859, 1T0S262003, 1T0S262007, 1T0S262011, 1T0S262013, 1T0S262015, 1T0S262016, 1T0S262020, 1T0S262022, 1T0S262031, 1T0S262032, 1T0S262036, 1T0S262040, 1T0S262041, 1T0S262042, 1T0S262052, 1T0S262054, 1T0S262060, 1T0S262175, 1T0S262199, 1T0S262202, 1T0S262206, 1T0S262207, 1T0S262212, 1T0S262215, 1T0S262216, 1T0S262219, 1T0S262230, 1T0S262231, 1T0S262237, 1T0S262241, 3T0S274852, 3T0S274963, 3T0S274967, 3T0S274973, 3T0S274975, 3T0S274978, 3T0S274994, 3T0S274999, 3T0S275001, 3T0S275002, 3T0S275003, 3T0S275005, 3T0S275007, 3T0S275012, 1T0S259402, 1T0S259406, 1T0S259412, 1T0S259418, 1T0S259425, 1T0S259433, 1T0S259436, 1T0S259443, 1T0S259601, 1T0S259643, 1T0S259655, 1T0S259671, 1T0S259673, 1T0S259723, 1T0S259752, 1T0S260045, 1T0S260063, 1T0S260064, 1T0S261180, 1T0S261266, 1T0S261282, 1T0S261316, 1T0S261322, 1T0S261349, 1T0S261436, 1T0S261438, 1T0S259164, 1T0S259177, 1T0S259387, 1T0S260142, 1T0S262617, 2T0S262752, 2T0S262754, 2T0S262755, 2T0S262756, 2T0S262768, 2T0S262770, 2T0S262773, 2T0S262775, 2T0S262777, 2T0S263396, 2T0S263408, 2T0S263414, 2T0S263450, 2T0S263458, 2T0S263479, 2T0S263484, 3T0S270612, 3T0S270637, 3T0S270638, 3T0S270639, 3T0S270640, 3T0S270648, 3T0S270650, 3T0S270684, 3T0S270705, 3T0S270727, 3T0S270741, 3T0S270750, 3T0S270753, 3T0S270754, 3T0S270769, 3T0S271844, 3T0S271895, 3T0S271899, 3T0S271900, 3T0S271929, 2T0S263151, 2T0S263160, 2T0S263167, 2T0S263169, 2T0S263179, 2T0S263183, 2T0S263189, 2T0S263191, 2T0S263192, 2T0S263193, 2T0S263194, 2T0S263195, 2T0S263196, 2T0S263197, 2T0S263199, 2T0S263201, 2T0S263202, 2T0S263203, 2T0S263205, 2T0S263206, 2T0S263207, 2T0S263208, 2T0S263468, 2T0S263507, 1T0S255785, 1T0S256806, 2T0S266578, 2T0S264696, 2T0S264707, 2T0S264716, 2T0S264868, 2T0S267601, 2T0S267611, 2T0S267616, 2T0S267628, 2T0S267739, 2T0S267843, 2T0S267850, 2T0S267853, 2T0S267893, 2T0S268100, 2T0S268112, 2T0S268113, 2T0S268117, 3T0S268702, 3T0S268705, 3T0S268706, 3T0S268733, 3T0S268745, 4T0S275912, 4T0S275928, 4T0S275938, 4T0S275945, 4T0S275948, 4T0S275953, 4T0S275971, 4T0S276320, 4T0S276350, 4T0S276396, 4T0S276848, 4T0S276865, 4T0S276914, 1T0S256711, 1T0S256732, 1T0S256744, 1T0S256748, 1T0S256758, 1T0S256762, 1T0S256769, 1T0S256771, 1T0S256773, 1T0S256777, 1T0S256778, 1T0S256779, 1T0S256784, 1T0S256791, 1T0S256794, 1T0S256796, 1T0S258818, 1T0S258831, 1T0S258834, 1T0S258838, 1T0S258840, 1T0S258844, 1T0S258851, 1T0S258863, 1T0S258865, 1T0S258871, 1T0S258879, 1T0S258898, 2T0S265422, 2T0S265429, 2T0S265465, 2T0S265479, 2T0S265511, 2T0S265515, 2T0S265576, 2T0S267156, 2T0S267197, 2T0S267208, 2T0S267211, 2T0S267228, 2T0S267265, 2T0S267290, 2T0S267837, 2T0S267840, 2T0S267842, 2T0S267854, 2T0S267864, 2T0S267866, 2T0S267869, 2T0S267870, 2T0S267872, 2T0S267886, 2T0S267888, 2T0S267890, 2T0S267895, 2T0S267898, 4T0S276321, 4T0S276323, 4T0S276329, 4T0S276339, 4T0S276343, 4T0S276346, 4T0S276349, 4T0S276352, 4T0S276367, 4T0S276379, 4T0S276380, 4T0S276388, 4T0S276399, 4T0S276405, 4T0S276407, 3T0S268704, 3T0S268709, 3T0S268716, 3T0S268719, 3T0S268726, 3T0S268729, 3T0S268737, 3T0S269663, 3T0S269679, 3T0S269688, 3T0S269690, 3T0S269691, 3T0S269692, 3T0S269694, 3T0S269697, 3T0S269704, 3T0S269707, 3T0S269708, 3T0S269714, 3T0S269715, 3T0S269718, 3T0S269730, 3T0S269736, 3T0S269744, 3T0S269752, 3T0S269763, 3T0S269764, 3T0S269781, 3T0S269787, 3T0S269788, 3T0S269794, 3T0S269796, 3T0S269797, 3T0S269800, 3T0S269801, 3T0S269807, 3T0S269810, 3T0S269811, 3T0S269814, 3T0S269817, 3T0S270393, 2T0S264294, 2T0S265455, 2T0S265528, 3T0S271724, 3T0S271820, 3T0S271840, 3T0S271849, 3T0S271852, 3T0S271857, 3T0S271860, 3T0S271872, 3T0S271873, 3T0S271876, 3T0S271879, 3T0S271889, 3T0S271892, 3T0S271893, 3T0S271903, 3T0S271925, 3T0S271938, 3T0S271939, 3T0S271943, 3T0S271947, 3T0S271949, 3T0S271960, 3T0S271968, 3T0S271970, 3T0S271973, 3T0S271974, 3T0S271977, 3T0S271980, 3T0S271984, 3T0S271989, 3T0S271990, 3T0S271991, 0T0S252389, 0T0S252602, 2T0S266863, 2T0S266869, 2T0S266899, 2T0S266909, 2T0S266912, 2T0S266913, 2T0S266916, 2T0S266931, 2T0S266938, 2T0S266942, 2T0S266968, 2T0S266980, 2T0S267804, 2T0S267818, 2T0S267832, 2T0S267838, 2T0S267846, 2T0S267863, 2T0S267871, 2T0S267887, 2T0S267891, 2T0S267892, 2T0S267894, 2T0S267896, 1T0S256083, 1T0S256141, 1T0S256150, 1T0S261409, 1T0S261516, 1T0S261583, 2T0S263108, 2T0S263122, 2T0S263134, 2T0S267056, 2T0S267062, 2T0S267068, 2T0S267080, 2T0S267083, 2T0S267087, 3T0S268754, 3T0S271174, 3T0S271187, 3T0S271189, 3T0S271198, 3T0S271203, 3T0S271208, 3T0S271210, 0T0S253965, 0T0S253974, 0T0S253986, 0T0S253988, 0T0S253992, 0T0S253994, 0T0S253998, 0T0S254000, 0T0S254001, 0T0S254003, 0T0S254006, 0T0S254009, 0T0S254010, 0T0S254014, 0T0S254016, 0T0S254018, 0T0S254044, 0T0S254046, 4T0S277534, 1T0S262398, 1T0S262400, 1T0S262404, 1T0S262427, 1T0S262446, 1T0S262512, 1T0S262514, 1T0S262515, 1T0S262519, 1T0S262527, 1T0S262529, 1T0S262544, 0T0S254640, 0T0S254677, 2T0S266568, 2T0S266579, 2T0S266608, 2T0S266616, 2T0S264371, 2T0S264751, 1T0S259367, 1T0S259372, 1T0S259376, 1T0S259378, 1T0S259380, 1T0S259382, 1T0S259384, 1T0S259390, 1T0S259392, 1T0S259393, 1T0S259395, 1T0S259397, 3T0S271804, 3T0S271824, 3T0S271940, 3T0S271944, 3T0S271946, 3T0S271954, 3T0S271956, 3T0S271964, 3T0S271982, 3T0S271987, 3T0S274000, 3T0S274128, 3T0S274150, 3T0S274846, 3T0S274936, 3T0S275010, 3T0S275211, 3T0S275213, 3T0S275214, 3T0S275216, 3T0S275217, 3T0S275219, 3T0S275225, 3T0S275227, 3T0S275233, 3T0S275236, 3T0S275237, 3T0S275241, 3T0S275242, 3T0S275246, 3T0S275248, 3T0S275254, 1T0S256723, 1T0S256755, 1T0S256763, 1T0S257193, 1T0S257226, 1T0S257237, 0T0S252193, 0T0S252213, 3T0S270047, 3T0S270048, 3T0S270049, 3T0S270052, 3T0S270053, 3T0S270054, 3T0S270057, 3T0S270071, 3T0S270107, 3T0S270108, 3T0S270110, 3T0S270112, 3T0S270114, 3T0S270387, 3T0S270388, 3T0S270389, 3T0S270390, 3T0S270391, 0T0S254033, 0T0S254035, 0T0S254054, 0T0S254058, 0T0S254060, 0T0S254080, 1T0S256575, 1T0S256698, 1T0S256699, 1T0S256725, 1T0S256729, 1T0S256736, 1T0S256737, 1T0S256745, 1T0S256970, 1T0S256980, 1T0S256988, 1T0S257175, 1T0S257255, 1T0S257269, 1T0S257270, 1T0S257286, 1T0S257287, 1T0S257288, 1T0S257294, 1T0S257297, 1T0S257301, 1T0S257302, 2T0S264565, 3T0S270757, 3T0S270833, 4T0S275936, 0T0S252030, 0T0S252031, 0T0S252032, 0T0S252057, 0T0S252061, 0T0S252079, 0T0S252084, 0T0S252239, 0T0S252259, 0T0S252262, 0T0S252278, 0T0S252280, 0T0S252285, 0T0S252314, 0T0S252324, 0T0S252326, 0T0S252327, 0T0S252329, 0T0S252330, 0T0S252332, 0T0S252333, 0T0S252336, 0T0S252337, 0T0S252338, 0T0S252342, 0T0S252343, 0T0S252346, 0T0S252349, 0T0S252353, 0T0S252356, 0T0S252361, 0T0S252369, 2T0S265348, 2T0S265393, 2T0S265397, 2T0S265412, 2T0S265420, 2T0S265424, 2T0S265447, 2T0S265448, 2T0S265468, 2T0S265469, 2T0S265482, 2T0S265494, 2T0S265495, 2T0S265504, 2T0S265506, 2T0S265516, 2T0S265517, 2T0S265519, 2T0S265522, 2T0S265527, 2T0S265544, 2T0S265546, 2T0S265550, 2T0S265627, 2T0S268019, 2T0S268020, 2T0S268024, 0T0S252236, 0T0S252774, 0T0S254049, 0T0S254052, 0T0S254071, 4T0S276353, 4T0S276361, 4T0S276370, 4T0S276377, 4T0S276417, 4T0S276429, 4T0S276430, 4T0S276436, 4T0S276437, 4T0S276438, 4T0S276443, 4T0S276444, 4T0S276449, 4T0S276450, 4T0S276451, 4T0S276452, 4T0S276454, 4T0S276455, 4T0S276456, 4T0S276461, 4T0S276464, 4T0S276470, 4T0S276471, 4T0S276476, 4T0S276487, 4T0S276490, 4T0S276494, 4T0S276508, 4T0S276549, 4T0S276554, 1T0S256085, 1T0S256100, 1T0S257010, 1T0S257015, 1T0S257073, 3T0S268677, 1T0S255741, 1T0S255776, 1T0S255796, 1T0S256081, 1T0S256087, 1T0S256105, 1T0S256108, 1T0S256113, 1T0S256118, 1T0S256122, 1T0S256124, 1T0S256126, 1T0S256129, 1T0S256130, 1T0S256131, 1T0S256132, 1T0S256133, 1T0S256134, 1T0S256135, 1T0S256136, 1T0S256137, 1T0S256138, 1T0S256139, 1T0S256140, 1T0S256143, 1T0S256147, 1T0S256151, 1T0S256159, 1T0S256168, 1T0S256169, 1T0S256171, 1T0S256175, 0T0S252384, 0T0S252590, 0T0S253985, 0T0S253959, 0T0S255042, 0T0S255043, 0T0S255058, 0T0S255061, 0T0S255067, 0T0S255069, 0T0S255079, 0T0S255084, 0T0S255087, 0T0S255088, 0T0S255089, 0T0S255091, 0T0S255094, 0T0S255096, 0T0S255097, 0T0S255098, 0T0S255108, 3T0S273960, 3T0S274023, 3T0S274031, 3T0S274041, 3T0S274044, 3T0S274049, 3T0S274053, 3T0S274078, 3T0S274080, 3T0S274876, 3T0S274902, 3T0S274916, 3T0S274925, 3T0S274927, 3T0S274932, 3T0S274937, 3T0S274961, 3T0S274964, 3T0S274968, 3T0S274976, 3T0S274977, 3T0S274987, 3T0S274998, 2T0S264506, 2T0S264549, 2T0S264564, 2T0S264577, 2T0S264940, 2T0S265245, 2T0S266560, 2T0S267061, 2T0S267085, 3T0S268775, 3T0S271165, 3T0S271186, 3T0S271201, 3T0S271646, 3T0S271679, 2T0S262751, 2T0S262753, 2T0S262758, 2T0S262760, 2T0S262763, 2T0S262764, 2T0S262766, 2T0S262769, 2T0S262771, 2T0S262772, 2T0S262774, 2T0S262776, 2T0S262779, 2T0S262784, 2T0S262787, 2T0S262790, 2T0S262925, 3T0S271933, 3T0S271966, 2T0S264871, 2T0S264943, 2T0S264951, 2T0S264957, 2T0S264959, 2T0S264960, 2T0S264961, 2T0S265392, 2T0S265403, 2T0S265411, 2T0S265439, 2T0S265458, 2T0S265475, 2T0S265508, 2T0S266131, 2T0S266151, 2T0S266158, 2T0S266160, 0T0S255105, 1T0S256125, 2T0S264954, 2T0S265385, 0T0S252304, 0T0S252316, 0T0S252319, 0T0S252320, 0T0S252321, 1T0S261587, 1T0S261593, 1T0S261597, 1T0S261599, 1T0S261606, 1T0S261607, 1T0S261616, 1T0S261619, 1T0S261622, 1T0S261625, 1T0S261628, 1T0S261631, 1T0S261634, 1T0S261636, 1T0S261638, 1T0S261641, 1T0S261643, 1T0S261645, 1T0S261647, 1T0S261649, 1T0S261650, 1T0S261651, 1T0S261656, 1T0S261673, 1T0S259538, 1T0S259543, 1T0S259548, 1T0S259554, 1T0S259558, 1T0S259575, 1T0S259581, 1T0S259582, 1T0S259584, 1T0S259586, 1T0S259589, 1T0S259590, 1T0S259591, 1T0S259592, 1T0S259595, 1T0S259597, 1T0S259603, 1T0S259604, 1T0S259605, 1T0S259606, 1T0S259608, 1T0S259612, 1T0S259738, 1T0S259784, 1T0S259788, 1T0S259791, 1T0S259802, 1T0S259803, 1T0S259830, 1T0S259843, 2T0S266865, 2T0S266877, 2T0S266903, 2T0S266924, 2T0S266926, 2T0S266932, 2T0S266935, 2T0S266960, 1T0S256439, 2T0S263634, 2T0S263821, 2T0S263836, 2T0S267050, 2T0S267065, 2T0S267073, 2T0S267074, 2T0S267075, 2T0S267077, 4T0S277387, 4T0S277466, 4T0S277486, 4T0S277496, 4T0S277500, 4T0S277511, 4T0S277517, 4T0S277521, 4T0S277550, 4T0S277551, 4T0S277561, 4T0S277569, 4T0S277572, 4T0S277575, 4T0S277577, 4T0S277580, 4T0S277584, 4T0S277586, 4T0S277591, 4T0S277612, 4T0S277615, 4T0S277624, 4T0S277630, 4T0S277640, 4T0S277644, 4T0S277646, 4T0S277657, 4T0S277658, 4T0S277667, 4T0S277696, 4T0S277698, 4T0S277712, 4T0S277719, 4T0S277722, 4T0S277725, 4T0S277745, 1T0S257393, 0T0S252209, 0T0S252218, 0T0S252220, 0T0S255032, 0T0S255037, 0T0S255040, 0T0S255047, 0T0S255048, 0T0S255050, 0T0S255052, 1T0S257071, 1T0S257164, 1T0S257214, 1T0S257228, 1T0S257229, 1T0S257241, 1T0S257242, 1T0S257244, 1T0S257251, 1T0S257256, 1T0S257259, 1T0S257261, 1T0S257264, 1T0S257267, 1T0S257271, 1T0S257272, 1T0S257290, 1T0S257293, 1T0S261732, 1T0S261743, 2T0S266953, 2T0S266989, 2T0S267011, 0T0S248640, 0T0S248731, 0T0S248753, 1T0S258139, 1T0S258182, 1T0S258230, 3T0S271676, 3T0S271686, 3T0S271696, 3T0S271796, 3T0S271891, 0T0S253979, 0T0S255030, 2T0S263496, 2T0S263531, 3T0S270588, 3T0S270590, 0T0S253196, 1T0S257385, 1T0S258210, 1T0S260050, 1T0S260055, 1T0S260069, 1T0S260090, 3T0S273270, 4T0S275931, 4T0S275973, 4T0S276324, 4T0S276327, 4T0S276330, 4T0S276340, 4T0S276348, 4T0S276356, 4T0S276357, 4T0S276358, 4T0S276359, 4T0S276365, 4T0S276371, 4T0S276372, 4T0S276373, 4T0S276378, 4T0S276383, 4T0S276387, 4T0S276392, 4T0S276393, 4T0S276394, 4T0S276397, 4T0S276400, 4T0S276404, 4T0S276419, 4T0S276420, 4T0S276422, 4T0S276423, 4T0S276425, 4T0S276441, 4T0S276465, 4T0S276469, 4T0S276502, 3T0S271608, 3T0S271682, 3T0S271689, 3T0S271723, 3T0S271823, 3T0S271825, 3T0S271831, 3T0S271832, 3T0S271835, 3T0S271855, 3T0S271863, 3T0S271865, 3T0S271868, 3T0S271869, 3T0S271871, 3T0S271883, 3T0S271887, 3T0S271890, 3T0S271904, 3T0S271906, 3T0S271909, 3T0S271911, 3T0S271918, 3T0S271921, 3T0S271923, 3T0S271924, 3T0S271927, 3T0S271932, 3T0S270642, 3T0S270670, 3T0S270676, 3T0S273844, 3T0S273845, 3T0S273851, 3T0S273854, 3T0S273855, 3T0S273857, 3T0S273859, 3T0S273860, 3T0S273861, 3T0S273863, 3T0S273869, 3T0S273871, 3T0S273876, 3T0S273877, 3T0S273880, 3T0S273885, 3T0S273889, 3T0S273890, 3T0S273902, 3T0S274052, 3T0S274156, 3T0S274159, 3T0S274162, 3T0S274163, 3T0S274168, 1T0S256849, 1T0S256873, 1T0S256945, 1T0S257095, 1T0S257140, 1T0S257144, 1T0S257176, 1T0S257177, 1T0S257196, 1T0S257199, 1T0S257222, 1T0S257234, 1T0S257236, 1T0S257246, 1T0S257249, 1T0S257254, 1T0S257285, 1T0S257307, 1T0S257326, 2T0S264675, 2T0S264703, 2T0S264958, 2T0S266196, 1T0S256927, 2T0S262916, 2T0S263413, 2T0S263419, 2T0S263430, 2T0S263434, 2T0S263435, 2T0S263441, 2T0S263445, 2T0S263446, 2T0S263447, 2T0S263451, 2T0S263480, 1T0S261096, 1T0S261212, 1T0S261219, 1T0S261230, 1T0S261245, 1T0S261246, 1T0S261248, 1T0S261249, 1T0S261250, 1T0S261252, 1T0S261258, 1T0S261269, 1T0S261277, 1T0S261278, 1T0S261297, 1T0S261301, 1T0S261303, 1T0S261324, 1T0S261435, 1T0S261453, 1T0S261469, 1T0S261476, 1T0S261489, 1T0S261490, 1T0S261496, 1T0S261509, 1T0S261510, 1T0S261512, 1T0S261514, 1T0S261530, 1T0S261531, 1T0S261533, 1T0S261540, 1T0S261541, 1T0S261542, 1T0S261546, 1T0S261549, 1T0S261551, 1T0S261553, 1T0S261562, 1T0S256395, 1T0S256398, 1T0S256401, 1T0S256410, 1T0S256411, 1T0S256415, 1T0S256421, 1T0S256436, 1T0S256437, 1T0S256438, 1T0S256440, 1T0S257349, 1T0S257356, 1T0S257371, 1T0S257382, 1T0S262372, 1T0S262442, 1T0S262542, 1T0S256327, 1T0S256335, 1T0S256371, 1T0S256377, 1T0S256378, 1T0S256380, 1T0S256381, 1T0S256388, 1T0S256390, 1T0S256393, 1T0S256397, 1T0S256400, 1T0S256601, 1T0S256602, 1T0S256617, 1T0S256639, 1T0S256457, 1T0S256538, 1T0S256662, 1T0S256673, 4T0S276847, 4T0S276856, 4T0S276869, 4T0S276875, 4T0S276893, 4T0S276904, 4T0S276432, 4T0S276475, 4T0S276480, 4T0S276486, 4T0S276493, 4T0S276509, 4T0S276516, 4T0S276518, 4T0S276526, 4T0S276532, 4T0S276538, 4T0S276557, 4T0S276559, 4T0S276561, 4T0S276563, 4T0S276578, 4T0S276580, 4T0S276583, 4T0S276584, 4T0S276595, 4T0S276600, 4T0S276603, 4T0S276614, 4T0S276621, 4T0S276624, 4T0S276626, 4T0S276631, 4T0S276633, 4T0S276659, 4T0S276812, 4T0S276813, 4T0S276821, 4T0S276835, 4T0S276981, 4T0S276985, 4T0S276990, 0T0S251625, 0T0S251629, 0T0S251636, 0T0S251637, 0T0S251643, 0T0S251653, 0T0S251658, 0T0S251660, 0T0S251665, 0T0S251682, 0T0S251684, 0T0S251686, 0T0S251688, 0T0S251694, 2T0S264287, 3T0S268675, 1T0S261624, 1T0S261640, 1T0S261652, 1T0S261654, 1T0S261657, 1T0S261659, 1T0S261676, 1T0S261677, 1T0S261678, 1T0S261679, 1T0S261683, 1T0S261688, 1T0S261689, 1T0S261690, 1T0S261691, 1T0S261693, 1T0S261706, 1T0S261882, 1T0S262009, 1T0S262017, 1T0S262034, 1T0S262093, 2T0S264283, 2T0S264694, 2T0S264722, 2T0S264735, 2T0S266873, 2T0S266893, 2T0S266901, 2T0S266927, 2T0S266939, 2T0S267272, 2T0S268648, 3T0S268650, 3T0S268658, 3T0S268667, 3T0S268671, 3T0S270437, 3T0S270461, 3T0S270473, 3T0S270496, 3T0S270497, 3T0S271656, 3T0S271705, 0T0S252018, 0T0S252060, 0T0S252065, 0T0S252069, 0T0S252075, 0T0S253461, 0T0S254041, 0T0S254077, 0T0S254082, 0T0S254089, 0T0S254092, 0T0S254097, 0T0S254100, 1T0S256636, 1T0S256642, 1T0S256646, 1T0S256649, 1T0S256653, 1T0S256655, 1T0S256657, 1T0S256666, 1T0S256671, 2T0S266472, 2T0S266479, 2T0S266486, 2T0S266488, 2T0S266492, 2T0S266493, 2T0S266499, 2T0S266503, 2T0S266506, 1T0S256337, 1T0S256394, 1T0S256403, 1T0S256404, 1T0S256434, 1T0S257306, 1T0S257316, 1T0S257342, 1T0S257345, 1T0S257348, 1T0S257354, 1T0S257355, 1T0S257838, 1T0S258205, 1T0S258521, 1T0S258529, 1T0S258566, 1T0S258572, 1T0S258580, 1T0S258955, 1T0S259012, 1T0S261262, 1T0S261402, 1T0S261418, 1T0S261429, 1T0S261443, 1T0S261446, 1T0S261448, 1T0S261451, 1T0S261452, 1T0S261459, 1T0S261463, 1T0S261464, 1T0S261466, 1T0S261467, 1T0S261468, 1T0S261473, 1T0S261474, 1T0S261475, 1T0S261478, 1T0S261480, 1T0S261482, 1T0S261484, 1T0S261498, 1T0S261502, 1T0S261503, 1T0S261507, 1T0S261633, 0T0S254005, 0T0S254040, 4T0S276421, 4T0S276460, 4T0S276523, 4T0S276556, 4T0S276558, 4T0S276571, 4T0S276590, 4T0S276599, 0T0S251633, 0T0S251634, 0T0S251638, 0T0S251639, 0T0S251640, 0T0S251642, 0T0S251644, 0T0S251647, 0T0S251648, 0T0S251649, 0T0S251650, 0T0S251651, 0T0S251652, 0T0S251655, 0T0S251664, 0T0S251666, 0T0S251667, 0T0S251668, 0T0S251669, 0T0S251671, 0T0S251678, 0T0S251679, 0T0S251680, 0T0S251689, 0T0S251691, 0T0S251695, 0T0S251724, 0T0S252088, 0T0S252205, 0T0S252211, 0T0S252215, 0T0S252217, 0T0S252222, 1T0S256946, 1T0S258571, 1T0S256813, 2T0S264690, 2T0S264698, 2T0S264700, 2T0S264701, 2T0S264709, 2T0S264719, 2T0S264733, 2T0S264745, 2T0S264769, 2T0S265440, 2T0S265533, 2T0S265549, 2T0S265582, 2T0S265598, 2T0S265606, 2T0S265624, 2T0S265625, 2T0S267642, 2T0S267674, 2T0S267696, 2T0S267705, 2T0S267803, 2T0S267810, 2T0S267815, 2T0S267828, 2T0S267845, 2T0S267865, 2T0S267889, 0T0S253186, 0T0S253314, 0T0S253337, 0T0S253345, 0T0S253349, 0T0S253359, 0T0S253364, 0T0S253368, 0T0S253379, 0T0S253380, 0T0S253381, 0T0S253387, 0T0S253388, 0T0S253391, 1T0S256807, 1T0S256812, 1T0S256814, 1T0S256817, 1T0S256819, 1T0S256820, 1T0S256822, 1T0S256828, 1T0S256832, 1T0S256833, 1T0S256835, 1T0S256836, 1T0S256839, 1T0S256840, 1T0S256841, 1T0S256842, 1T0S256843, 1T0S256854, 1T0S256859, 1T0S256869, 2T0S264105, 2T0S264113, 2T0S264114, 2T0S264119, 2T0S264121, 2T0S264153, 2T0S264157, 2T0S264164, 2T0S264166, 2T0S264167, 2T0S264168, 2T0S264170, 2T0S264173, 2T0S264196, 4T0S275659, 4T0S275667, 4T0S275672, 4T0S275675, 4T0S275676, 4T0S275680, 4T0S275690, 4T0S275693, 4T0S275694, 4T0S275701, 4T0S275704, 4T0S275705, 4T0S275709, 4T0S275714, 4T0S275716, 4T0S275718, 4T0S275719, 4T0S275724, 4T0S275725, 4T0S275726, 4T0S275727, 4T0S275741, 1T0S258438, 1T0S258451, 1T0S258452, 1T0S258471, 1T0S258473, 1T0S258478, 1T0S258479, 1T0S258481, 1T0S258482, 1T0S258501, 0T0S253478, 0T0S254095, 0T0S253256, 0T0S253278, 0T0S253282, 0T0S253302, 0T0S253303, 0T0S253305, 0T0S253308, 0T0S255305, 1T0S256096, 1T0S256097, 1T0S256099, 0T0S254136, 0T0S254142, 0T0S254145, 0T0S254146, 0T0S254148, 0T0S254153, 0T0S254154, 0T0S254158, 0T0S254159, 0T0S254160, 0T0S254161, 0T0S254163, 0T0S254165, 0T0S254166, 2T0S264295, 2T0S264300, 2T0S264342, 2T0S264348, 2T0S264351, 2T0S264354, 2T0S264360, 2T0S264361, 2T0S264384, 2T0S265473, 2T0S265498, 2T0S265505, 2T0S265536, 2T0S265577, 2T0S265607, 3T0S274905, 3T0S274918, 0T0S253424, 0T0S253433, 0T0S254636, 0T0S254652, 1T0S259529, 1T0S259552, 1T0S259562, 1T0S259570, 1T0S260032, 1T0S260057, 1T0S260059, 1T0S260062, 1T0S260066, 1T0S260071, 1T0S260073, 1T0S260076, 1T0S260079, 1T0S260081, 1T0S260083, 1T0S260084, 1T0S260085, 1T0S260086, 1T0S260087, 1T0S260088, 1T0S260089, 1T0S260092, 1T0S260093, 1T0S260094, 1T0S260095, 1T0S260096, 1T0S260098, 1T0S260100, 1T0S260101, 1T0S260102, 1T0S260103, 1T0S260104, 1T0S260105, 1T0S260107, 1T0S260108, 1T0S260110, 1T0S260111, 1T0S260112, 1T0S260113, 1T0S260114, 1T0S260115, 1T0S260118, 1T0S260123, 1T0S260127, 1T0S260134, 0T0S253212, 0T0S253419, 0T0S254119, 0T0S254128, 2T0S263462, 2T0S263486, 0T0S251622, 0T0S251623, 0T0S251624, 0T0S251626, 0T0S251627, 0T0S251628, 0T0S251630, 0T0S251631, 0T0S251632, 0T0S251635, 0T0S251641, 0T0S251645, 0T0S251646, 0T0S251661, 0T0S251662, 0T0S251663, 0T0S251672, 4T0S277032, 4T0S277101, 4T0S277107, 4T0S277113, 4T0S277116, 4T0S277119, 4T0S277352, 4T0S277358, 4T0S277373, 4T0S277377, 4T0S277384, 4T0S277406, 4T0S277410, 4T0S277411, 4T0S277412, 4T0S277430, 4T0S277432, 4T0S277442, 4T0S277447, 4T0S277448, 4T0S277456, 4T0S277479, 4T0S277484, 4T0S277506, 4T0S277512, 4T0S277522, 4T0S277529, 4T0S277530, 0T0S255055, 4T0S277421, 4T0S277428, 0T0S253413, 0T0S253481, 0T0S253502, 0T0S253518, 0T0S253519, 0T0S253522, 0T0S253533, 0T0S253536, 0T0S253539, 0T0S253540, 0T0S253541, 0T0S253545, 0T0S253546, 0T0S253566, 0T0S253574, 0T0S253600, 0T0S253607, 0T0S253614, 0T0S253615, 0T0S253639, 1T0S261974, 1T0S261987, 1T0S261999, 1T0S262005, 1T0S262018, 1T0S262021, 1T0S262028, 1T0S262043, 1T0S262044, 1T0S262045, 1T0S262047, 1T0S262049, 1T0S262050, 1T0S262053, 1T0S262055, 1T0S262058, 1T0S262061, 1T0S262062, 1T0S262063, 1T0S262064, 1T0S262071, 1T0S262080, 1T0S262089, 4T0S276907, 4T0S276922, 4T0S276948, 4T0S276986, 4T0S277009, 4T0S277011, 4T0S277015, 4T0S277020, 4T0S277021, 4T0S277022, 4T0S277024, 4T0S277027, 4T0S277028, 4T0S277029, 4T0S277030, 4T0S277034, 4T0S277035, 4T0S277037, 4T0S277038, 4T0S277039, 4T0S277041, 4T0S277042, 4T0S277043, 4T0S277046, 4T0S277052, 4T0S277056, 4T0S277057, 4T0S277063, 4T0S277065, 4T0S277067, 4T0S277069, 4T0S277073, 4T0S277074, 4T0S277075, 4T0S277076, 4T0S277086, 4T0S277087, 2T0S266474, 2T0S266478, 2T0S266480, 2T0S266485, 2T0S266966, 2T0S267009, 2T0S267018, 2T0S267020, 2T0S267023, 2T0S267029, 2T0S267033, 2T0S267039, 2T0S267041, 2T0S267042, 2T0S267043, 2T0S267044, 2T0S267045, 2T0S267047, 2T0S267051, 2T0S267052, 2T0S267053, 2T0S267054, 2T0S267057, 2T0S267620, 2T0S267625, 2T0S267629, 2T0S267631, 1T0S256953, 1T0S257324, 1T0S257333, 1T0S258181, 1T0S258207, 0T0S253938, 0T0S253949, 1T0S256713, 1T0S256764, 1T0S256804, 1T0S256810, 1T0S257019, 1T0S257026, 1T0S257043, 1T0S257087, 1T0S257122, 1T0S257265, 1T0S257373, 1T0S257399, 1T0S257403, 1T0S257414, 4T0S276328, 4T0S276332, 4T0S276336, 4T0S276338, 4T0S276341, 4T0S276345, 4T0S276354, 4T0S276355, 4T0S276360, 4T0S276368, 4T0S276389, 4T0S276391, 4T0S276401, 4T0S276402, 4T0S276411, 4T0S276413, 4T0S276414, 4T0S276424, 4T0S276431, 4T0S276468, 4T0S275732, 4T0S275766, 4T0S275775, 4T0S275779, 4T0S275782, 4T0S275785, 4T0S275799, 4T0S275803, 4T0S275805, 4T0S275812, 4T0S275814, 4T0S275822, 4T0S275824, 4T0S275833, 4T0S275842, 4T0S276458, 4T0S276520, 4T0S276521, 4T0S276536, 4T0S276539, 4T0S276552, 4T0S276565, 4T0S276570, 4T0S276574, 4T0S276577, 4T0S276586, 4T0S276589, 4T0S276596, 4T0S276606, 4T0S276607, 4T0S276609, 4T0S276616, 4T0S276635, 4T0S276636, 4T0S276639, 4T0S276644, 4T0S276645, 4T0S276648, 4T0S276649, 4T0S276651, 4T0S276654, 4T0S276655, 4T0S276656, 4T0S276661, 4T0S276662, 1T0S256276, 1T0S256295, 1T0S256303, 1T0S256312, 1T0S256314, 1T0S256315, 1T0S256316, 1T0S256320, 1T0S256321, 1T0S256323, 1T0S256326, 1T0S256332, 1T0S256333, 1T0S256339, 1T0S256342, 4T0S275611, 4T0S275612, 4T0S275621, 4T0S275625, 4T0S275630, 4T0S275631, 4T0S275632, 4T0S275633, 4T0S275644, 4T0S275645, 4T0S275648, 4T0S275661, 4T0S275663, 4T0S275665, 4T0S275669, 4T0S275674, 4T0S275678, 4T0S275686, 4T0S275695, 4T0S275697, 4T0S275698, 3T0S274826, 3T0S274940, 3T0S275215, 3T0S275221, 3T0S275230, 3T0S275234, 3T0S275243, 3T0S275244, 3T0S275245, 3T0S275247, 3T0S275251, 3T0S275253, 3T0S275256, 3T0S275257, 3T0S275258, 3T0S275259, 3T0S275260, 3T0S275261, 3T0S275262, 3T0S275263, 3T0S275264, 3T0S275266, 3T0S275268, 3T0S275270, 3T0S275273, 3T0S275275, 3T0S275278, 1T0S257252, 1T0S256717, 1T0S256937, 1T0S257009, 1T0S257036, 1T0S257038, 1T0S257092, 1T0S257139, 1T0S257165, 1T0S257204, 1T0S257303, 1T0S257364, 1T0S257374, 1T0S257386, 1T0S257404, 1T0S257411, 4T0S275622, 4T0S275651, 4T0S275744, 4T0S275761, 4T0S275771, 4T0S275784, 4T0S275790, 4T0S275795, 4T0S275796, 4T0S275806, 4T0S275809, 4T0S275810, 4T0S275811, 4T0S275815, 4T0S275816, 4T0S275817, 4T0S275818, 4T0S275820, 4T0S275821, 4T0S275825, 4T0S275826, 4T0S275830, 4T0S275831, 3T0S273843, 3T0S273846, 3T0S273850, 3T0S273872, 3T0S273874, 3T0S273881, 3T0S273891, 3T0S273893, 3T0S273894, 3T0S273898, 3T0S273900, 3T0S273901, 3T0S273903, 3T0S273904, 3T0S273905, 3T0S273906, 3T0S273908, 3T0S273909, 3T0S273910, 3T0S273914, 3T0S273915, 3T0S273916, 3T0S273917, 3T0S273918, 3T0S273919, 3T0S273922, 3T0S273924, 3T0S273927, 3T0S273928, 3T0S273929, 3T0S273930, 3T0S273932, 3T0S273935, 3T0S273936, 3T0S273937, 3T0S273938, 3T0S273939, 3T0S273940, 3T0S273941, 3T0S273967, 3T0S274447, 1T0S257812, 1T0S257829, 1T0S257853, 1T0S257874, 1T0S257900, 1T0S257910, 1T0S257916, 1T0S257919, 1T0S257921, 1T0S257926, 1T0S257927, 1T0S257930, 1T0S257932, 1T0S257936, 1T0S257937, 1T0S257941, 1T0S257944, 1T0S257947, 1T0S257950, 1T0S257951, 1T0S257952, 1T0S257958, 1T0S257963, 1T0S257965, 1T0S258116, 1T0S258117, 1T0S258118, 1T0S258123, 1T0S258124, 1T0S258129, 1T0S258855, 1T0S258862, 1T0S258896, 2T0S266491, 2T0S266498, 2T0S266504, 2T0S266507, 2T0S266921, 2T0S266929, 2T0S266941, 2T0S266954, 2T0S266959, 2T0S266962, 2T0S266963, 2T0S266965, 2T0S266971, 2T0S266975, 2T0S266977, 2T0S266979, 2T0S266981, 2T0S266982, 2T0S266985, 2T0S266986, 2T0S266988, 2T0S266993, 2T0S266996, 2T0S267005, 2T0S267008, 2T0S267012, 2T0S267017, 4T0S275921, 4T0S275932, 4T0S275934, 4T0S275942, 4T0S275952, 4T0S275954, 4T0S275958, 4T0S275959, 4T0S275960, 4T0S275962, 4T0S275967, 4T0S275970, 4T0S275974, 4T0S275977, 4T0S275980, 4T0S275982, 4T0S275983, 4T0S275984, 1T0S258783, 1T0S258787, 1T0S258799, 1T0S258802, 1T0S258806, 1T0S258813, 1T0S258824, 1T0S258826, 1T0S258833, 1T0S258842, 1T0S258859, 1T0S258860, 1T0S258878, 1T0S256633, 1T0S256638, 1T0S257291, 1T0S258169, 1T0S258778, 1T0S258788, 1T0S258975, 1T0S259005, 1T0S261242, 1T0S261292, 1T0S261304, 1T0S261305, 1T0S261317, 1T0S261318, 1T0S261321, 1T0S261327, 1T0S261328, 1T0S261337, 1T0S261345, 1T0S261346, 1T0S261350, 1T0S261351, 1T0S261353, 1T0S261364, 1T0S261367, 1T0S261373, 1T0S261381, 1T0S261387, 1T0S261405, 1T0S261406, 1T0S261408, 1T0S261442, 1T0S256665, 1T0S256668, 1T0S256689, 1T0S262618, 1T0S262632, 1T0S262633, 2T0S262867, 2T0S262881, 2T0S262882, 2T0S262886, 2T0S262888, 2T0S262896, 2T0S262915, 2T0S262928, 2T0S263018, 2T0S263019, 2T0S263032, 2T0S263047, 2T0S263057, 2T0S263060, 2T0S263066, 2T0S263075, 2T0S263082, 2T0S263092, 2T0S263099, 2T0S263102, 2T0S263103, 2T0S263112, 2T0S263113, 2T0S263124, 2T0S263130, 2T0S263131, 2T0S263141, 2T0S263149, 2T0S263152, 2T0S263153, 2T0S263159, 2T0S263161, 2T0S263163, 2T0S263164, 2T0S263165, 2T0S263166, 2T0S263168, 2T0S263170, 2T0S263171, 2T0S263172, 2T0S263173, 2T0S263176, 2T0S263178, 2T0S263181, 2T0S263182, 2T0S263184, 2T0S263186, 2T0S263187, 2T0S263188, 2T0S263198, 2T0S264714, 2T0S264916, 2T0S264955, 2T0S265055, 2T0S266125, 2T0S266177, 2T0S266191, 2T0S266207, 2T0S267712, 2T0S267808, 2T0S267812, 2T0S267816, 2T0S268042, 1T0S261049, 1T0S261054, 1T0S261109, 1T0S261116, 1T0S261136, 1T0S261140, 1T0S261141, 1T0S261151, 1T0S261159, 1T0S261169, 1T0S261172, 1T0S261184, 1T0S261191, 1T0S261194, 1T0S261208, 1T0S261216, 1T0S261228, 1T0S261232, 1T0S261234, 1T0S261236, 1T0S261237, 1T0S261238, 1T0S261613, 1T0S261686, 1T0S261694, 1T0S261695, 1T0S261701, 1T0S261702, 1T0S261703, 1T0S261704, 1T0S261708, 1T0S261709, 1T0S261734, 1T0S261740, 2T0S266495, 2T0S266502, 2T0S266508, 2T0S266510, 2T0S266513, 2T0S266516, 2T0S266517, 2T0S266519, 2T0S266523, 2T0S266526, 2T0S266527, 2T0S266529, 2T0S266531, 2T0S266532, 2T0S266534, 2T0S266535, 2T0S266538, 2T0S266539, 1T0S259467, 1T0S259524, 1T0S259526, 1T0S259528, 1T0S259546, 1T0S259560, 1T0S259561, 1T0S259569, 1T0S259585, 1T0S259593, 1T0S259596, 1T0S259599, 1T0S259600, 1T0S259602, 1T0S259607, 1T0S259609, 1T0S259610, 1T0S259611, 1T0S259613, 1T0S259614, 1T0S259615, 1T0S259616, 1T0S259617, 1T0S259619, 1T0S259620, 1T0S259621, 1T0S259622, 1T0S259623, 1T0S259628, 1T0S259636, 1T0S259638, 1T0S259639, 1T0S259644, 1T0S259646, 1T0S259652, 1T0S259654, 1T0S261078, 1T0S261082, 1T0S261095, 1T0S261801, 1T0S261804, 1T0S261807, 1T0S261808, 1T0S261809, 1T0S261817, 1T0S261818, 1T0S261820, 1T0S261825, 1T0S261827, 1T0S261837, 1T0S261838, 1T0S261839, 1T0S261842, 1T0S261843, 1T0S261847, 1T0S261850, 1T0S261855, 1T0S261857, 1T0S261873, 2T0S263876, 2T0S263878, 2T0S263891, 2T0S263901, 2T0S263905, 2T0S263906, 2T0S263907, 2T0S263995, 2T0S263996, 2T0S263999, 2T0S264008, 2T0S264013, 2T0S264017, 2T0S264027, 2T0S264029, 2T0S264031, 2T0S264032, 2T0S264035, 2T0S264037, 2T0S264038, 2T0S264040, 2T0S264041, 2T0S264042, 2T0S264045, 2T0S264046, 2T0S264048, 2T0S264049, 2T0S264050, 3T0S268721, 3T0S268739, 3T0S268740, 3T0S268742, 3T0S268744, 3T0S268746, 3T0S268749, 3T0S268752, 3T0S268756, 3T0S268758, 3T0S268761, 3T0S268762, 3T0S268763, 3T0S268766, 3T0S268768, 3T0S268769, 3T0S268771, 3T0S268774, 3T0S268776, 3T0S268778, 3T0S268783, 3T0S270024, 3T0S270050, 3T0S270059, 3T0S270065, 3T0S270067, 3T0S270069, 3T0S270072, 3T0S270073, 3T0S270074, 3T0S270075, 3T0S270076, 3T0S270077, 3T0S270078, 3T0S270082, 3T0S270089, 3T0S270090, 3T0S270094, 3T0S270597, 3T0S270609, 3T0S270613, 3T0S270616, 3T0S271103, 3T0S271109, 3T0S271111, 3T0S271114, 3T0S271128, 3T0S271134, 3T0S271139, 3T0S271145, 3T0S271146, 3T0S271147, 3T0S271149, 3T0S271151, 3T0S271156, 3T0S271157, 3T0S273849, 3T0S273858, 3T0S273862, 3T0S273868, 3T0S273870, 3T0S273878, 3T0S273879, 3T0S273882, 3T0S273884, 3T0S273886, 3T0S273888, 3T0S273895, 3T0S273897, 3T0S273907, 3T0S273911, 0T0S253530, 2T0S263399, 2T0S263404, 2T0S263423, 2T0S263440, 2T0S263452, 2T0S263454, 2T0S263461, 2T0S263519, 2T0S263621, 2T0S263797, 2T0S263799, 2T0S263800, 2T0S263805, 2T0S263816, 2T0S263822, 2T0S263827, 2T0S263828, 2T0S263829, 2T0S263830, 2T0S263831, 2T0S263832, 2T0S263838, 2T0S263841, 2T0S263842, 2T0S263852, 2T0S263854, 2T0S263863, 2T0S263865, 2T0S263867, 2T0S263868, 2T0S263886, 2T0S263904, 1T0S261539, 1T0S261707, 1T0S261711, 1T0S261712, 1T0S261725, 1T0S261728, 1T0S261729, 1T0S261730, 1T0S261733, 1T0S261735, 1T0S261737, 1T0S261742, 1T0S261746, 1T0S261748, 1T0S261752, 1T0S261779, 1T0S261782, 1T0S261783, 1T0S261785, 1T0S261786, 1T0S256774, 1T0S256797, 1T0S256800, 1T0S260023, 1T0S260024, 1T0S260025, 1T0S260029, 1T0S260033, 1T0S260037, 1T0S260038, 1T0S260040, 1T0S260041, 1T0S260042, 1T0S260043, 1T0S260048, 1T0S260051, 1T0S260054, 1T0S260056, 1T0S260061, 1T0S260075, 1T0S260082, 1T0S260099, 3T0S271118, 3T0S271168, 3T0S271172, 3T0S271188, 3T0S271190, 3T0S271194, 3T0S271202, 0T0S248750, 0T0S253361, 0T0S253385, 0T0S254530, 1T0S256799, 1T0S257182, 3T0S268695, 3T0S268708, 3T0S268727, 2T0S266473, 2T0S266972, 2T0S266991, 2T0S267021, 2T0S267046, 2T0S267066, 2T0S267081, 2T0S267084, 2T0S267089, 2T0S267102, 2T0S267103, 2T0S267111, 2T0S267112, 2T0S267115, 2T0S267117, 2T0S267118, 2T0S267121, 2T0S267126, 2T0S267127, 2T0S267130, 2T0S267132, 2T0S267139, 2T0S267141, 2T0S267149, 2T0S267157, 2T0S268017, 2T0S268018, 2T0S268021, 2T0S268023, 2T0S268025, 1T0S262292, 1T0S262303, 1T0S261104, 1T0S261119, 1T0S261824, 1T0S261846, 1T0S261848, 1T0S261856, 1T0S261858, 1T0S261860, 1T0S261863, 1T0S261898, 1T0S261904, 1T0S261910, 1T0S261919, 1T0S261920, 1T0S261921, 1T0S261924, 1T0S261926, 1T0S261938, 1T0S261939, 1T0S261947, 1T0S261952, 1T0S261955, 1T0S261957, 1T0S261961, 1T0S258905, 1T0S258908, 1T0S258923, 1T0S258924, 1T0S258929, 1T0S258930, 1T0S258934, 1T0S258937, 1T0S258938, 1T0S258940, 1T0S258942, 1T0S258950, 1T0S258951, 1T0S258953, 1T0S258954, 1T0S258956, 1T0S258958, 1T0S258960, 1T0S258962, 1T0S258964, 1T0S258968, 1T0S258969, 1T0S258971, 1T0S258973, 1T0S258978, 1T0S259002, 1T0S259003, 3T0S271600, 3T0S271739, 4T0S275892, 4T0S275897, 4T0S275898, 4T0S275903, 4T0S275906, 4T0S275910, 4T0S275923, 4T0S275940, 2T0S264763, 3T0S268653, 3T0S268659, 3T0S268676, 3T0S268684, 3T0S268687, 3T0S268691, 3T0S268693, 3T0S268694, 3T0S268697, 3T0S268698, 3T0S268699, 3T0S268700, 3T0S268701, 3T0S268724, 3T0S268725, 3T0S268728, 3T0S268755, 3T0S270095, 3T0S270098, 3T0S270448, 3T0S270463, 3T0S270494, 3T0S270653, 3T0S270662, 3T0S270663, 3T0S270665, 3T0S270667, 3T0S270671, 3T0S270674, 3T0S270682, 3T0S271908, 0T0S253512, 0T0S253514, 0T0S254023, 0T0S254030, 0T0S254031, 0T0S254034, 0T0S254036, 0T0S254039, 0T0S254042, 0T0S254050, 0T0S254053, 0T0S254055, 0T0S254056, 0T0S254059, 0T0S254061, 0T0S254066, 0T0S254067, 0T0S254073, 0T0S254075, 0T0S254079, 0T0S254081, 0T0S254084, 0T0S254085, 0T0S254499, 0T0S254504, 0T0S254520, 0T0S254532, 0T0S254548, 0T0S253521, 0T0S253526, 0T0S253548, 2T0S263826, 2T0S263834, 2T0S263837, 2T0S263840, 3T0S268663, 3T0S268670, 3T0S268672, 3T0S268681, 3T0S268686, 3T0S268689, 3T0S268690, 3T0S268692, 0T0S253446, 0T0S253557, 0T0S253559, 0T0S253577, 0T0S253583, 0T0S253430, 1T0S259635, 1T0S260047, 3T0S271937, 3T0S271945, 0T0S254227, 0T0S255391, 0T0S255395, 0T0S255396, 0T0S255399, 0T0S255406, 0T0S255407, 0T0S255412, 0T0S255426, 0T0S255427, 0T0S255428, 0T0S255431, 0T0S255435, 0T0S255438, 0T0S255442, 0T0S255640, 0T0S255641, 0T0S255642, 0T0S255646, 0T0S255662, 1T0S255718, 1T0S255797, 0T0S251654, 0T0S251681, 0T0S251685, 0T0S251692, 0T0S251693, 0T0S251697, 0T0S251698, 0T0S251700, 0T0S251701, 0T0S251705, 0T0S251707, 0T0S251708, 0T0S251710, 0T0S251720, 2T0S266583, 2T0S266595, 2T0S266609, 2T0S266614, 2T0S266618, 2T0S266621, 2T0S266624, 2T0S266625, 2T0S266854, 2T0S266858, 2T0S266859, 2T0S266862, 2T0S266881, 2T0S266884, 2T0S266984, 3T0S270587, 3T0S271141, 3T0S271142, 3T0S271161, 0T0S251659, 0T0S251673, 0T0S251675, 0T0S251677, 0T0S251702, 0T0S251711, 0T0S251717, 0T0S251721, 0T0S251723, 0T0S251726, 0T0S251728, 0T0S251739, 0T0S251748, 0T0S251752, 0T0S252644, 0T0S252647, 0T0S252649, 0T0S252651, 0T0S252654, 0T0S252659, 0T0S252667, 0T0S252668, 0T0S252670, 0T0S252691, 0T0S252694, 0T0S252229, 3T0S271766, 3T0S271772, 3T0S271775, 3T0S271808, 3T0S271838, 3T0S271930, 3T0S271951, 1T0S259721, 0T0S252729, 0T0S252744, 0T0S252752, 2T0S264866, 2T0S264873, 2T0S265502, 2T0S265521, 2T0S265537, 2T0S265561, 2T0S265573, 2T0S265593, 2T0S265596, 2T0S265601, 2T0S265603, 2T0S265611, 2T0S266882, 2T0S266898, 2T0S264160, 2T0S264206, 0T0S253234, 0T0S253444, 0T0S253532, 0T0S253563, 0T0S253570, 0T0S253582, 0T0S252066, 0T0S252074, 0T0S252086, 0T0S254539, 0T0S254540, 0T0S254553, 0T0S254556, 1T0S257062, 1T0S257085, 1T0S257185, 1T0S257195, 1T0S258462, 1T0S258476, 1T0S258484, 1T0S258495, 1T0S258503, 1T0S258507, 1T0S258509, 1T0S258522, 0T0S252592, 0T0S252601, 0T0S252612, 0T0S252613, 0T0S252616, 0T0S252623, 0T0S252624, 0T0S252625, 0T0S252630, 0T0S252631, 0T0S252634, 0T0S252639, 0T0S252641, 0T0S252655, 0T0S252685, 0T0S252689, 0T0S252693, 0T0S252703, 0T0S252704, 0T0S252707, 0T0S252709, 0T0S252716, 0T0S252718, 0T0S252721, 0T0S252731, 0T0S252734, 0T0S252760, 0T0S253527, 0T0S253564, 0T0S253573, 0T0S253595, 0T0S253597, 0T0S253598, 0T0S253599, 0T0S253601, 0T0S253602, 0T0S253603, 0T0S253604, 0T0S253605, 0T0S253606, 0T0S253609, 0T0S253613, 0T0S253619, 0T0S253627, 0T0S253631, 0T0S253637, 0T0S253641, 0T0S253642, 1T0S256541, 1T0S256680, 1T0S256710, 1T0S256716, 1T0S256728, 1T0S256733, 1T0S256734, 1T0S256741, 1T0S256743, 1T0S256749, 1T0S256750, 1T0S256751, 1T0S256753, 1T0S256754, 1T0S256756, 1T0S256757, 1T0S256760, 1T0S256767, 1T0S256768, 1T0S256770, 1T0S256782, 1T0S256783, 1T0S256785, 1T0S256798, 1T0S257017, 1T0S257031, 1T0S257090, 3T0S274005, 3T0S274027, 3T0S274067, 3T0S274099, 3T0S274103, 3T0S274104, 3T0S274106, 3T0S274114, 3T0S274115, 3T0S274119, 3T0S274122, 3T0S274142, 3T0S274849, 3T0S274856, 3T0S274860, 3T0S274926, 3T0S274931, 3T0S274934, 3T0S274947, 3T0S274951, 3T0S274958, 3T0S274960, 3T0S274962, 3T0S274971, 3T0S274972, 3T0S274979, 3T0S274981, 3T0S274983, 3T0S274984, 3T0S274991, 3T0S274995, 3T0S275000, 1T0S261561, 1T0S261569, 1T0S261576, 1T0S261580, 1T0S261586, 1T0S261591, 1T0S261596, 1T0S261602, 1T0S261610, 1T0S261623, 1T0S261626, 1T0S261627, 1T0S261629, 1T0S261630, 1T0S261632, 1T0S261637, 1T0S261639, 1T0S261642, 1T0S261646, 1T0S261653, 3T0S268794, 3T0S268795, 3T0S269665, 3T0S269676, 3T0S269677, 3T0S269681, 3T0S269682, 3T0S269683, 3T0S269687, 3T0S269689, 3T0S269696, 3T0S269706, 3T0S269748, 3T0S269799, 3T0S269821, 3T0S269824, 3T0S269825, 3T0S269832, 3T0S269833, 3T0S269840, 3T0S269845, 3T0S269846, 3T0S269847, 3T0S269967, 1T0S261051, 1T0S261071, 1T0S261075, 1T0S261076, 1T0S261077, 1T0S261079, 1T0S261081, 1T0S261083, 1T0S261087, 1T0S261088, 1T0S261090, 1T0S261092, 1T0S261094, 1T0S261098, 1T0S261100, 1T0S261102, 1T0S261105, 1T0S261107, 1T0S261111, 1T0S261118, 1T0S261123, 1T0S261124, 1T0S261130, 1T0S261138, 1T0S261139, 1T0S261145, 1T0S261153, 1T0S261155, 1T0S261156, 3T0S269962, 3T0S269969, 3T0S269970, 3T0S269971, 3T0S269973, 3T0S269974, 3T0S269975, 3T0S269977, 3T0S269980, 3T0S269981, 3T0S269982, 3T0S269983, 3T0S269984, 3T0S269992, 3T0S269994, 3T0S270004, 3T0S270009, 3T0S270019, 1T0S261608, 1T0S261614, 1T0S261648, 1T0S261655, 1T0S261658, 1T0S261661, 1T0S261662, 1T0S261663, 1T0S261664, 1T0S261665, 1T0S261667, 1T0S261668, 1T0S261669, 1T0S261670, 1T0S261671, 1T0S261672, 1T0S261674, 1T0S261682, 1T0S261684, 1T0S261687, 1T0S261697, 1T0S261698, 1T0S261699, 1T0S261700, 0T0S251985, 0T0S251998, 0T0S251999, 0T0S252001, 0T0S252003, 0T0S252004, 0T0S252005, 0T0S252019, 0T0S252047, 0T0S252064, 0T0S252067, 0T0S252070, 0T0S252076, 0T0S252078, 0T0S252080, 0T0S252082, 0T0S252083, 0T0S252202, 0T0S251981, 0T0S251995, 0T0S252002, 0T0S252006, 0T0S252007, 0T0S252008, 0T0S252009, 0T0S252012, 0T0S252014, 0T0S252015, 0T0S252017, 0T0S252020, 0T0S252021, 0T0S252025, 0T0S252027, 0T0S252039, 0T0S252040, 0T0S252046, 0T0S252049, 0T0S252050, 0T0S252051, 0T0S252056, 3T0S273931, 3T0S273949, 3T0S273971, 3T0S273974, 3T0S273982, 3T0S273983, 3T0S273986, 3T0S273991, 3T0S274002, 3T0S274013, 3T0S274022, 3T0S274028, 3T0S274033, 3T0S274038, 3T0S274063, 3T0S274070, 3T0S274073, 3T0S274144, 3T0S274855, 3T0S274889, 3T0S274910, 3T0S274939, 0T0S252688, 0T0S252722, 0T0S252737, 0T0S252740, 0T0S252743, 0T0S252746, 0T0S252749, 0T0S252751, 0T0S252753, 0T0S252757, 0T0S252758, 0T0S252765, 0T0S252766, 4T0S277344, 4T0S277346, 4T0S277348, 4T0S277353, 4T0S277361, 4T0S277364, 4T0S277365, 4T0S277366, 4T0S277368, 4T0S277376, 4T0S277378, 4T0S277379, 4T0S277386, 4T0S277395, 4T0S277396, 4T0S277400, 4T0S277402, 4T0S277405, 4T0S277407, 4T0S277415, 4T0S277423, 4T0S277425, 4T0S277427, 4T0S277440, 4T0S277441, 4T0S277443, 4T0S277450, 4T0S277457, 4T0S277460, 4T0S277463, 4T0S277467, 4T0S277468, 4T0S277469, 4T0S277482, 4T0S277490, 4T0S277492, 1T0S259461, 1T0S259495, 1T0S259523, 1T0S259535, 1T0S259563, 1T0S259564, 1T0S259565, 1T0S259571, 1T0S260052, 1T0S260065, 1T0S260068, 1T0S260070, 1T0S261148, 1T0S261840, 1T0S261880, 1T0S261894, 1T0S262181, 1T0S262186, 1T0S262200, 1T0S262224, 1T0S262233, 1T0S262236, 1T0S262238, 1T0S262239, 1T0S262242, 1T0S262243, 1T0S261948, 1T0S261953, 1T0S261969, 1T0S261972, 1T0S261976, 1T0S261977, 1T0S261978, 1T0S261981, 1T0S261985, 1T0S261986, 1T0S261989, 1T0S261991, 1T0S261992, 1T0S261995, 1T0S261996, 1T0S261997, 1T0S262001, 1T0S262004, 1T0S262008, 1T0S262012, 0T0S255065, 0T0S255118, 0T0S255121, 0T0S255133, 4T0S275946, 4T0S275951, 4T0S275956, 4T0S275957, 4T0S276333, 4T0S276334, 4T0S276347, 4T0S276376, 4T0S276382, 4T0S276385, 4T0S276390, 4T0S276395, 4T0S276398, 4T0S276412, 4T0S276418, 4T0S276548, 4T0S276585, 4T0S276594, 4T0S276632, 3T0S268732, 3T0S268736, 3T0S268782, 3T0S268797, 3T0S269740, 3T0S269750, 3T0S269753, 3T0S269768, 3T0S269771, 3T0S269772, 3T0S269783, 3T0S269786, 3T0S269790, 3T0S269791, 3T0S269795, 3T0S269802, 3T0S269815, 3T0S269818, 3T0S269820, 3T0S269831, 3T0S269839, 3T0S269848, 3T0S269849, 3T0S269850, 2T0S264202, 2T0S264209, 2T0S264212, 2T0S264213, 2T0S264217, 2T0S264219, 1T0S256568, 1T0S256589, 1T0S256600, 1T0S256606, 1T0S256619, 1T0S256620, 1T0S256621, 1T0S256623, 1T0S256624, 1T0S256626, 1T0S256627, 1T0S256629, 1T0S256630, 1T0S256631, 1T0S256632, 1T0S256640, 1T0S256643, 1T0S256644, 1T0S256648, 4T0S277409, 4T0S277579, 4T0S277608, 4T0S277619, 4T0S277627, 4T0S277631, 4T0S277645, 4T0S277654, 4T0S277663, 4T0S277666, 4T0S277673, 4T0S277674, 4T0S277685, 4T0S277687, 4T0S277727, 4T0S277747, 4T0S277748, 4T0S277751, 1T0S262499, 1T0S262503, 1T0S262504, 1T0S262505, 1T0S262507, 1T0S262508, 1T0S262510, 1T0S262511, 1T0S262513, 1T0S262516, 1T0S262517, 1T0S262523, 1T0S262524, 1T0S262533, 1T0S262537, 1T0S262539, 1T0S262543, 1T0S262547, 1T0S262562, 4T0S277389, 4T0S277462, 4T0S277472, 4T0S277498, 4T0S277502, 4T0S277508, 4T0S277524, 4T0S277525, 4T0S277535, 4T0S277536, 4T0S277537, 4T0S277539, 4T0S277540, 4T0S277543, 4T0S277544, 4T0S277548, 4T0S277549, 4T0S277553, 4T0S277567, 4T0S277568, 4T0S277589, 4T0S275710, 4T0S275723, 4T0S275746, 4T0S275752, 4T0S275756, 4T0S275772, 4T0S275773, 4T0S275777, 4T0S275792, 4T0S275801, 4T0S275808, 4T0S275819, 4T0S275823, 4T0S275834, 4T0S275835, 4T0S275836, 4T0S275838, 4T0S275839, 4T0S275840, 4T0S275841, 4T0S275843, 4T0S275844, 4T0S275847, 4T0S275848, 4T0S275849, 4T0S275850, 4T0S275851, 4T0S275853, 4T0S275854, 4T0S275855, 4T0S275867, 4T0S275873, 4T0S275876, 4T0S275877, 4T0S275880, 4T0S275881, 0T0S252013, 0T0S252022, 0T0S252024, 0T0S252026, 0T0S252035, 0T0S252037, 0T0S252052, 0T0S252053, 0T0S252059, 0T0S252062, 0T0S252063, 0T0S252068, 0T0S252072, 0T0S252073, 0T0S252077, 0T0S252081, 0T0S252085, 0T0S252087, 2T0S264374, 2T0S264681, 2T0S264734, 2T0S264750, 2T0S264764, 2T0S264949, 2T0S264953, 2T0S265407, 2T0S266140, 2T0S268645, 3T0S268652, 3T0S268654, 3T0S268655, 3T0S268656, 3T0S268657, 3T0S268660, 3T0S268662, 3T0S268664, 3T0S268666, 3T0S268668, 3T0S268669, 3T0S268685, 3T0S268696, 3T0S268741, 3T0S269746, 0T0S255045, 0T0S253383, 2T0S264697, 2T0S264713, 2T0S264876, 2T0S264892, 2T0S265311, 2T0S266110, 2T0S266127, 2T0S266150, 2T0S266155, 2T0S266162, 2T0S266164, 2T0S266167, 2T0S266168, 2T0S266173, 2T0S266179, 2T0S266181, 2T0S266187, 2T0S266197, 2T0S266205, 2T0S266208, 2T0S267709, 2T0S267737, 4T0S275575, 4T0S275576, 4T0S275577, 4T0S275581, 4T0S275583, 4T0S275584, 4T0S275585, 4T0S275592, 4T0S275594, 4T0S275596, 4T0S275598, 4T0S275599, 4T0S275600, 4T0S275602, 4T0S275603, 4T0S275605, 4T0S275606, 4T0S275607, 1T0S258877, 1T0S261908, 1T0S261930, 1T0S261931, 1T0S261963, 0T0S255114, 1T0S258839, 1T0S258850, 1T0S258852, 1T0S258853, 1T0S258856, 1T0S258875, 1T0S258886, 1T0S258887, 1T0S258888, 1T0S258889, 1T0S258890, 1T0S258892, 1T0S258894, 1T0S258897, 1T0S258901, 1T0S258906, 1T0S258909, 1T0S258915, 1T0S258933, 3T0S270051, 3T0S270080, 3T0S270081, 3T0S270083, 3T0S270085, 3T0S270086, 3T0S270088, 3T0S270099, 3T0S270102, 3T0S270105, 3T0S270394, 3T0S270405, 3T0S270407, 3T0S270414, 2T0S267646, 2T0S267666, 2T0S267667, 2T0S267680, 2T0S267688, 2T0S267691, 2T0S267695, 2T0S267703, 2T0S267704, 2T0S267706, 2T0S267722, 2T0S267731, 2T0S267734, 2T0S267741, 2T0S267744, 2T0S267745, 2T0S267754, 2T0S267759, 2T0S267763, 2T0S267768, 2T0S267780, 2T0S267797, 2T0S267809, 2T0S267847, 2T0S267857, 2T0S267867, 4T0S276808, 4T0S276816, 4T0S276819, 4T0S276820, 4T0S276825, 4T0S276828, 4T0S276831, 4T0S276836, 4T0S276840, 4T0S276844, 4T0S276851, 4T0S276852, 4T0S276859, 4T0S276862, 4T0S276864, 4T0S276866, 0T0S251734, 0T0S251737, 2T0S267713, 2T0S267761, 2T0S267792, 2T0S267821, 2T0S267827, 2T0S267830, 2T0S267834, 2T0S267835, 2T0S267836, 2T0S267841, 2T0S267848, 2T0S267849, 2T0S267852, 2T0S267862, 1T0S256446, 1T0S256877, 1T0S256898, 1T0S256917, 1T0S256918, 1T0S256924, 1T0S256948, 1T0S256965, 1T0S256969, 1T0S256972, 1T0S256973, 1T0S256984, 1T0S256985, 1T0S256992, 1T0S257006, 1T0S257008, 1T0S257037, 2T0S263633, 2T0S263798, 2T0S263813, 2T0S263819, 2T0S263839, 2T0S263844, 2T0S263848, 2T0S263850, 2T0S263871, 2T0S263872, 2T0S263884, 2T0S263887, 2T0S263892, 2T0S263896, 2T0S263900, 2T0S263908, 2T0S263992, 2T0S264002, 2T0S264005, 2T0S264006, 2T0S264007, 2T0S264012, 2T0S264024, 2T0S264036, 2T0S267724, 2T0S267749, 2T0S267757, 2T0S267779, 2T0S267786, 2T0S267805, 2T0S267806, 2T0S267813, 2T0S267823, 2T0S267824, 2T0S267826, 2T0S267829, 2T0S267833, 2T0S267844, 2T0S267851, 2T0S267859, 2T0S267868, 1T0S258183, 1T0S258185, 1T0S258192, 1T0S258193, 1T0S258209, 1T0S258211, 1T0S258221, 1T0S258224, 1T0S258227, 1T0S258229, 1T0S258234, 1T0S258236, 1T0S258239, 1T0S258241, 1T0S258247, 1T0S258253, 1T0S258260, 1T0S258261, 1T0S258264, 1T0S258461, 1T0S258490, 1T0S258499, 1T0S259238, 1T0S259240, 1T0S259241, 1T0S259263, 1T0S259267, 1T0S259270, 1T0S259276, 1T0S259344, 1T0S259345, 1T0S259346, 1T0S259348, 1T0S259349, 1T0S259352, 1T0S259353, 1T0S259354, 1T0S259355, 1T0S259363, 1T0S259364, 1T0S259365, 1T0S259368, 1T0S256845, 1T0S256876, 1T0S256901, 1T0S256908, 1T0S256933, 1T0S256940, 1T0S256959, 1T0S256961, 1T0S256964, 1T0S256966, 1T0S256974, 1T0S256978, 1T0S256989, 1T0S256990, 1T0S256408, 1T0S259415, 1T0S259424, 1T0S259452, 1T0S259454, 1T0S259466, 1T0S259468, 1T0S259472, 1T0S259476, 1T0S259480, 1T0S259482, 1T0S259486, 1T0S259490, 1T0S259493, 1T0S259496, 1T0S259498, 1T0S259501, 1T0S259503, 1T0S259509, 1T0S259510, 1T0S259511, 1T0S259513, 1T0S259516, 1T0S259517, 1T0S256423, 1T0S256968, 1T0S257310, 1T0S257325, 1T0S257346, 1T0S257347, 1T0S257351, 1T0S257352, 1T0S257353, 1T0S257358, 1T0S257368, 1T0S257372, 1T0S257375, 1T0S257377, 1T0S257384, 1T0S257387, 1T0S257392, 1T0S257398, 2T0S267060, 2T0S267063, 2T0S267064, 2T0S267078, 2T0S267079, 2T0S267082, 2T0S267086, 2T0S267090, 2T0S267091, 2T0S267092, 2T0S267093, 2T0S267094, 2T0S267095, 2T0S267096, 2T0S267101, 2T0S267104, 2T0S267108, 2T0S267109, 2T0S267110, 1T0S256975, 1T0S256994, 1T0S257323, 1T0S257331, 1T0S258128, 1T0S258147, 1T0S258167, 1T0S258168, 1T0S258173, 1T0S258174, 1T0S258176, 1T0S258180, 1T0S258184, 1T0S258189, 1T0S258194, 1T0S258198, 1T0S258199, 1T0S258200, 1T0S258203, 1T0S258206, 1T0S258208, 1T0S258213, 1T0S258866, 1T0S261091, 1T0S261114, 1T0S261126, 1T0S261135, 1T0S261144, 1T0S261205, 1T0S261397, 1T0S261430, 1T0S261445, 1T0S261826, 1T0S261917, 1T0S261923, 1T0S261932, 1T0S261935, 1T0S261936, 1T0S261940, 1T0S261942, 1T0S261945, 1T0S261949, 1T0S261954, 1T0S261959, 1T0S261960, 1T0S261966, 1T0S255794, 1T0S255798, 1T0S256077, 1T0S256078, 1T0S256082, 1T0S256084, 1T0S256086, 1T0S256088, 1T0S256089, 1T0S256090, 1T0S256091, 1T0S256092, 1T0S256095, 1T0S256098, 1T0S256103, 3T0S273944, 3T0S273966, 3T0S273973, 3T0S273992, 3T0S273995, 3T0S274006, 3T0S274043, 3T0S274046, 3T0S274048, 3T0S274051, 3T0S274054, 3T0S274055, 3T0S274071, 3T0S274074, 3T0S274079, 3T0S274081, 3T0S274094, 3T0S274095, 3T0S274102, 3T0S274110, 3T0S274112, 3T0S274149, 3T0S274151, 3T0S274165, 3T0S274171, 3T0S274172, 3T0S274173, 4T0S277350, 4T0S277444, 4T0S277474, 4T0S277491, 4T0S277519, 4T0S277526, 4T0S277538, 4T0S277541, 4T0S277571, 4T0S277578, 4T0S277593, 4T0S277594, 4T0S277600, 4T0S277602, 4T0S277626, 4T0S277632, 4T0S277639, 4T0S277648, 4T0S277652, 4T0S277665, 4T0S277680, 4T0S277686, 4T0S277740, 2T0S264107, 2T0S264200, 2T0S264203, 2T0S264215, 2T0S264216, 2T0S264222, 2T0S264223, 2T0S264228, 2T0S264233, 2T0S264238, 2T0S264244, 2T0S264253, 2T0S264254, 2T0S264257, 1T0S259374, 1T0S259375, 1T0S259381, 1T0S259401, 1T0S259405, 1T0S259410, 1T0S259413, 1T0S259421, 1T0S259428, 1T0S259437, 1T0S259439, 1T0S259449, 1T0S259453, 1T0S259469, 4T0S276962, 4T0S277026, 4T0S277033, 4T0S277053, 4T0S277055, 4T0S277058, 4T0S277061, 4T0S277062, 4T0S277079, 4T0S277083, 4T0S277118, 3T0S273887, 3T0S273923, 3T0S273942, 3T0S273943, 3T0S273945, 3T0S273946, 3T0S273955, 3T0S273957, 3T0S273959, 3T0S273962, 3T0S273963, 3T0S273965, 3T0S273975, 3T0S273977, 3T0S273980, 3T0S273984, 3T0S273987, 3T0S273988, 3T0S273989, 3T0S273990, 3T0S273997, 3T0S274001, 0T0S253426, 0T0S253489, 0T0S253497, 0T0S253507, 0T0S253535, 1T0S262065, 1T0S262072, 1T0S262078, 1T0S262094, 1T0S262295, 1T0S262311, 1T0S262326, 4T0S275641, 4T0S275683, 4T0S275687, 4T0S275699, 4T0S275700, 4T0S275707, 4T0S275708, 4T0S275712, 4T0S275717, 4T0S275720, 4T0S275728, 4T0S275733, 4T0S275736, 4T0S275743, 4T0S275747, 4T0S275748, 4T0S275749, 4T0S275751, 4T0S275753, 4T0S275754, 4T0S275755, 4T0S275759, 4T0S275780, 4T0S275781, 4T0S275786, 1T0S256547, 1T0S256573, 1T0S256635, 1T0S256477, 1T0S256686, 1T0S256721, 0T0S253429, 0T0S253496, 4T0S276466, 3T0S274818, 3T0S274988, 3T0S275013, 3T0S275232, 3T0S275249, 3T0S275250, 3T0S275265, 3T0S275267, 3T0S275269, 3T0S275271, 3T0S275277, 4T0S275580, 4T0S275597, 4T0S275601, 4T0S275604, 3T0S270499, 3T0S270617, 3T0S270628, 3T0S270634, 3T0S270636, 3T0S270816, 3T0S270817, 3T0S270820, 3T0S270821, 3T0S270822, 3T0S270823, 3T0S270834, 3T0S270835, 3T0S270838, 3T0S271720, 3T0S271758, 3T0S271783, 3T0S271785, 3T0S271811, 3T0S271837, 3T0S271910, 3T0S271915, 3T0S271920, 0T0S253180, 0T0S255033, 0T0S255056, 0T0S255057, 0T0S255063, 0T0S255066, 0T0S255074, 0T0S255115, 0T0S255120, 0T0S255143, 0T0S255149, 4T0S276588, 4T0S276880, 4T0S276888, 4T0S276891, 4T0S276899, 4T0S276900, 4T0S276906, 4T0S276910, 4T0S276913, 4T0S276917, 4T0S276918, 4T0S276925, 4T0S276931, 4T0S276932, 4T0S276933, 4T0S276936, 4T0S276939, 4T0S276940, 4T0S276943, 4T0S276947, 4T0S276954, 4T0S276972, 4T0S277044, 3T0S273972, 3T0S273981, 3T0S273985, 3T0S273994, 3T0S273998, 3T0S273999, 3T0S274003, 3T0S274004, 3T0S274008, 3T0S274010, 3T0S274015, 3T0S274019, 3T0S274020, 3T0S274029, 3T0S274039, 3T0S274050, 3T0S274109, 3T0S274117, 3T0S274157, 3T0S274864, 3T0S274879, 3T0S274965, 1T0S262171, 1T0S262173, 1T0S262178, 1T0S262183, 1T0S262195, 1T0S262197, 1T0S262208, 1T0S262214, 1T0S262220, 1T0S262223, 1T0S262227, 1T0S262228, 1T0S262229, 1T0S262234, 0T0S253291, 0T0S253405, 0T0S253415, 0T0S253428, 0T0S253480, 0T0S253510, 2T0S263412, 2T0S263459, 2T0S263460, 2T0S263467, 2T0S263469, 2T0S263473, 2T0S263474, 2T0S263475, 2T0S263481, 2T0S263488, 2T0S263489, 2T0S263491, 2T0S263498, 2T0S263499, 2T0S263504, 2T0S263506, 2T0S263508, 2T0S263512, 2T0S263516, 2T0S263525, 2T0S263538, 2T0S263540, 2T0S263542, 2T0S263543, 0T0S254122, 0T0S254134, 2T0S267647, 2T0S267658, 2T0S267660, 2T0S267671, 2T0S267685, 2T0S267698, 2T0S267725, 2T0S267727, 2T0S267729, 2T0S267748, 2T0S267751, 2T0S267753, 2T0S267778, 2T0S267794, 3T0S268770, 3T0S268788, 3T0S270436, 3T0S270458, 3T0S270582, 3T0S270584, 3T0S270593, 3T0S270615, 3T0S270686, 3T0S270688, 3T0S270691, 3T0S270693, 3T0S270740, 3T0S270751, 3T0S270759, 3T0S270760, 3T0S270761, 3T0S270763, 3T0S270768, 3T0S270770, 3T0S270771, 3T0S270772, 3T0S270774, 3T0S270776, 3T0S270778, 3T0S270781, 3T0S270782, 3T0S270783, 3T0S270784, 3T0S270785, 3T0S270792, 3T0S270798, 3T0S270800, 3T0S270801, 3T0S271138, 3T0S271143, 0T0S252584, 0T0S252591, 0T0S252593, 0T0S252595, 0T0S252596, 0T0S252604, 0T0S252606, 0T0S252607, 0T0S252608, 0T0S252609, 0T0S252618, 0T0S252619, 0T0S252622, 1T0S256821, 1T0S256866, 1T0S256899, 1T0S256952, 1T0S256957, 1T0S256971, 1T0S256987, 1T0S256396, 1T0S256850, 1T0S258120, 1T0S258252, 1T0S258420, 1T0S258502, 1T0S258513, 1T0S258518, 1T0S258519, 1T0S258530, 1T0S258533, 1T0S258536, 1T0S258538, 1T0S258545, 1T0S258546, 1T0S258555, 1T0S258556, 1T0S258557, 1T0S258562, 1T0S258576, 1T0S258577, 1T0S258581, 1T0S258582, 1T0S258584, 1T0S258586, 1T0S258589, 1T0S258593, 1T0S258594, 1T0S256704, 1T0S256712, 1T0S256714, 1T0S256738, 1T0S256759, 1T0S256761, 1T0S256765, 1T0S256766, 1T0S256772, 1T0S256775, 1T0S256776, 1T0S256780, 1T0S256781, 1T0S256786, 1T0S256787, 1T0S256788, 1T0S256789, 1T0S256790, 1T0S256792, 1T0S256795, 1T0S256801, 1T0S256803, 1T0S256805, 1T0S256808, 1T0S257007, 1T0S257023, 1T0S257028, 1T0S257029, 1T0S257030, 1T0S257033, 1T0S257044, 1T0S257050, 1T0S257063, 1T0S257070, 1T0S257072, 1T0S257101, 4T0S276889, 4T0S276955, 4T0S276970, 4T0S276992, 4T0S277000, 4T0S277007, 4T0S277008, 4T0S277010, 4T0S277012, 4T0S277013, 4T0S277017, 4T0S277018, 4T0S277019, 4T0S277031, 4T0S277049, 3T0S269673, 3T0S269693, 3T0S269699, 3T0S269705, 3T0S269716, 3T0S269717, 3T0S269719, 3T0S269725, 3T0S269727, 3T0S269733, 3T0S269734, 3T0S269735, 3T0S269738, 3T0S269741, 3T0S269757, 3T0S269758, 3T0S269766, 3T0S269779, 3T0S269789, 3T0S269792, 3T0S269803, 3T0S269812, 0T0S253468, 0T0S253529, 0T0S253534, 0T0S253556, 0T0S253571, 0T0S253581, 0T0S253594, 4T0S276837, 4T0S276929, 4T0S276944, 4T0S276946, 4T0S277495, 4T0S277513, 4T0S277523, 4T0S277546, 4T0S276318, 4T0S276322, 4T0S276362, 4T0S276374, 4T0S276409, 4T0S276428, 4T0S276445, 4T0S276446, 4T0S276448, 4T0S276453, 4T0S276459, 4T0S276472, 4T0S276473, 4T0S276477, 4T0S276478, 4T0S276481, 4T0S276482, 4T0S276484, 4T0S276485, 4T0S276491, 4T0S276496, 4T0S276497, 4T0S276499, 4T0S276500, 4T0S276501, 4T0S276503, 4T0S276505, 4T0S276507, 4T0S276510, 4T0S276511, 4T0S276524, 4T0S276525, 4T0S276529, 4T0S276582, 0T0S251656, 0T0S251674, 0T0S251715, 0T0S251716, 0T0S251722, 0T0S251727, 0T0S251729, 0T0S251735, 0T0S251736, 0T0S251740, 0T0S251741, 0T0S251742, 0T0S251746, 0T0S251749, 0T0S251753, 0T0S251757, 0T0S251758, 0T0S251759, 0T0S251761, 0T0S251763, 0T0S251768, 0T0S251769, 0T0S251770, 0T0S252583, 0T0S252585, 0T0S252594, 1T0S256452, 1T0S256460, 1T0S256484, 1T0S256490, 1T0S256493, 1T0S256495, 1T0S256529, 1T0S256534, 1T0S256543, 1T0S256551, 1T0S256565, 1T0S256586, 1T0S256587, 1T0S256592, 1T0S256598, 1T0S256702, 1T0S256707, 1T0S256719, 0T0S253253, 3T0S268764, 3T0S268772, 3T0S270400, 3T0S270402, 3T0S270445, 3T0S270456, 3T0S270485, 3T0S270486, 3T0S270501, 3T0S270583, 3T0S270585, 3T0S270586, 3T0S270589, 3T0S270592, 3T0S270595, 3T0S271105, 3T0S271108, 3T0S271115, 3T0S271116, 3T0S271120, 3T0S271121, 3T0S271122, 3T0S271123, 3T0S271124, 3T0S271127, 3T0S271129, 3T0S271130, 3T0S271133, 3T0S271144, 3T0S271154, 3T0S270787, 3T0S270818, 3T0S270826, 3T0S270851, 3T0S270853, 3T0S270855, 3T0S270858, 3T0S270860, 3T0S270861, 3T0S270865, 3T0S270866, 3T0S270870, 3T0S270871, 3T0S270872, 3T0S270880, 3T0S270881, 3T0S270883, 3T0S270886, 3T0S270888, 3T0S270889, 3T0S270890, 3T0S270895, 1T0S257825, 1T0S257832, 1T0S257846, 1T0S257857, 1T0S257861, 1T0S257871, 1T0S257880, 1T0S257886, 1T0S257887, 1T0S257888, 1T0S257889, 1T0S257893, 1T0S257894, 1T0S257895, 1T0S257898, 1T0S257901, 1T0S257903, 1T0S257905, 1T0S257906, 1T0S257907, 1T0S257908, 1T0S257909, 1T0S257911, 1T0S257912, 1T0S257913, 1T0S257915, 1T0S257918, 1T0S257920, 1T0S257922, 1T0S257924, 1T0S257928, 1T0S257929, 1T0S257931, 1T0S257933, 1T0S257934, 1T0S257935, 1T0S257938, 1T0S257940, 1T0S257942, 1T0S257943, 1T0S257945, 1T0S257946, 1T0S257949, 1T0S257954, 1T0S257955, 1T0S257956, 1T0S257957, 1T0S257960, 1T0S257961, 1T0S257962, 1T0S257964, 1T0S259190, 2T0S265451, 2T0S266165, 2T0S268027, 2T0S268029, 2T0S268031, 2T0S268040, 2T0S268041, 2T0S268046, 2T0S268047, 2T0S268048, 2T0S268056, 2T0S268057, 2T0S268059, 2T0S268063, 2T0S268064, 2T0S268065, 2T0S268081, 2T0S268082, 2T0S268090, 2T0S268092, 2T0S268093, 2T0S268096, 2T0S268101, 2T0S268102, 2T0S268105, 2T0S268107, 2T0S268114, 3T0S270765, 3T0S270836, 3T0S270842, 3T0S270844, 3T0S270845, 3T0S270846, 3T0S270848, 3T0S270850, 3T0S270854, 3T0S270857, 3T0S270859, 3T0S270862, 3T0S270863, 3T0S270867, 3T0S270873, 3T0S270875, 3T0S270876, 3T0S270877, 3T0S270878, 3T0S270879, 3T0S270884, 3T0S270885, 3T0S270887, 3T0S270891, 3T0S270892, 3T0S270893, 3T0S270894, 2T0S266530, 2T0S266574, 2T0S266577, 2T0S267212, 2T0S267239, 2T0S267252, 3T0S271164, 3T0S271169, 3T0S271176, 3T0S271177, 3T0S271178, 3T0S271179, 3T0S271181, 3T0S271182, 3T0S271183, 3T0S271184, 3T0S271185, 3T0S271200, 3T0S271206, 3T0S271207, 3T0S271209, 3T0S271588, 3T0S271593, 3T0S271604, 0T0S255648, 0T0S255657, 0T0S255667, 0T0S255683, 0T0S255690, 0T0S255692, 1T0S255725, 1T0S255749, 1T0S255751, 1T0S255752, 1T0S255753, 1T0S255756, 1T0S255758, 1T0S255760, 1T0S255761, 1T0S255762, 1T0S255763, 1T0S255764, 1T0S255765, 1T0S255766, 1T0S255769, 1T0S255770, 3T0S273956, 3T0S274101, 3T0S274121, 3T0S274124, 3T0S274131, 3T0S274133, 3T0S274135, 3T0S274136, 3T0S274139, 3T0S274140, 3T0S274141, 3T0S274148, 3T0S274154, 3T0S274164, 3T0S274448, 2T0S267684, 2T0S267690, 2T0S267700, 2T0S267701, 2T0S267738, 2T0S267746, 2T0S267752, 2T0S267756, 2T0S267765, 2T0S267770, 2T0S267776, 2T0S267795, 2T0S267798, 2T0S267799, 2T0S267800, 2T0S267802, 2T0S267807, 2T0S267814, 2T0S267817, 2T0S267819, 2T0S267820, 2T0S267822, 2T0S267831, 1T0S256816, 1T0S256826, 1T0S256855, 1T0S256857, 1T0S256874, 1T0S256878, 1T0S256885, 1T0S256890, 1T0S256894, 1T0S256903, 1T0S256904, 1T0S256906, 1T0S256907, 1T0S256915, 1T0S256920, 1T0S256928, 1T0S256932, 1T0S256944, 1T0S256960, 1T0S256962, 1T0S256979, 1T0S256981, 1T0S257001, 1T0S257312, 1T0S258148, 1T0S258154, 1T0S258156, 1T0S258166, 1T0S258171, 1T0S258172, 1T0S258179, 1T0S258186, 1T0S258187, 1T0S258196, 1T0S258204, 1T0S258212, 1T0S258214, 1T0S258219, 1T0S258237, 1T0S258240, 1T0S257105, 1T0S257109, 1T0S257192, 1T0S257205, 1T0S257209, 1T0S257795, 1T0S257798, 1T0S257826, 1T0S257835, 1T0S257836, 1T0S257837, 1T0S257867, 1T0S257868, 1T0S257875, 1T0S257882, 1T0S257891, 1T0S257896, 1T0S257925, 4T0S277111, 4T0S277120, 4T0S277349, 4T0S277369, 4T0S277388, 4T0S277433, 4T0S277437, 4T0S278022, 4T0S278023, 4T0S278024, 4T0S278025, 4T0S278026, 4T0S278027, 4T0S278032, 4T0S278033, 4T0S278036, 4T0S278038, 4T0S278041, 4T0S278042, 4T0S278043, 4T0S278045, 4T0S278046, 4T0S278049, 4T0S278052, 4T0S278053, 4T0S278054, 4T0S278058, 4T0S278060, 4T0S278061, 4T0S278063, 4T0S278066, 4T0S278082, 4T0S278139, 4T0S278141, 4T0S278142, 3T0S268714, 3T0S270482, 3T0S271150, 3T0S271167, 3T0S271191, 3T0S271192, 3T0S271195, 3T0S271623, 3T0S271685, 3T0S271690, 3T0S271693, 3T0S271695, 3T0S271698, 3T0S271718, 3T0S271736, 3T0S271746, 3T0S271753, 3T0S271769, 3T0S271778, 3T0S271780, 3T0S271797, 3T0S271813, 3T0S271888, 3T0S271894, 3T0S271912, 3T0S271914, 3T0S271922, 3T0S271967, 2T0S267133, 2T0S267135, 2T0S267136, 2T0S267158, 2T0S267168, 2T0S267179, 2T0S267184, 2T0S267186, 2T0S267189, 2T0S267195, 2T0S267199, 2T0S267203, 2T0S267206, 2T0S267215, 2T0S267217, 2T0S267218, 2T0S267219, 2T0S267224, 2T0S267226, 2T0S267229, 2T0S267232, 2T0S267233, 2T0S267245, 2T0S267249, 1T0S262038, 1T0S262192, 1T0S262205, 1T0S262235, 2T0S267134, 2T0S267138, 2T0S267152, 2T0S267180, 2T0S267181, 2T0S267191, 2T0S267205, 2T0S267209, 2T0S267213, 2T0S267220, 2T0S267221, 2T0S267236, 2T0S267237, 2T0S267242, 2T0S267243, 2T0S267251, 2T0S267262, 4T0S277416, 4T0S277509, 4T0S277527, 4T0S277570, 4T0S277605, 4T0S277611, 4T0S277617, 4T0S277621, 4T0S277625, 4T0S277628, 4T0S277634, 4T0S277636, 4T0S277638, 4T0S277642, 4T0S277660, 4T0S277661, 4T0S277662, 4T0S277664, 4T0S277675, 4T0S277676, 4T0S277679, 4T0S277689, 4T0S277690, 4T0S277705, 4T0S277709, 4T0S277720, 4T0S277738, 1T0S258222, 1T0S258225, 1T0S258226, 1T0S258235, 1T0S258238, 1T0S258244, 1T0S258246, 1T0S258248, 1T0S258254, 1T0S258255, 1T0S258256, 1T0S258258, 1T0S258259, 1T0S258265, 1T0S262128, 1T0S262136, 1T0S262138, 1T0S262139, 1T0S262143, 1T0S262144, 1T0S262145, 1T0S262146, 1T0S262147, 1T0S262148, 1T0S262149, 1T0S262150, 1T0S262151, 1T0S262152, 1T0S262156, 1T0S262158, 1T0S262159, 1T0S262160, 1T0S262162, 1T0S262163, 1T0S262165, 1T0S262166, 1T0S262167, 1T0S262169, 1T0S262172, 1T0S262174, 1T0S262176, 1T0S262177, 1T0S262179, 1T0S262180, 1T0S262182, 1T0S262184, 1T0S262185, 1T0S262187, 1T0S262188, 1T0S262190, 1T0S262191, 1T0S262198, 3T0S270091, 3T0S270431, 3T0S270435, 3T0S270444, 3T0S270598, 3T0S270599, 3T0S270600, 3T0S270603, 3T0S270604, 3T0S270607, 3T0S270620, 3T0S270622, 3T0S270624, 3T0S270626, 3T0S271104, 3T0S271113, 3T0S271117, 3T0S271125, 3T0S271126, 3T0S271135, 3T0S271136, 3T0S271140, 3T0S271158, 3T0S271159, 3T0S271590, 3T0S271591, 3T0S271592, 0T0S253465, 0T0S254021, 0T0S254026, 0T0S254057, 0T0S254063, 0T0S254068, 0T0S254074, 0T0S254078, 0T0S254083, 0T0S254105, 0T0S254495, 0T0S254503, 0T0S254510, 0T0S254512, 0T0S254514, 0T0S254529, 0T0S254533, 0T0S254536, 0T0S254541, 0T0S254542, 0T0S254543, 0T0S254551, 0T0S252249, 0T0S252302, 0T0S252347, 0T0S252358, 0T0S252359, 0T0S252373, 0T0S252385, 0T0S252388, 0T0S252582, 0T0S252589, 0T0S253177, 0T0S253183, 2T0S265463, 2T0S265484, 4T0S275673, 4T0S275706, 4T0S275713, 4T0S275730, 4T0S275735, 4T0S275737, 4T0S275750, 4T0S275763, 4T0S275764, 4T0S275770, 4T0S275774, 4T0S275776, 4T0S275783, 4T0S275788, 4T0S275789, 4T0S275791, 4T0S275793, 4T0S275794, 4T0S275797, 4T0S275798, 4T0S275802, 4T0S275807, 4T0S275813, 4T0S275827, 4T0S275828, 4T0S275829, 4T0S275832, 0T0S254632, 0T0S254641, 0T0S254644, 0T0S254645, 0T0S254651, 3T0S271631, 3T0S271931, 3T0S271936, 3T0S271942, 3T0S271948, 3T0S271950, 3T0S271957, 3T0S271969, 3T0S271971, 3T0S271972, 3T0S271978, 3T0S271979, 3T0S271986, 3T0S268759, 3T0S270488, 3T0S270606, 3T0S271193, 3T0S271196, 3T0S271204, 1T0S257320, 1T0S257389, 1T0S257394, 1T0S257396, 1T0S257407, 1T0S257416, 1T0S257417, 1T0S257420, 1T0S257788, 1T0S257790, 1T0S257801, 1T0S257805, 1T0S257807, 1T0S257808, 1T0S257811, 1T0S257819, 1T0S257840, 1T0S261280, 1T0S261291, 1T0S261300, 1T0S261332, 1T0S261347, 1T0S261357, 1T0S261362, 1T0S261370, 1T0S261376, 1T0S261383, 1T0S261384, 1T0S261388, 1T0S261392, 1T0S261394, 1T0S261396, 1T0S261398, 1T0S261401, 1T0S261412, 1T0S261427, 1T0S261441, 1T0S261635, 1T0S261644, 1T0S261660, 1T0S261675, 1T0S261287, 1T0S261329, 1T0S261352, 1T0S261354, 1T0S261368, 1T0S261379, 1T0S261419, 1T0S261440, 1T0S261449, 1T0S261460, 1T0S261483, 1T0S261499, 1T0S261513, 1T0S261518, 2T0S266910, 0T0S253176, 0T0S253191, 0T0S253193, 0T0S253209, 1T0S258815, 1T0S258828, 0T0S253195, 1T0S262395, 1T0S262402, 1T0S262412, 1T0S262422, 1T0S262437, 1T0S262443, 1T0S262521, 1T0S262564, 1T0S262600, 1T0S262614, 1T0S262624, 1T0S262629, 2T0S267114, 2T0S267120, 2T0S267122, 2T0S267123, 2T0S267125, 2T0S267144, 2T0S267146, 2T0S267155, 2T0S267161, 2T0S267163, 2T0S267165, 2T0S267175, 2T0S267201, 2T0S267250, 2T0S263017, 2T0S263020, 2T0S263021, 2T0S263024, 2T0S263043, 2T0S263048, 2T0S263050, 2T0S263052, 2T0S263055, 2T0S263058, 2T0S263067, 2T0S263072, 2T0S266585, 2T0S266606, 2T0S267022, 3T0S268789, 3T0S269666, 1T0S257186, 1T0S257189, 0T0S254107, 0T0S254112, 0T0S254117, 0T0S254126, 2T0S264386, 2T0S264398, 2T0S264449, 2T0S264540, 2T0S264551, 2T0S264552, 2T0S264563, 2T0S264576, 2T0S264578, 2T0S264917, 2T0S265059, 2T0S265298, 3T0S273952, 3T0S273976, 3T0S273978, 3T0S274901, 3T0S274904, 3T0S274917, 4T0S276426, 4T0S276522, 4T0S276637, 4T0S276668, 4T0S276805, 4T0S276806, 4T0S276810, 4T0S276811, 4T0S276817, 4T0S276822, 4T0S276824, 4T0S276827, 4T0S276830, 4T0S276841, 4T0S276857, 4T0S276956, 4T0S276980, 4T0S277023, 4T0S277047, 1T0S256499, 1T0S256502, 1T0S256503, 1T0S256504, 1T0S256508, 1T0S256509, 1T0S256517, 1T0S256522, 1T0S256544, 1T0S256545, 1T0S256552, 1T0S256553, 1T0S256555, 1T0S256559, 1T0S256560, 1T0S256566, 1T0S256567, 1T0S256580, 1T0S257005, 1T0S257012, 1T0S257013, 1T0S257014, 1T0S257024, 1T0S257045, 1T0S257055, 1T0S257059, 2T0S266518, 2T0S266541, 2T0S266559, 2T0S266569, 1T0S256448, 1T0S257336, 1T0S258162, 1T0S258514, 1T0S258595, 1T0S259009, 1T0S259434, 1T0S259710, 1T0S259727, 1T0S259747, 1T0S259809, 1T0S259837, 1T0S259857, 1T0S259859, 0T0S254654, 1T0S259350, 1T0S259351, 1T0S259358, 1T0S259359, 1T0S259361, 1T0S259366, 1T0S259379, 1T0S259385, 1T0S259386, 1T0S259394, 1T0S259396, 1T0S259399, 1T0S259404, 1T0S256512, 1T0S256514, 1T0S256525, 1T0S256526, 1T0S256528, 1T0S256540, 1T0S256549, 1T0S256558, 4T0S275608, 4T0S275609, 4T0S275614, 4T0S275616, 4T0S275617, 4T0S275618, 4T0S275620, 4T0S275624, 4T0S275626, 4T0S275635, 4T0S275636, 4T0S275638, 4T0S275640, 4T0S275646, 4T0S275670, 4T0S275671, 4T0S275684, 4T0S275689, 4T0S275696, 1T0S259744, 0T0S253414, 0T0S253565, 0T0S253612, 3T0S274021, 3T0S274068, 3T0S274085, 3T0S274086, 3T0S274091, 3T0S274107, 3T0S274113, 3T0S274145, 3T0S274817, 3T0S274833, 3T0S274838, 3T0S274848, 3T0S274863, 3T0S274872, 3T0S274882, 3T0S271160, 0T0S251713, 0T0S251725, 0T0S251738, 0T0S251745, 0T0S251747, 0T0S251750, 0T0S251755, 0T0S251756, 0T0S251762, 0T0S252626, 0T0S252643, 0T0S253357, 0T0S253371, 0T0S253377, 0T0S253399, 0T0S253402, 0T0S253409, 0T0S252206, 0T0S252228, 0T0S252242, 0T0S252251, 1T0S256610, 1T0S256615, 1T0S256622, 1T0S256641, 1T0S256645, 1T0S256650, 1T0S256652, 1T0S256654, 1T0S256656, 1T0S256660, 1T0S256663, 1T0S256664, 1T0S256667, 1T0S256669, 1T0S256670, 1T0S256672, 1T0S256674, 1T0S256676, 1T0S256677, 1T0S256678, 1T0S256679, 1T0S256682, 1T0S256683, 1T0S256685, 1T0S256687, 1T0S256688, 1T0S256690, 3T0S270625, 3T0S270633, 3T0S270641, 3T0S270645, 3T0S270647, 3T0S270649, 3T0S270654, 3T0S270657, 3T0S270659, 3T0S270661, 3T0S270664, 3T0S270666, 3T0S270668, 3T0S270669, 3T0S270672, 3T0S270675, 3T0S270677, 3T0S270678, 3T0S270680, 3T0S270681, 3T0S270683, 3T0S270689, 3T0S274018, 3T0S274034, 3T0S270679, 3T0S270706, 3T0S270708, 3T0S270720, 3T0S270724, 3T0S270728, 3T0S270730, 3T0S270732, 3T0S270737, 3T0S270738, 3T0S270739, 3T0S270743, 3T0S270744, 3T0S270746, 3T0S270747, 3T0S270749, 3T0S270752, 3T0S270756, 1T0S261058, 1T0S261062, 1T0S261086, 1T0S261089, 1T0S261097, 1T0S261103, 1T0S261115, 1T0S261121, 1T0S261125, 1T0S261133, 1T0S261143, 1T0S261165, 1T0S261166, 1T0S261170, 1T0S261187, 1T0S261189, 1T0S261193, 1T0S261937, 1T0S261968, 3T0S270406, 3T0S270439, 3T0S270484, 3T0S270490, 3T0S270500, 3T0S270596, 3T0S270601, 3T0S270608, 3T0S270610, 3T0S270614, 3T0S270619, 3T0S270623, 3T0S271697, 3T0S271735, 3T0S271748, 3T0S271757, 3T0S271790, 3T0S271817, 3T0S271819, 3T0S271847, 2T0S264277, 2T0S264307, 1T0S258790, 2T0S267253, 2T0S267261, 2T0S267270, 2T0S267287, 2T0S267292, 2T0S267299, 2T0S267301, 2T0S267304, 2T0S267308, 3T0S270602, 3T0S270629, 3T0S270630, 3T0S270631, 3T0S270635, 3T0S270646, 3T0S270651, 3T0S270696, 3T0S270698, 3T0S270707, 3T0S270715, 3T0S273993, 3T0S274025, 3T0S274035, 3T0S274126, 3T0S274933, 3T0S274935, 3T0S274942, 3T0S274959, 3T0S274974, 3T0S274986, 3T0S274992, 3T0S275014, 0T0S253459, 0T0S253462, 1T0S257046, 1T0S257060, 1T0S257068, 1T0S257128, 0T0S251657, 0T0S251670, 0T0S251683, 0T0S251687, 0T0S251690, 0T0S251696, 0T0S251699, 0T0S251703, 0T0S251704, 0T0S251706, 0T0S251709, 0T0S251712, 0T0S251714, 0T0S251718, 0T0S251719, 0T0S251730, 0T0S251731, 0T0S251732, 0T0S251733, 0T0S251743, 0T0S251751, 0T0S251754, 3T0S273996, 3T0S274024, 3T0S274026, 3T0S274030, 3T0S274032, 3T0S274036, 3T0S274037, 3T0S274045, 3T0S274061, 3T0S274062, 1T0S258201, 1T0S258245, 1T0S258415, 1T0S258446, 1T0S258458, 1T0S258460, 1T0S258463, 0T0S254511, 1T0S257112, 1T0S257335, 1T0S261127, 1T0S256858, 1T0S256938, 4T0S276519, 4T0S276579, 3T0S268688, 3T0S268707, 3T0S268711, 3T0S268734, 3T0S268748, 1T0S262415, 1T0S262493, 1T0S261454, 1T0S258464, 1T0S258587, 0T0S255205, 0T0S255230, 0T0S255240, 0T0S255244, 0T0S255247, 0T0S255250, 0T0S255254, 0T0S255271, 3T0S270420, 3T0S270474, 3T0S270477, 3T0S271713, 3T0S271742, 1T0S257839, 1T0S257860, 1T0S261925, 1T0S261927, 1T0S261481, 3T0S270755, 2T0S263877, 2T0S263880, 2T0S263881, 2T0S263889, 2T0S263909, 2T0S263991, 2T0S263993, 2T0S263994, 2T0S264003, 2T0S264004, 2T0S264010, 2T0S264011, 2T0S264014, 2T0S264015, 2T0S264016, 2T0S264018, 2T0S264030, 2T0S264033, 0T0S254517, 0T0S252663, 0T0S252666, 4T0S276839, 4T0S276868, 4T0S276878, 4T0S276879, 4T0S276882, 4T0S276919, 4T0S276923, 4T0S276941, 4T0S276951, 1T0S260049, 1T0S260053, 1T0S261073, 1T0S261074, 1T0S261099, 1T0S261112, 1T0S261134, 1T0S261161, 1T0S261168, 1T0S261197, 1T0S261422, 1T0S261458, 1T0S261841, 1T0S261922, 1T0S261933, 1T0S261941, 1T0S261951, 1T0S261956, 1T0S261958, 1T0S261962, 1T0S260060, 0T0S253967, 0T0S253972, 1T0S256913, 1T0S256943, 1T0S256983, 1T0S256998, 1T0S257003, 0T0S253307, 0T0S253331, 0T0S253366, 0T0S253367, 2T0S266487, 2T0S266496, 2T0S266497, 2T0S266515, 2T0S266524, 2T0S266533, 0T0S252203, 0T0S252335, 0T0S252632, 0T0S252635, 0T0S252700, 0T0S252702, 0T0S252710, 0T0S252711, 0T0S252715, 0T0S252719, 0T0S252724, 0T0S252726, 0T0S252727, 0T0S252730, 0T0S252732, 0T0S253207, 0T0S253213, 0T0S253224, 0T0S253225, 0T0S253228, 0T0S253229, 0T0S253241, 0T0S253242, 0T0S253243, 0T0S253246, 0T0S253247, 0T0S253259, 0T0S253270, 1T0S262553, 1T0S262567, 1T0S262595, 1T0S262598, 1T0S262604, 1T0S262610, 1T0S262613, 1T0S262616, 1T0S262619, 1T0S262628, 1T0S262630, 1T0S262631, 1T0S262634, 1T0S262638, 1T0S262640, 1T0S262642, 2T0S262750, 2T0S262853, 2T0S262865, 2T0S262879, 2T0S262880, 2T0S262895, 2T0S262899, 2T0S262904, 2T0S262908, 2T0S262919, 2T0S262923, 3T0S270409, 3T0S270418, 3T0S270611, 3T0S270643, 3T0S270644, 3T0S270655, 3T0S271635, 3T0S271710, 3T0S271743, 3T0S271750, 3T0S271752, 3T0S271764, 3T0S271768, 3T0S271771, 3T0S271803, 3T0S271859, 3T0S271874, 3T0S271877, 3T0S271878, 3T0S271905, 3T0S271926, 3T0S271958, 3T0S271959, 3T0S271961, 3T0S271962, 3T0S271963, 1T0S258436, 1T0S258450, 1T0S258457, 1T0S258485, 1T0S258492, 1T0S258497, 1T0S258498, 1T0S258504, 1T0S258506, 1T0S258508, 3T0S271171, 3T0S271577, 3T0S271579, 3T0S271580, 3T0S271582, 3T0S271583, 3T0S271585, 3T0S271586, 3T0S271587, 3T0S271595, 3T0S271597, 3T0S271602, 3T0S271606, 3T0S271607, 3T0S271609, 3T0S271610, 3T0S271614, 3T0S271616, 3T0S271625, 3T0S271633, 3T0S271640, 3T0S271642, 3T0S271721, 3T0S271777, 3T0S271802, 3T0S271809, 3T0S271812, 3T0S271816, 3T0S271834, 0T0S255102, 2T0S263053, 2T0S263056, 2T0S263062, 2T0S263069, 2T0S263071, 2T0S263073, 2T0S263074, 2T0S263085, 2T0S263087, 2T0S263089, 2T0S263090, 2T0S263091, 2T0S263093, 2T0S263094, 2T0S263096, 2T0S263097, 2T0S263098, 2T0S263100, 2T0S263101, 2T0S263104, 2T0S263106, 2T0S263107, 2T0S263111, 2T0S263114, 2T0S263119, 2T0S263121, 2T0S263123, 2T0S263127, 2T0S263128, 2T0S263129, 2T0S263132, 2T0S263133, 2T0S263135, 2T0S263137, 2T0S263139, 2T0S263144, 2T0S264179, 2T0S264271, 2T0S264291, 2T0S264368, 2T0S264388, 2T0S264399, 2T0S264400, 2T0S264581, 2T0S264687, 2T0S264691, 2T0S264692, 2T0S264712, 2T0S264717, 2T0S264736, 2T0S264738, 2T0S264759, 2T0S264762, 2T0S264864, 2T0S264908, 2T0S264921, 2T0S264966, 2T0S265052, 2T0S265058, 2T0S265401, 2T0S265402, 2T0S265416, 2T0S265419, 2T0S265426, 2T0S265483, 2T0S265486, 2T0S265500, 2T0S265529, 2T0S265532, 2T0S265542, 2T0S265562, 2T0S265578, 2T0S265588, 2T0S265602, 2T0S265605, 2T0S265608, 2T0S265621, 2T0S266062, 2T0S266130, 2T0S266138, 2T0S266139, 2T0S266154, 2T0S266157, 2T0S266159, 2T0S266169, 2T0S266180, 2T0S266183, 2T0S266184, 2T0S266185, 2T0S266194, 2T0S266961, 2T0S267097, 2T0S267107, 2T0S267145, 2T0S267147, 2T0S267148, 2T0S267150, 2T0S267151, 2T0S267153, 2T0S267159, 2T0S267160, 2T0S267162, 2T0S267164, 2T0S267173, 2T0S267174, 2T0S267177, 2T0S267178, 2T0S267182, 2T0S267183, 2T0S267187, 2T0S267190, 2T0S267192, 2T0S267193, 2T0S267198, 2T0S267200, 2T0S267202, 2T0S267210, 2T0S267216, 2T0S268022, 2T0S268028, 2T0S268037, 2T0S268043, 2T0S268049, 2T0S268050, 2T0S268051, 2T0S268054, 2T0S268055, 2T0S268058, 2T0S268060, 2T0S268067, 2T0S268070, 2T0S268075, 2T0S268078, 2T0S268079, 2T0S268083, 2T0S268087, 2T0S268094, 2T0S268095, 2T0S268108, 2T0S268109, 2T0S268115, 2T0S268116, 2T0S268119, 3T0S268767, 3T0S268779, 3T0S268781, 3T0S268791, 3T0S269668, 3T0S269672, 3T0S269684, 3T0S269698, 3T0S269711, 3T0S269720, 3T0S269721, 3T0S269726, 3T0S269729, 3T0S269731, 3T0S269745, 3T0S269751, 3T0S269773, 3T0S269784, 3T0S269785, 3T0S269805, 3T0S269809, 3T0S269813, 0T0S255027, 0T0S255029, 0T0S255034, 0T0S255035, 0T0S255038, 0T0S255039, 0T0S255041, 0T0S255044, 0T0S255049, 0T0S255051, 0T0S255060, 0T0S255070, 0T0S255135, 0T0S255155, 0T0S255158, 0T0S255159, 0T0S255163, 1T0S256383, 1T0S256413, 1T0S256414, 1T0S256449, 1T0S256453, 1T0S256479, 1T0S256480, 1T0S256483, 1T0S256815, 1T0S256861, 1T0S256880, 1T0S256888, 1T0S256895, 1T0S257207, 0T0S252684, 0T0S252728, 0T0S252733, 0T0S252735, 0T0S252736, 0T0S252741, 0T0S252745, 0T0S252747, 0T0S252755, 0T0S252756, 0T0S252759, 0T0S252762, 0T0S252763, 0T0S252767, 0T0S252768, 0T0S252769, 0T0S252770, 0T0S252771, 0T0S252772, 0T0S252773, 0T0S252776, 0T0S252778, 0T0S252782, 0T0S252783, 0T0S252784, 0T0S252785, 0T0S253374, 0T0S253505, 0T0S255432, 0T0S255643, 1T0S255784, 1T0S255801, 2T0S267758, 1T0S261692, 0T0S252600, 0T0S252605, 0T0S252614, 0T0S252615, 0T0S252617, 0T0S252621, 0T0S252628, 0T0S252637, 0T0S252676, 0T0S252708, 2T0S266885, 2T0S266891, 2T0S266896, 2T0S266897, 2T0S266900, 2T0S266905, 2T0S266911, 2T0S266917, 2T0S266930, 2T0S266933, 2T0S266934, 2T0S266936, 2T0S266940, 2T0S266943, 2T0S266944, 2T0S266946, 2T0S266948, 2T0S266949, 2T0S266973, 2T0S266974, 2T0S266976, 2T0S266983, 1T0S261385, 3T0S271106, 0T0S254582, 0T0S254613, 2T0S267048, 2T0S267076, 2T0S266864, 2T0S266866, 2T0S266875, 2T0S266892, 2T0S266894, 2T0S266907, 2T0S266914, 2T0S266919, 2T0S266951, 0T0S253174, 0T0S251948, 0T0S251953, 0T0S251960, 0T0S251962, 0T0S251963, 0T0S251965, 0T0S251968, 0T0S251969, 0T0S251970, 0T0S251974, 0T0S251975, 0T0S251976, 0T0S251977, 0T0S251978, 0T0S251979, 0T0S251980, 0T0S251982, 0T0S251984, 0T0S251986, 0T0S251987, 0T0S251990, 0T0S251991, 0T0S251992, 0T0S251993, 0T0S251994, 0T0S251996, 3T0S270748, 3T0S270764, 3T0S270779, 1T0S256852, 1T0S256856, 1T0S256860, 1T0S256872, 1T0S256923, 1T0S256947, 0T0S252652, 0T0S252677, 1T0S256353, 1T0S256392, 1T0S257367, 1T0S257850, 1T0S257881, 1T0S258133, 1T0S258526, 1T0S258569, 1T0S258570, 1T0S258976, 1T0S259011, 1T0S259174, 1T0S259175, 1T0S259183, 1T0S259189, 1T0S259196, 0T0S253185, 0T0S253269, 0T0S253279, 0T0S253280, 0T0S253283, 0T0S253290, 0T0S253297, 0T0S253239, 0T0S253340, 0T0S253346, 0T0S253352, 0T0S254544, 1T0S257388, 1T0S257843, 1T0S257890, 1T0S257904, 1T0S257939, 1T0S257948, 1T0S257953, 1T0S258125, 1T0S258131, 1T0S258132, 1T0S258136, 1T0S258138, 1T0S258439, 1T0S258470, 1T0S258483, 1T0S258493, 1T0S258496, 1T0S258520, 1T0S258524, 1T0S258525, 1T0S258531, 1T0S258534, 1T0S258542, 1T0S258560, 1T0S258563, 1T0S258564, 1T0S258567, 1T0S258568, 1T0S259420, 1T0S259487, 1T0S259494, 1T0S259530, 1T0S259537, 1T0S259541, 1T0S259542, 1T0S259549, 1T0S259551, 1T0S259553, 1T0S259556, 1T0S259557, 1T0S259566, 1T0S259567, 1T0S259572, 1T0S259573, 1T0S259574, 1T0S259576, 1T0S259577, 1T0S259578, 1T0S259580, 1T0S259583, 1T0S260147, 1T0S261055, 1T0S261085, 1T0S261106, 1T0S261108, 1T0S261120, 0T0S255175, 0T0S255180, 0T0S255198, 0T0S255199, 0T0S255200, 0T0S255202, 0T0S255204, 0T0S255222, 0T0S255233, 0T0S255234, 0T0S255243, 0T0S255248, 0T0S255251, 0T0S255253, 0T0S255257, 0T0S255261, 1T0S262591, 1T0S262593, 1T0S262601, 2T0S262900, 2T0S262914, 2T0S262918, 2T0S263015, 2T0S263016, 2T0S263022, 2T0S263023, 2T0S263025, 2T0S263026, 2T0S263028, 2T0S263029, 2T0S263030, 2T0S263033, 2T0S263034, 2T0S263035, 2T0S263037, 2T0S263038, 2T0S263039, 2T0S263041, 2T0S263042, 2T0S263045, 2T0S263046, 2T0S263049, 2T0S263051, 2T0S263059, 2T0S263061, 2T0S263063, 2T0S263068, 2T0S263070, 2T0S263076, 2T0S263077, 2T0S263078, 2T0S263079, 2T0S263080, 2T0S263081, 2T0S263083, 2T0S263084, 2T0S263086, 2T0S263088, 0T0S250065, 0T0S250066, 0T0S250068, 0T0S250073, 0T0S250074, 0T0S253941, 0T0S253958, 0T0S253966, 0T0S253971, 0T0S253993, 0T0S253995, 0T0S253996, 0T0S253997, 0T0S254002, 0T0S254004, 0T0S254007, 0T0S254008, 0T0S254011, 0T0S254015, 0T0S254017, 0T0S254019, 0T0S254022, 0T0S254024, 0T0S254025, 0T0S254028, 0T0S254029, 0T0S254032, 0T0S254037, 0T0S254038, 0T0S254045, 0T0S254048, 0T0S254051, 0T0S254065, 0T0S255345, 0T0S255356, 0T0S255357, 0T0S255359, 0T0S255362, 0T0S255364, 0T0S255367, 0T0S255370, 0T0S255371, 0T0S255376, 0T0S255377, 0T0S255385, 3T0S274966, 0T0S255686, 0T0S255691, 0T0S255700, 0T0S255701, 0T0S255705, 0T0S255707, 0T0S255708, 0T0S255710, 1T0S255716, 1T0S255720, 1T0S255733, 1T0S255734, 1T0S255736, 1T0S255739, 1T0S255781, 0T0S253442, 0T0S253516, 0T0S255130, 0T0S255166, 0T0S255167, 0T0S255172, 0T0S255176, 0T0S255182, 0T0S255183, 0T0S255188, 0T0S255192, 0T0S255193, 0T0S255196, 0T0S255210, 0T0S255211, 0T0S255217, 0T0S255220, 0T0S255223, 0T0S255224, 0T0S255229, 0T0S255235, 0T0S255237, 0T0S255241, 0T0S255245, 0T0S255252, 0T0S255256, 0T0S255266, 0T0S250473, 0T0S250476, 0T0S250477, 0T0S251940, 0T0S251944, 0T0S251945, 0T0S251947, 0T0S250475, 0T0S251942, 0T0S251943, 0T0S251946, 0T0S251952, 0T0S251954, 0T0S251959, 0T0S251961, 0T0S251964, 0T0S251966, 0T0S251967, 0T0S253620, 0T0S253633, 0T0S253640, 0T0S253644, 0T0S254521, 0T0S254523, 0T0S254525, 0T0S254527, 0T0S254531, 0T0S254534, 0T0S254537, 0T0S254545, 0T0S252640, 0T0S252675, 0T0S252714, 0T0S252761, 0T0S252764, 0T0S252779, 0T0S252780, 0T0S252787, 0T0S252790, 0T0S253167, 0T0S253170, 0T0S253171, 0T0S253173, 0T0S253175, 0T0S253189, 0T0S252597, 0T0S252599, 0T0S252611, 0T0S252636, 0T0S252642, 0T0S252650, 0T0S252658, 0T0S252660, 0T0S252661, 0T0S252662, 0T0S252665, 0T0S252671, 0T0S252687, 0T0S252697, 0T0S252699, 1T0S256539, 1T0S256557, 1T0S256563, 0T0S253168, 0T0S253181, 0T0S253198, 0T0S253199, 0T0S253200, 0T0S253210, 0T0S253214, 0T0S253238, 0T0S253524, 0T0S253543, 0T0S253560, 0T0S253568, 0T0S253575, 0T0S253580, 0T0S253585, 0T0S253618, 0T0S253623, 0T0S253628, 0T0S253629, 0T0S253635, 0T0S252226, 0T0S252237, 0T0S252238, 0T0S252244, 0T0S252246, 0T0S252248, 0T0S252254, 0T0S252258, 0T0S252261, 0T0S252271, 0T0S252272, 0T0S252276, 0T0S252293, 0T0S252294, 3T0S268651, 3T0S268674, 3T0S268678, 3T0S268679, 3T0S268680, 3T0S268682, 0T0S251955, 0T0S251983, 0T0S252000, 0T0S252016, 0T0S252041, 0T0S252042, 0T0S252044, 0T0S252048, 0T0S252194, 0T0S252195, 0T0S252196, 0T0S252197, 0T0S252199, 0T0S252201, 2T0S266556, 2T0S266630, 2T0S266631, 0T0S253318, 0T0S253432, 0T0S253436, 0T0S253438, 0T0S253490, 0T0S253610, 0T0S253624, 0T0S253634, 0T0S253648, 0T0S253943, 0T0S253953, 0T0S253955, 0T0S253960, 0T0S253964, 0T0S254124, 0T0S254125, 0T0S254130, 0T0S254135, 0T0S254139, 0T0S254141, 0T0S254143, 0T0S254147, 0T0S254149, 0T0S254150, 0T0S254155, 0T0S254157, 0T0S254162, 0T0S254164, 0T0S252586, 0T0S252603, 0T0S252234, 0T0S252307, 0T0S252308, 0T0S252322, 0T0S252351, 0T0S252355, 0T0S252357, 0T0S252360, 0T0S252364, 0T0S252365, 0T0S252367, 0T0S252370, 0T0S252372, 0T0S252374, 0T0S252375, 0T0S252376, 0T0S252377, 0T0S252378, 0T0S252379, 0T0S252380, 0T0S252381, 0T0S252382, 0T0S252383, 0T0S252386, 0T0S252387, 0T0S252390, 0T0S252391, 1T0S257202, 1T0S257211, 1T0S257235, 0T0S255031, 1T0S259627, 1T0S259690, 2T0S266604, 2T0S266629, 0T0S252739, 0T0S252748, 0T0S252750, 0T0S252754, 0T0S252777, 0T0S252781, 0T0S252786, 0T0S252788, 0T0S252789, 0T0S252791, 0T0S253339, 0T0S253348, 0T0S253355, 0T0S253373, 0T0S253382, 0T0S253389, 0T0S253390, 0T0S253392, 0T0S253396, 0T0S253397, 0T0S253398, 0T0S253406, 0T0S253407, 0T0S252638, 0T0S253660, 0T0S253665, 0T0S254127, 0T0S254129, 0T0S254132, 0T0S254138, 0T0S254140, 0T0S254144, 0T0S254151, 0T0S254152, 0T0S254156, 0T0S254497, 0T0S254507, 0T0S254518, 0T0S251744, 0T0S251760, 0T0S252033, 0T0S252192, 0T0S252198, 0T0S252200, 0T0S252204, 0T0S252207, 0T0S252208, 0T0S252210, 0T0S252212, 0T0S252214, 0T0S252216, 0T0S252221, 0T0S252223, 0T0S252225, 0T0S252227, 0T0S252230, 0T0S252231, 0T0S252232, 0T0S252233, 0T0S252235, 0T0S252240, 0T0S252243, 0T0S252247, 0T0S252250, 0T0S252252, 0T0S252253, 0T0S252255, 0T0S252256, 0T0S252263, 0T0S252264, 0T0S252265, 0T0S252267, 0T0S252268, 0T0S252269, 0T0S252273, 0T0S252274, 0T0S252275, 0T0S252279, 0T0S252281, 0T0S252282, 0T0S252284, 0T0S252286, 0T0S252287, 0T0S252288, 0T0S252289, 0T0S252295, 0T0S252298, 0T0S252300, 0T0S252301, 0T0S252305, 0T0S252309, 0T0S252310, 0T0S252311, 0T0S252312, 0T0S252313, 0T0S252318, 0T0S252323, 0T0S252325, 0T0S252328, 0T0S252331, 0T0S252334, 0T0S252339, 0T0S252340, 0T0S252341, 0T0S252344, 0T0S252345, 0T0S252348, 0T0S252350, 0T0S252352, 0T0S252354, 0T0S252362, 0T0S252363, 0T0S252366, 0T0S252368, 0T0S252371, 1T0S258119, 1T0S258122, 1T0S258126, 1T0S258127, 1T0S258130, 1T0S258135, 1T0S258137, 1T0S258141, 1T0S258142, 1T0S258143, 1T0S258144, 1T0S258145, 1T0S258146, 1T0S258149, 1T0S258150, 1T0S258153, 1T0S258155, 1T0S258157, 1T0S258158, 1T0S258159, 1T0S258163, 1T0S258164, 1T0S258165, 1T0S258170, 1T0S258175, 1T0S258177, 1T0S258517, 1T0S258535, 0T0S253617, 0T0S253621, 0T0S254076, 0T0S254087, 0T0S254088, 0T0S254090, 0T0S254091, 0T0S254093, 0T0S254094, 0T0S254096, 0T0S254098, 0T0S254099, 0T0S254101, 0T0S254102, 0T0S254104, 0T0S254108, 0T0S254109, 0T0S254110, 0T0S254111, 0T0S254113, 0T0S254114, 0T0S254115, 0T0S254116, 0T0S254118, 0T0S254120, 0T0S254121, 1T0S255715, 1T0S255724, 1T0S255726, 1T0S255735, 1T0S255737, 1T0S255738, 1T0S255740, 1T0S255744, 1T0S255745, 1T0S255747, 1T0S255779, 1T0S255788, 1T0S255791, 1T0S255792, 1T0S255793, 0T0S253942, 0T0S255028, 0T0S255360, 0T0S255685, 0T0S255697, 1T0S257142, 1T0S257263, 1T0S257266, 1T0S257289, 1T0S257296, 1T0S257298, 1T0S256450, 1T0S256455, 1T0S256458, 1T0S256466, 1T0S256469, 1T0S256471, 1T0S256473, 1T0S256542, 1T0S256548, 1T0S256550, 1T0S256554, 1T0S256561, 1T0S256569, 1T0S256574, 1T0S256578, 1T0S256581, 1T0S256607, 1T0S256608, 1T0S256609, 1T0S256614, 1T0S256616, 1T0S256618, 1T0S256628, 1T0S256637, 1T0S256647, 1T0S256651, 1T0S256709, 1T0S256793, 1T0S256809, 0T0S253254, 0T0S253284, 0T0S253315, 0T0S253323, 0T0S253324, 0T0S253333, 0T0S253334, 0T0S253341, 0T0S253342, 0T0S253343, 0T0S253350, 0T0S253351, 0T0S253353, 0T0S253360, 0T0S253362, 0T0S253363, 0T0S253365, 0T0S253386, 0T0S253401, 0T0S253404, 0T0S253410, 0T0S253417, 0T0S253420, 0T0S253421, 0T0S253422, 0T0S253425, 0T0S253437, 0T0S253439, 0T0S254577, 0T0S254597, 0T0S255275, 0T0S255289, 0T0S255298, 0T0S255303, 0T0S255309, 0T0S255312, 0T0S255315, 0T0S255317, 0T0S255320, 0T0S255323, 0T0S255325, 0T0S255333, 0T0S255334, 0T0S255337, 0T0S255347, 0T0S255348, 0T0S255352, 0T0S252224, 0T0S252241, 0T0S252245, 0T0S252257, 0T0S252260, 0T0S252266, 0T0S252270, 0T0S252277, 0T0S252283, 0T0S252290, 0T0S252291, 0T0S252292, 0T0S252296, 0T0S252297, 0T0S252299, 0T0S252303, 0T0S252306, 0T0S252315, 0T0S252317, 0T0S255068, 0T0S255085, 0T0S255092, 0T0S255095, 0T0S255100, 0T0S255101, 0T0S255106, 0T0S255107, 0T0S255110, 0T0S255112, 0T0S255116, 0T0S255123, 0T0S255137, 0T0S255144, 0T0S255154, 3T0S271668, 3T0S271677, 3T0S271782, 1T0S259598, 1T0S259645, 1T0S259670, 1T0S259674, 1T0S259678, 1T0S259679, 1T0S259682, 1T0S259684, 1T0S259686, 1T0S259687, 1T0S259688, 1T0S259689, 1T0S259691, 1T0S259693, 1T0S259694, 1T0S259701, 1T0S259702, 1T0S259704, 1T0S259709, 1T0S259711, 1T0S259719, 1T0S259728, 1T0S259729, 1T0S259816, 1T0S259820, 1T0S259822, 1T0S259834, 3T0S271619, 3T0S271647, 3T0S271650, 3T0S271651, 3T0S271652, 3T0S271658, 3T0S271670, 3T0S271671, 3T0S271675, 3T0S271681, 3T0S271683, 3T0S271684, 3T0S271687, 3T0S271688, 3T0S271691, 3T0S271694, 3T0S271699, 3T0S271703, 3T0S271706, 3T0S271708, 3T0S271711, 3T0S271715, 3T0S271716, 3T0S271719, 3T0S271729, 3T0S271732, 3T0S271734, 3T0S271745, 3T0S271776, 0T0S253369, 0T0S253434, 0T0S253937, 0T0S253963, 0T0S253981, 0T0S253982, 0T0S253983, 0T0S253984, 0T0S253989, 0T0S253999, 0T0S254012, 0T0S254013, 0T0S254020, 0T0S255644, 0T0S255647, 0T0S255658, 0T0S255665, 0T0S255666, 0T0S255671, 0T0S255672, 0T0S255673, 0T0S255675, 0T0S255676, 0T0S255678, 0T0S255679, 0T0S255680, 0T0S255681, 0T0S255688, 0T0S255689, 0T0S255694, 0T0S255711, 1T0S255727, 1T0S255754, 1T0S255757, 1T0S255767, 0T0S254043, 1T0S256271, 1T0S256296, 1T0S256298, 1T0S256318, 1T0S256319, 1T0S256325, 1T0S256330, 1T0S256520, 1T0S262334, 1T0S262341, 1T0S262432, 1T0S262453, 1T0S256322, 1T0S256324, 1T0S256328, 1T0S256336, 1T0S256344, 1T0S256347, 1T0S256351, 1T0S256359, 1T0S256478, 1T0S256488, 1T0S256497, 1T0S256501, 1T0S256505, 1T0S256506, 1T0S256507, 1T0S256511, 1T0S256515, 1T0S256518, 1T0S256519, 1T0S256521, 1T0S256523, 1T0S256524, 1T0S256527, 1T0S256531, 1T0S256536, 1T0S256537, 1T0S261830, 1T0S261845, 1T0S261849, 1T0S261853, 1T0S261854, 1T0S261861, 1T0S261862, 1T0S261866, 1T0S261867, 1T0S261868, 1T0S261869, 1T0S261871, 1T0S261872, 1T0S261875, 1T0S261876, 1T0S261877, 1T0S261878, 1T0S261879, 1T0S261881, 1T0S261884, 1T0S261885, 1T0S261886, 1T0S261887, 1T0S261888, 1T0S261889, 1T0S261890, 1T0S261891, 1T0S261893, 1T0S261895, 1T0S261896, 1T0S261899, 1T0S261900, 1T0S261901, 1T0S261902, 1T0S261903, 1T0S261905, 1T0S261906, 1T0S261911, 1T0S261913, 1T0S261914, 1T0S261915, 1T0S261916, 1T0S261918, 1T0S261928, 1T0S261929, 1T0S261964, 1T0S261971, 1T0S261973, 1T0S261975, 1T0S261979, 1T0S261980, 1T0S261983, 1T0S261984, 1T0S261988, 1T0S261994, 1T0S261998, 1T0S262000, 1T0S262056, 1T0S262057, 1T0S262066, 1T0S262068, 1T0S262069, 1T0S262070, 1T0S262073, 1T0S262075, 1T0S262079, 1T0S262082, 1T0S262083, 1T0S262086, 1T0S262087, 1T0S262091, 1T0S262092, 1T0S262098, 1T0S256993, 1T0S256996, 1T0S257160, 1T0S257238, 1T0S257260, 1T0S257268, 1T0S257274, 1T0S257277, 1T0S257284, 1T0S257292, 1T0S257311, 1T0S257330, 1T0S257332, 1T0S257339, 1T0S257363, 2T0S266543, 2T0S266547, 2T0S266549, 2T0S266551, 2T0S266552, 2T0S266563, 0T0S251764, 0T0S251765, 0T0S251766, 0T0S251771, 0T0S253226, 0T0S253250, 0T0S253251, 0T0S253260, 0T0S253265, 0T0S253285, 0T0S253288, 0T0S253294, 0T0S253295, 0T0S254496, 0T0S254498, 0T0S254502, 0T0S254506, 0T0S255122, 0T0S255262, 1T0S256459, 1T0S256496, 1T0S256535, 1T0S256546, 1T0S256692, 1T0S256703, 1T0S256727, 1T0S256730, 1T0S256752, 1T0S257042, 1T0S257048, 1T0S257064, 1T0S257088, 1T0S257124, 1T0S257146, 4T0S276408, 4T0S276427, 4T0S276434, 4T0S276439, 4T0S276447, 4T0S276457, 4T0S276463, 4T0S276467, 4T0S276474, 4T0S276479, 4T0S276483, 4T0S276488, 4T0S276495, 4T0S276498, 4T0S276504, 4T0S276506, 4T0S276513, 4T0S276517, 4T0S276531, 4T0S276533, 4T0S276537, 4T0S276541, 4T0S276544, 4T0S276545, 4T0S276546, 4T0S276547, 4T0S276555, 4T0S276587, 1T0S256465, 1T0S256472, 1T0S256475, 1T0S256485, 1T0S256486, 1T0S256487, 1T0S256489, 1T0S256491, 1T0S256492, 1T0S256498, 1T0S256510, 1T0S256513, 1T0S256516, 1T0S256530, 1T0S256533, 2T0S265051, 2T0S266182, 3T0S268757, 1T0S257148, 1T0S257220, 1T0S257791, 1T0S257827, 1T0S257834, 1T0S257841, 1T0S257845, 1T0S257851, 1T0S257852, 1T0S257855, 1T0S257856, 1T0S257862, 1T0S257863, 1T0S257864, 1T0S257866, 1T0S257869, 1T0S257870, 1T0S257872, 1T0S257884, 2T0S266855, 2T0S266856, 2T0S266870, 2T0S266871, 0T0S255414, 0T0S255645, 0T0S255649, 0T0S255650, 0T0S255653, 0T0S255656, 0T0S255660, 0T0S255661, 0T0S255664, 0T0S255670, 0T0S255677, 0T0S255687, 0T0S255696, 0T0S255698, 0T0S255699, 0T0S255703, 0T0S255704, 0T0S255706, 0T0S255709, 1T0S255719, 1T0S255786, 0T0S253657, 0T0S255394, 0T0S255410, 0T0S255413, 0T0S255430, 0T0S255434, 0T0S255441, 0T0S255651, 0T0S255652, 0T0S255654, 0T0S255655, 0T0S255663, 0T0S255668, 0T0S255682, 0T0S255693, 1T0S255713, 1T0S257066, 1T0S257080, 1T0S257169, 1T0S257179, 1T0S257350, 1T0S257408, 1T0S257415, 1T0S257419, 1T0S256142, 1T0S256146, 1T0S256148, 1T0S256149, 1T0S256152, 1T0S256153, 1T0S256155, 1T0S256156, 1T0S256161, 1T0S256172, 1T0S256173, 1T0S256174, 1T0S256317, 1T0S256334, 1T0S256340, 1T0S256270, 1T0S256274, 1T0S256297, 1T0S256301, 1T0S256302, 1T0S256304, 1T0S256308, 1T0S256310, 1T0S256311, 1T0S256313, 1T0S257117, 1T0S257147, 1T0S257224, 1T0S257227, 1T0S257232, 1T0S257239, 1T0S257243, 1T0S257245, 1T0S257247, 1T0S257248, 1T0S257250, 1T0S257262, 1T0S257279, 0T0S253479, 0T0S253591, 0T0S253663, 0T0S255119, 0T0S255129, 0T0S255146, 0T0S255151, 0T0S255153, 0T0S255160, 0T0S255373, 0T0S255388, 0T0S255393, 0T0S255397, 0T0S255398, 0T0S255405, 0T0S255411, 0T0S255417, 0T0S255418, 0T0S255421, 0T0S255433, 0T0S255439, 0T0S255440, 1T0S257093, 1T0S257123, 1T0S257158, 1T0S257159, 1T0S257170, 1T0S257230, 1T0S257233, 1T0S257240, 1T0S257253, 1T0S257257, 1T0S257273, 1T0S257276, 1T0S257278, 1T0S257282, 1T0S257299, 1T0S257300, 1T0S257304, 3T0S268683, 1T0S256272, 1T0S256293, 1T0S256307, 1T0S256329, 1T0S256331, 1T0S256338, 1T0S256341, 1T0S256343, 1T0S256345, 1T0S256348, 1T0S256349, 1T0S256350, 1T0S256352, 1T0S256354, 1T0S256355, 1T0S256356, 1T0S256357, 1T0S256358, 1T0S256360, 1T0S256361, 1T0S256362, 1T0S256363, 1T0S256364, 1T0S256365, 1T0S256366, 1T0S256367, 1T0S256368, 1T0S256369, 1T0S256370, 1T0S256372, 1T0S256373, 1T0S256374, 1T0S256375, 1T0S256376, 1T0S256379, 1T0S256384, 1T0S256385, 1T0S256386, 1T0S256387, 1T0S256389, 1T0S256500, 3T0S270093, 3T0S270404, 3T0S270433, 3T0S270475, 3T0S270483, 2T0S267143, 2T0S267244, 2T0S267246, 2T0S267247, 2T0S267248, 2T0S267260, 2T0S267266, 2T0S267271, 2T0S267273, 2T0S267276, 2T0S267277, 2T0S267278, 2T0S267280, 2T0S267284, 2T0S267289, 2T0S267294, 2T0S267296, 2T0S267297, 2T0S267298, 2T0S267307, 2T0S268646, 2T0S268647, 2T0S268649, 1T0S256418, 1T0S256443, 1T0S256851, 1T0S256853, 1T0S256863, 1T0S256868, 1T0S256870, 1T0S256879, 1T0S256883, 1T0S256889, 1T0S256892, 1T0S256893, 1T0S256897, 1T0S256905, 1T0S256909, 1T0S256910, 1T0S256941, 1T0S256949, 1T0S256976, 0T0S250070, 0T0S253451, 0T0S253464, 0T0S253552, 0T0S253590, 0T0S253650, 0T0S253653, 0T0S253655, 0T0S253667, 0T0S253668, 0T0S253669, 0T0S253674, 0T0S254103, 0T0S254106, 0T0S254137, 0T0S254226, 0T0S255331, 0T0S255353, 0T0S255355, 0T0S255361, 0T0S255366, 0T0S255368, 0T0S255374, 0T0S255375, 0T0S255378, 0T0S255379, 0T0S255380, 0T0S255381, 0T0S255383, 0T0S255384, 4T0S277347, 4T0S277357, 4T0S277363, 4T0S277370, 4T0S277383, 4T0S277390, 4T0S277403, 4T0S277404, 4T0S277413, 4T0S277426, 0T0S253456, 0T0S253457, 0T0S253511, 0T0S253513, 0T0S253551, 0T0S253555, 0T0S253587, 0T0S253589, 0T0S253652, 0T0S253654, 0T0S255369, 0T0S255390, 0T0S255401, 0T0S255416, 0T0S253460, 0T0S253658, 0T0S253677, 0T0S255139, 0T0S255389, 0T0S255402, 0T0S255403, 0T0S255404, 0T0S255408, 0T0S255415, 0T0S255419, 0T0S255420, 0T0S255422, 0T0S255423, 0T0S255424, 0T0S255425, 0T0S255429, 0T0S255436, 0T0S255437, 0T0S255443, 0T0S253549, 0T0S253550, 0T0S253651, 0T0S253656, 0T0S253659, 0T0S253666, 0T0S253671, 0T0S253672, 0T0S253673, 0T0S253676, 0T0S255327, 0T0S255336, 0T0S255340, 0T0S255365, 0T0S255372, 0T0S255387, 0T0S255392, 0T0S255400, 0T0S255409, 1T0S259370, 1T0S259371, 1T0S259383, 1T0S259391, 1T0S259409, 1T0S259347, 1T0S259458, 1T0S259479, 1T0S259481, 1T0S259483, 1T0S259497, 1T0S259499, 1T0S259500, 1T0S259507, 1T0S259508, 1T0S259512, 1T0S259514, 1T0S259515, 1T0S259519, 2T0S263406, 2T0S263420, 2T0S263432, 2T0S263439, 2T0S263453, 2T0S263483, 2T0S263802, 2T0S263810, 2T0S263815, 2T0S263820, 2T0S263849, 2T0S263856, 1T0S256433, 1T0S256884, 1T0S256896, 1T0S256902, 1T0S256912, 1T0S256963, 1T0S256986, 1T0S256991, 1T0S256997, 1T0S257002, 1T0S257280, 1T0S257283, 1T0S257305, 1T0S257308, 1T0S257313, 1T0S257329, 1T0S257337, 1T0S257357, 1T0S259763, 1T0S259771, 1T0S259772, 1T0S259775, 1T0S259776, 1T0S259785, 1T0S259786, 1T0S259787, 1T0S259798, 1T0S259799, 1T0S259805, 1T0S259808, 1T0S259814, 1T0S259817, 1T0S259823, 1T0S259825, 1T0S259826, 1T0S259831, 1T0S259749, 1T0S259751, 1T0S259759, 1T0S259773, 1T0S259777, 1T0S259783, 1T0S259789, 1T0S259790, 1T0S259795, 1T0S259692, 1T0S259718, 1T0S259739, 1T0S259745, 1T0S259750, 1T0S259755, 1T0S259758, 1T0S259760, 1T0S259761, 1T0S259765, 1T0S259766, 1T0S259767, 1T0S259774, 1T0S259778, 1T0S259780, 1T0S259781, 1T0S259792, 1T0S259796, 1T0S256435, 1T0S257360, 1T0S257362, 1T0S258140, 1T0S258178, 1T0S258195, 1T0S258548, 1T0S259008, 1T0S258418, 1T0S258426, 1T0S258431, 1T0S258432, 1T0S258441, 1T0S258443, 1T0S258447, 1T0S258453, 1T0S258454, 1T0S258459, 1T0S258467, 1T0S258468, 1T0S258475, 1T0S258477, 1T0S258480, 1T0S258486, 1T0S258487, 1T0S258488, 1T0S258489, 1T0S258491, 1T0S258494, 1T0S258505, 1T0S256405, 1T0S256416, 1T0S256995, 1T0S257314, 1T0S257341, 1T0S257361, 1T0S257365, 1T0S258152, 1T0S258202, 1T0S258216, 1T0S258510, 1T0S258512, 1T0S258527, 1T0S258528, 1T0S258532, 1T0S258547, 1T0S258551, 1T0S258559, 1T0S259004, 4T0S276809, 4T0S276834, 4T0S276846, 4T0S276850, 4T0S276855, 4T0S276860, 4T0S276871, 4T0S276885, 4T0S276887, 4T0S276890, 4T0S276892, 4T0S276896, 4T0S276898, 4T0S276903, 4T0S276905, 4T0S276915, 4T0S276930, 4T0S276945, 4T0S276949, 4T0S276950, 1T0S256818, 1T0S256865, 1T0S256881, 1T0S256882, 1T0S256911, 1T0S256921, 1T0S256922, 1T0S256930, 1T0S256942, 1T0S256951, 1T0S257344, 1T0S257366, 1T0S257376, 1T0S257379, 1T0S257380, 1T0S257383, 1T0S257390, 1T0S257401, 1T0S257418, 1T0S256468, 1T0S257018, 1T0S257025, 1T0S257027, 1T0S257034, 1T0S257035, 1T0S257049, 1T0S257052, 1T0S257054, 1T0S257056, 1T0S257057, 1T0S257067, 1T0S257074, 1T0S257076, 1T0S257077, 1T0S257084, 1T0S257086, 1T0S257098, 1T0S257104, 1T0S257133, 1T0S257184, 1T0S257328, 2T0S267231, 2T0S267241, 2T0S267264, 2T0S267267, 2T0S267268, 2T0S267269, 2T0S267286, 2T0S267291, 2T0S267295, 2T0S267654, 2T0S267661, 2T0S267663, 2T0S267677, 2T0S267678, 2T0S267694, 2T0S267702, 2T0S267730, 2T0S267747, 1T0S256706, 1T0S256731, 1T0S256735, 1T0S257016, 1T0S257051, 1T0S257053, 1T0S257058, 1T0S257065, 1T0S257069, 1T0S257075, 1T0S257079, 1T0S257081, 1T0S257091, 1T0S257106, 1T0S257119, 1T0S257120, 1T0S257121, 1T0S257130, 1T0S257131, 1T0S257135, 1T0S257136, 1T0S257137, 1T0S257145, 1T0S257094, 1T0S257100, 1T0S257102, 1T0S257118, 1T0S257126, 1T0S257134, 1T0S257149, 1T0S257150, 1T0S257152, 1T0S257155, 1T0S257168, 1T0S257174, 1T0S257181, 1T0S257187, 1T0S257188, 1T0S257190, 1T0S257201, 1T0S257203, 1T0S257206, 1T0S256715, 1T0S256740, 1T0S257032, 1T0S257040, 1T0S257096, 1T0S257151, 1T0S257167, 1T0S257409, 1T0S258734, 1T0S258735, 1T0S258736, 1T0S258737, 1T0S258738, 1T0S258739, 1T0S258740, 1T0S258741, 1T0S258742, 1T0S258743, 1T0S258744, 1T0S258745, 1T0S258746, 1T0S258747, 1T0S258748, 1T0S258749, 1T0S258750, 1T0S258752, 1T0S258754, 1T0S258756, 1T0S258757, 1T0S258758, 1T0S258759, 1T0S258760, 1T0S258761, 1T0S258765, 1T0S258771, 1T0S258798, 1T0S256811, 1T0S257011, 1T0S257020, 1T0S257021, 1T0S257154, 1T0S257172, 1T0S257200, 1T0S257208, 1T0S257210, 1T0S257218, 1T0S257402, 1T0S257410, 1T0S257823, 1T0S257883, 1T0S257041, 1T0S257216, 1T0S257413, 1T0S257099, 1T0S257180, 1T0S257191, 1T0S257197, 1T0S257213, 1T0S257215, 1T0S257225, 1T0S257816, 1T0S257820, 1T0S257847, 1T0S257876, 1T0S257877, 1T0S257879, 1T0S257885, 3T0S268773, 1T0S259192, 1T0S259237, 1T0S259244, 1T0S259249, 1T0S259252, 1T0S259253, 1T0S259255, 1T0S259256, 1T0S259259, 1T0S259272, 1T0S259277, 1T0S259343, 1T0S259356, 1T0S259357, 1T0S259389, 1T0S260097, 1T0S260116, 1T0S260117, 1T0S260119, 1T0S260121, 1T0S260124, 1T0S260126, 1T0S260129, 1T0S260130, 1T0S260132, 1T0S260133, 1T0S260137, 1T0S260138, 1T0S260139, 1T0S260140, 2T0S264267, 2T0S264274, 2T0S264278, 2T0S264284, 2T0S264285, 2T0S264304, 2T0S264306, 2T0S264320, 2T0S264699, 2T0S264708, 2T0S264723, 2T0S264739, 2T0S264747, 2T0S264771, 2T0S264870, 2T0S264947, 2T0S264950, 2T0S264952, 2T0S265421, 2T0S265437, 2T0S265453, 2T0S265456, 2T0S265459, 2T0S265491, 2T0S266081, 2T0S266161, 2T0S266206, 2T0S268026, 2T0S268030, 2T0S268032, 2T0S268034, 2T0S268035, 2T0S268036, 2T0S268038, 2T0S268039, 2T0S268053, 2T0S268062, 2T0S268066, 2T0S268069, 2T0S268071, 2T0S268073, 2T0S268074, 2T0S268076, 2T0S268080, 2T0S268084, 2T0S268086, 2T0S268088, 2T0S268089, 2T0S268097, 2T0S268099, 2T0S268103, 2T0S268104, 2T0S268106, 2T0S268110, 2T0S268111, 2T0S268118, 1T0S258407, 1T0S259180, 2T0S264322, 2T0S264349, 2T0S264352, 2T0S264356, 2T0S264359, 2T0S264378, 2T0S264379, 2T0S264426, 2T0S265414, 2T0S265418, 2T0S265433, 2T0S265434, 2T0S265435, 2T0S265441, 2T0S265444, 2T0S265449, 2T0S265452, 2T0S265454, 2T0S265467, 2T0S265472, 2T0S265485, 2T0S265492, 2T0S265510, 2T0S265543, 2T0S265552, 2T0S265567, 2T0S265575, 1T0S259162, 1T0S259167, 1T0S259232, 1T0S259233, 1T0S259242, 1T0S259243, 1T0S259245, 1T0S259248, 1T0S259250, 1T0S259254, 1T0S259258, 1T0S259260, 1T0S259261, 1T0S259262, 1T0S259266, 1T0S258861, 1T0S258872, 1T0S258874, 1T0S258880, 1T0S258883, 1T0S258884, 1T0S258885, 1T0S258895, 1T0S258899, 1T0S258902, 1T0S258903, 1T0S258904, 1T0S258907, 1T0S258910, 1T0S258912, 1T0S258913, 1T0S258916, 1T0S258917, 1T0S258918, 1T0S258921, 1T0S258925, 1T0S258926, 1T0S258928, 1T0S258932, 1T0S258944, 1T0S258947, 1T0S258974, 1T0S258773, 1T0S258795, 1T0S258804, 1T0S258827, 1T0S258836, 1T0S258869, 1T0S258870, 1T0S258876, 1T0S259656, 1T0S259722, 1T0S259740, 1T0S259741, 1T0S259746, 1T0S259756, 1T0S259757, 1T0S259764, 1T0S259768, 1T0S259769, 1T0S261052, 1T0S261093, 1T0S261101, 1T0S261110, 1T0S261122, 1T0S261128, 1T0S261129, 1T0S261152, 1T0S261154, 1T0S261157, 1T0S261162, 1T0S261163, 1T0S261167, 1T0S261171, 1T0S261176, 1T0S261181, 1T0S261183, 1T0S261185, 1T0S261188, 1T0S261199, 1T0S261200, 1T0S261201, 1T0S261203, 1T0S261207, 1T0S258858, 1T0S258893, 1T0S259160, 1T0S259182, 1T0S259184, 1T0S259187, 1T0S259191, 1T0S259193, 1T0S259194, 1T0S259195, 1T0S259199, 1T0S259230, 1T0S259231, 1T0S259265, 1T0S259632, 1T0S259640, 1T0S260058, 1T0S261281, 1T0S261289, 1T0S261411, 1T0S261426, 1T0S261431, 1T0S261432, 1T0S261433, 1T0S256382, 1T0S256412, 1T0S256419, 1T0S256422, 1T0S256427, 1T0S256429, 1T0S256956, 1T0S257047, 1T0S257319, 1T0S257338, 1T0S258160, 1T0S258191, 1T0S258197, 1T0S258215, 1T0S258511, 1T0S258515, 1T0S258516, 1T0S258537, 1T0S258540, 1T0S258543, 1T0S258554, 1T0S258561, 1T0S258574, 1T0S258949, 1T0S259006, 1T0S259007, 1T0S256425, 1T0S256428, 1T0S256444, 1T0S256445, 1T0S256447, 1T0S257004, 1T0S257143, 1T0S257317, 1T0S257334, 1T0S257378, 1T0S257892, 1T0S258417, 1T0S258500, 1T0S258541, 1T0S258550, 1T0S258578, 1T0S258579, 1T0S258922, 1T0S258952, 1T0S261132, 1T0S261158, 1T0S261263, 1T0S261372, 1T0S261378, 1T0S261403, 1T0S261423, 1T0S261425, 1T0S261434, 1T0S261457, 2T0S264731, 2T0S264919, 2T0S264962, 2T0S265056, 2T0S265384, 2T0S265413, 2T0S265415, 2T0S265427, 2T0S265428, 2T0S265430, 2T0S265436, 2T0S265559, 2T0S266092, 2T0S266133, 2T0S266141, 2T0S266148, 2T0S266153, 2T0S266156, 2T0S266163, 2T0S267634, 2T0S267656, 2T0S267665, 2T0S267670, 2T0S267675, 2T0S267697, 2T0S267699, 2T0S267707, 2T0S267720, 2T0S267728, 2T0S267735, 2T0S267764, 2T0S267782, 2T0S267784, 2T0S267790, 2T0S267791, 1T0S262525, 1T0S262528, 1T0S262532, 1T0S262546, 1T0S262549, 1T0S262550, 1T0S262551, 1T0S262554, 1T0S262555, 1T0S262556, 1T0S262557, 1T0S262558, 1T0S262559, 1T0S262560, 1T0S262561, 1T0S262563, 1T0S262566, 1T0S262569, 1T0S262570, 1T0S262572, 1T0S262574, 1T0S262575, 1T0S262577, 1T0S262586, 1T0S262587, 1T0S262588, 1T0S262590, 1T0S262592, 1T0S262594, 1T0S262596, 1T0S262599, 1T0S262605, 1T0S262607, 1T0S262354, 1T0S262363, 1T0S262381, 1T0S262408, 1T0S262420, 1T0S262424, 1T0S262461, 1T0S262478, 1T0S262509, 1T0S262518, 1T0S262520, 1T0S262526, 1T0S262530, 1T0S262531, 1T0S262534, 1T0S262536, 1T0S262540, 1T0S262545, 2T0S266610, 2T0S266633, 1T0S258774, 1T0S258782, 1T0S258820, 1T0S258821, 1T0S258822, 1T0S258837, 1T0S258843, 1T0S258848, 1T0S258849, 1T0S258854, 1T0S258857, 1T0S258864, 1T0S258867, 1T0S258868, 1T0S258881, 1T0S258882, 1T0S256417, 1T0S256424, 1T0S256430, 1T0S257318, 1T0S257327, 1T0S257340, 1T0S257370, 1T0S258151, 1T0S258161, 1T0S258188, 1T0S258190, 1T0S258465, 1T0S258523, 1T0S258539, 1T0S258544, 1T0S258552, 1T0S258558, 1T0S258591, 1T0S258592, 1T0S262444, 1T0S262449, 1T0S262454, 1T0S262458, 1T0S262464, 1T0S262467, 1T0S262471, 1T0S262473, 1T0S262477, 1T0S262485, 1T0S262496, 1T0S262497, 1T0S262501, 1T0S262502, 1T0S259342, 1T0S259398, 1T0S259408, 1T0S259417, 1T0S259426, 1T0S259430, 1T0S259431, 1T0S259432, 1T0S259435, 1T0S259438, 1T0S259440, 1T0S259444, 1T0S259703, 1T0S259713, 1T0S259714, 1T0S259716, 1T0S259724, 1T0S259725, 1T0S259737, 1T0S261883, 1T0S261909, 1T0S261950, 2T0S263398, 2T0S263410, 2T0S263431, 2T0S263472, 3T0S271148, 1T0S259013, 1T0S259618, 1T0S259631, 1T0S259642, 1T0S259650, 1T0S259676, 1T0S261308, 1T0S261348, 1T0S261360, 1T0S261374, 1T0S261404, 1T0S261407, 1T0S261410, 1T0S261414, 1T0S261416, 1T0S261428, 1T0S261439, 1T0S261444, 1T0S261447, 1T0S261462, 2T0S264732, 2T0S264737, 2T0S264749, 2T0S264760, 2T0S264768, 2T0S264862, 2T0S264863, 2T0S264905, 2T0S265488, 2T0S266851, 2T0S266867, 2T0S266868, 2T0S266872, 2T0S266876, 2T0S266879, 2T0S266880, 2T0S266888, 2T0S266895, 2T0S266906, 1T0S259648, 1T0S259651, 1T0S259653, 1T0S259659, 1T0S259661, 1T0S259662, 1T0S259665, 1T0S259672, 1T0S259675, 1T0S259677, 1T0S259696, 1T0S259664, 1T0S260034, 1T0S260046, 2T0S264224, 2T0S264226, 2T0S264227, 2T0S264229, 2T0S264230, 2T0S264231, 2T0S264232, 2T0S264234, 2T0S264235, 2T0S264236, 2T0S264239, 2T0S264256, 2T0S264258, 2T0S264259, 2T0S264260, 2T0S264263, 2T0S264264, 2T0S264265, 1T0S262565, 2T0S262759, 2T0S262785, 2T0S262788, 2T0S262793, 2T0S262799, 2T0S262802, 2T0S262816, 2T0S262819, 2T0S262826, 2T0S262827, 2T0S262829, 2T0S262832, 2T0S262835, 2T0S262838, 2T0S262839, 2T0S262844, 2T0S262850, 2T0S262855, 2T0S262861, 2T0S262863, 1T0S259446, 1T0S259475, 1T0S259489, 1T0S259502, 1T0S259518, 1T0S259522, 1T0S259544, 1T0S259555, 1T0S259568, 1T0S259579, 1T0S261800, 1T0S262240, 1T0S262244, 1T0S262245, 1T0S262246, 1T0S262247, 1T0S262248, 1T0S262250, 1T0S262251, 1T0S262252, 1T0S262253, 1T0S262255, 1T0S262258, 1T0S262259, 1T0S262260, 1T0S262261, 1T0S262262, 1T0S262263, 1T0S262264, 1T0S262270, 1T0S262277, 1T0S262281, 1T0S262291, 1T0S262298, 2T0S263403, 2T0S263418, 2T0S263457, 2T0S263464, 2T0S263477, 2T0S263478, 2T0S263495, 2T0S263501, 2T0S263502, 2T0S263503, 2T0S263505, 2T0S263510, 2T0S263511, 2T0S263515, 2T0S263524, 1T0S262312, 1T0S262314, 1T0S262320, 1T0S262323, 1T0S262324, 1T0S262325, 1T0S262327, 1T0S262328, 1T0S262330, 1T0S262331, 1T0S262332, 1T0S262333, 1T0S262336, 1T0S262337, 1T0S262338, 1T0S262339, 1T0S262340, 1T0S262342, 1T0S262343, 1T0S262344, 1T0S262345, 1T0S262346, 1T0S262347, 1T0S262348, 1T0S262349, 1T0S262352, 1T0S262353, 1T0S262355, 1T0S262356, 1T0S262358, 1T0S262359, 1T0S262364, 1T0S262365, 1T0S262366, 1T0S262368, 1T0S262369, 1T0S262397, 1T0S262399, 1T0S262401, 1T0S262405, 1T0S262409, 1T0S262411, 1T0S262414, 1T0S262417, 1T0S262418, 1T0S262421, 1T0S262425, 1T0S262426, 1T0S262428, 1T0S262430, 1T0S262431, 1T0S262435, 1T0S262436, 1T0S262438, 1T0S262439, 1T0S262441, 1T0S262447, 1T0S262448, 1T0S262450, 1T0S262451, 1T0S262452, 1T0S262456, 1T0S262457, 1T0S262460, 1T0S262463, 1T0S262465, 1T0S262466, 1T0S262468, 1T0S262469, 1T0S262470, 1T0S262472, 1T0S262476, 1T0S262481, 1T0S262482, 1T0S262489, 1T0S262494, 1T0S262495, 2T0S263405, 2T0S263409, 2T0S263411, 2T0S263426, 2T0S263427, 2T0S263428, 2T0S263436, 2T0S263437, 2T0S263442, 2T0S263444, 2T0S263449, 2T0S263455, 2T0S263456, 2T0S263463, 2T0S263466, 2T0S263470, 2T0S263471, 2T0S263487, 2T0S263492, 2T0S263500, 2T0S263513, 2T0S263521, 2T0S263532, 2T0S263533, 2T0S263534, 2T0S263535, 2T0S263536, 2T0S263539, 2T0S263541, 2T0S263544, 2T0S263545, 2T0S263547, 2T0S262871, 2T0S262930, 2T0S263040, 2T0S263105, 2T0S263110, 2T0S263116, 2T0S263117, 2T0S263118, 2T0S263125, 2T0S263126, 2T0S263136, 2T0S263138, 2T0S263140, 2T0S263142, 2T0S263143, 2T0S263145, 2T0S263146, 2T0S263147, 2T0S263148, 2T0S263150, 2T0S263154, 2T0S263155, 2T0S263156, 2T0S263157, 2T0S263158, 2T0S263162, 2T0S263174, 2T0S263175, 2T0S263177, 2T0S263180, 2T0S263185, 1T0S262506, 1T0S262522, 1T0S262535, 1T0S262538, 1T0S262541, 1T0S262548, 1T0S262552, 1T0S262568, 1T0S262571, 1T0S262576, 1T0S262578, 1T0S262582, 1T0S262583, 1T0S262584, 1T0S262597, 2T0S262761, 2T0S262767, 2T0S262815, 2T0S262817, 2T0S262825, 2T0S262830, 2T0S262834, 2T0S262836, 2T0S262837, 2T0S262840, 2T0S262842, 2T0S262843, 2T0S262845, 2T0S262846, 2T0S262847, 2T0S262848, 2T0S262849, 2T0S262851, 2T0S262852, 2T0S262854, 2T0S262856, 2T0S262857, 2T0S262858, 2T0S262859, 2T0S262860, 2T0S262862, 2T0S262864, 2T0S262866, 2T0S262868, 2T0S262869, 2T0S262870, 2T0S262872, 2T0S262873, 2T0S262874, 2T0S262875, 2T0S262876, 2T0S262883, 2T0S262884, 2T0S262885, 2T0S262887, 2T0S262889, 2T0S262890, 2T0S262892, 2T0S262894, 2T0S262897, 2T0S262898, 2T0S263875, 2T0S264026, 2T0S264028, 2T0S264039, 2T0S264095, 2T0S264096, 2T0S264103, 2T0S264104, 2T0S264108, 2T0S264110, 2T0S264111, 2T0S264116, 2T0S264118, 2T0S264120, 2T0S264148, 2T0S264149, 2T0S264152, 2T0S264155, 2T0S264156, 2T0S264158, 2T0S264159, 2T0S264161, 2T0S264162, 2T0S264165, 2T0S264172, 2T0S264174, 2T0S264192, 2T0S264194, 2T0S264195, 2T0S264197, 2T0S264205, 2T0S266500, 2T0S266883, 2T0S266904, 2T0S266920, 2T0S266923, 2T0S266937, 2T0S266955, 2T0S266957, 2T0S266969, 2T0S267715, 2T0S267769, 2T0S267811, 2T0S267825, 2T0S264695, 2T0S264702, 2T0S264706, 2T0S264872, 2T0S264893, 2T0S265249, 2T0S265457, 2T0S265503, 2T0S265530, 2T0S265539, 2T0S265564, 2T0S265565, 2T0S265572, 2T0S265597, 2T0S265609, 2T0S265619, 2T0S265626, 2T0S266874, 2T0S266886, 2T0S266889, 2T0S266902, 2T0S266922, 2T0S266925, 2T0S263204, 2T0S263397, 2T0S263400, 2T0S263401, 2T0S263422, 2T0S263425, 2T0S263443, 2T0S263482, 2T0S263485, 2T0S263490, 2T0S263497, 2T0S263517, 2T0S263518, 2T0S263520, 2T0S263522, 2T0S263523, 2T0S263527, 2T0S263528, 2T0S263529, 2T0S263537, 2T0S263546, 2T0S263619, 2T0S263626, 2T0S263627, 2T0S263628, 2T0S263631, 2T0S263632, 2T0S262905, 2T0S262906, 2T0S262907, 2T0S262910, 2T0S262913, 2T0S262917, 2T0S262920, 2T0S262921, 2T0S262922, 2T0S262924, 2T0S262929, 0T0S248745, 2T0S262841, 2T0S262877, 2T0S262878, 2T0S262891, 2T0S262893, 2T0S262901, 2T0S262902, 2T0S262903, 2T0S262909, 2T0S262911, 2T0S262912, 2T0S262926, 2T0S262927, 2T0S267595, 2T0S267602, 2T0S267613, 2T0S267624, 2T0S267630, 2T0S267633, 2T0S267640, 2T0S267643, 2T0S267644, 2T0S267652, 2T0S267655, 2T0S267659, 2T0S267681, 2T0S263027, 2T0S263031, 2T0S263036, 2T0S263044, 2T0S263054, 2T0S263064, 2T0S263065, 2T0S263095, 2T0S263115, 2T0S263120, 2T0S265446, 2T0S266152, 2T0S266172, 2T0S266189, 2T0S267760, 2T0S267772, 2T0S268045, 3T0S269675, 2T0S263416, 2T0S263417, 2T0S263424, 2T0S263429, 2T0S263433, 2T0S263438, 2T0S263465, 2T0S263476, 2T0S263509, 2T0S263514, 2T0S266490, 2T0S266514, 2T0S266521, 2T0S266525, 2T0S266537, 2T0S266542, 2T0S266544, 2T0S266567, 2T0S264210, 2T0S264211, 2T0S264220, 2T0S264221, 2T0S264240, 2T0S264241, 2T0S264243, 2T0S264245, 2T0S264246, 2T0S264247, 2T0S264248, 2T0S264249, 2T0S264250, 2T0S264255, 2T0S266494, 2T0S266505, 2T0S266512, 2T0S266520, 2T0S266528, 2T0S266536, 2T0S263811, 2T0S263997, 2T0S264000, 2T0S264009, 2T0S264025, 2T0S264034, 2T0S264043, 2T0S264097, 2T0S264098, 2T0S264099, 2T0S264100, 2T0S264101, 2T0S264102, 2T0S264106, 2T0S264109, 2T0S264112, 2T0S264115, 2T0S264117, 2T0S264151, 4T0S277066, 4T0S277408, 4T0S277414, 4T0S277417, 4T0S277418, 4T0S277439, 4T0S277449, 4T0S277454, 4T0S277455, 4T0S277464, 4T0S277476, 4T0S277478, 4T0S277480, 4T0S277487, 4T0S277488, 4T0S277493, 4T0S277497, 4T0S277501, 4T0S277503, 4T0S277505, 4T0S277507, 4T0S277510, 4T0S277528, 4T0S277533, 4T0S277555, 4T0S277556, 4T0S277557, 4T0S277574, 2T0S266575, 2T0S266581, 2T0S266601, 2T0S266607, 2T0S266619, 2T0S266853, 2T0S267234, 2T0S267293, 4T0S277070, 4T0S277084, 4T0S277085, 4T0S277105, 4T0S277117, 4T0S275868, 4T0S275875, 4T0S275890, 4T0S275901, 4T0S275902, 4T0S275905, 4T0S275911, 4T0S275915, 4T0S275916, 4T0S275917, 4T0S275920, 4T0S275927, 4T0S275930, 4T0S275933, 4T0S275935, 4T0S275937, 4T0S275941, 4T0S275943, 4T0S275947, 4T0S275950, 4T0S275955, 4T0S275963, 4T0S275964, 4T0S275965, 4T0S275966, 4T0S275968, 4T0S275969, 4T0S275972, 4T0S275975, 4T0S275976, 4T0S275985, 4T0S275986, 2T0S266915, 2T0S266956, 2T0S266958, 2T0S266964, 2T0S266970, 2T0S266997, 2T0S266999, 2T0S267013, 2T0S267019, 2T0S267028, 2T0S267031, 2T0S267034, 2T0S267067, 2T0S267098, 2T0S267100, 2T0S267106, 2T0S267169, 2T0S267196, 2T0S267214, 3T0S269664, 3T0S269667, 3T0S269669, 3T0S269670, 3T0S269671, 3T0S269678, 3T0S269680, 3T0S269685, 4T0S275862, 4T0S275886, 4T0S275899, 4T0S275900, 4T0S275904, 4T0S275907, 4T0S275918, 2T0S266908, 2T0S266928, 2T0S266945, 2T0S266950, 2T0S266952, 2T0S266987, 2T0S266990, 2T0S266994, 2T0S266995, 2T0S267000, 2T0S267001, 2T0S267004, 2T0S267007, 2T0S267014, 2T0S267016, 2T0S267024, 2T0S267025, 2T0S267026, 2T0S267027, 3T0S268722, 3T0S268735, 3T0S268743, 3T0S268753, 3T0S268785, 2T0S264693, 2T0S264755, 2T0S265478, 2T0S265534, 2T0S265558, 2T0S265568, 2T0S265599, 2T0S265617, 3T0S268703, 3T0S268710, 3T0S268715, 3T0S268720, 3T0S268723, 3T0S268731, 3T0S268747, 3T0S268751, 3T0S268760, 3T0S268765, 3T0S268780, 3T0S268784, 3T0S268786, 3T0S268787, 3T0S268790, 3T0S268792, 3T0S268793, 2T0S264266, 2T0S264276, 2T0S264301, 2T0S264370, 2T0S264387, 2T0S264461, 2T0S264673, 2T0S264684, 2T0S264686, 2T0S264688, 2T0S264748, 2T0S264753, 2T0S264758, 2T0S264761, 2T0S264956, 2T0S264963, 2T0S265054, 2T0S265442, 2T0S264290, 2T0S264312, 2T0S264316, 2T0S264382, 2T0S264579, 2T0S264683, 2T0S264741, 2T0S264914, 2T0S264948, 2T0S265053, 2T0S265438, 2T0S265551, 2T0S267668, 2T0S267714, 2T0S267723, 2T0S267740, 2T0S267767, 2T0S268033, 2T0S268044, 2T0S268052, 2T0S268061, 2T0S268068, 2T0S268072, 2T0S268077, 2T0S268085, 2T0S268091, 2T0S268098, 2T0S264286, 2T0S264372, 2T0S264752, 2T0S264766, 2T0S264918, 2T0S264944, 2T0S266147, 2T0S266171, 2T0S266174, 2T0S266176, 2T0S266202, 2T0S267593, 2T0S267594, 2T0S267614, 2T0S267626, 2T0S267627, 2T0S267632, 2T0S267635, 2T0S267641, 2T0S264289, 2T0S264860, 2T0S265366, 3T0S270777, 3T0S270786, 3T0S270789, 3T0S270790, 3T0S270791, 3T0S270794, 3T0S270797, 3T0S270799, 3T0S270803, 3T0S270809, 3T0S270810, 3T0S270812, 3T0S270813, 3T0S270815, 3T0S270819, 3T0S270824, 3T0S270830, 3T0S270831, 3T0S270839, 3T0S270843, 3T0S270849, 3T0S270079, 3T0S270096, 3T0S270097, 3T0S270103, 3T0S270396, 3T0S270398, 3T0S270403, 3T0S270411, 3T0S270412, 3T0S270413, 3T0S270415, 3T0S270416, 3T0S270417, 3T0S270421, 3T0S270422, 3T0S270423, 3T0S270424, 3T0S270426, 3T0S270427, 3T0S270430, 3T0S270440, 3T0S270453, 3T0S270457, 3T0S270462, 3T0S270476, 3T0S270479, 3T0S270481, 2T0S264968, 2T0S265443, 2T0S266175, 2T0S266178, 2T0S266918, 3T0S268738, 3T0S270087, 3T0S270104, 3T0S270397, 3T0S270399, 3T0S270408, 3T0S270425, 3T0S270428, 3T0S270429, 3T0S270432, 3T0S270442, 3T0S270446, 3T0S270450, 3T0S270466, 3T0S270468, 3T0S270469, 3T0S270472, 3T0S274819, 3T0S274839, 3T0S275205, 3T0S275206, 3T0S275207, 3T0S275208, 3T0S275209, 3T0S275210, 3T0S275218, 3T0S275222, 3T0S275223, 3T0S275224, 3T0S275226, 3T0S275228, 3T0S275229, 3T0S275238, 3T0S275239, 3T0S275240, 3T0S269988, 3T0S270002, 3T0S270005, 3T0S270012, 3T0S270016, 3T0S270018, 3T0S270055, 3T0S270056, 3T0S270066, 3T0S270070, 3T0S270084, 3T0S270101, 3T0S270111, 3T0S270113, 3T0S270392, 3T0S270618, 3T0S270581, 3T0S270591, 3T0S270594, 3T0S271666, 3T0S268713, 3T0S268717, 3T0S268730, 3T0S269737, 3T0S269749, 3T0S269755, 3T0S269798, 3T0S269804, 3T0S269806, 3T0S269808, 3T0S269816, 3T0S269819, 3T0S269823, 3T0S269834, 3T0S269844, 2T0S264704, 2T0S264705, 2T0S264711, 2T0S264726, 2T0S264729, 2T0S264740, 2T0S264744, 2T0S264754, 2T0S264757, 2T0S264772, 2T0S264869, 2T0S264874, 2T0S265525, 2T0S265538, 2T0S265586, 2T0S265589, 2T0S265595, 2T0S265600, 2T0S265614, 2T0S267185, 2T0S267227, 2T0S267255, 2T0S267256, 2T0S267259, 2T0S267285, 2T0S267288, 2T0S267300, 2T0S267303, 2T0S267309, 2T0S267590, 2T0S267591, 2T0S267592, 2T0S267596, 2T0S267597, 2T0S267598, 2T0S267599, 2T0S267600, 2T0S267603, 2T0S267604, 2T0S267605, 2T0S267606, 2T0S267607, 2T0S267608, 2T0S267609, 2T0S267610, 2T0S267615, 2T0S267617, 2T0S267619, 2T0S267621, 2T0S267623, 2T0S267664, 3T0S270605, 3T0S274820, 3T0S274822, 3T0S274830, 3T0S274832, 3T0S274946, 3T0S274950, 3T0S269986, 3T0S269997, 3T0S270008, 3T0S270014, 3T0S270026, 3T0S270109, 3T0S269769, 3T0S269822, 3T0S269837, 3T0S273864, 3T0S273865, 3T0S273866, 3T0S273873, 4T0S275978, 4T0S276325, 4T0S276326, 4T0S276331, 4T0S276335, 4T0S276337, 4T0S276342, 4T0S276344, 4T0S276363, 4T0S276364, 4T0S276366, 4T0S276381, 4T0S276384, 4T0S276403, 4T0S276406, 4T0S276410, 4T0S276415, 4T0S276416, 4T0S276462, 3T0S274821, 3T0S274828, 3T0S274840, 3T0S274841, 3T0S273896, 3T0S273913, 3T0S273925, 4T0S277596, 4T0S277607, 4T0S277618, 4T0S277651, 4T0S277672, 4T0S277697, 4T0S277699, 4T0S277702, 4T0S277715, 4T0S277717, 4T0S277718, 4T0S277733, 4T0S277735, 4T0S277737, 4T0S277743, 4T0S277759, 3T0S274105, 3T0S274118, 3T0S274815, 3T0S274816, 3T0S274842, 3T0S274869, 3T0S274875, 3T0S274877, 3T0S274883, 3T0S274891, 3T0S274893, 3T0S274894, 3T0S274897, 3T0S274909, 3T0S274930, 3T0S274948, 3T0S274952, 3T0S274953, 3T0S274982, 3T0S274017, 3T0S274072, 3T0S274075, 3T0S274087, 3T0S274116, 3T0S274123, 3T0S274134, 3T0S274152, 3T0S274834, 3T0S274851, 3T0S274853, 3T0S274857, 3T0S274867, 3T0S274895, 3T0S274906, 3T0S271717, 3T0S271763, 3T0S271767, 3T0S271807, 3T0S271826, 3T0S271853, 4T0S275656, 4T0S275715, 4T0S275721, 4T0S275722, 4T0S275731, 4T0S275734, 4T0S275739, 4T0S275740, 4T0S275757, 4T0S275758, 4T0S275762, 4T0S275765, 4T0S275767, 3T0S274843, 3T0S274907, 3T0S274924, 3T0S274929, 3T0S274941, 3T0S274943, 3T0S274944, 3T0S274945, 3T0S274955, 3T0S274956, 3T0S273867, 3T0S273920, 3T0S273933, 3T0S273934, 3T0S273958, 3T0S274836, 3T0S274014, 3T0S274047, 3T0S274056, 3T0S274064, 3T0S274076, 3T0S274084, 3T0S274098, 3T0S274111, 3T0S274065, 3T0S274069, 4T0S275922, 4T0S276351, 4T0S276386, 4T0S276440, 4T0S276442, 4T0S276489, 4T0S276512, 4T0S276515, 4T0S276528, 4T0S276535, 4T0S276550, 4T0S276551, 4T0S276573, 4T0S276575, 4T0S276598, 4T0S276610, 4T0S276611, 4T0S276617, 4T0S276620, 4T0S276623, 4T0S276628, 4T0S276629, 4T0S276638, 4T0S276641, 4T0S276643, 4T0S276646, 4T0S276657, 4T0S276667, 4T0S277050, 4T0S277060, 4T0S277091, 4T0S277092, 4T0S277351, 4T0S277354, 4T0S277356, 4T0S277362, 4T0S277372, 4T0S277375, 4T0S277382, 4T0S277385, 4T0S277391, 4T0S277394, 3T0S274835, 3T0S274854, 3T0S274861, 3T0S274871, 3T0S274887, 4T0S276369, 4T0S276433, 4T0S276435, 4T0S276540, 4T0S276542, 4T0S276560, 4T0S276562, 4T0S276566, 4T0S276567, 4T0S276568, 4T0S276569, 4T0S276576, 4T0S276581, 4T0S276591, 4T0S276592, 4T0S276597, 4T0S276608, 4T0S276613, 4T0S276619, 4T0S276625, 4T0S276630, 4T0S276634, 4T0S276640, 4T0S276642, 4T0S276653, 4T0S276658, 4T0S276663, 4T0S276665, 4T0S276666, 4T0S277040, 4T0S277064, 4T0S277088, 4T0S277097, 4T0S277098, 4T0S277471, 4T0S277473, 4T0S277494, 4T0S277499, 4T0S277514, 4T0S277515, 4T0S277531, 4T0S277542, 4T0S277559, 4T0S276988, 4T0S277359, 4T0S277436, 4T0S277438, 4T0S277453, 4T0S277470, 4T0S277477, 4T0S277516, 4T0S277518, 4T0S277520, 4T0S277532, 4T0S277552, 4T0S277554, 4T0S277558, 4T0S277560, 4T0S277566, 4T0S277573, 4T0S277581, 4T0S277583, 4T0S277590, 4T0S277597, 4T0S277599, 4T0S276967, 4T0S276978, 4T0S277014, 1T0S256916, 1T0S256958, 1T0S256967, 1T0S256982, 1T0S257198, 1T0S257258, 1T0S257275, 1T0S257295, 1T0S257315, 1T0S257321, 2T0S266501, 2T0S266992, 2T0S267006, 2T0S267010, 2T0S267030, 2T0S267035, 2T0S267036, 2T0S267037, 2T0S267070, 3T0S270829, 3T0S270837, 3T0S270847, 3T0S270852, 3T0S270856, 3T0S270864, 3T0S270868, 3T0S270869, 3T0S270874, 3T0S270882, 3T0S271578, 3T0S271581, 3T0S271598, 3T0S271599, 3T0S271601, 3T0S271605, 3T0S271613, 3T0S271634, 3T0S271643, 3T0S271645, 0T0S253375, 0T0S253473, 0T0S253482, 0T0S253487, 0T0S253493, 0T0S253498, 0T0S253500, 0T0S253504, 0T0S253508, 0T0S253509, 0T0S253569, 1T0S256494, 1T0S261113, 1T0S261892, 1T0S261907, 1T0S261912, 1T0S261943, 1T0S261944, 2T0S263817, 2T0S263824, 2T0S263835, 2T0S263845, 2T0S263846, 2T0S263847, 2T0S267032, 2T0S267040, 2T0S267055, 2T0S267069, 3T0S274837, 3T0S274844, 3T0S274845, 3T0S274850, 3T0S274858, 3T0S274862, 3T0S274865, 3T0S274866, 3T0S274874, 3T0S274881, 3T0S274884, 3T0S274886, 3T0S274888, 3T0S274896, 3T0S274899, 3T0S274913, 3T0S274915, 3T0S274919, 3T0S274921, 3T0S274949, 1T0S256120, 1T0S256170, 2T0S263109, 2T0S263190, 2T0S266511, 2T0S266522, 0T0S254509, 0T0S254522, 0T0S254524, 0T0S254526, 0T0S254555, 0T0S254557, 1T0S256825, 1T0S256829, 1T0S256838, 1T0S256846, 1T0S256848, 1T0S256862, 1T0S256867, 1T0S256875, 1T0S256886, 1T0S256887, 1T0S256900, 1T0S256950, 1T0S256954, 3T0S270690, 3T0S270694, 3T0S270702, 3T0S270709, 3T0S270714, 3T0S270716, 3T0S270718, 3T0S270719, 3T0S270721, 3T0S270722, 3T0S270723, 3T0S270725, 3T0S270731, 3T0S270733, 3T0S270734, 3T0S270735, 3T0S270736, 3T0S270742, 1T0S261488, 1T0S261497, 1T0S261500, 1T0S261506, 1T0S261523, 1T0S261524, 1T0S261528, 1T0S261534, 1T0S261535, 1T0S261544, 1T0S261545, 1T0S261547, 1T0S261550, 1T0S261555, 1T0S261568, 1T0S261571, 1T0S261582, 3T0S270106, 3T0S270471, 3T0S270478, 3T0S270502, 3T0S270621, 3T0S270656, 3T0S270773, 3T0S270775, 3T0S270804, 3T0S270807, 2T0S266509, 2T0S266561, 0T0S255124, 1T0S256613, 1T0S256625, 1T0S256634, 3T0S270438, 3T0S270498, 2T0S264252, 2T0S264742, 2T0S267612, 2T0S267839, 2T0S267855, 2T0S267856, 2T0S267860, 2T0S267861, 2T0S267885, 2T0S267897, 2T0S264044, 2T0S264047, 2T0S264122, 2T0S264154, 2T0S264193, 2T0S264199, 2T0S264201, 2T0S264204, 2T0S263807, 2T0S263855, 2T0S263861, 2T0S263862, 2T0S263869, 2T0S263870, 2T0S263895, 2T0S263897, 2T0S264001, 0T0S255141, 0T0S255157, 0T0S255181, 2T0S264337, 2T0S264685, 2T0S264718, 2T0S267131, 2T0S267254, 2T0S267258, 2T0S267263, 2T0S267618, 3T0S268661, 3T0S268673, 3T0S271107, 3T0S271110, 3T0S271112, 3T0S271119, 3T0S271131, 3T0S271741, 3T0S271774, 3T0S271798, 3T0S271830, 3T0S271843, 0T0S253187, 0T0S253192, 0T0S253197, 0T0S253216, 0T0S253218, 0T0S253219, 0T0S253233, 0T0S253236, 0T0S253244, 0T0S253248, 0T0S253249, 0T0S253252, 0T0S253258, 0T0S253261, 0T0S253262, 0T0S253263, 0T0S253266, 0T0S253267, 0T0S253277, 0T0S252723, 0T0S252725, 0T0S254588, 0T0S254590, 0T0S254594, 0T0S254596, 0T0S254608, 0T0S254609, 0T0S254610, 0T0S254611, 0T0S254612, 0T0S254614, 0T0S254615, 0T0S254617, 0T0S254618, 0T0S254619, 0T0S254620, 0T0S254621, 0T0S254622, 0T0S254623, 0T0S254624, 0T0S254625, 0T0S254626, 0T0S254627, 0T0S254628, 0T0S254629, 0T0S254630, 0T0S254631, 0T0S254634, 0T0S254635, 0T0S254637, 0T0S254638, 0T0S254639, 0T0S254643, 0T0S254648, 0T0S254649, 0T0S254650, 0T0S254659, 0T0S254660, 0T0S254671, 0T0S253542, 0T0S253562, 0T0S253608, 0T0S253622, 0T0S253626, 0T0S253630, 0T0S254528, 0T0S253961, 0T0S255046, 0T0S255054, 0T0S255059, 0T0S255062, 0T0S255072, 0T0S255075, 0T0S255076, 0T0S255080, 0T0S255081, 0T0S255082, 0T0S255083, 0T0S255086, 0T0S255093, 0T0S255099, 0T0S255103, 0T0S255111, 0T0S255113, 0T0S255136, 0T0S255140, 0T0S251951, 0T0S251957, 0T0S251972, 0T0S251988, 0T0S251997, 0T0S252010, 0T0S252011, 0T0S252023, 0T0S252029, 0T0S252054, 0T0S251676, 0T0S251767, 2T0S267038, 2T0S267049, 0T0S252598, 0T0S252627, 0T0S252633, 0T0S252645, 0T0S252646, 0T0S252648, 0T0S252653, 0T0S252656, 0T0S252657, 0T0S252664, 0T0S252669, 0T0S252672, 0T0S252673, 0T0S252674, 0T0S252678, 0T0S252679, 0T0S252680, 0T0S252681, 0T0S252682, 0T0S252686, 0T0S252692, 0T0S252695, 0T0S252696, 0T0S252698, 0T0S252701, 0T0S252705, 0T0S252706, 0T0S252713, 0T0S251989, 0T0S252028, 0T0S252034, 0T0S252036, 0T0S252038, 0T0S252043, 0T0S252055, 0T0S252058, 1T0S258753, 1T0S258755, 1T0S258763, 1T0S258767, 1T0S258768, 1T0S258772, 1T0S258776, 1T0S258779, 1T0S258780, 1T0S258781, 1T0S258800, 0T0S251941, 0T0S253326, 0T0S253327, 0T0S253344, 0T0S253347, 0T0S253370, 0T0S253376, 0T0S253395, 0T0S253412, 0T0S253447, 0T0S253495, 0T0S253499, 0T0S253506, 0T0S253520, 0T0S253523, 0T0S253525, 0T0S253528, 0T0S253531, 0T0S253537, 0T0S253538, 0T0S253544, 0T0S253547, 0T0S253561, 0T0S253567, 0T0S253572, 0T0S253576, 0T0S253579, 0T0S253584, 0T0S253586, 0T0S253596, 0T0S254500, 0T0S254501, 0T0S254505, 0T0S254508, 0T0S254513, 0T0S254516, 0T0S254519, 0T0S253952, 0T0S253956, 0T0S253957, 0T0S253970, 0T0S253977, 0T0S253939, 0T0S253940, 0T0S253945, 0T0S253946, 0T0S253947, 0T0S253950, 0T0S253951, 0T0S253976, 0T0S254653, 0T0S254655, 0T0S254656, 0T0S254657, 0T0S254658, 0T0S254662, 0T0S254663, 0T0S254664, 0T0S254666, 0T0S254667, 0T0S254669, 0T0S254674, 0T0S254675, 0T0S254676, 0T0S254678, 0T0S254679, 0T0S254680, 0T0S254681, 0T0S254682, 0T0S251971, 0T0S252045, 0T0S252071, 0T0S252610, 0T0S252690, 0T0S252717, 0T0S252720, 0T0S252738, 0T0S253201, 0T0S253255, 0T0S253264, 0T0S253273, 0T0S253275, 0T0S253276, 0T0S253281, 0T0S253287, 0T0S253289, 0T0S253292, 0T0S253293, 0T0S253300, 0T0S253301, 0T0S253304, 0T0S253306, 0T0S253309, 0T0S253310, 0T0S253311, 0T0S253312, 0T0S253313, 0T0S253316, 0T0S253317, 0T0S253319, 0T0S253320, 0T0S253321, 0T0S253322, 0T0S253325, 0T0S253328, 0T0S253329, 0T0S253330, 0T0S253332, 0T0S253335, 0T0S253336, 0T0S253338, 0T0S253354, 0T0S253358, 1T0S256532, 1T0S256556, 1T0S256562, 1T0S256564, 1T0S256570, 1T0S256571, 1T0S256572, 1T0S256576, 1T0S256577, 1T0S256579, 1T0S256582, 1T0S256583, 1T0S256584, 1T0S256585, 1T0S256588, 1T0S256590, 1T0S256591, 1T0S256593, 1T0S256594, 1T0S256595, 1T0S256596, 1T0S256597, 1T0S256599, 1T0S256603, 1T0S256604, 1T0S256605, 1T0S256611, 1T0S256612, 1T0S256346, 1T0S256391, 1T0S256399, 1T0S256426, 1T0S256431, 1T0S256442, 1T0S256451, 1T0S256454, 1T0S256456, 1T0S256461, 1T0S256463, 1T0S256467, 1T0S256481, 0T0S253450, 0T0S255195, 0T0S255197, 0T0S255203, 0T0S255207, 0T0S255232, 0T0S255267, 0T0S255272, 0T0S255281, 0T0S255286, 0T0S255292, 0T0S255302, 0T0S255316, 0T0S255335, 0T0S255349, 0T0S253476, 0T0S255127, 0T0S255187, 0T0S255201, 0T0S255212, 0T0S255213, 0T0S255215, 0T0S255216, 0T0S255226, 0T0S255228, 0T0S255231, 0T0S255239, 0T0S255255, 0T0S255259, 0T0S255260, 0T0S254559, 0T0S254562, 0T0S254565, 0T0S254569, 0T0S254570, 0T0S254571, 0T0S254572, 0T0S254575, 0T0S254576, 0T0S254578, 0T0S254580, 0T0S254581, 0T0S254583, 0T0S254584, 0T0S254585, 0T0S254586, 0T0S254587, 0T0S254589, 0T0S254592, 0T0S254593, 0T0S254595, 0T0S254598, 0T0S254601, 0T0S254602, 0T0S254603, 0T0S254604, 0T0S254605, 0T0S254607, 0T0S254535, 0T0S254546, 0T0S254547, 0T0S254549, 0T0S254550, 0T0S254552, 0T0S254554, 0T0S254558, 0T0S254560, 0T0S254561, 0T0S254563, 0T0S254564, 0T0S254566, 0T0S254567, 0T0S254568, 0T0S254573, 0T0S254574, 0T0S254579, 0T0S254599, 0T0S253517, 0T0S253973, 0T0S255073, 0T0S255208, 0T0S255214, 0T0S255218, 0T0S255225, 0T0S255236, 0T0S255238, 0T0S255242, 0T0S255258, 0T0S255263, 0T0S255268, 0T0S255269, 0T0S255276, 0T0S255279, 0T0S255280, 0T0S255284, 0T0S255287, 0T0S253458, 0T0S253662, 0T0S255274, 0T0S255294, 0T0S255295, 0T0S255296, 0T0S255299, 0T0S255300, 0T0S255311, 0T0S255322, 0T0S255324, 0T0S255326, 0T0S255330, 0T0S255338, 0T0S255342, 0T0S253448, 0T0S253452, 0T0S253477, 0T0S253515, 0T0S253553, 0T0S253554, 0T0S253593, 0T0S253661, 0T0S253675, 0T0S255221, 0T0S255249, 0T0S255264, 0T0S255265, 0T0S255273, 0T0S255277, 0T0S255278, 0T0S255282, 0T0S255288, 0T0S255290, 0T0S255293, 0T0S255297, 0T0S255301, 0T0S255304, 0T0S255306, 0T0S255307, 0T0S255313, 0T0S255314, 0T0S255318, 1T0S256079, 1T0S256080, 1T0S256094, 1T0S256101, 1T0S256102, 1T0S256104, 1T0S256106, 1T0S256107, 1T0S256109, 1T0S256110, 1T0S256111, 1T0S256112, 1T0S256117, 1T0S256119, 1T0S256123, 1T0S256824, 1T0S256844, 1T0S256864, 1T0S256871, 1T0S256891, 1T0S256914, 1T0S256925, 1T0S256926, 1T0S256929, 1T0S256931, 1T0S256934, 1T0S256935, 1T0S256936, 1T0S256955, 1T0S256999, 0T0S254616, 0T0S254633, 0T0S254642, 0T0S254646, 0T0S254647, 0T0S254661, 0T0S254665, 0T0S254668, 0T0S254670, 0T0S254672, 0T0S254673, 0T0S254683, 1T0S256309, 1T0S256114, 1T0S256127, 1T0S256158, 1T0S256160, 1T0S256162, 1T0S256163, 1T0S256164, 1T0S256165, 1T0S256273, 1T0S256275, 1T0S256294, 1T0S256299, 1T0S256300, 1T0S256305, 1T0S256306, 1T0S257391, 1T0S257789, 1T0S257792, 1T0S257793, 1T0S257796, 1T0S257797, 1T0S257799, 1T0S257803, 1T0S257804, 1T0S257806, 1T0S257809, 1T0S257810, 1T0S257813, 1T0S257814, 1T0S257815, 1T0S257817, 1T0S257821, 1T0S257822, 1T0S257828, 1T0S257830, 1T0S257831, 1T0S257833, 1T0S257844, 1T0S257848, 1T0S257849, 1T0S257854, 1T0S257865, 1T0S257873, 1T0S259159, 1T0S259161, 1T0S259163, 1T0S259165, 1T0S259166, 1T0S259168, 1T0S259169, 1T0S259170, 1T0S259171, 1T0S259172, 1T0S259173, 1T0S259176, 1T0S259178, 1T0S259179, 1T0S259181, 1T0S259197, 1T0S259198, 2T0S262749, 2T0S262757, 2T0S262762, 2T0S262765, 2T0S262778, 2T0S262782, 2T0S262783, 2T0S262786, 2T0S262789, 2T0S262792, 2T0S262795, 2T0S262796, 2T0S262797, 2T0S262798, 2T0S262800, 2T0S262801, 2T0S262803, 2T0S262804, 2T0S262805, 2T0S262806, 2T0S262807, 2T0S262808, 2T0S262809, 2T0S262810, 2T0S262811, 2T0S262812, 2T0S262813, 2T0S262814, 2T0S262818, 2T0S262821, 2T0S262822, 2T0S262824, 2T0S262828, 2T0S262831, 2T0S262833, 1T0S259388, 1T0S259411, 1T0S259423, 1T0S259484, 1T0S259485, 1T0S259492, 1T0S259550, 1T0S259559, 1T0S259588, 1T0S259594, 1T0S260146, 1T0S260148, 1T0S261050, 1T0S261053, 1T0S261056, 1T0S261057, 1T0S261059, 1T0S261060, 1T0S261061, 1T0S261063, 1T0S261064, 1T0S261065, 1T0S261066, 1T0S261067, 1T0S261068, 1T0S261069, 1T0S261070, 1T0S261072, 1T0S261080, 1T0S261084, 1T0S261117, 1T0S261147, 1T0S261149, 1T0S261160, 1T0S256658, 1T0S256659, 1T0S256661, 1T0S256681, 1T0S256684, 1T0S256691, 1T0S256693, 1T0S256694, 1T0S256695, 1T0S256696, 1T0S256697, 1T0S256700, 1T0S256701, 1T0S256705, 1T0S256708, 1T0S256718, 1T0S256720, 1T0S256722, 1T0S256724, 1T0S256726, 1T0S256739, 1T0S256742, 1T0S256746, 1T0S256747, 1T0S259630, 1T0S259634, 1T0S259695, 1T0S259697, 1T0S259707, 1T0S259708, 1T0S259720, 1T0S259731, 1T0S259732, 1T0S259733, 1T0S259743, 1T0S258900, 1T0S258911, 1T0S258914, 1T0S258919, 1T0S258927, 1T0S258939, 1T0S258943, 1T0S258945, 1T0S258946, 1T0S258963, 1T0S258967, 1T0S258970, 2T0S264242, 1T0S257858, 1T0S257899, 1T0S257902, 1T0S257914, 1T0S257959, 1T0S258920, 1T0S258931, 1T0S258935, 1T0S258936, 1T0S258948, 1T0S258957, 1T0S258959, 1T0S258965, 1T0S258966, 1T0S258972, 1T0S258977, 1T0S258979, 1T0S258999, 1T0S259000, 1T0S257359, 1T0S258565, 1T0S258585, 1T0S261283, 1T0S261295, 1T0S261306, 1T0S261309, 1T0S261310, 1T0S261323, 1T0S261325, 1T0S261330, 1T0S261331, 1T0S261333, 1T0S261334, 1T0S261335, 1T0S261336, 1T0S261341, 1T0S261342, 1T0S261343, 1T0S261344, 1T0S261355, 1T0S261356, 1T0S261361, 1T0S261365, 1T0S261366, 1T0S261369, 1T0S261375, 1T0S261377, 1T0S261382, 1T0S261386, 1T0S261389, 1T0S258409, 1T0S258410, 1T0S258425, 1T0S258428, 1T0S258430, 1T0S258433, 1T0S258434, 1T0S258435, 1T0S258437, 1T0S258444, 1T0S258445, 1T0S258448, 1T0S258449, 1T0S258455, 1T0S258456, 1T0S258466, 1T0S258469, 1T0S258472, 1T0S258474, 1T0S258764, 1T0S258766, 1T0S258769, 1T0S258775, 1T0S258777, 1T0S258784, 1T0S258785, 1T0S258791, 1T0S258792, 1T0S258803, 1T0S258805, 1T0S258807, 1T0S258808, 1T0S258809, 1T0S258810, 1T0S258811, 1T0S258812, 1T0S258814, 1T0S258816, 1T0S258823, 1T0S258825, 1T0S258829, 1T0S258830, 1T0S258832, 1T0S258841, 1T0S258846, 1T0S258847, 1T0S259587, 1T0S259625, 1T0S259626, 1T0S259633, 1T0S259637, 1T0S259649, 1T0S259717, 1T0S261538, 1T0S261552, 1T0S261564, 1T0S261565, 1T0S261567, 1T0S261570, 1T0S261584, 1T0S261585, 1T0S261589, 1T0S261590, 1T0S261592, 1T0S261594, 1T0S261595, 1T0S261598, 1T0S261600, 1T0S261601, 1T0S261603, 1T0S261604, 1T0S261605, 1T0S261609, 1T0S261611, 1T0S261612, 1T0S261615, 1T0S261617, 1T0S261618, 1T0S261620, 1T0S261621, 1T0S258817, 1T0S258835, 1T0S260143, 1T0S260144, 1T0S258408, 1T0S258411, 1T0S258412, 1T0S258413, 1T0S258414, 1T0S258416, 1T0S258419, 1T0S258421, 1T0S258422, 1T0S258423, 1T0S258424, 1T0S258427, 1T0S258429, 1T0S258440, 1T0S258442, 1T0S259234, 1T0S259235, 1T0S259236, 1T0S259239, 1T0S259246, 1T0S259247, 1T0S259257, 1T0S259264, 1T0S259268, 1T0S259269, 1T0S259271, 1T0S259273, 1T0S259275, 1T0S259504, 1T0S260145, 1T0S259762, 1T0S259779, 1T0S259782, 1T0S259793, 1T0S259794, 1T0S259797, 1T0S259800, 1T0S259804, 1T0S259806, 1T0S259807, 1T0S259811, 1T0S259812, 1T0S259818, 1T0S259819, 1T0S259824, 1T0S259827, 1T0S259828, 1T0S259829, 1T0S259832, 1T0S259407, 1T0S259414, 1T0S259416, 1T0S259419, 1T0S259422, 1T0S259427, 1T0S259429, 1T0S259441, 1T0S259442, 1T0S259447, 1T0S259448, 1T0S259450, 1T0S259451, 1T0S259455, 1T0S259456, 1T0S259459, 1T0S259460, 1T0S259462, 1T0S259463, 1T0S259464, 1T0S259465, 1T0S259470, 1T0S259471, 1T0S259473, 1T0S259474, 1T0S259477, 1T0S259478, 1T0S259506, 1T0S262076, 1T0S262104, 1T0S262106, 1T0S262108, 1T0S262109, 1T0S262112, 1T0S262118, 1T0S262119, 1T0S262120, 1T0S262121, 1T0S262126, 1T0S262130, 1T0S262132, 1T0S262134, 1T0S262137, 1T0S262142, 1T0S262153, 1T0S262161, 1T0S262164, 1T0S259730, 1T0S262302, 1T0S262329, 1T0S262335, 1T0S262350, 1T0S262351, 1T0S262357, 1T0S262360, 1T0S262361, 1T0S262362, 1T0S262367, 1T0S262370, 1T0S262371, 1T0S262373, 1T0S262374, 1T0S262375, 1T0S262376, 1T0S262377, 1T0S262379, 1T0S262380, 1T0S262383, 1T0S262384, 1T0S262385, 1T0S262386, 1T0S262387, 1T0S262389, 1T0S262390, 1T0S262391, 1T0S262393, 1T0S262394, 1T0S262396, 1T0S262403, 1T0S262407, 1T0S262410, 1T0S262416, 1T0S262423, 1T0S262429, 1T0S262445, 1T0S262455, 1T0S262459, 1T0S262462, 1T0S262474, 1T0S262475, 1T0S262479, 1T0S262480, 1T0S262483, 1T0S262484, 1T0S262486, 1T0S262487, 1T0S262488, 1T0S262490, 1T0S262491, 1T0S262492, 1T0S262498, 1T0S262500, 2T0S263402, 2T0S263407, 2T0S263415, 2T0S263421, 2T0S263448, 2T0S263804, 2T0S263809, 2T0S263814, 2T0S263833, 2T0S263851, 2T0S263853, 2T0S263859, 2T0S263860, 2T0S263866, 2T0S263873, 2T0S263879, 2T0S263882, 2T0S263883, 2T0S263885, 2T0S263888, 2T0S263890, 2T0S263893, 2T0S263894, 2T0S263898, 2T0S263899, 2T0S263902, 2T0S263903, 2T0S263998, 1T0S262573, 1T0S262589, 1T0S262602, 1T0S262603, 1T0S262606, 1T0S262608, 1T0S262609, 1T0S262612, 1T0S262615, 1T0S262620, 1T0S262621, 1T0S262622, 1T0S262623, 1T0S262625, 1T0S262627, 1T0S262635, 1T0S262636, 1T0S262637, 1T0S262641, 1T0S262643, 1T0S262378, 1T0S262382, 1T0S262388, 1T0S262392, 1T0S262406, 1T0S262413, 1T0S262419, 1T0S262434, 1T0S262440, 1T0S262611, 1T0S262626, 1T0S262639, 1T0S262644, 2T0S262780, 2T0S262781, 2T0S262791, 2T0S262794, 2T0S262820, 2T0S262823, 2T0S263493, 2T0S263494, 2T0S263526, 2T0S263530, 2T0S263617, 2T0S263618, 2T0S263620, 2T0S263622, 2T0S263623, 2T0S263624, 2T0S263625, 2T0S263629, 2T0S263630, 2T0S263635, 2T0S263636, 2T0S263801, 2T0S263803, 2T0S263806, 2T0S263808, 2T0S263812, 2T0S263818, 2T0S263823, 2T0S263825, 2T0S263843, 2T0S263857, 2T0S263858, 2T0S263864, 2T0S263874, 2T0S267622, 2T0S267636, 2T0S267637, 2T0S267638, 2T0S267645, 2T0S267648, 2T0S267649, 2T0S267650, 2T0S267651, 2T0S267653, 2T0S267662, 2T0S267669, 2T0S267672, 2T0S267673, 2T0S267676, 2T0S267679, 2T0S267682, 2T0S267683, 2T0S267686, 2T0S267689, 2T0S267692, 2T0S267693, 2T0S267708, 2T0S267710, 2T0S267711, 2T0S267716, 2T0S267718, 2T0S267719, 2T0S267721, 2T0S267726, 2T0S267733, 2T0S267736, 2T0S267742, 2T0S267743, 2T0S267755, 2T0S267762, 3T0S271941, 3T0S271952, 3T0S271953, 3T0S271965, 3T0S271975, 3T0S271976, 3T0S271981, 3T0S271983, 3T0S271985, 3T0S273847, 3T0S273848, 3T0S273883, 3T0S273899, 3T0S273912, 3T0S273926, 3T0S273948, 3T0S273954, 3T0S273964, 3T0S273979, 2T0S266476, 2T0S266477, 2T0S266481, 2T0S266482, 2T0S266483, 2T0S266484, 2T0S266489, 2T0S267059, 2T0S267099, 2T0S267105, 2T0S267113, 2T0S267116, 2T0S267119, 2T0S267124, 2T0S267128, 2T0S267129, 2T0S267140, 2T0S267154, 2T0S266540, 2T0S266545, 2T0S266548, 2T0S266550, 2T0S266554, 2T0S266555, 2T0S266557, 2T0S266558, 2T0S266562, 2T0S266564, 2T0S266565, 2T0S266566, 2T0S266570, 2T0S266571, 2T0S266572, 2T0S266576, 2T0S266580, 2T0S266582, 2T0S266584, 2T0S266586, 2T0S266587, 2T0S266588, 2T0S266589, 2T0S266591, 2T0S266592, 2T0S266593, 2T0S266594, 2T0S266596, 2T0S266598, 2T0S266603, 2T0S266612, 2T0S266617, 2T0S266620, 2T0S266632, 4T0S276514, 4T0S276527, 4T0S276530, 4T0S276564, 4T0S276572, 4T0S276593, 4T0S276615, 4T0S276622, 2T0S267732, 2T0S267750, 3T0S270100, 3T0S270395, 3T0S270401, 3T0S270410, 3T0S270419, 3T0S270434, 3T0S270441, 3T0S270443, 3T0S270447, 3T0S270449, 3T0S270451, 3T0S270452, 3T0S270454, 3T0S270455, 3T0S270459, 3T0S270460, 3T0S270464, 3T0S270465, 3T0S270467, 3T0S270470, 3T0S270480, 3T0S270487, 3T0S270489, 3T0S270491, 3T0S270492, 3T0S270493, 3T0S270495, 3T0S270503, 2T0S267657, 2T0S267687, 2T0S267717, 2T0S267766, 2T0S267771, 2T0S267773, 2T0S267774, 2T0S267775, 2T0S267777, 2T0S267781, 2T0S267783, 2T0S267785, 2T0S267787, 2T0S267788, 2T0S267789, 2T0S267793, 2T0S267796, 2T0S267801, 2T0S267858, 3T0S269829, 3T0S269830, 3T0S269835, 3T0S269843, 3T0S271589, 3T0S271594, 3T0S271596, 3T0S271603, 3T0S271611, 3T0S271617, 3T0S271626, 3T0S271627, 3T0S271629, 3T0S271632, 3T0S271636, 3T0S271638, 3T0S271639, 3T0S271644, 3T0S271648, 3T0S271653, 3T0S271655, 3T0S271657, 3T0S271660, 3T0S271663, 3T0S271664, 3T0S271667, 3T0S271673, 3T0S271674, 3T0S271678, 3T0S271702, 3T0S271712, 3T0S271792, 3T0S271833, 3T0S271836, 3T0S271841, 3T0S271846, 3T0S271851, 3T0S271881, 3T0S271935, 3T0S271621, 3T0S271661, 3T0S271692, 3T0S271700, 3T0S271722, 3T0S271726, 3T0S271728, 3T0S271730, 3T0S271731, 3T0S271733, 3T0S271737, 3T0S271738, 3T0S271744, 3T0S271751, 3T0S271754, 3T0S271755, 3T0S271756, 3T0S271759, 3T0S271770, 3T0S271773, 3T0S271779, 3T0S271784, 3T0S271789, 3T0S271791, 3T0S271793, 3T0S271805, 3T0S271810, 3T0S275212, 3T0S275220, 3T0S275231, 3T0S275235, 3T0S275252, 3T0S275255, 3T0S275272, 3T0S275274, 3T0S275276, 3T0S270660, 3T0S270685, 3T0S270692, 3T0S270695, 3T0S270697, 3T0S270699, 3T0S270701, 3T0S270703, 3T0S270704, 3T0S270710, 3T0S270711, 3T0S270712, 3T0S270713, 3T0S270717, 3T0S270825, 3T0S270827, 3T0S270828, 3T0S270832, 3T0S270840, 3T0S270841, 3T0S271180, 3T0S271707, 3T0S271725, 3T0S271749, 3T0S271760, 3T0S271761, 3T0S271799, 3T0S271800, 3T0S271815, 3T0S271862, 3T0S271880, 3T0S271885, 3T0S271896, 3T0S271902, 3T0S271988, 3T0S273852, 3T0S273921, 3T0S273947, 3T0S273950, 3T0S273951, 3T0S273961, 2T0S264443, 2T0S264514, 2T0S264573, 2T0S264575, 2T0S264580, 2T0S264756, 2T0S264894, 2T0S264909, 2T0S264941, 2T0S264964, 2T0S265057, 3T0S269674, 3T0S269686, 3T0S274127, 3T0S274132, 3T0S274137, 3T0S274993, 4T0S276902, 4T0S276957, 4T0S276966, 4T0S276968, 4T0S276983, 4T0S277045, 4T0S277051, 4T0S277059, 4T0S277068, 4T0S277071, 4T0S277072, 4T0S277078, 4T0S277080, 4T0S277081, 4T0S277082, 4T0S277089, 4T0S277093, 4T0S277096, 4T0S277100, 4T0S277103, 4T0S277104, 4T0S277106, 4T0S277108, 4T0S277109, 4T0S277110, 4T0S277112, 4T0S277114, 4T0S277121, 4T0S275896, 4T0S275944, 4T0S277647, 4T0S277726, 4T0S277729, 4T0S277504, 4T0S277562, 4T0S277563, 4T0S277564, 4T0S277585, 4T0S277588, 4T0S277592, 4T0S277595, 4T0S277598, 4T0S277601, 4T0S277603, 4T0S277604, 4T0S277610, 4T0S277613, 4T0S277614, 4T0S277620, 4T0S277622, 4T0S277623, 4T0S277641, 4T0S277659, 4T0S277668, 4T0S277670, 4T0S277677, 4T0S277678, 4T0S277681, 4T0S277683, 4T0S277684, 4T0S277688, 4T0S277691, 4T0S277692, 4T0S277694, 4T0S277695, 4T0S277700, 4T0S277701, 4T0S277703, 4T0S277704, 4T0S277708, 4T0S277710, 4T0S277728, 4T0S277750.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- 2008T BlueStar Hemodialysis Machine 191126Fresenius Medical Care Holdings, Inc. · Dialyzer, high permeability with or without sealed dialysate system5 recalls10 deaths441 reports
Product lines
Other devices under the same FDA submission
2 devices cleared under K173972, K222952, K231125 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- 2008T BlueStar Hemodialysis Machine 191130Fresenius Medical Care Holdings, Inc. · Dialyzer, high permeability with or without sealed dialysate system4 recalls7 deaths272 reports
- 2008T BlueStar Hemodialysis Machine 191124Fresenius Medical Care Holdings, Inc. · Dialyzer, high permeability with or without sealed dialysate system4 recalls32 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 94463).
- Z-1867-2024 · Open, Classified2008T BlueStar Hemodialysis Machine, Model 191124 with Component: Blood Pump Rotor, Model Number: F40015481 Rev A.
- Z-1869-2024 · Open, Classified2008T BlueStar Hemodialysis Machine, Model 191128 with Component: Blood Pump Rotor, Model Number: F40015481 Rev A.
- Z-1870-2024 · Open, Classified2008T BlueStar Hemodialysis Machine, Model 191130 with Component: Blood Pump Rotor, Model Number: F40015481 Rev A
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page