ONE-CUF recallZ-2027-2024 · GE Medical Systems Information Technologies Inc · initiated Apr 9, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from GE Medical Systems Information Technologies Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Use of the blood pressure cuffs could result in inaccurate non-invasive blood pressure measurements.
FDA root cause category: Under Investigation by firm
What the company is doing
Recall letters dated and issued on 4/9/2024 by the recalling firm explained the safety issue and actions to be taken by the customer/user. Those actions were to not use the affected products listed in the Affected Product Details section of the letter, use an alternate cuff for NIBP monitoring, and destroy all affected product per the customer facility procedures. A photograph of the product was embedded in the letter to show where to find the product name, model number, and GTIN number. The letter informs the consignee that GE HealthCare will replace the affected products with CRITIKON SOFT-CUF Blood Pressure Cuffs and a GE HealthCare representative will contact the consignee to arrange for this correction. An Acknowledgement Form was enclosed for completion and return.
Recall details
- Recall number
- Z-2027-2024
- Status
- Open, Classified
- Initiated by the company
- Apr 9, 2024
- Posted by the FDA
- Jun 6, 2024
- Recalling firm
- GE Medical Systems Information Technologies Inc
- Quantity
- 23,180/20-cuff boxes
- Product code
- DXQ
- FDA submissions
- K230145
- Distribution
- US Nationwide distribution in the states of AL, AR, AZ, CA, FL, GA, IA, IL, IN, LA, MD, MO, MS, NC, NJ, NM, OH, PA, TN, and TX.
Product description
GE HealthCare CRITIKON ONE-CUF blood pressure cuffs, Model numbers: 1. ONE-A1-1B, ONE-CUF, SMALL ADULT, 1 TB BAYONET, 17 - 25 CM, 20/BOX; 2. ONE-A1-2A, ONE-CUF, SMALL ADULT, DINACLICK 80369-5, 17 - 25 CM, 20/BOX; 3. ONE-A2-1B, ONE-CUF, ADULT, 1 TB BAYONET, 23 - 33 CM, 20/BOX; 4. ONE-A1-1B-L, ONE-CUF, ADULT LONG, 1 TB BAYONET, 23 - 33 CM, 20/BOX; 5. ONEA2-2A, ONE-CUF, ADULT, DINACLICK 80369-5, 23 - 33 CM, 20/BOX; 6. ONE-A2-2A-L, ONE-CUF, ADULT LONG, DINACLICK 80369-5, 23 - 33 CM, 20/BOX; 7. ONE-A3-1B, ONE-CUF, LARGE ADULT, 1 TB BAYONET, 31 - 40 CM, 20/BOX; 8. ONE-A3-2A, ONE-CUF, LARGE ADULT, DINACLICK 80369-5, 31 - 40 CM, 20/BOX; 9. ONE-P1-1B, ONE-CUF, INFANT, 1 TB BAYONET, 08 - 13 CM, 20/BOX; 10. ONE-P1-2A, ONE-CUF, INFANT, DINACLICK 80369-5, 08 - 13 CM, , 20/BOX; 11. ONE-P2-1B, ONE-CUF, CHILD, 1 TB BAYONET, 12 - 19 CM, 20/BOX; and 12. ONE-P2-2A, ONE-CUF, CHILD, DINACLICK 80369-5, 12 - 19 CM, 20/BOX.
Code information (models, lots, serial numbers)
No serial numbers are on the blood pressure cuffs. GTIN NUMBERS: ONE-A1-1B - GTIN 00195278573803; ONE-A1-2A - GTIN 00195278576460; ONE-A2-1B - GTIN 00195278573797; ONE-A1-1B-L - GTIN 00195278573780; ONEA2-2A - GTIN 00195278576477; ONE-A2-2A-L GTIN 00195278576484; ONE-A3-1B - GTIN 00195278573810; ONE-A3-2A - GTIN 00195278576491; ONE-P1-1B - GTIN 00195278573834; ONE-P1-2A - GTIN 00195278576507; ONE-P2-1B - GTIN 00195278573827; and ONE-P2-2A - GTIN 00195278576514.
Devices named in this recall
12 devices in the FDA device registry whose identifier appears in the recall's code information.
- ONE-CUF ONE-A2-2AGE Medical Systems Information Technologies, Inc. · Blood Pressure Cuff1 recall1 report
- ONE-CUF ONE-A1-1BGE Medical Systems Information Technologies, Inc. · Blood Pressure Cuff1 recall0 reports
- ONE-CUF ONE-P1-1BGE Medical Systems Information Technologies, Inc. · Blood Pressure Cuff1 recall0 reports
- ONE-CUF ONE-A2-2A-LGE Medical Systems Information Technologies, Inc. · Blood Pressure Cuff1 recall0 reports
- ONE-CUF ONE-P2-1BGE Medical Systems Information Technologies, Inc. · Blood Pressure Cuff1 recall0 reports
- ONE-CUF ONE-A1-2AGE Medical Systems Information Technologies, Inc. · Blood Pressure Cuff1 recall0 reports
- ONE-CUF ONE-P1-2AGE Medical Systems Information Technologies, Inc. · Blood Pressure Cuff1 recall0 reports
- ONE-CUF ONE-A2-1BGE Medical Systems Information Technologies, Inc. · Blood Pressure Cuff1 recall0 reports
- ONE-CUF ONE-P2-2AGE Medical Systems Information Technologies, Inc. · Blood Pressure Cuff1 recall0 reports
- ONE-CUF ONE-A2-1B-LGE Medical Systems Information Technologies, Inc. · Blood Pressure Cuff1 recall0 reports
- ONE-CUF ONE-A3-2AGE Medical Systems Information Technologies, Inc. · Blood Pressure Cuff1 recall0 reports
- ONE-CUF ONE-A3-1BGE Medical Systems Information Technologies, Inc. · Blood Pressure Cuff1 recall0 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page