ADVANTA SLIDER GDS recallZ-1956-2024 · Atrium Medical Corporation · initiated May 3, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Atrium Medical Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Complaints of reported separation of the Slider GDS Swivel Rod from the Swivel Core and of a notable gap between the two pieces.
FDA root cause category: Equipment maintenance
What the company is doing
Consignees were sent an URGENT MEDICAL DEVICE- REMOVAL notification by mail, dated 5/3/24. Consignees are instructed to examine their inventory immediately to determine if any recalled product is in their possession. Recalled product possesses the REF number on device labeling and has an expiration date prior to 24FEB2027. All consignees are asked to complete and return the provided response form by email to [email protected] or by fax to 1-866-560-4062. Consignees that are distributors are asked to forward the provided notification to their customers. Consignees are to post a copy of the recall notification in all inventory locations where affected devices are stored and ensure all appropriate personnel are aware. If consignees possess recalled devices, they are to contact their local Atrium/Getinge Customer Support Department at 888-943-8872 (option 2) to receive instructions for returning any affected devices. The recalling firm will provide replacement product within 5-8 weeks of return of devices. The recalling firm states that if there is a medical need to use the device and no alterative device is available, then the graft may be used per the provided instructions. Customers can contact Customer support with any questions at 888-943-8872 (option 2) from Monday through Friday, between 8:00 AM and 6:00 PM.
Recall details
- Recall number
- Z-1956-2024
- Status
- Open, Classified
- Initiated by the company
- May 3, 2024
- Posted by the FDA
- Jun 4, 2024
- Recalling firm
- Atrium Medical Corporation
- Quantity
- 53,308 units (US: 11,236 ; OUS: 42,072)
- Product code
- DSY
- FDA submissions
- K992960
- Distribution
- Worldwide distribution - US Nationwide including Puerto Rico, and the countries of AE, AR, AT, AU, BE, BG, BO, BR, CA, CH, CL, CN, CO, CY, CZ, DE, DZ, EC, EE, EG, ES, FI, FR, GB, GR, HK, ID, IE, IL, IQ, IT, JO, JP, KR, KW, LT, LU, LV, ME, MN, MU, MX, MY, NI, NL, NO, NZ, OM, PA, PE, PH, PK, PL, PT, PY, RO, RU, SA, SE, SG, SI, SV, TH, TN, TR, TW, UA, VN, XK, YE, ZA.
Product description
ADVANTA VXT, 7X50, 1GDS, NH, STR-SW. Single-Ended Slider. Vascular graft
Code information (models, lots, serial numbers)
Product Code: 22018; UDI-DI: 00650862220186.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- ADVANTA SLIDER GDS 22018Atrium Medical Corporation · PROSTHESIS, VASCULAR GRAFT, OF 6MM AND GREATER DIAMETER1 recall0 reports
Product lines
Other devices under the same FDA submission
1 device cleared under K992960 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- ADVANTA 21000Atrium Medical Corporation · PROSTHESIS, VASCULAR GRAFT, OF 6MM AND GREATER DIAMETER1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 94614).
- Z-1951-2024 · Open, ClassifiedADVANTA VXT, 5X40, 1GDS, NH, STR-SW. Single-Ended Slider. Vascular graft
- Z-1952-2024 · Open, ClassifiedADVANTA VXT, 6X40, 1GDS, NH, STR-SW. Single-Ended Slider. Vascular graft
- Z-1953-2024 · Open, ClassifiedADVANTA VXT, 8X40, 1GDS, NH, STR-SW. Single-Ended Slider. Vascular graft
- Z-1954-2024 · Open, ClassifiedADVANTA VXT, 5X50, 1GDS, NH, STR-SW. Single-Ended Slider. Vascular graft
- Z-1955-2024 · Open, ClassifiedADVANTA VXT, 6X50, 1GDS, NH, STR-SW. Single-Ended Slider. Vascular graft
- Z-1957-2024 · Open, ClassifiedADVANTA VXT, 8X50, 1GDS, NH, STR-SW. Single-Ended Slider. Vascular graft
- Z-1958-2024 · Open, ClassifiedADVANTA VXT, 5X70, 1GDS, NH, STR-SW. Single-Ended Slider. Vascular graft
- Z-1959-2024 · Open, ClassifiedADVANTA VXT, 6X70, 1GDS, NH, STR-SW. Single-Ended Slider. Vascular graft
- Z-1960-2024 · Open, ClassifiedADVANTA VXT, 7X70, 1GDS, NH, STR-SW. Single-Ended Slider. Vascular graft
- Z-1961-2024 · Open, ClassifiedADVANTA VXT, 8X70, 1GDS, NH, STR-SW. Single-Ended Slider. Vascular graft
- Z-1962-2024 · Open, ClassifiedADVANTA VXT, 5X40, 1GDS, FH, STR-SW. Single-Ended Slider GDS. Vascular graft
- Z-1963-2024 · Open, ClassifiedADVANTA VXT, 6X40, 1GDS, FH, STR-SW. Single-Ended Slider GDS. Vascular graft
- Z-1964-2024 · Open, ClassifiedADVANTA VXT, 5X50, 1GDS, FH, STR-SW. Single-Ended Slider GDS. Vascular graft
- Z-1965-2024 · Open, ClassifiedADVANTA VXT, 6X50, 1GDS, FH, STR-SW. Single-Ended Slider GDS. Vascular graft
- Z-1966-2024 · Open, ClassifiedADVANTA VXT, 7X50, 1GDS, FH, STR-SW. Single-Ended Slider GDS. Vascular graft
- Z-1967-2024 · Open, ClassifiedADVANTA VXT, 8X50, 1GDS, FH, STR-SW. Single-Ended Slider GDS. Vascular graft
- Z-1968-2024 · Open, ClassifiedADVANTA VXT, 6X70, 1GDS, FH, STR-SW. Single-Ended Slider GDS. Vascular graft
- Z-1969-2024 · Open, ClassifiedADVANTA VXT, 7X70, 1GDS, FH, STR-SW. Single-Ended Slider GDS. Vascular graft
- Z-1970-2024 · Open, ClassifiedADVANTA VXT, 8X70, 1GDS, FH, STR-SW. Single-Ended Slider GDS. Vascular graft
- Z-1971-2024 · Open, ClassifiedADVANTA VXT, 6X80, 1GDS, FH, STR-SW. Single-Ended Slider GDS. Vascular graft
- Z-1972-2024 · Open, ClassifiedADVANTA VXT, 7X80, 1GDS, FH, STR-SW. Single-Ended Slider GDS. Vascular graft
- Z-1973-2024 · Open, ClassifiedADVANTA VXT, 6X40, 1GDS, NH, STR-TW. Single-Ended Slider GDS. Vascular graft
- Z-1974-2024 · Open, ClassifiedADVANTA VXT, 8X40, 1GDS, NH, STR-TW. Single-Ended Slider GDS. Vascular graft
- Z-1975-2024 · Open, ClassifiedADVANTA VXT, 6X50, 1GDS, NH, STR-TW. Single-Ended Slider GDS. Vascular graft
- Z-1976-2024 · Open, ClassifiedADVANTA VXT, 7X50, 1GDS, NH, STR-TW. Single-Ended Slider GDS. Vascular graft
- Z-1977-2024 · Open, ClassifiedADVANTA VXT, 6X70, 1GDS, NH, STR-TW. Single-Ended Slider GDS. Vascular graft
- Z-1978-2024 · Open, ClassifiedADVANTA VXT, 7X70, 1GDS, NH, STR-TW. Single-Ended Slider GDS. Vascular graft
- Z-1979-2024 · Open, ClassifiedADVANTA VXT, 8X70, 1GDS, NH, STR-TW. Single-Ended Slider GDS. Vascular graft
- Z-1980-2024 · Open, ClassifiedADVANTA VXT, 6X80, 1GDS, NH, STR-TW. Single-Ended Slider GDS. Vascular graft
- Z-1981-2024 · Open, ClassifiedADVANTA VXT, 8X80, 1GDS, NH, STR-TW. Single-Ended Slider GDS. Vascular graft
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page