BD Pyxis: MedStation ES (Main), REF: 323 recallZ-2374-2024 · CareFusion 303, Inc. · initiated May 2, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from CareFusion 303, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
If automated dispensing cabinets have specific software versions, and Component Manager was configured to "installed mode", and software patch KB 5033688 was installed in the production environment, then this will will lead to an error that prevents the user from accessing the dispensing software application, which may delay accessing medication.
FDA root cause category: Software change control
What the company is doing
On 6/6/24, correction notices were distributed to customers who were informed that their devices had been remediated. That they had been contacted by phone in May 2024 to schedule device remediation. The firm re-configured the component manager from "installed mode" to "disabled mode" and checked if the Microsoft patch was present on each device. If the patch was identified on a device, it was removed. All devices were then configured to "managed mode" and rebooted to enable the system appropriately. Customers are asked to complete and return the response form to Email [email protected] For safety information about the system not covered in this document, call the firm's Technical Support Center at 1-800-727-6102, Phone hours: 24 hours, 7 days a week. Additional contacts include: 1) Regional Complaint Center, 1-844-823-5433 Say "Product Complaints" when prompted Mon-Fri 8:00am and 5:00pm CT, Email: [email protected] 2) Post Market Quality Field Action Operations, [email protected]
Recall details
- Recall number
- Z-2374-2024
- Status
- Open, Classified
- Initiated by the company
- May 2, 2024
- Posted by the FDA
- Jul 17, 2024
- Recalling firm
- CareFusion 303, Inc.
- Quantity
- 326
- Product code
- BRY
- Distribution
- Worldwide - US Nationwide distribution including in the states of OR, KS, NE, LA, ND, OK, AL, AK, WI, WV, WA, NV, PR, TX, CO, OH, CA, MO, IA, AZ, IN, AR, IL, MN, HI, NM, FL, SD, UT, NY, VA, GA, MA, NC and the countries of TW, CA.
Product description
BD Pyxis: MedStation ES (Main), REF: 323; MedStation ES Tower, REF: 352; Anesthesia Station ES, REF: 327; CII Safe ES, REF: 1116-00; CII Safe ES, Desktop PC, Wired Scanner, REF: 107-254-01; CII Safe ES, Desktop PC, Wireless Scanner, REF: 107-255-01; CII Safe ES, Desktop PC, Wired Scanner 1.7.3, REF: 107-256-01; CII Safe ES, Desktop PC, Wireless Scanner 1.7.3, REF: 107-257-01
Code information (models, lots, serial numbers)
Show the full code information (3k characters)
REF/UDI-DI: 323/10885403512667, 352/10885403512674, 327/10885403477836, 1116-00/10885403512605 Software Version: ES 1.7.x Serial Numbers: 13534745, 16220738, 16446299, 16446303, 16446302, 16247443, 16247440, 15962990, 16417275, 16446297, 16446296, 16242783, 16234266, 15425093, 16446295, 16446291, 16242784, 16417277, 15644176, 14019332, 16212329, 16221517, 16395857, 16059358, 16254191, 15319466, 15154945, 15153602, 16215325, 15154942, 14959051, 16395855, 16395853, 16215313, 16361893, 14826969, 14826972, 14826971, 14826973, 14232856, 16241342, 13877154, 15148881, 15148860, 15148880, 15148882, 15854601, 15148883, 15148865, 15148885, 15160417, 16212961, 15854595, 15148859, 15148862, 15148861, 15148863, 15148866, 15148868, 15854596, 15854600, 15148870, 15854597, 15854598, 15854599, 15148873, 15148864, 15148867, 15148884,15148878, 15148879, 16324054, 16223557, 15136666, 13841400, 14138516, 15519749, 16446315, 16141586, 16416617, 16416618, 16416622, 16416619, 16416621, 16416623, 16416616, 16416615, 13781497, 14557551, 16212312, 15500549, 14700155, 16212346, 16214431, 16214432, 16214430, 16214429, 16212344, 16247441, 16247444, 16320157, 16335215, 16204472, 15154938, 14813899, 14697245, 15642731, 16209866, 15135154, 16214435, 16446294, 16227849, 16446311, 16010961, 16375509, 15303370, 16415510, 14668921, 16239128, 16346806, 16242800, 16204479, 16352591, 16324052, 16222213, 16212322, 16271477, 15015227, 16214427, 16373774, 15978927, 16358533, 16215391, 16446313, 16446308, 16247439, 16247445, 16247442, 16347745, 16347746, 14157588, 16271458, 16346807, 15196235, 16322403, 16352593, 14728645, 16212338, 15519703, 16217225, 14806556, 16352598, 16352590, 16352595, 16352596, 15263493, 16216425, 16416639, 16416102, 16416104, 16416632, 16416108, 16416637, 16416110, 16416111, 16416106, 16416638, 14987500, 16234192, 16405456, 16405459, 15136661, 16242578, 16271457, 15911127, 15954728, 16271463, 16271465, 14966741, 16214437, 16324055, 16225327, 16407028, 14613489, 15997584, 14442210, 16357615, 16357616, 16307328, 13839287, 13839299, 15136665, 15045814, 16054085, 15642724, 16244068, 16346335, 16283945, 16231509, 16277495, 16240201, 16211337, 16291566, 16212335, 15933030, 16242781, 16346808, 15135153, 16204480, 16271460, 16212330, 16324056, 16215395, 16214434, 15454500, 16173369, 16423187, 15895362, 16242797, 16204482, 16274624, 16223058, 16223059, 16274625, 15540317, 16223060, 16274626, 16257759, 16352588, 16446301, 12345678, 16271461, 16352594, 16212342, 16120462, 16235542, 16242799, 16212336, 16468210, 16195795, 16211170, 15570657, 16223556, 16412617, 16412618, 16412615, 16412616, 16321316, 16022041, 16242801, 16267770, 15136667, 15136668, 16174009, 14956271, 15236068, 15236067, 16218170, 16061265, 14572826, 16128483, 16128482, 16143915, 15398129, 15449307, 16174023, 16174021, 16143913, 16174036, 16174008, 16174019, 16174020, 15449302, 15691588, 15449303, 16227720, 16446290, 16242780, 16242802, 16405462, 16116814, 16212339, 16008937, 16215393, 16215389, 16215398, 16223558, 15147150, 16342242, 16253003, 16253008, 15495195, 16149377, 16094701, 15560137, 16296089, 16296090, 16296091, 16296092, 13347728, 14826153, 16417216, 16417243, 16417242, 16324044, 16415642, 16419185, 16415641, 16415639, 16415640, 16242803, 16242804, 16212319, 15136664, 16262416, 15136663, 16317160, 16212343, 15136670
Devices named in this recall
3 devices in the FDA device registry whose identifier appears in the recall's code information.
- BD Pyxis 323Carefusion 303, Inc. · CABINET, TABLE AND TRAY, ANESTHESIA14 recalls2 deaths24,375 reports
- Pyxis 327Carefusion 303, Inc. · CABINET, TABLE AND TRAY, ANESTHESIA10 recalls3,137 reports
- BD Pyxis 352Carefusion 303, Inc. · CABINET, TABLE AND TRAY, ANESTHESIA12 recalls898 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.