Software version 5.3 in recallZ-2214-2024 · Siemens Healthcare Diagnostics Inc · initiated May 17, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Siemens Healthcare Diagnostics Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Siemens Healthineers has identified an issue with the RAPIDPoint500e system software version 5.3. This issue affects how the sample source is identified at the LIS when using the capillary mode. With this software version, samples run in the capillary mode are incorrectly labeled as arterial when displayed on the LIS, leading to the potential for results to be flagged according to an established arterial range. Capillary sample results are correctly identified as capillary on both the RAPIDPoint 500e system display and the instrument printouts. The issue only affects the new software version 5.3 and only capillary sample mode is impacted. The inability to distinguish specimen sources for blood gas may lead to incorrect diagnoses and management decisions.
FDA root cause category: Software Design Change
What the company is doing
Siemens Healthineers issued Urgent Medical Device Correction Letter (POC 24-015.A.US) on 5/1/7/24 to US customers via Fed'x. An Urgent Field Safety Notice (UFSN) POC 24-015.A.OUS was issued to all OUS Siemens reporting sites with affected customers for distribution and regional reporting to their local competent authorities. Letter states reason for recall, health risk and action to take: " Please review this letter with your Medical Director to determine the appropriate course of action. " Siemens will convert all RAPIDPoint 500e systems at software v5.3 to v5.2 regardless of whether the capillary sample mode is used. The software conversion will be performed by Siemens Service group. Please contact Siemens customer support for the system software conversion. " Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days. " Adverse reactions or quality problems experienced with the use of the affected products listed in the above table may be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail or by fax. " Please retain this letter with your laboratory records and forward this letter to those who may have received this product. " For any questions, you may call our 24/7 Customer Service USA line at 877-229-3711.
Recall details
- Recall number
- Z-2214-2024
- Status
- Open, Classified
- Initiated by the company
- May 17, 2024
- Posted by the FDA
- Jun 26, 2024
- Recalling firm
- Siemens Healthcare Diagnostics Inc
- Quantity
- 9 units US
- Product code
- CHL
- FDA submissions
- K192240
- Distribution
- Worldwide - US Nationwide distribution in the states of FL, GA, LA, MI, NC and the countries of Chile, Costa Rica, France, Germany, Japan, Lebanon, Mexico, Netherlands, New Zealand, Poland, Spain, Turkey, United Kingdom.
Product description
Software version 5.3 in: RAPIDPoint 500e Blood Gas System (USA) Siemens Material Number: 11416751 RAPIDPoint 500e Blood Gas System (China) 11416752; RAPIDPoint 500e Blood Gas System (Japan) 11416754; RAPIDPoint 500e Blood Gas System (ROW) 11416755 ;
Code information (models, lots, serial numbers)
UDI-DI: (1)RAPIDPoint 500e Blood Gas System (USA) 00630414286143; (2)RAPIDPoint 500e Blood Gas System (China) 00630414286150; (3) RAPIDPoint 500e Blood Gas System (Japan) 00630414286167; (4) RAPIDPoint 500e Blood Gas System (ROW) 00630414286174 Software Version 5.3.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- RAPIDPoint® 500e Blood Gas System 11416751Siemens Healthcare Diagnostics Inc. · ELECTRODE, ION SPECIFIC, CALCIUM2 recalls7 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page