INSTRUMENT recallZ-2153-2024 · Howmedica Osteonics Corp. · initiated May 23, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Howmedica Osteonics Corp. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The design of the HRIS Storage & Sterilization Case and Tray P/N 6210-9-900 and the Gray Revision Instrument System Accessory Tray/Case P/N 6210-9-910 may lead to misuse of the Gray Revision Instruments including re-use and re-sterilization of the single-use, sterile-packaged instruments.

FDA root cause category: Device Design

What the company is doing

On May 23, 2024 MEDICAL DEVICE RECALL letters were sent to customers. Actions needed: Our records indicate that you may have received the affected product(s). It is Stryker s responsibility as the manufacturer to ensure that customers who may have received these affected products also receive this important communication. We therefore request that you read this notice carefully and complete the following actions. 1. Please inform users of this Medical Device Recall and forward this notice to all individuals who need to be made aware. 2. Immediately check all stock areas and/or operating room storage at your facility to determine if any Gray Revision Instruments from the affected product list in Table 2 have been removed from sterile packaging (Gray Revision Instruments are single-use, sterile-packaged manual surgical instruments). 3. If any Gray Revision Instruments have been removed from sterile packaging, discard/dispose of those instruments. 4. Gray Revision Instruments found within their sterile packaging may be used as intended (single-use, sterile-packaged). 5. Do not use the HRIS Storage & Sterilization Case and Tray or the Gray Revision Instrument System Accessory Tray/Case to re-process (re-sterilize, clean, or re-use) the Gray Revision Instruments, which are single-use, sterile-packaged manual surgical instruments. 6. You may continue to use the HRIS Storage & Sterilization Case and Tray or the Gray Revision Instrument System Accessory Tray/Case for all other instruments as intended. 7. Complete and sign the enclosed Medical Device Recall Business Reply Form and email to [email protected]/ fax (844) 412-9273. 8. Please contact your Local Sales Office or your Stryker Sales Representative directly for product replacement and inventory questions. Please assist us in meeting our regulatory obligation by emailing back the attached Medical Device Recall Business Reply Form within 5 days. A response is required, even though you may not have any physic

Recall details

Recall number
Z-2153-2024
Status
Open, Classified
Initiated by the company
May 23, 2024
Posted by the FDA
Jun 18, 2024
Recalling firm
Howmedica Osteonics Corp.
Quantity
209 units
Product code
LXH
Distribution
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Brazil, Canada, Chile, Denmark, Hong Kong, Iran, Italy, Japan, Kuwait, Mexico, New Zealand, Singapore, Spain, Sweden, Switzerland, Thailand, Turkey, UK, Venlo.

Product description

Gray Revision Instrument System Accessory Tray/Case, Catalog/Part Number: 6210-9-910

Code information (models, lots, serial numbers)

Part 6210-9-910, UDI-DI: 07613327217780, Lot#: All lots back to 2006.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 94708).