INSTRUMENT recallZ-2153-2024 · Howmedica Osteonics Corp. · initiated May 23, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Howmedica Osteonics Corp. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The design of the HRIS Storage & Sterilization Case and Tray P/N 6210-9-900 and the Gray Revision Instrument System Accessory Tray/Case P/N 6210-9-910 may lead to misuse of the Gray Revision Instruments including re-use and re-sterilization of the single-use, sterile-packaged instruments.
FDA root cause category: Device Design
What the company is doing
On May 23, 2024 MEDICAL DEVICE RECALL letters were sent to customers. Actions needed: Our records indicate that you may have received the affected product(s). It is Stryker s responsibility as the manufacturer to ensure that customers who may have received these affected products also receive this important communication. We therefore request that you read this notice carefully and complete the following actions. 1. Please inform users of this Medical Device Recall and forward this notice to all individuals who need to be made aware. 2. Immediately check all stock areas and/or operating room storage at your facility to determine if any Gray Revision Instruments from the affected product list in Table 2 have been removed from sterile packaging (Gray Revision Instruments are single-use, sterile-packaged manual surgical instruments). 3. If any Gray Revision Instruments have been removed from sterile packaging, discard/dispose of those instruments. 4. Gray Revision Instruments found within their sterile packaging may be used as intended (single-use, sterile-packaged). 5. Do not use the HRIS Storage & Sterilization Case and Tray or the Gray Revision Instrument System Accessory Tray/Case to re-process (re-sterilize, clean, or re-use) the Gray Revision Instruments, which are single-use, sterile-packaged manual surgical instruments. 6. You may continue to use the HRIS Storage & Sterilization Case and Tray or the Gray Revision Instrument System Accessory Tray/Case for all other instruments as intended. 7. Complete and sign the enclosed Medical Device Recall Business Reply Form and email to [email protected]/ fax (844) 412-9273. 8. Please contact your Local Sales Office or your Stryker Sales Representative directly for product replacement and inventory questions. Please assist us in meeting our regulatory obligation by emailing back the attached Medical Device Recall Business Reply Form within 5 days. A response is required, even though you may not have any physic
Recall details
- Recall number
- Z-2153-2024
- Status
- Open, Classified
- Initiated by the company
- May 23, 2024
- Posted by the FDA
- Jun 18, 2024
- Recalling firm
- Howmedica Osteonics Corp.
- Quantity
- 209 units
- Product code
- LXH
- Distribution
- Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Brazil, Canada, Chile, Denmark, Hong Kong, Iran, Italy, Japan, Kuwait, Mexico, New Zealand, Singapore, Spain, Sweden, Switzerland, Thailand, Turkey, UK, Venlo.
Product description
Gray Revision Instrument System Accessory Tray/Case, Catalog/Part Number: 6210-9-910
Code information (models, lots, serial numbers)
Part 6210-9-910, UDI-DI: 07613327217780, Lot#: All lots back to 2006.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- INSTRUMENT 6210-9-910Howmedica Osteonics Corp. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1 recall0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 94708).
- Z-2152-2024 · Open, ClassifiedHRIS Storage & Sterilization Case and Tray, Catalog/Part Number: 6210-9-900
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page