Centurion procedure kits labeled as NEURO ANGIO PACK, Pack… recallZ-2515-2024 · MEDLINE INDUSTRIES, LP - Northfield · initiated Apr 8, 2024
Class II Open, Classified All recalls posted in August 2024
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from MEDLINE INDUSTRIES, LP - Northfield or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health.
FDA root cause category: Under Investigation by firm
What the company is doing
Medline Industries LP issued a recall notice to its consignees on 04/08/2024 via letter delivered via USPS and email. The notice explained the issue, provided a link to for information pertaining to the risk involved https://www.fda.gov/medical-devices/safety-communications/update-evaluating-plastic-syringes-made-china-potential-device-failures-fda-safety-communication) and request the the syringes within the convenience be discarded, except for circumstances where no alternative syringes are readily available. If these syringes must be used, closely monitor for leaks, breakage, and any other problems, and report any quality issues to Medline and the FDA. Distributors were directed to notify their customers. The firm issued a second recall notice (an expansion) on 05/02/2024 using the same methods. Link for the posting: https://www.medline.com/media/mkt/pdf/Shenli%20Syringe%20Kitting%20Recall%20info.pdf
Recall details
- Recall number
- Z-2515-2024
- Recall class
- Class II
- Status
- Open, Classified
- Initiated by the company
- Apr 8, 2024
- Posted by the FDA
- Aug 7, 2024
- Recalling firm
- MEDLINE INDUSTRIES, LP - Northfield
- Product code
- LRO
- FDA submissions
- K213481
- Distribution
- Worldwide distribution.
Product description
Centurion procedure kits labeled as NEURO ANGIO PACK, Pack Number DYNDA1431A
Code information (models, lots, serial numbers)
Lot Number 2019072590
Other devices under the same FDA submission
321 devices cleared under K213481 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Centurion ART690Centurion Medical Products Corporation · General Surgery Tray4 recalls0 reports
- Centurion CVI4505Centurion Medical Products Corporation · General Surgery Tray3 recalls0 reports
- MEDLINE ECVC6355Medline Industries, Inc. · General surgery tray3 recalls0 reports
- MEDLINE ECVC1045Medline Industries, Inc. · General surgery tray3 recalls0 reports
- Centurion ART255Centurion Medical Products Corporation · General surgery tray3 recalls0 reports
- Centurion CVI3880Centurion Medical Products Corporation · General surgery tray3 recalls0 reports
- MEDLINE ECVC1375Medline Industries, Inc. · General surgery tray3 recalls0 reports
- MEDLINE DYNDV2520Medline Industries, Inc. · General surgery tray3 recalls0 reports
- Centurion ST11670ACenturion Medical Products Corporation2 recalls0 reports
- Medline Industries, Inc. ECVC6370Centurion Medical Products Corporation · General Surgery Tray2 recalls0 reports
- MEDLINE ECVC6840Medline Industries, Inc. · General surgery tray2 recalls0 reports
- MEDLINE ECVC6265Medline Industries, Inc. · General Surgery Tray2 recalls0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.