Skip to content

MNA-420 ENC HDMI, Material Number K9303311 recallZ-2332-2024 · Barco N.V. · initiated Apr 24, 2024

Class II Completed All recalls posted in July 2024

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Barco N.V. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

We have identified a potential issue with the SFP+ module supplied by Molex (Part Number: B620050) when connected to the ETH0 slot of either the MNA-6x0 ENC HDMI or MNA-4x0 ENC HDMI. In some instances, this connection may render the first HDMI channel inaccessible, making both the channel and the respective MNA ENC unit unusable. Additionally, readings from the module become unavailable when this issue arises. Similarly, when this Molex module is connected to the ETH0 slot of an MNA-620 ENC ANA, readings may also become inaccessible.

FDA root cause category: Other

What the company is doing

Barco NV notified its sole consignee on 04/29/2024 via email. The notice explained the issue, potential risk, and requested the following: "Actions to be taken by the customer: 1. Identify the location and amount of all Molex modules that are listed in pdf document Molex_SN_Partner . 2. Fill out and send back Response Form 1 to your Barco Key Account Manager, including the number of distributed units per country. 3. Disable or discard all used Molex modules. 4. Fill out and send back Response Form 2 to your Barco Key Account Manager after completion of the actions. For question, contact your Barco Key Account Manager.

Recall details

Recall number
Z-2332-2024
Recall class
Class II
Status
Completed
Initiated by the company
Apr 24, 2024
Posted by the FDA
Jul 11, 2024
Recalling firm
Barco N.V.
Quantity
88 units
Product code
DXJ
FDA submissions
K173381
Distribution
US: FL

Product description

MNA-420 ENC HDMI, Material Number K9303311; cathode ray tube display

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

serial numbers: 2530986299, 2530986300, 2530986301, 2530986302, 2530986303, 2530986304, 2530986305, 2530986306, 2530986307, 2530986308, 2530986309, 2530986310, 2530986311, 2530986312, 2530986313, 2530986314, 2530986315, 2530986316, 2530986317, 2530986318, 2530986319, 2530986320, 2530986321, 2530986322, 2530986323, 2530986324, 2530986325, 2530986326, 2530986327, 2531041594, 2531041591, 2531041578, 2531041585, 2531041595, 2531041596, 2531041586, 2531041588, 2531041587, 2531041577, 2531041584, 2531041581, 2531041612, 2531041608, 2531041600, 2531041601, 2531041604, 2531041599, 2531041602, 2531041605, 2531041606, 2531041607, 2531041609, 2531041610, 2531041580, 2531041583, 2531041598, 2531041593, 2531041590, 2531041614, 2531041582, 2531041613, 2531041589, 2531041616, 2531041579, 2531041615, 2531041603, 2531041597, 2531041592, 2530985205, 2530986280, 2530986281, 2530986282, 2530986283, 2530986284, 2530986285, 2530986286, 2530986287, 2530986288, 2530986289, 2530986290, 2530986291, 2530986292, 2530986293, 2530986294, 2530986295, 2530986296, 2530986297, 2530986298

Other devices under the same FDA submission

4 devices cleared under K173381 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates