The D*Clot HD Rotational Thrombectomy System is a… recallZ-2213-2024 · Mermaid Medical A/S · initiated May 1, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Mermaid Medical A/S or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Internal product testing failed bubble leak testing (ASTM F2096), which indicates the sterility barrier may not be fully intact.

FDA root cause category: Under Investigation by firm

What the company is doing

The consignees and distributors were sent out the notification via email on 05/01/2024. The letter instructs the consignees to: 1. Quarantine and Discontinue Use of Affected Product: Identify, segregate, and discontinue use of the affected product in your inventory. 2. Complete and Return the attached Response Form: Complete the attached Response Form and return to Mermaid Medical by May 10, 2024 via email or by post. 3. Distributors/ If You Further Distributed Affected Product: Forward this communication to any customers that were distributed affected product. Instruct any of your customers to complete the response form and return to Mermaid Medical. Once they complete and return the response form, Mermaid Medical will provide instructions on how to return or destroy any affected product remaining in their inventory.

Recall details

Recall number
Z-2213-2024
Status
Open, Classified
Initiated by the company
May 1, 2024
Posted by the FDA
Jun 26, 2024
Recalling firm
Mermaid Medical A/S
Quantity
190 devices
Product code
QEW
FDA submissions
K021641
Distribution
Worldwide - US Nationwide distribution in the states of OH, MO, CO, NC, TX, MA, GA, FL, IL, IN, MD, PA, VA, and the country of Chile.

Product description

The D*Clot HD Rotational Thrombectomy System is a percutaneous catheter-based system available in 6 French. The disposable system consists of an outer sheath through which a rotatable radiopaque spiral shaft extends ending in an eccentric blunt distal tip. Once activated, the shaft rotates at approximately 10.000 RPMs to break thrombus into smaller particles, which can be simultaneously aspirated using a dedicated vacuum syringe. The catheter is a stand-alone system and is powered by a 4.5 VDC alkaline battery pack in the control unit.

Code information (models, lots, serial numbers)

Unique Device Identifier: 15711055004333; Lot codes: S2306001; Model Number: T10106065