RINGLOC BI-POLAR HIP SYSTEM recallZ-2179-2024 · Biomet, Inc. · initiated May 17, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Biomet, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
41 mm acetabular outer heads were potentially assembled with a 42 mm locking ring, may lead to joint instability leading requiring surgical intervention
FDA root cause category: Labeling mix-ups
What the company is doing
Zimmer Biomet issued Urgent Medical Device Recall Letter on 5/17/24 to Distributors and End-Users via FedEx and email. Letter states reason for recall, health risk and action to take: Risk Manager Responsibilities: 1. Review this notification and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3.If the product has been further distributed, provide your customers with the recall notice and ensure documentation. 4. Complete Attachment 1 Certificate of Acknowledgement and send to [email protected]. This form shall be returned even if you do not have affected products at your facility. 5. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facility s documentation. 6. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to [email protected]. Surgeon Responsibilities: 1. Review this notification for awareness of the contents. 2. There are no specific patient monitoring instructions related to this recall that are recommended beyond your existing follow-up schedule. 3. Complete Attachment 1 Certificate of Acknowledgement and send to [email protected]. 4. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facility s documentation. 5. If you have further questions or concerns after reviewing this notice, please call custom
Recall details
- Recall number
- Z-2179-2024
- Status
- Open, Classified
- Initiated by the company
- May 17, 2024
- Posted by the FDA
- Jun 25, 2024
- Recalling firm
- Biomet, Inc.
- Quantity
- 12 units
- Product code
- JDI
- FDA submissions
- K051569
- Distribution
- Worldwide - US Nationwide distribution in the states of AZ, CA, NJ, PA, TX, VA and the country of Chile.
Product description
RingLoc Bi-Polar Hip System Acetabular Cup, 41 mm Outer Diameter - Used in hip hemi-arthroplasty Item Number: 11-165206
Code information (models, lots, serial numbers)
Lot Number / UDI Number (1) 66540443 (01)00880304001923(17)290408(10)66540443; (2) 66601057 (01)00880304001923(17)290405(10)66601057; (3) 66623486 (01)00880304001923(17)290327(10)66623486; (4) 66635486 (01)00880304001923(17)290327(10)66635486
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- RINGLOC BI-POLAR HIP SYSTEM 11-165206Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented2 recalls3 reports
Product lines
Other devices under the same FDA submission
15 devices cleared under K051569 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- RINGLOC BI-POLAR HIP SYSTEM 11-165228Biomet Orthopedics, LLC · Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented1 recall1 death12 reports
- RINGLOC BI-POLAR HIP SYSTEM 11-165218Biomet Orthopedics, LLC · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED1 recall2 deaths9 reports
- RINGLOC BI-POLAR HIP SYSTEM 11-165212Biomet Orthopedics, LLC · Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented1 death12 reports
- RINGLOC BI-POLAR HIP SYSTEM 11-165226Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained, uncemented, metal/polymer, non-porous, calicum-phosphate1 death10 reports
- RINGLOC BI-POLAR HIP SYSTEM 11-165214Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained, uncemented, metal/polymer, non-porous, calicum-phosphate1 death9 reports
- RINGLOC BI-POLAR HIP SYSTEM 11-165232Biomet Orthopedics, LLC · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED8 reports
- RINGLOC BI-POLAR HIP SYSTEM 11-165224Biomet Orthopedics, LLC · Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented7 reports
- RINGLOC BI-POLAR HIP SYSTEM 11-165222Biomet Orthopedics, LLC · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED1 death6 reports
- RINGLOC BI-POLAR HIP SYSTEM 11-165208Biomet Orthopedics, LLC · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED6 reports
- RINGLOC BI-POLAR HIP SYSTEM 11-165216Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained, uncemented, metal/polymer, non-porous, calicum-phosphate5 reports
- RINGLOC BI-POLAR HIP SYSTEM 11-165210Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained, uncemented, metal/polymer, non-porous, calicum-phosphate4 reports
- RINGLOC BI-POLAR HIP SYSTEM 11-165230Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained, uncemented, metal/polymer, non-porous, calicum-phosphate3 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.