Skip to content

Centurion procedure packs, containing Caina syringes… recallZ-2775-2024 · MEDLINE INDUSTRIES, LP - Northfield · initiated Apr 8, 2024

Class II Open, Classified All recalls posted in August 2024

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from MEDLINE INDUSTRIES, LP - Northfield or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health.

FDA root cause category: Under Investigation by firm

What the company is doing

Medline Industries LP issued a recall notice to its consignees on 04/08/2024 via letter delivered via USPS and email. The notice explained the issue, provided a link to for information pertaining to the risk involved https://www.fda.gov/medical-devices/safety-communications/update-evaluating-plastic-syringes-made-china-potential-device-failures-fda-safety-communication) and request that the syringes within the convenience be discarded, except for circumstances where no alternative syringes are readily available. If these syringes must be used, closely monitor for leaks, breakage, and any other problems, and report any quality issues to Medline and the FDA. Distributors were directed to notify their customers. The firm issued a second recall notice (an expansion) on 05/02/2024 using the same methods.

Recall details

Recall number
Z-2775-2024
Recall class
Class II
Status
Open, Classified
Initiated by the company
Apr 8, 2024
Posted by the FDA
Aug 28, 2024
Recalling firm
MEDLINE INDUSTRIES, LP - Northfield
Product code
LRO
FDA submissions
K213481
Distribution
Worldwide distribution.

Product description

Centurion procedure packs, containing Caina syringes, labeled as: 1) ARTHROGRAPH TRAY, Pack Name MNS12670A; 2) BREAST BIOPSY KIT, Pack Name DYNDH1512A; 3) CIRCUMCISION TRAY, Pack Name CIT6720; 4) CIRCUMCISION TRAY, Pack Name CIT6715; 5) CIRCUMCISION TRAY, Pack Name CIT7020; 6) CIRCUMCISION TRAY, Pack Name DYNDA2271; 7) NICU CIRCUMCISION TRAY W/OUT CLAMP, Pack Name CIT6920; 8) NURSERY CIRCUMCISION TRAY, Pack Name CIT4845; 9) PUNCH BIOPSY KIT, Pack Name MNS8245; 10) STERILE 1CC SYR W/ 27GX1/2 NDL, Pack Name DYNDA2496

Code information (models, lots, serial numbers)

MNS12670A, Lot Number 2022011090; MNS12670A, Lot Number 2022070690; MNS12670A, Lot Number 2023022190; MNS12670A, Lot Number 2023101890; MNS12670A, Lot Number 2024021590; DYNDH1512A, Lot Number 2022121990; DYNDH1512A, Lot Number 2023101890; CIT6720, Lot Number 2021061490; CIT6715, Lot Number 2022100490; CIT7020, Lot Number 2023010690; DYNDA2271, Lot Number 2023022190; CIT6715, Lot Number 2023082290; CIT7020, Lot Number 2023101890; CIT6715, Lot Number 2023102090; CIT6715, Lot Number 2023120590; CIT6920, Lot Number 2021082390; CIT4845, Lot Number 2021122150; CIT4845, Lot Number 2022052450; MNS8245, Lot Number 2023082490; DYNDA2496, Lot Number 2023012790

Other devices under the same FDA submission

321 devices cleared under K213481 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page