Transfer Device, REF 2008S. Packaged on a form, fill and… recallZ-2643-2024 · Microtek Medical Inc. · initiated Jun 3, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Microtek Medical Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

During sterile barrier testing performed on the decanter product line, the samples were identified to have pin holes and tears in the packaging film.

FDA root cause category: Device Design

What the company is doing

Consignees were notified via email and letter on 06/03/2024. The consignees were instructed to "Segregate Recalled Product/immediately examine the inventory for the lots listed above and remove all products from inventory regardless of its location and segregate this product in a secure location. The lot number can be found on the individual package or case label. If the consignee has affected product in their possession, please dispose of the product and packaging in accordance with hospital, administrative and /or local government policy. If the consignee is unable to dispose of the product, please contact the Customer Service team at the information provided below in this letter. The consignee is requested to please complete the enclosed Ecolab Customer Response Form and e-mail it to [email protected], whether they have affected product in their possession or not. If they need assistance with filling out the form, please contact the sales representative. If the consignee has further distributed this product, instead complete the enclosed Ecolab Distributor Response Form and e-mail it to [email protected] whether their customers have affected product in their possession or not. Please identify the consignee's customers and notify [email protected] so Ecolab may notify them of this Field Recall. It is important that the firm receive the completed Ecolab Customer Response Form or Ecolab Distributor Response Form by June 14, 2024. The consignee is instructed to please contact the Customer Service at 1-800-824-3027 from 8:30 a.m. to 5 p.m. (CST) to obtain a Returned Material Authorization (RMA) number if they are unable to dispose of the affected product, or email at [email protected] or [email protected]. The consignee is to package the decanter(s) that are being returned in an appropriate shipping box. Place a copy of the response form in the box, appropriately seal the box, and send it on or before June 14, 2024, to: ATTN: Belissa del Val

Recall details

Recall number
Z-2643-2024
Status
Open, Classified
Initiated by the company
Jun 3, 2024
Posted by the FDA
Aug 16, 2024
Recalling firm
Microtek Medical Inc.
Quantity
11,781 cases (589,050 units)
Product code
KPE
FDA submissions
K811268
Distribution
Distributed nationwide and internationally to Canada, Mexico, New Zealand, Belize, and Hong Kong.

Product description

Transfer Device, REF 2008S. Packaged on a form, fill and seal machine with a poly/Tyvek material then placed 50 products to a case (cs).

Code information (models, lots, serial numbers)

Show the full code information (52k characters)

SKU/UDI-Device Identifier: 2008S/00748426002535; Lot Numbers: 210924,190828,190828,190312,190527,190507,190528,190507,181022,190527,181224,190717,181226,181225,190312,190828,190410,180828,190527,190528,180828,181226,211115,210924,210922,210922,210922,210923,210924,210922,210807,210805,210706,210706,210922,210706,210804,210804,210419,210419,210419,210415,210415,210415,210416,210416,210419,210104,210104,210104,201111,201111,201110,201110,201113,201113,201113,201113,201112,201112,201112,201112,201112,201113,201019,201019,201019,200818,200818,200818,200818,200818,200702,200702,200702,200702,190528,190312,181022,190527,190828,181225,181224,190410,190829,181224,180828,190527,190312,181022,181022,181226,190718,180828,190528,190313,190527,181224,181226,181226,190717,181225,181226,230628,230626,230608,230623,230608,230626,230607,230509,230605,230605,230605,230509,230509,230324,230324,230326,230325,230325,221117,221117,221112,221112,221112,221006,211116,211115,211115,211115,220525,220525,220417,220406,220527,220527,220527,220527,220530,220527,220530,220407,220405,220407,220103,220103,220102,220102,220102,181226,190410,190507,190410,211116,211117,210924,210922,210924,210922,210922,210923,210923,210807,210924,210807,210807,210924,210706,210924,210706,210706,210805,210419,210804,210706,210419,210419,210415,210415,210416,210416,210415,210416,210416,210104,210104,210419,210104,201111,201111,201111,201110,201110,201110,201113,201113,201112,201020,201019,200818,190527,190410,230628,230628,230626,230629,230626,230629,230623,230623,230623,230324,230325,230326,221117,220526,211115,220406,220526,220530,220527,220527,220407,220405,220406,220405,220103,220103,220103,220103,200424,190507,181226,230628,230629,230608,230509,220103,220103,210415,181225,211117,210924,201111,201112,200702,181225,190718,230628,190312,181225,181226,180828,181225,181225,211117,211117,211117,211117,211116,210924,210923,210416,200424,210924,210923,210805,210805,210706,210706,210805,210804,210419,210419,210416,210416,210104,210104,201111,201111,201111,201110,201113,201113,201112,201112,201112,201020,201020,201020,200818,200702,200702,200702,200425,200702,200702,190507,190718,181226,181226,190312,190528,181022,181022,181225,190828,181224,181226,190717,230628,230627,230629,230623,230629,230608,230629,230626,230605,230607,230605,230324,230326,230325,230325,230325,221117,221112,211116,211115,220526,220525,220417,220406,220527,220530,220530,220407,220405,220405,220405,220103,220103,220103,220103,190527,181224,190507,190527,181225,190828,180828,190410,181022,190410,211117,211116,210924,210922,210922,210922,210922,210923,210923,210807,210706,210805,210805,210706,210706,210706,210419,210419,210415,210415,210416,210416,210104,210104,210104,201111,201111,201111,201111,201111,201110,201113,201110,201113,201113,201112,201112,201112,201020,201019,201019,201019,201019,200818,200818,200702,200702,200702,190528,181225,181224,181225,190828,190527,190528,181225,190828,180828,190410,181022,190527,190528,190528,190527,181226,190312,190312,181224,190718,190312,180828,190828,181224,181224,180828,181225,190312,190528,230623,230623,230623,230623,230608,230608,230629,230607,230607,230607,230629,230607,230605,230605,230509,230324,230326,230326,230326,230326,230326,230326,230325,230325,221117,221112,221112,221112,221112,220528,211116,220528,220526,211115,211115,211115,211115,220525,220417,220417,220417,220530,220527,220527,220530,220530,220527,220530,220406,220405,220405,220102,220103,220103,220103,220102,220103,220103,211117,211117,210807,210924,230629,230608,230629,230509,230326,230326,230326,221117,221112,211115,220527,220405,220103,200702,210416,201020,180828,211117,181225,190312,211117,211115,210924,210807,210419,210415,210415,210416,210416,210104,201111,201111,201110,201112,201112,201019,200818,200818,200702,190528,181225,181022,190527,180828,190312,190718,181224,190528,181226,180828,181022,190828,181225,181022,210104,201111,201112,201019,201019,181225,221117,230628,230608,230626,230626,230623,230623,230608,230626,230607,230607,230605,230509,230324,230326,230325,221117,221112,211116,220526,211115,220526,220527,220530,220407,220406,220103,210706,201112,190828,180828,230605,230324,221011,220525,220525,220527,190507,181224,181226,211115,210924,210807,201110,201110,181224,190717,220405,220103,220406,220103,210923,210923,210807,210807,210807,210805,210706,201019,201019,230623,210922,210924,210706,201111,201111,200702,210805,210706,210804,210804,210706,210419,210415,210415,210419,210415,210415,210416,210416,210416,230326,211115,211115,220405,210416,210104,210104,210104,210104,210419,201111,201111,201111,201111,201111,201111,201111,201111,201111,201110,201110,201110,211116,181022,210924,210706,230605,230605,220103,220103,210924,210807,181022,230326,220525,220525,230628,230628,230626,230629,230627,230626,230629,230623,230623,230623,230623,230623,230608,230608,230608,230626,230629,230626,230626,230605,230607,230605,230605,230605,230509,230509,230509,230509,230509,230324,230324,230326,230326,230326,221117,221112,221112,221112,221112,221006,211115,220417,220525,220417,220406,220406,220526,220526,220527,220530,220527,220527,220527,220407,220407,220407,220103,211116,211115,210924,210922,210922,210922,210923,210923,210922,210923,210807,210807,210807,210805,210706,210706,210805,210805,210706,210804,210804,210804,210706,210419,210415,230626,230626,230623,230605,230608,230607,181224,200702,190527,190718,181225,190507,181226,180828,181225,190410,201019,200818,181224,180828,190527,190909,181226,181225,220417,220530,220407,220407,220103,210922,210923,210416,210415,210415,210415,210416,210416,210419,210104,201111,201111,201111,201110,201110,201113,201113,201113,201112,201112,201112,201019,201020,201020,201019,201019,200818,190528,190410,181022,181224,190507,190717,190312,181022,181225,180828,190528,190410,200818,200818,200818,200425,200702,200702,190527,181226,190312,190410,190528,181224,190718,180828,181226,190312,181225,190528,190717,181226,180828,181226,230628,220103,190527,201110,230628,230628,230626,230626,230627,230627,230627,230623,230623,230629,230623,230608,230607,230608,230629,230608,230629,230607,230607,230629,230607,230605,230605,230509,230509,230326,230324,230325,230325,230325,230325,221117,211115,220417,220407,220527,220530,220530,220527,220407,220407,220407,211117,220407,220405,220103,220102,220102,220102,211117,211117,210804,210415,181224,190410,190312,211117,211117,210924,210922,210922,210922,210807,210706,210419,210415,210416,210419,201111,201113,201113,201110,201020,200818,200818,200818,200702,200702,190410,181224,181022,190312,181224,190312,190312,180828,181224,190507,180828,230605,230605,230605,230626,230629,230605,230629,230509,230326,230326,221117,211116,211115,220417,220530,220407,220405,220103,220102,230629,230608,230626,230623,230623,230608,230629,230607,230605,230509,230324,230326,230325,230325,221112,221011,211116,211115,220525,220417,220526,220527,220527,220405,220103,220103,211117,181022,181226,181225,181022,190828,190410,190527,190312,190507,190528,211116,210924,210922,210923,210807,210805,210706,210804,210804,210419,210415,210416,210104,201111,201110,201112,201020,200818,200425,201110,201113,201113,201111,201111,201111,200702,200702,190527,230628,230605,230326,220527,201113,201113,201113,201112,201112,201112,201112,201112,201112,201112,201112,201020,201020,201020,201020,201020,201020,201020,201020,201019,201019,201019,201019,201019,201019,201019,210706,210416,201111,201113,201020,201020,201019,201019,201019,200818,211115,210922,210923,210706,210419,210419,210419,210415,201112,201112,210415,181022,190313,190410,190828,190527,181225,181022,190528,181224,190717,190507,190507,190828,181226,181226,190312,181225,181022,181022,190527,190527,180828,181224,190528,181226,190410,190410,200818,200818,200818,200702,200702,200702,200702,200702,200702,181226,181226,190828,190527,190527,181022,181225,190312,190312,190410,190410,181224,181224,181224,190528,190718,180828,190507,180828,190507,190312,181022,190312,190313,190507,190528,181224,181224,190717,180828,180828,211117,210922,210807,210804,210415,210104,201110,201020,201020,200818,230626,230623,230629,230324,221117,221112,211116,220525,220406,220530,220103,220103,211117,190507,190312,181224,181226,190828,190527,190528,210924,210924,210923,210804,210416,201110,201113,201019,201019,201019,181225,181022,190410,190507,190828,190828,190527,190312,181022,190507,190528,181224,190312,190717,181225,190528,220527,220407,220405,220103,181225,210805,201112,201110,201113,201112,201019,200702,200702,190527,181226,181225,190410,190313,230608,230608,230605,230607,230607,230626,230605,230605,230509,230509,230324,230325,230326,230325,230325,221117,221117,220417,220417,220406,220406,220417,220526,220527,220527,220527,220527,220407,181226,181226,190828,190410,210416,201020,181022,180828,230628,230628,230626,230629,230629,230623,230629,230623,230608,230608,230626,230626,230607,230607,230607,230605,230605,230605,230509,230509,230509,230509,230326,230324,230326,221117,221112,221112,221112,220528,211116,220526,211115,220417,220530,220407,220405,211117,211117,211116,210922,210924,210807,210807,210807,210807,210805,210706,210807,210706,210804,210804,210804,210706,210419,210419,210419,210419,210415,210415,210415,210416,210416,210416,210104,210419,201111,201111,201111,201113,201113,201112,201112,201112,200818,181225,190717,190828,190527,181225,181022,180828,181224,190410,200818,200702,200702,200702,190528,181226,190828,190312,181022,190410,190718,190718,190312,181225,181224,190717,180828,181226,201111,201110,181224,190312,190527,181224,181226,181226,190507,190507,211117,210805,210805,210416,210104,201111,201110,201113,201112,201019,201020,190507,181225,190528,190718,190828,190312,181022,181224,190717,190717,181225,190528,230628,230626,230608,230608,221112,220407,220528,181224,181225,190828,190528,190507,181226,190828,190507,190527,181225,181022,181224,190528,181226,190527,200702,190528,181225,190312,181224,180828,180828,181226,190312,190410,180828,190528,190410,181226,190718,190312,181224,181226,190507,190507,181022,190410,190528,190507,190528,190313,190718,180828,190717,180828,181226,181022,181225,181226,220407,220405,180828,211117,211117,211115,210922,210922,210922,210923,210807,210807,210924,210706,210805,210419,210419,210415,210416,210416,210104,210104,201111,201111,201111,201110,201113,201113,201112,201020,200818,200702,181022,230628,230628,230608,230629,230627,230626,230629,230623,230623,230623,230629,230608,230608,230626,230626,230607,230605,230607,230607,230605,230605,230509,230509,221117,221117,221112,211115,220525,220525,220526,220527,220530,220527,220407,220407,201019,190313,181225,181022,181022,181022,180828,181224,181224,181226,190312,190312,181224,181022,181226,200818,190527,190528,180828,190527,181224,190313,190507,190828,190507,190717,190312,181225,181022,220530,220530,220407,220405,220405,220405,220103,220103,190528,190507,190527,190828,190312,181226,190527,190410,190828,180828,190527,190527,190528,181224,190527,190313,181226,211116,211115,211117,210924,210924,210922,210924,210923,210807,210807,210706,210922,210805,210706,210706,210805,210805,210706,210804,210706,210419,210419,210419,210419,210419,210415,210415,210415,210416,210416,210416,210416,210416,210104,210104,201111,201110,201110,201110,201113,201112,201020,201020,201020,201020,201019,201019,200818,200818,200818,200702,200702,190528,190528,181225,190828,190312,190312,181224,181226,190527,190507,190312,190410,190410,190528,190528,181224,190718,190507,220527,220103,230628,230626,230626,230608,230627,230627,230626,230623,220406,220405,220103,220102,211117,210923,210923,210807,210807,210922,210805,210924,210419,210415,210415,210416,210416,210416,210416,201113,201113,201113,201112,201020,201020,201019,201019,201019,200818,190410,190528,190717,181225,190527,200818,190828,190312,180828,190718,181225,180828,190528,181224,190828,180828,190507,190312,190410,181226,200702,180828,220525,190528,181022,181224,190312,190312,181226,190828,181226,190527,181022,181224,190828,190528,181226,190528,190410,190410,211117,211117,211117,210924,210924,210924,210922,210922,210922,210922,210922,210923,210923,210923,210922,210923,210807,210807,210924,210706,210807,210805,210706,210804,210804,210804,210804,210804,210419,210419,210419,210415,210415,210415,210415,210416,210416,210415,210104,210104,210104,210104,210104,210104,201111,201111,201110,201110,201110,201110,201113,201113,201113,201113,201112,201113,201112,201112,201112,201112,201112,201020,201020,201020,201019,201019,201019,201019,201019,200818,201019,200818,200818,200818,200702,200702,200702,200702,200702,190527,190828,190528,181226,181224,190410,180828,180828,190828,180828,180828,190527,190312,190828,190312,190528,190528,190410,180828,190718,190507,181022,190528,181022,181226,180828,190312,190507,190528,181224,181226,190718,181225,190312,181224,181226,181226,181226,181226,180828,230628,230628,230627,230623,230623,230608,230629,230626,230605,230605,230509,230324,230326,230326,230326,230326,230325,230325,221117,221117,221117,221117,221117,221117,221117,221112,221112,221006,211116,211116,220525,220526,220526,220526,220526,211115,211115,211115,220815,220528,220525,220525,220525,220525,220525,220417,220417,220525,220406,220527,220527,220527,220530,220527,220527,220530,220407,220406,220103,220102,220102,220102,220103,220103,220102,230623,230629,230608,230608,230608,230629,230626,230607,230629,230607,230607,230605,230605,230605,230509,230324,230324,230326,230326,230325,221117,221117,221112,221112,221006,211116,211116,220528,211115,211115,220525,220525,220525,220417,220417,220406,220526,220530,220407,220407,220405,220405,220103,220103,220103,220102,181225,190410,181022,181022,190527,190527,181224,190828,181022,190717,190312,190410,190410,190312,190527,181022,181226,211117,211117,210924,210922,210922,210923,210923,210923,210922,210805,210923,210805,210706,210804,210706,210419,210419,210415,210416,210416,210416,210416,210416,210104,210104,210104,201111,201111,201111,190828,180828,190528,181224,181226,181225,190527,190410,190528,190718,180828,190312,190718,190410,190312,181226,190507,190528,190528,181224,181224,180828,180828,230626,211117,190528,190528,190507,181224,190718,190507,181022,181225,190528,181224,181022,181226,181224,181225,190507,200818,181225,181224,181226,190528,190507,181224,190828,190828,210924,210923,210706,210805,210706,210924,220417,220525,220417,220406,220526,220103,220103,220102,211116,211115,210922,210923,210924,210923,210923,210419,210415,210419,210415,210416,210416,210104,210104,201110,201113,201113,201112,201112,200818,190313,190507,200818,200702,200702,200702,181226,190410,181226,190313,190410,230628,230628,230628,230628,230628,230626,230626,230608,230627,230627,230627,230629,230623,230623,230623,230608,230608,230608,230608,230626,230629,230629,230626,230607,230607,230629,230607,230607,230607,230607,230605,230605,230509,230509,230509,230509,230509,230509,230509,230509,230325,230326,230326,230324,230324,230326,230326,230325,230325,221117,221117,221117,221117,221117,221117,221112,221112,221006,221006,220528,211116,211116,211116,220526,211115,220815,220525,220417,220417,220417,220417,220417,220417,220406,220526,220527,220530,220527,220527,220527,220407,220407,220407,220407,220406,211117,220407,220407,220102,220407,220405,220405,220103,220103,220102,220102,210924,210922,210924,210922,210922,210922,210923,210923,210922,210923,210923,210923,210923,210807,210807,210807,210807,210805,210807,210706,210706,210805,210805,210706,210706,210706,210804,210804,210804,210804,210804,210804,210706,210706,210419,210419,210419,210419,210415,210415,210419,210415,210416,210416,210416,210416,210416,210415,210416,210419,210416,210104,201111,201111,201110,201110,201110,201110,201110,201113,201113,201113,201113,201019,201020,201020,201020,201020,201019,201020,201019,201019,201019,200818,200818,181225,190528,190507,190717,190717,190312,181226,180713,180828,190527,190410,200818,200818,200702,200702,200702,200818,200702,190528,180828,190527,181226,190828,190828,190312,181022,190410,190528,181224,190718,190718,180828,190312,181022,181022,181022,181225,190410,190507,190528,190717,180828,181226,181225,181022,181224,190507,190717,190312,190312,190527,181224,210924,210922,210922,210922,210923,210922,210923,210924,210923,210807,210807,210807,210807,210805,210805,210805,210706,210706,210706,210804,210706,210805,210419,210419,210419,210415,210415,210415,210415,210416,210416,210416,210415,210416,210104,210104,210104,210104,201111,201111,201111,201111,201110,201110,201110,201113,201112,201113,201112,201112,201110,201020,201110,201020,200818,201019,201019,200818,200818,200702,200702,190527,190528,181225,190507,190828,190410,190528,181022,181224,181226,190528,180828,201110,190507,211117,210924,210922,210805,210706,210706,210419,210415,210416,210104,201111,201110,201113,201112,201112,201019,200818,200702,180828,190507,190312,230623,230605,230326,221112,220525,211115,220525,220417,220527,220530,220405,210706,220102,230326,211116,230326,220407,201019,190507,230623,211116,220103,210922,210807,210805,210415,210416,201111,201111,201113,201112,201020,200818,190828,190527,190312,181022,181226,190312,190313,190312,201112,200818,181225,180828,210922,210804,210419,201020,200818,220406,220102,181022,181226,181225,180828,190527,190528,211117,211116,210924,210922,210922,201111,210807,210807,210807,210805,210805,210924,210706,210923,210804,210419,210415,210415,210415,210416,210416,210104,201112,201111,201113,201113,201113,201112,201019,201112,201112,201020,201113,201019,201019,201019,200818,200818,200818,200818,200702,200702,200702,190527,190312,181224,190507,190507,190312,190507,190313,230628,230628,230626,230629,230627,230627,230626,230629,230623,230623,230629,230623,230629,230608,230608,230608,230608,230626,230509,230626,230629,230626,230626,230607,230607,230607,230628,230509,230605,230605,230509,230509,230509,230324,230324,230324,230326,230324,230326,221117,230325,221117,221117,221117,221112,221117,221112,220528,211116,211116,220526,220526,220525,220525,211115,220525,220527,220530,220407,220103,211117,220102,220407,220405,220405,220102,220102,220102,190528,210416,201111,201111,201110,201112,201019,200818,230628,200818,181225,210923,210706,210706,210419,210415,201110,201020,190312,181022,230608,230325,220525,220417,220527,220103,200702,181224,190717,181226,190312,181225,211115,210922,210923,210807,210805,210706,210804,210419,210415,210416,210104,210104,201111,201110,201113,201112,201110,201020,200818,190312,190718,180828,230628,230626,230629,230626,230623,230623,230623,230608,230608,230629,230629,230605,230605,230509,230324,230326,230325,221117,221117,220526,220525,220417,220527,220527,211117,211117,220407,220102,181224,180828,211117,210922,210923,210104,181226,181226,181226,190527,181224,190410,190527,190528,211117,210924,210807,210807,210805,210706,210804,210419,210415,201112,210415,201110,201110,201113,201112,201020,201020,201020,201020,201019,200818,200818,200702,200702,200702,190527,181224,180828,190528,190527,190410,180828,190312,181226,181226,190528,190528,180711,181226,181226,181226,230627,230608,230605,230605,230326,221117,221112,211116,220528,220526,211115,211115,220525,220527,220407,220103,211117,200702,200702,230628,230628,230628,230626,230626,230608,230627,230627,230627,230623,230623,230623,230629,230623,230629,230629,230608,230608,230608,230626,230626,230626,230626,230629,230629,230607,230607,230607,230607,230605,230605,230605,230509,230509,230509,230509,230326,230324,230324,230324,230324,230326,230326,230325,221117,221117,221117,221117,221117,221117,221112,221112,221112,221112,221117,221112,221011,221006,220528,211116,211116,211116,220526,211115,220526,220525,211115,220525,220525,220525,220417,220417,220406,220528,220526,220527,220527,220527,220527,220527,220530,220530,220527,220527,220527,220407,220407,220407,220103,220103,220103,220103,220103,220102,220102,211117,211115,210924,210922,210922,210922,210922,210922,210922,210923,210923,210924,210922,210807,210706,210805,210706,210706,210805,210706,210804,210805,210804,210706,210419,210419,210419,210415,210419,210415,210416,210416,210416,201111,201110,201113,201113,201113,201113,201112,201112,201019,201019,201019,201019,200818,200818,190527,181022,181224,190313,190717,190828,190507,190527,181226,190527,181225,181022,200702,200702,200702,200702,190528,190410,181224,190828,190527,190527,190312,181225,190410,180828,181022,190507,190528,190507,181224,190718,180801,181022,180828,180828,190312,190507,190528,190410,181224,190717,181226,201110,200702,230608,230605,230325,220102,181224,180828,190507,181022,190528,200702,181225,181022,190528,190717,190717,190312,190410,190828,181022,190527,181224,190410,181226,181226,211117,210924,210922,210922,210923,210923,210807,210922,201111,210419,210419,210415,210415,210416,210416,210104,210104,201111,201111,201110,201110,201113,201112,201020,201020,200702,201019,201019,201019,200818,200818,200818,201020,200702,200702,190828,181225,180828,190410,190528,181224,190527,190718,181226,190312,190528,190507,190528,190528,181224,220103,211117,211117,210922,210923,210922,210419,210416,210416,210104,201111,201113,201019,230324,211116,220525,220525,220417,220417,220526,220527,220405,220103,220103,181022,210805,220406,220406,220102,230626,230626,230623,230607,230605,230325,221112,211116,220526,220525,220525,220405,181225,190527,210924,210922,210922,210807,210706,210706,210804,210706,210416,210416,210104,210104,201110,201113,201113,201112,201020,201020,201020,200818,200425,190507,181224,190527,180828,190527,190507,181225,190717,181225,180828,181224,181022,190312,181226,190410,180828,190527,181224,181226,181225,190312,181022,190410,181224,180828,190507,190718,181226,190507,181022,181224,181226,180828,181226,210924,230629,230629,230608,230608,230607,230605,221117,221117,221112,221112,220815,220527,220527,220530,220103,211117,181225,190410,190828,210922,210807,210415,210419,210415,210416,210104,201113,201020,200818,200818,180727,181224,181022,181022,230626,190528,190312,190528,190410,210922,210923,210706,210804,210805,210419,210419,210419,201110,200702,201112,200702,181225,181224,181022,230608,230607,230607,230607,230509,230509,221112,221117,220525,220417,220405,211117,181224,181225,190313,181225,180828,190528,190410,181224,190828,211117,210924,210922,210923,210807,210706,210804,210804,210415,210415,210416,201111,201110,201110,201113,201112,201020,201020,201019,200818,200818,200818,200818,200702,200702,190528,190527,190312,181224,181022,190507,220526,220525,220528,220527,220530,220103,181224,181022,181224,210922,210807,210924,210804,210415,210416,210104,201111,201110,201113,201112,201019,200818,190410,190312,190828,190312,190507,180828,181226,230628,230623,230607,221117,220525,220405,220405,220102,190528,190903,181224,181022,201113,201112,190828,181226,220103,210924,210923,210807,210804,210706,210104,210416,201110,201112,230626,230629,230629,230626,230629,230626,230605,230509,230605,230605,230326,230509,221117,221112,211116,211116,211116,211115,220525,220526,220527,220530,220405,220103,190527,190507,211117,211117,211115,210922,210805,210924,210923,201112,201020,181225,190507,181226,180828,230628,230629,230629,230623,230623,230608,230608,230626,230626,230607,230607,230607,230629,230605,230629,230326,230325,221117,221112,221112,220528,211116,220525,220525,211115,220525,220417,220525,220417,220406,220406,190507,190828,190527,190528,181022,211117,211115,210922,210924,210922,210922,210923,210923,210923,210805,210924,210805,210804,210104,210104,201111,201111,201110,201111,201113,201113,201112,201020,200424,181224,181226,190410,230608,230623,230608,230626,230626,230628,230628,230605,230605,230509,230509,230325,230324,230324,230326,230326,230325,230325,221117,221112,211116,220528,211116,220417,220406,220526,220527,210922,230608,230326,230326,190507,190507,190828,190828,181226,181224,180828,210807,210805,210706,210706,210706,210706,210415,190527,181022,181022,190410,190528,181224,190527,190527,190718,190312,220525,220526,220530,220530,220102,220528,230626,230623,230626,230607,230605,230324,221117,221117,211116,220526,211115,220406,220527,220530,220405,220103,220102,211116,210807,210807,210807,210706,210805,210804,210804,210419,210416,210416,210416,201111,201111,201112,201112,190312,190507,190717,190410,180828,230627,230605,230325,211116,220526,211115,220525,220406,220527,220407,220407,220103,211116,210419,201113,200702,230605,220417,181225,190312,181225,190410,181224,190312,190312,190717,190312,181224,181224,190828,190528,190312,190828,190507,211117,211115,210924,210922,210922,210922,210807,210923,210923,210807,210706,210805,210922,210805,210706,210706,210805,210804,210706,210419,210419,210419,210415,210415,210415,210416,210416,210104,210104,210104,210104,210104,201111,201111,201111,201111,201111,201111,201110,201110,201113,201113,201113,201113,201113,201112,201112,201020,201020,201019,201019,201019,201019,201019,201019,201019,200818,200818,200702,200702,200702,200702,200702,181225,190410,181225,181224,230628,230628,230629,230626,230623,230629,230629,230608,230626,230626,230607,230605,230607,230607,230605,230605,230509,230324,230324,230326,221117,221117,221117,221112,221112,221117,221011,211116,211115,220525,220406,220406,220406,220406,220406,220526,220527,220530,220530,220407,220405,220405,220407,220405,220103,220103,220102,201110,201110,210416,210416,201020,230326,201020,200818,200702,181226,230626,190312,190410,181022,210923,210923,211116,190313,181225,181022,190527,190312,190717,190410,190528,190410,190527,181224,190717,211117,210922,211115,211115,210924,210924,210924,210924,210924,210924,210924,210922,210922,210923,210923,210923,210807,210807,210807,210706,210805,210706,210706,210805,210706,210805,210804,210419,210419,210419,210419,210415,210415,210415,210415,201112,210104,210416,210416,210416,210416,210416,210104,210104,210104,201111,201111,201110,201113,201113,201113,201113,201112,201020,201020,201020,201112,201019,201019,201019,200818,200818,200702,190903,190828,190527,190312,190410,181224,190312,190718,190410,190507,180727,180801,190507,190507,190717,180828,190528,190312,190312,181022,181226,180828,230623,230623,230629,230623,230608,230608,230608,230509,230324,211116,220528,220527,220527,220527,220405,220405,220103,230628,230628,230626,230629,230629,230629,230627,230629,230623,230629,230629,230608,230608,230626,230629,230607,230605,230605,230509,230509,230509,230324,230324,230326,230326,230326,230326,221117,221117,221117,221117,221117,221117,221112,221112,221112,221112,221006,211116,211116,211116,211116,211116,211116,211116,220525,220525,220525,220525,220525,220417,220525,220528,220528,220526,220526,220526,220527,220530,220530,220530,220530,220530,220407,220405,220405,220405,220406,220103,220103,220103,220102,220103,220103,220103,220102,211115,211117,230626,230509,221112,220103,220103,220103,210923,181225,210924,210923,210923,181224,190313,190312,190527,180828,190410,200702,190507,181022,190410,211117,201110,201111,200818,190527,230608,220417,181224,190507,190527,181022,181226,210419,210415,201111,210104,210104,201111,201110,201110,201113,201113,201113,201112,201020,201020,201019,201019,200818,200818,201019,200702,181022,190312,190828,181226,180828,181022,190528,190718,190717,181225,181022,210805,230608,220525,220525,220417,220526,220407,190312,190528,180828,190528,181022,190312,190828,181022,190312,190527,190828,211115,210924,210922,210922,210922,210924,210416,210923,210807,210805,210807,210807,210706,210706,210804,210706,210419,210419,210416,210416,210416,210104,210104,210104,210104,210104,201111,201111,201110,201110,201110,201113,201113,201112,201112,201112,201020,201020,201112,201020,201019,201019,200818,200702,200702,200702,200818,190527,190410,181226,180828,181224,180828,230626,230608,230623,230623,230623,230629,230623,230626,230607,230607,230605,230325,230326,230324,230326,230326,230326,230325,221117,221117,221112,221112,221006,211116,220526,220528,220406,220406,220406,220103,220103,220102,220102,230626,230626,230626,230626,230629,230608,230607,230607,230607,230509,230326,221112,221112,221117,211116,220525,220530,220103,181022,181022,181022,181224,180828,181022,180828,210807,210804,221117,221117,221112,220528,220525,220417,220527,190528,210924,210922,210922,210804,210415,210807,210104,201111,201113,181225,190718,190410,230607,190527,190507,181022,181022,211117,210922,210922,210924,210923,210923,210807,210923,210923,210419,201113,201113,201112,201112,201112,201112,201020,201112,201112,201019,190527,190718,190528,190717,230623,230629,230608,230608,230626,230626,230607,230607,230509,221112,221112,221112,221006,211116,220406,220530,220530,220527,220405,220103,220103,220103,220102,181022,190312,181225,181224,210922,211117,210922,210807,210804,210415,210416,201111,201112,200818,200818,200818,190528,190313,180828,181225,190828,190410,181226,190528,181224,180828,181226,230623,230607,230325,211115,220406,220527,220406,230623,230605,200702,230607,181022,190717,190312,190313,190828,190528,180423,190507,210924,210922,210922,210807,210807,210924,210706,210419,210415,210415,210416,210416,201113,201112,201020,201020,201019,200818,200702,200818,200424,181022,190312,190718,190312,190313,190507,181224,230626,230626,230608,230629,230626,230626,230623,230509,230324,211116,220525,220527,220527,210924,210805,230627,230629,230608,210922,210807,210415,201110,201112,181225,230626,230605,230608,201019,200818,200702,201019,200702,200702,200818,200818,210419,210415,210104,230608,230326,211116,211115,220417,220405,220103,190528,190507,181226,181224,190312,190527,180828,211117,211116,210924,210922,210923,210807,210924,210805,210804,210419,210415,210415,210415,210416,210104,201111,201111,201111,201110,201113,201113,201112,201112,201020,201020,201019,200818,200702,200424,181225,190903,190410,190527,190528,181224,190312,181022,181226,180828,181226,190717,190410,181225,181022,190507,190717,211115,210922,210924,210922,210922,210807,210706,210706,210804,210415,210416,210104,201110,201113,201112,201020,201020,200702,200702,200702,190828,190410,190528,181224,190312,190312,181226,230628,230623,230608,230626,230607,230605,230325,230326,221117,221112,221112,211116,220525,220406,220406,220102,231027,230629,230626,230629,230608,230608,230629,230607,230509,230326,230325,221112,220526,211115,220525,220530,220103,210807,210805,201112,201112,200818,190527,190717,190717,190312,190507,181022,211117,210922,210922,210922,210922,210807,210706,210419,210415,210415,210416,210416,201111,201111,201110,201113,201112,201110,201020,200818,200702,181224,181224,190312,181224,180828,230628,230628,230608,230608,230607,230607,230629,230629,230509,230509,230324,230326,230325,221117,221112,211116,211115,220525,220417,220527,220407,220407,220407,220103,190828,181022,181224,181224,190717,190828,190312,190718,190828,190828,211117,210922,210923,210807,210923,210924,210805,210807,210805,210805,210805,210706,210804,210419,210419,210415,210415,201111,210104,210104,181225,181022,190410,190410,190410,190718,190828,181224,180828,181224,230608,230626,230629,230626,230629,230608,230608,230605,230605,230325,230326,230325,221117,221117,221112,221112,221112,221006,220525,221006,221011,221006,220528,220525,220528,220526,220528,221006,221006,211115,220525,220525,220526,220527,220530,220407,220102,230626,230629,220526,220102,210415,201112,200702,190528,190528,181022,210807,210706,210415,201111,201113,200818,200702,230626,230608,230607,230605,230325,221117,211115,220103,220103,220102,210706,210807,190507,230628,230627,230623,230629,230626,230605,230509,230326,230326,230326,230326,221117,221117,221112,221006,211116,211115,220815,220525,220417,220406,220530,220530,220527,220530,220407,211117,220103,220102,181022,181225,190312,190507,210922,210922,210807,210807,210805,210924,210706,210804,210419,210419,210415,210416,210416,210104,210104,201110,201110,201113,201113,201112,201112,201112,201020,201020,201019,201020,201019,201019,201020,200818,200818,200702,200702,190528,190527,181022,181226,181225,190410,181224,181226,180828,190312,190312,190312,190717,180828,190528,190528,190828,201110,201113,201112,201112,201112,201112,201112,201020,201020,201112,201019,200818,190313,190528,190528,181224,181022,181022,190312,190828,190527,190507,190528,181224,190312,190527,190828,210419,201111,201110,201020,200818,201020,180828,190410,190507,180828,181225,190410,230626,230623,230626,230605,230509,221117,221112,220525,220407,220407,220405,230629,230607,181224,190312,190312,190507,181226,181225,211117,211117,211117,210923,210807,210807,210706,210419,210419,210415,201111,201110,201113,201112,201020,200818,200702,190527,181226,181225,190718,190828,181224,181226,180828,230605,230509,230325,230324,211116,220530,220530,210804,220815,201111,190312,181225,181022,181224,190717,190717,180828,210924,210924,210922,210923,210805,210804,190528,201113,201112,181225,190527,181224,180828,210104,210416,180828,181022,180828,181022,181022,230605,230324,230326,230326,230326,230326,230326,181226,190528,181225,181022,190410,190717,190507,190507,181226,211115,210922,210922,210924,210805,210805,210804,210419,210419,210415,210415,210416,210104,210104,201111,201111,201111,201110,201113,201112,201112,201020,201020,201019,201019,200818,200818,200702,200702,181022,180828,190528,190527,190828,190312,180828,181225,181224,181226,221117,221112,221112,221011,221006,211116,220526,211115,220417,220406,220526,0002621105,220527,220407,220405,220103,220102,181022,181224,190528,211115,210922,210923,210419,210805,210706,210419,181225,190527,190312,190528,190717,230628,230623,230608,230626,230605,230608,230326,230509,230326,230326,230325,221112,220406,220530,210922,210415,201113,200818,230626,230629,230607,230605,230509,230325,221117,230325,221117,220527,181224,190507,190313,190507,190313,210924,210807,201112,201019,200818,190507,230607,230509,220525,221117,231027,221112,221112,220530,181022,210104,201112,200818,190528,190507,190718,181225,190528,210415,201111,201112,200818,200702,180828,181224,181226,190828,230607,181225,181022,181022,190527,190828,181224,190507,190528,190312,181226,190717,190828,190828,181224,190313,210922,210807,210807,210807,210706,201113,201113,190527,230608,230605,230509,221117,221112,211116,220525,230629,230623,230629,230608,230626,230509,230324,221117,221117,220528,221006,211116,190527,190507,190507,180828,190528,210706,190829,180828,190527,181226,190528,181224,230628,230626,230626,230628,230626,230607,230607,230607,230629,230509,230509,230509,230324,230324,230326,221117,221117,211115,220525,220525,220417,220406,220526,220407,220407,220407,220405,220406,210922,210706,211117,211116,211115,210922,210923,210924,210807,210805,210805,210706,210804,210804,210419,210419,210415,210415,210415,210415,210416,210104,201111,201110,201113,201113,201113,201112,201112,201020,201020,201020,201019,201019,201019,190528,190527,200818,200702,190528,190527,190828,190312,190718,210923,210923,210706,210706,210706,210415,210415,210415,230626,230623,230608,220527,220527,220530,220405,220103,220103,220406,190717,181225,190828,201019,200818,200818,190507,190507,190528,180828,190312,181022,190507,190313,190410,181226,181224,181224,190717,230608,230509,211117,200702,230629,230607,230509,230325,221117,221112,211116,220406,220530,220405,220527,220103,181225,181022,210804,201111,201019,190528,230608,230626,230629,230605,230509,230324,230325,221006,220405,220405,220103,220102,211115,181022,181224,211117,211116,210924,210922,210923,210923,210807,210807,210807,210807,210924,210706,210706,210804,210804,210419,210419,210419,210415,210415,210416,210104,201111,201111,201110,201113,200702,201113,201112,201112,201020,201020,201020,201019,201019,201019,200818,200818,200702,200702,181224,180828,190507,181226,190528,181225,181225,190717,180713,230628,230628,230626,230608,230629,230623,230623,230629,230626,230626,230626,230605,230605,230605,230605,230326,230324,230326,230325,221117,221117,221112,220815,211116,211116,211115,220525,220525,220417,220406,220526,220530,220527,220530,220405,220405,210922,211117,210924,210923,210923,210415,210415,201112,201020,230626,230607,230607,230605,230509,230324,230325,221117,221112,221006,211115,220417,220526,220530,220407,190312,210415,201111,201111,201112,200702,200702,181225,181225,181022,190718,190528,181022,181224,190312,181226,190313,190410,190528,190828,190528,190828,181022,181224,190410,181022,181226,210924,210924,210807,210807,210706,210706,210419,210419,201111,201110,201110,201113,201112,201019,200818,200424,200818,200702,200702,181226,181022,190410,181224,180828,190528,230608,230629,230629,230509,230509,221117,221117,221117,221112,220528,220528,220526,211115,220525,211115,220417,220417,220525,220417,201113,230324,230324,230628,230509,220525,220530,220103,210924,210419,201020,201019,190528,181224,220417,190507,210923,210804,201110,200818,190828,190313,200818,181225,181226,190507,210419,181022,180828,210924,210922,210923,210807,210805,210706,210804,210415,210415,210415,210416,210416,210416,210104,201111,201110,201110,201113,201113,201112,201113,201019,200818,200702,200702,230605,221112,221112,220417,220417,220417,220526,220530,220527,220405,220102,220103,220102,220405,220407,220103,230628,230628,230626,230629,230627,230627,230629,230623,230623,230608,230626,230608,230626,230626,230626,230607,230607,230607,230605,230605,230509,230324,230326,221117,220528,220406,220530,220406,220103,220103,211117,210923,210104,220405,190828,211117,210924,210922,210922,210923,210924,210415,210415,210416,210104,210104,201111,201110,201112,201020,201020,201020,200818,200702,200702,200702,180828,190528,181224,190507,230623,230623,230608,230626,230626,230605,230324,230325,230325,230325,221112,211115,220528,220530,220527,220407,220405,220103,211117,220530,181224,190717,181226,190528,181226,190410,210706,210415,210415,210104,201111,201019,201019,201020,190527,181224,190312,190312,190410,190313,190528,190528,181022,190507,210922,210104,201110,201019,190528,190312,230626,230608,220528,211116,220530,190507,211115,210922,210922,210805,210415,201111,190528,230607,230629,230324,221117,221006,220526,220530,220407,220405,211117,181225,190717,190527,180828,181226,190528,210924,210923,210807,210805,210419,210419,210415,210416,210104,201113,201113,201020,201019,200702,190410,181225,190527,190527,190527,181022,181224,190718,190718,230608,230605,230605,230509,230326,221112,211116,220525,220527,220407,220103,220103,190313,190312,190312,190507,190528,181022,210922,210923,210922,210807,210805,210805,210804,210805,210804,210415,210415,201112,210416,210416,210104,201111,210416,201111,201110,201113,201112,201112,201020,201020,201019,201019,200818,200702,180828,190410,190507,190528,181226,190718,190312,181226,190312,230628,230626,230629,230608,230626,230626,230607,230605,230605,211116,220102,220103,220102,190717,180828,210924,201111,201113,190527,190528,190527,190507,190507,190527,190507,201020,181225,220407,210804,201110,201110,190527,190828,190312,190507,210924,210922,210923,210807,210922,210805,210804,210804,210706,210419,210419,210415,210415,210416,210415,210104,201111,201113,201112,201020,200818,200702,181225,190718,231027,230623,230626,230607,230509,230326,221112,221112,211116,220525,220417,220526,220530,220527,220405,220103,220102,211117,210922,210804,210415,210416,201112,201020,201019,200818,190312,190507,181022,190717,181226,180828,180828,210924,210923,230623,230629,230608,230626,230607,230325,230325,221117,211116,220525,220526,220527,220530,220103,220103,190507,190410,210924,210805,210924,210804,210804,210706,210706,210419,201113,201112,201112,201020,180828,190313,190528,230629,230605,230607,230326,220526,220527,220406,220102,230623,210807,210807,210804,210415,201020,200702,190528,190828,190312,190718,181226,230623,230605,230605,230509,220526,220526,220102,190507,201110,190312,190717,230623,230623,230629,230608,230629,230605,230607,230607,230605,230605,230509,230324,230326,230326,220526,190528,190507,180828,190410,180828,190718,181225,181224,211117,210924,210922,210807,210807,210706,210706,210416,201113,201019,181225,190528,181224,190527,190312,190312,190507,190313,190410,181226,190718,190717,190717,181225,230608,230509,230326,221112,220406,220530,220405,220407,190313,180828,190410,220528,211116,210922,220103,220406,211117,220103,190828,210807,210104,201020,190718,181224,181022,181022,190527,190410,190528,211117,211117,211115,210924,210924,210922,210922,210922,210922,210923,210922,210923,210807,210807,210923,210923,210922,210805,210706,210419,210804,210804,210706,210706,210419,210419,210415,210415,210415,210416,210416,210416,201111,210104,201111,201111,201111,201111,201110,201110,201113,201113,201112,201112,201112,201112,201112,201020,201020,201020,201112,201020,201019,201019,200818,200818,201019,200702,200818,200702,200424,200818,190527,181224,181226,181225,230628,230628,230626,230608,230626,230629,230623,230623,230629,230608,230608,230626,230626,230607,230607,230607,230627,230605,230605,230629,230509,230509,230509,230509,230509,230326,230326,230326,230326,230325,221117,221117,221117,221112,221112,221112,221112,221112,221011,220528,211116,211116,220528,220526,211115,220525,220525,220417,220525,220406,220526,220530,220406,220526,220527,220527,220407,220102,220405,220103,220406,220103,200702,210104,220525,211117,210924,210805,210923,210924,210805,210804,210419,210706,210922,210805,210804,210805,210419,210419,210415,210416,210104,210419,201111,201111,210104,201111,210104,201111,201110,210104,201112,201113,201020,201110,201019,201112,200818,201019,200425,200702,200702,200702,200818,190507,190312,190717,181022,181224,181226,190313,190410,190313,181226,190528,190410,181226,181226,231027,230626,230629,230623,230608,230629,230626,230627,230607,230629,230605,230509,230509,230326,230326,230326,230325,221112,220528,211116,211115,211115,220417,220526,220526,220406,220530,220406,220406,220102,220102,190528,181225,190828,181225,181224,190312,190312,190717,190507,181022,190527,211116,210923,210924,210922,210924,210923,210805,210805,210706,201110,210805,210706,210419,210419,210419,210415,210415,210416,210104,210104,201111,201111,201111,201110,201113,201113,200425,200702,201112,201112,201113,201020,201019,201019,201019,201019,201019,200818,200818,200702,200702,200702,190528,190312,181224,190828,181022,190828,190527,190312,181224,181226,190718,190528,190527,180828,220102,181225,211117,210924,210922,210922,210923,210923,210807,210924,210419,210416,201111,201113,201020,181224,190718,190828,230623,230608,230629,230623,230608,230608,230608,230608,230608,230629,230626,230607,230607,230509,230605,230605,230509,230325,230324,230326,221117,230325,221117,221117,221117,221112,221112,211115,220528,211116,220528,220526,220526,220526,220527,220530,220530,220530,220407,220405,220103,211117,230626,230626,230627,230608,230629,230509,230509,230326,221117,221112,211116,220417,220526,220527,220527,220103,211117,230626,230605,230626,220525,220407,230628,230626,230629,230626,230629,230608,230607,230605,230509,221117,221117,221112,221006,220525,220525,220525,220103,220102,230326,181224,181224,181022,181022,210807,210416,230605,221112,221112,221112,220526,220405,220405,230509,230509,181022,190410,181226,211117,211115,210922,210924,210805,210706,210804,210804,210419,210415,210104,201111,201111,201110,201112,201020,200818,200818,190312,190903,190528,190410,190313,230628,230629,230623,230608,230629,230605,230324,230325,221117,211116,220526,220525,220526,220530,220103,220103,230605,190528,190528,230623,230626,230605,230324,221112,211115,220530,220406,190527,180828,211116,210415,210419,210104,201110,201112,201019,200424,190527,190718,181224,190313,200702,190528,200818,190528,180828,190527,190718,180828,180712,201019,181022,201112,230629,230623,230629,230608,230626,230509,230325,221117,220528,220525,221006,221006,220528,220407,220407,220407,220103,220103,220103,181225,190507,181224,181226,190313,181224,211117,211117,210922,210807,210922,210805,210805,210706,210415,210104,201111,201113,201020,201020,200818,200702,200702,200702,190828,190528,190312,190312,190312,181224,181022,181224,180828,181225,180828,230628,230628,230626,231027,230627,230627,230626,230623,230623,230623,230608,230608,230608,230608,230605,230605,230509,230509,230509,230509,230324,230324,230324,221112,221112,221112,221112,221006,211115,211115,211115,220528,220525,220417,220525,220417,220406,220406,220527,220527,220406,220103,190828,190717,181225,220530,230629,230608,230608,210104,190410,190507,220103,220103,220103,220102,230623,230626,230326,230325,210706,210415,201020,180305,181226,230607,221112,220527,210706,210416,200818,190312,220526,230626,210104,201111,201113,230608,221117,221112,220103,220102,220102,190528,190507,181226,201111,201020,200818,190828,190507,190528,220103,210706,210807,210706,220530,220103,220103,220102,210922,210807,210805,210706,210805,210804,210804,210804,210706,210419,210104,210419,201112,201112,201020,201020,201020,201019,201019,200818,200818,190527,181224,190507,190507,190717,190717,190828,190312,181226,181225,180828,190527,181226,190410,200818,200818,200702,200702,200702,200702,200702,190528,190527,190828,181225,180828,190312,181022,181022,190410,190528,181224,190718,190507,181226,190312,181022,190410,181225,190528,190717,181226,180828,181022,181226,230509,211116,211115,220417,220526,220530,220527,220406,220103,220103,211116,210922,210923,210923,210805,210805,210804,210415,210104,201019,220527,201020,201020,230626,230608,230626,230605,230509,221117,221112,211115,220525,220525,211115,220407,220405,220405,220103,210923,210923,210924,210807,210807,210807,210805,210923,210923,210805,210706,210804,210706,210419,210416,210419,201111,201111,201110,201113,201113,201019,201019,201019,200818,190717,181022,200818,180828,190828,190718,190312,190717,180828,190507,181226,190828,190828,190312,181022,181022,190527,211115,210922,210923,210807,210804,210419,210104,201111,201110,201112,200818,181022,180828,190528,181224,181022,190717,181225,230628,230628,230627,230629,230607,230607,230623,230605,230509,230326,230326,221117,221112,211116,220525,211115,220525,220406,220528,220527,220530,220407,220102,220102,190312,201019,200818,220527,211116,210924,210922,210923,210807,210805,210807,210706,210804,210419,210419,210419,210415,210416,210416,210104,210104,201110,201110,201113,201113,201112,221006,220528,211116,221006,221011,221006,220528,220528,211116,211116,220528,211115,211115,221006,220406,220526,220527,220527,220530,220527,220102,220103,220103,190507,181226,181225,181225,190410,180828,181022,190527,190528,190718,181224,190312,181224,181225,181022,190527,190507,181226,180828,190312,181225,181022,211117,211117,210922,211115,210924,210924,210924,210922,210922,210922,210924,210923,210924,210807,210706,210807,210805,210804,210923,210419,210419,210419,210419,210104,210104,210416,210104,201111,201111,201110,201113,201113,200424,200425,201020,201020,201020,201020,201019,201020,200818,200702,200425,180828,190312,190528,190507,190718,181226,190312,180828,180801,190507,190313,211116,220406,220526,220527,220530,220405,220405,220103,220103,210415,201112,201112,201019,190528,181226,190312,190527,210924,210924,210807,210706,210419,210415,210104,201113,201112,201019,200818,180828,190312,190410,181224,181226,181022,190718,190718,230626,230629,230608,230607,230605,230509,230509,230326,230326,221117,211116,220527,220407,220407,220102,220102,210924,210807,210706,181226,190528,190527,190828,190528,181226,210415,201113,201112,201020,200702,230509,230509,221117,230623,230626,230605,230324,210924,210419,210415,221112,220528,220102,220405,211117,210706,201113,201020,201019,201019,200702,200702,230607,230608,230509,230326,220406,220530,220406,230607,230509,230326,221112,220525,220526,220102,220102,210922,210807,210416,210104,201113,201112,200702,181226,190312,181224,181022,190717,190717,181226,190507,190507,190507,190507,210922,211117,210922,210923,210923,210805,210804,210419,210419,210415,210415,210416,210416,210416,210104,210104,201111,201111,201110,201113,201113,201020,201020,201019,201019,200818,200702,200702,190527,190828,181225,180828,190312,181022,181022,190528,190528,190828,190828,190507,181224,181225,181225,230628,230623,230626,230623,230623,230623,230608,230626,230626,230607,230607,230605,230605,230326,230325,230324,221117,221117,221112,211116,211115,211115,220417,220406,220527,220527,220527,220103,220103,211117,190312,211117,210924,210924,210922,210923,210807,210805,210924,210804,210804,210805,210419,210415,210415,210415,210415,210416,210416,210415,210104,201110,201113,201112,201020,201020,201019,201019,200818,200702,200702,200818,181225,190410,190828,180828,181224,181226,190528,230626,230626,230623,230629,230608,230626,230626,230607,230607,230605,230509,230324,230326,230326,230326,221117,221117,221112,221006,211116,220528,211115,220525,220417,220525,220406,220406,220526,220527,220530,220530,220407,220103,220102,220103,190828,190507,181022,210924,210924,210922,210924,210805,210805,210706,201112,201019,180305,180828,190528,230509,221011,211115,220525,220526,220527,220527,220103,190527,190828,190528,181226,181226,181225,181225,181022,181226,210419,210419,210415,210415,210416,210416,210104,210104,201111,201111,201111,201113,201113,201113,201112,201112,201020,201020,201019,201019,201019,200818,200818,200818,200702,190528,180828,181022,190410,181224,190507,190718,190312,181226,190312,190312,190312,181225,190528,181224,180828,181224,181022,190507,190717,190312,190528,190528,211117,211115,210922,210923,210807,210922,210706,210419,210415,210415,210104,210104,210104,201111,201113,201113,201113,201112,201112,201020,201020,200818,200424,200702,200702,200424,190410,190718,190312,190410,190313,181224,230628,230627,230629,230629,230629,230605,230605,230509,230324,230324,230324,230326,230325,221117,220528,220526,220525,220530,220407,220405,211117,220405,220102,210924,210924,200702,230607,211115,210924,210922,210922,210923,210923,210924,210419,210415,210415,210416,210415,210416,210416,210416,210104,210104,210104,210419,201111,201111,201111,201111,201110,201019,200818,230628,230628,230626,230627,230627,230626,230623,230623,230629,230608,230626,230629,230509,230605,230605,230509,230509,220528,211115,221006,220526,220527,220527,220530,220103,220103,181225,181022,190528,181224,190507,181226,190313,190828,190312,181022,181225,190527,190410,200818,200818,200702,200702,200818,180828,190527,181022,190410,181224,180828,181022,181225,190528,181224,180828,190312,201112,190312,190312,190528,190828,181225,210419,201020,190507,190507,190312,211117,210924,210922,181226,181226,180828,230628,230608,230627,230623,230626,230629,230605,221117,220417,220527,220103,181022,190312,181022,180828,181224,181022,180828,210807,210706,201111,201110,201110,201113,201113,201113,201113,200702,201019,201113,201112,201112,201112,201110,201020,201020,201020,201020,201020,201019,201019,201019,201019,200818,200818,200818,200702,200702,200702,200702,211117,210923,210924,210805,210805,210804,210419,210104,201110,201112,230626,230608,230629,230626,230607,230509,230324,221117,221112,211116,220528,220527,220530,220527,220407,220406,220103,181022,230628,230623,230629,230608,230608,230623,230608,181225,190828,181022,190527,211115,210924,210922,210923,210807,210805,210804,210415,210415,210415,210415,210416,210104,201111,201111,201110,201113,201113,201112,201112,201020,201019,201019,200702,200702,200702,190828,190828,180828,190312,181022,181022,190410,180828,181022,190528,190313,190507,181226,190313,230509,230605,230605,230325,221117,221117,221112,211116,220526,220525,220417,220526,220530,220530,220407,220102,220405,220103,220103,230628,210416,201111,201110,201113,201113,201112,201019,181226,181226,190410,190507,230626,230623,230509,230605,230326,230326,230326,221117,221112,211116,211115,220417,220417,220406,220527,211117,211117,181224,210922,210922,210807,210805,210415,201111,201113,201113,201019,181225,190718,190507,190312,181225,181225,181224,181022,180828,181022,190507,190717,190312,190828,190527,190527,181225,181022,181224,181224,190410,210922,211116,210922,210923,210923,210807,210706,210805,210419,210419,210415,210415,210416,210416,210104,210416,210104,201111,201111,201111,201113,201110,201110,201113,201113,201113,201112,201112,201112,201020,201020,201020,201020,201020,201019,200818,200424,200818,200818,200702,190528,180828,190528,190527,181224,190507,180828,180828,190312,181225,230623,230623,230623,230608,230608,230629,230607,230607,230607,230605,230605,230605,230509,230509,230324,230326,230326,230325,230325,221117,221117,221112,221112,211115,220417,220525,220417,220406,220527,220530,220527,220527,220407,220407,220407,220103,220103,220102,220102