Skip to content

Beckman Coulter, REF: OSR61173, IGM 4x14 mL R1, 4x11 mL R2 recallZ-2428-2024 · Beckman Coulter Inc. · initiated May 27, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Beckman Coulter Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

LIH Influence check settings for Lipemia are incorrect which may result patient samples being reported as false low results or cause a high result to report as normal

FDA root cause category: Process control

What the company is doing

On May 23, 2024, Beckman Coulter issued a "Urgent Medical Device Recall" notification via: xxx. 1. Update the LIH Influence Check settings for Lipemia on AU480, AU680, DxC 500 AU and DxC 700 AU analyzers if they are enabled for IgM. 2. To determine if the LIH Influence Check is enabled for IgM and to update your Lipemia setting, see customer communication. 3. Beckman Coulter recommends sharing the content of this letter with your laboratory and/or Medical Director to determine if a review of previous patient test results should be conducted. . Discontinuance or disposal of this product is not necessary. . Retain a copy of this letter as it serves as current labelling. . Per the IFU, avoid highly Lipemia samples when using the IgM assay. 4. Please complete the response form within 10 days in one of the following ways: - Electronically, if you received this communication via email./Manually, complete and return the enclosed Response Form. 5. If you have any questions regarding this notice, please contact our Customer Support Center: From our website: http://www.beckmancoulter.com/By phone: call 1-800-854-3633 in the United States. 6. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail or by fax.

Recall details

Recall number
Z-2428-2024
Status
Open, Classified
Initiated by the company
May 27, 2024
Posted by the FDA
Jul 24, 2024
Recalling firm
Beckman Coulter Inc.
Quantity
13,565 units
Product code
CFN
FDA submissions
K073487
Distribution
Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, IL, IN, KS, KY, LA, MD, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, WA, WI, WV, WY and the countries of Albania, Algeria, Argentina, Australia, Austria, Bangladesh, Belarus, Belgium, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, China (DX), Colombia, Croatia, Cyprus, Czechia, Denmark, Ecuador, Egypt, Estonia, France, Georgia, Germany, Gibraltar, Greece, Hong Kong, Hungary, India, Iraq, Ireland, Israel, Italy, Jordan, Kazakhstan, Korea, Republic of, Lebanon, Lithuania, Malaysia, Mayotte, Mexico, Morocco, Mozambique, Netherlands, New Zealand, Oman, Pakistan, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania, Russian Federation, San Marino, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, Spain, Sri Lanka, Switzerland, Taiwan, Province of China, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom of Great Britain and Northern Ireland, United States of America, Viet Nam, Yemen.

Product description

Beckman Coulter, REF: OSR61173, IGM 4x14 mL R1, 4x11 mL R2

Code information (models, lots, serial numbers)

UDI: 15099590011598/ All Lots: 2548/Exp. 2024-06-01, 2549/Exp.2024-10-01, 2550/Exp. 2024-11-01, 2551/Exp. 2025-04-01, 2552/Exp. 2025-07-01, 2553/Exp. 2025-10-01, 2554/ Exp. 2026-03-01

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page