Infinity recallZ-2432-2024 · Abbott Medical · initiated May 22, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Abbott Medical or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Their is a potential that the implantable pulse generator replacement indicator and end of service date may be shorter than indicated in the product labeling potentially resulting in loss of therapy.
FDA root cause category: Error in labeling
What the company is doing
On June 6, 2024, Abbott, issued a "Urgent Medical Device Correction" notification to affected consignees via mail. Abbott asked consignees to take the following actions: 1. This issue will continue to affect your patients for the foreseeable future. Abbott is working on an update to align the product labeling with actual device performance. 2. For information on when an ERI notification appears and how to make a determination with your patients on timing of IPG replacement, see the customer communication. 3. For recommended strategies for preventing lapse in therapy, see the customer communication. 4. Please inform your patients about this medical device correction letter. 5. Please return a completed Acknowledgement Form and maintain a record of this notice along with a copy of the completed Acknowledgement Form. 6. For questions about this issue or reported complaints, please contact your Abbott representative, or Abbott Technical Support at 1-800-727-7846. 7. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail or by fax.
Recall details
- Recall number
- Z-2432-2024
- Status
- Open, Classified
- Initiated by the company
- May 22, 2024
- Posted by the FDA
- Jul 25, 2024
- Recalling firm
- Abbott Medical
- Quantity
- 1198 units
- Product code
- MHY
- Distribution
- Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, DC, FL, GA, IL, IN, KY, LA, MA, MD, MI, MN, MO, MT, NC, ND, NE, NH, NJ, NY, OH, OK, OR, PA, PR, SC, TX, UT, VA, WA, WI, WY and the countries of Algeria, Argentina, Australia, Austria, Belgium, Brazil, Canada, Colombia, Cyprus, Czech Republic, Ecuador, Estonia, Finland, France, Germany, Greece, Hungary, India, Ireland, Israel, Italy, Japan, Jordan, Korea, Lithuania, Mexico, Netherlands, Norway, Pakistan, Poland, Portugal, Russian Federation, Saudi Arabia, South Africa, Spain, Sweden, Taiwan, Turkey, United Arab Emirates, United Kingdom.
Product description
Abbott Infinity 7, Implantable Pulse Generator, REF: 6663, SterileEO, Rx Only
Code information (models, lots, serial numbers)
All Serial numbers/UDI: 05415067020277
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Infinity™ 7St. Jude Medical, Inc. · STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR1 recall1 death47 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 94796).
- Z-2429-2024 · Open, ClassifiedAbbott Infinity 5, Implantable Pulse Generator, REF: 6660, SterileEO, Rx Only
- Z-2430-2024 · Open, ClassifiedAbbott Infinity 5, Implantable Pulse Generator, REF: 6661, SterileEO, Rx Only
- Z-2431-2024 · Open, ClassifiedAbbott Infinity 7, Implantable Pulse Generator, REF: 6662, SterileEO, Rx Only
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page