Skip to content

Presource kits and trays, labeled as recallZ-2329-2024 · Cardinal Health 200, LLC · initiated May 29, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Cardinal Health 200, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Cardinal Health is following the FDA's recommendations regarding the disposition of affected Jiangsu Shenli Medical Production Co. Ltd (Shenli) Syringes in the Merit Medical supplied finished kits. Merit Medical supplied finished kits are included as components in Presource kits. According to the FDA s communication, these Shenli syringes, located inside a Merit Medical kit component, cannot be distributed for use as they could contain a significant change or modification in design which could significantly alter the safety or effectiveness of the finished kit product, resulting in the risk of patient harm such as inaccurate dosing and a leaking device.

FDA root cause category: Device Design

What the company is doing

Cardinal Health issued an URGENT: PRODUCT RECALL notice to its consignees on 05/28/2024 via letter. The notice explained the issue, potential risks involved, and requested the following: "If you locate affected product within your facility, please complete the following: 1. QUARANTINE affected kits & ALERT your clinicians of this action 2. AFFIX a WARNING LABEL to the front of each Presource kit so that it is clearly visible to clinicians, instructing them to remove and discard all syringes within the affected Merit Medical kit in the Presource product. 3. RETURN a copy of the enclosed Acknowledgment Form confirming your receipt of this action and the completion of labeling product via fax to 614.652.9648. Additional labels will be sent to your facility upon request. Please send an email to: [email protected] For component credit or replacement, please contact the following: " Hospitals Cardinal Customer Service at 866.551.0533 " Distributors Distributor Management at 800.635.6021 " Federal Government facilities Government Customer Service at 800.444.1166 If you are a Distributor and have further distributed affected kits, please notify your customers of this recall." Expansion letter was sent via FedEx Express on 6/20/2024

Recall details

Recall number
Z-2329-2024
Status
Open, Classified
Initiated by the company
May 29, 2024
Posted by the FDA
Jul 10, 2024
Recalling firm
Cardinal Health 200, LLC
Quantity
17445 units
Product code
OEZ
Distribution
Nationwide

Product description

Presource kits and trays, labeled as: 1) Cardiovascular Procedure Kit, Catalog Number SAN29ANCC3; 2) Cardiovascular Procedure Kit, Catalog Number SAN29ANCC4; 3) Cardiovascular Procedure Kit, Catalog Number SAN29ANCCL; 4) Cardiovascular Procedure Kit, Catalog Number SAN29ANCCM; 5) Cardiovascular Procedure Kit, Catalog Number SAN33CDCAS; 6) Cardiovascular Procedure Kit, Catalog Number SAN33CDCAT; 7) Cardiovascular Procedure Kit, Catalog Number SANOCAENSA; 8) Cardiovascular Procedure Kit, Catalog Number SCV29ANCCA; 9) Left Heart Kit, Catalog Number SAN73LHKCC; 10) Left Heart Kit, Catalog Number SAN73LHKC6

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

1) SAN29ANCC3, UDI/DI (CS/EA) 50888439692868/10888439692860, Lot Numbers: 377162; 2) SAN29ANCC4, UDI/DI (CS/EA) 50888439791929/10888439791921, Lot Numbers: 404590; 3) SAN29ANCCL, UDI/DI (CS/EA) 50887488572077/10887488572079, Lot Numbers: 227924, 248336, 268411, 295823, 330354, 441771, 470382, 485647; 4) SAN29ANCCM, UDI/DI (CS/EA) 50888439916933/10888439916935, Lot Numbers: 538664; 5) SAN33CDCAS, UDI/DI (CS/EA) 50195594100887/10195594100889, Lot Numbers: 7500, 27481, 47777, 66619, 84103, 98759, 654804, 664274, 670490, 690813, 719128, 742731, 769997, 789979, 806872, 821751, 853574, 892089, 905085, 929736, 955110, 974399, 989732; 6) SAN33CDCAT, UDI/DI (CS/EA) 50197106394401/10197106394403, Lot Numbers: 105901, 124750, 166138, 184313, 213752, 228035, 247566; 7) SANOCAENSA, UDI/DI (CS/EA) 50195594650863/10195594650865, Lot Numbers: 60783, 75070, 104936, 117262, 149856, 164959, 192358, 195004, 265705, 842090, 846704, 868814; 8) SCV29ANCCA, UDI/DI (CS/EA) 50888439972533/10888439972535, Lot Numbers: 35435, 51272, 106300, 186438, 257038, 583559, 619303, 639545, 668446, 721926, 744092, 783887, 806636, 868787, 896632, 933968, 969941, 985419; 9) SAN73LHKCC, UDI/DI (CS/EA) 50195594056344/10195594056346, Lot Numbers: 25179, 29199, 52789, 124487, 137706, 203053, 756293, 826474; 10) SAN73LHKC6, UDI/DI (CS/EA) 50197106375561/10197106375563, Lot Numbers: 89496

Devices named in this recall

10 devices in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 94763).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page