Skip to content

AURORA Surgiscope recallZ-2331-2024 · Integra LifeSciences Corp. · initiated Jun 7, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Integra LifeSciences Corp. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Potential sheath damage that could lead to tissue damage and/or track hemorrhage.

FDA root cause category: Process change control

What the company is doing

A "1st NOTIFICATION - URGENT: VOLUNTARY MEDICAL DEVICE RECALL" letter, dated 6/7/24, was mailed to consignees. The notification instructs consignees to immediately remove any affected devices from service and place them in quarantine. The recall notification is to be forwarded to users of the product within your facility for their awareness. Consignees are to complete the provided response form and return it via email to [email protected] or by fax to 1-609-750-4220. Integra Customer Service will reach out to consignees regarding a Return Materials Authorization and return directions upon receipt of the response form. Consignees with any questions are to contact Customer Service from 8:00 AM to 8:00 PM from Monday to Friday by phone at 1-800-654-2873 or by email at [email protected]

Recall details

Recall number
Z-2331-2024
Status
Open, Classified
Initiated by the company
Jun 7, 2024
Posted by the FDA
Jul 11, 2024
Recalling firm
Integra LifeSciences Corp.
Quantity
14 units
Product code
GWG
FDA submissions
K232618
Distribution
Domestic: NY, OH, & DC.

Product description

AURORA Surgiscope System. Sterile, single use device that contains a Sheath, Obturator, and Imager.

Code information (models, lots, serial numbers)

Model No. ASX9/130; UDI# 00850002332278; Lot No. 7336314.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

1 device cleared under K232618 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page