DxI 9000 Access Immunoassay Analyzer, Catalog #C11137 recallZ-2735-2024 · Beckman Coulter, Inc. · initiated May 22, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Beckman Coulter, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

When a DxI 9000 Access Immunoassay Analyzer is connected to a Laboratory Automation System (LAS) and is running with the system software versions 1.17.0 and below, the analyzer does not send a message to the LAS when the sample wheel is full, and the LAS continues sending samples despite the limited capacity to process new samples. Therefore, sample tests may be aliquoted but not processed when the number of tests associated with onboard patient samples exceeds the throughput limit. If this issue occurs test results are flagged with a sample wheel timeout (SWT) error flag, and there may be a delay in reporting test results which could lead to a delay in patient treatment.

FDA root cause category: Software design

What the company is doing

The recalling firm issued letters dated 5/16/2024 on 5/22/2024 via postal mail and email notification. The letter described the issue, impact, action, and resolution. Beckman Coulter will implement the correction with a future software release and the consignee will be contacted by their Beckman Coulter service representative to schedule the software upgrade. If the device has been forwarded to another laboratory, the consignee is requested to provide a copy of the letter to that location. A response was requested within 10 days electronically via email or through completion and return of the enclosed Response Form.

Recall details

Recall number
Z-2735-2024
Status
Open, Classified
Initiated by the company
May 22, 2024
Posted by the FDA
Aug 27, 2024
Recalling firm
Beckman Coulter, Inc.
Quantity
183 analyzers
Product code
JJE
Distribution
Worldwide - US Nationwide distribution in the states of CA, FL, GA, IN, KS, ME, NH, NJ, OK, PA, TX, and WA. The countries of Australia, Austria, Bahrain, Belgium, Brazil, Croatia, Czechia, Egypt, France, Germany, Hungary, Ireland, Israel, Italy, Netherlands, New Zealand, Poland, Portugal, Slovakia, South Africa, Spain, Switzerland, Taiwan, and United Kingdom of Great Britain and Northern Ireland.

Product description

DxI 9000 Access Immunoassay Analyzer, Catalog #C11137.

Code information (models, lots, serial numbers)

Software versions 1.17.0 or below, UDI-DI 15099590732103.