Dxl 9000 Access Immunoassay Analyzer, Catalog #C11137 recallZ-2414-2024 · Beckman Coulter, Inc. · initiated Jun 4, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Beckman Coulter, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The analyzers were built with nonconforming USB cables made with brass contacts instead of specified gold contacts. Brass contacts may corrode over time, leading to loss of internal instrument communication and delayed patient results.
FDA root cause category: Nonconforming Material/Component
What the company is doing
Letters dated 5/29/2024 were issued via email and postal mail on 6/4/2024. The letter described the issue, the impact, and the action which was that the consignee's Beckman Coulter service representative will contact the consignee to replace parts affected by the issue. The firm will replace all affected USB 2.0 cables. The information in the letter is to be shared with the consignees laboratory staff and if the device has been forwarded to another laboratory, the consignee is requested to send them a copy of the letter. The consignee is requested to respond to the letter within 10 days either via email or by manually completing the enclosed Response Form.
Recall details
- Recall number
- Z-2414-2024
- Status
- Open, Classified
- Initiated by the company
- Jun 4, 2024
- Posted by the FDA
- Jul 23, 2024
- Recalling firm
- Beckman Coulter, Inc.
- Quantity
- 143 devices
- Product code
- JJE
- FDA submissions
- K221225
- Distribution
- Worldwide - US Nationwide distribution in the states of FL, IN, KS, NH, PA, and TX. The countries of Australia, Austria, Belgium, Brazil, Croatia, Czechia, Egypt, France, Germany, Hungary, Ireland, Israel, Italy, Netherlands, New Zealand, Poland, Portugal, Slovakia, South Africa, Spain, Switzerland, Taiwan, Province of China, and United Kingdom of Great Britain and Northern Ireland.
Product description
Dxl 9000 Access Immunoassay Analyzer, Catalog #C11137.
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
Serial numbers: 300117, 300116, 300137, 300138, 300139, 300129, 300141, 300143, 300142, 300151, 300152, 300128, 300156, 300153, 300154, 300135, 300164, 300180, 300155, 300167, 300166, 300158, 300171, 300179, 300169, 300182, 300176, 300183, 300198, 300192, 300133, 300188, 300161, 300134, 300190, 300209, 300212, 300211, 300149, 300148, 300150, 300130, 300172, 300197, 300160, 300189, 300170, 300181, 300195, 300194, 300224, 300222, 300168, 300210, 300233, 300207, 300205, 300221, 300203, 300240, 300199, 300200, 300136, 300196, 300178, 300220, 300208, 300126, 300213, 300206, 300219, 300214, 300217, 300215, 300229, 300255, 300257, 300202, 300254, 300223, 300258, 300259, 300218, 300201, 300234, 300237, 300249, 300260, 300238, 300243, 300261, 300231, 300191, 300131, 300236, 300264, 300263, 300193, 300239, 300230, 300235, 300245, 300216, 300247, 300242, 300226, 300270, 300271, 300269, 300174, 300251, 300241, 300232, 300204, 300227, 300228, 300256, 300266, 300281, 300282, 300253, 300244, 300272, 300246, 300145, 300123, 300147, 300297, 300157, 300275, 300284, 300286, 300248, 300273, 300252, 300144, 300146, 300165, 300177, 300173, 300250, 300268, and 300140; UDI-DI 15099590732103.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- DxI 9000 Access Immunoassay Analyzer C11137Beckman Coulter, Inc. · Analyzer, Chemistry (Photometric, Discrete), For Clinical Use17 recalls1 death3 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page