Skip to content

Meridian recallZ-2357-2024 · Seaspine Orthopedics Corporation · initiated Jun 4, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Seaspine Orthopedics Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Insufficient weld around the cap component of the instrument is insufficient to withstand impaction forces.

FDA root cause category: Process design

What the company is doing

On June 4, 2024, SeaSpine Orthopedics issued a "Urgent Medical Device Recall" Notification via E-Mail to affected consignees. SeaSpine Orthopedics ask consignees to take the following actions: We request that you return the affected part to SeaSpine s Carlsbad facility for evaluation and potential rework. To support your upcoming cases, a replacement Vu aPod Prime NM Angled Awl will be sent (PN: 33-40-1028). Additionally, please continue to utilize the loaner pool. Please use the attached label to return the inventory in your possession at your earliest convenience. Once the inventory has been returned, please acknowledge via DocuSign.

Recall details

Recall number
Z-2357-2024
Status
Open, Classified
Initiated by the company
Jun 4, 2024
Posted by the FDA
Jul 15, 2024
Recalling firm
Seaspine Orthopedics Corporation
Quantity
100 units
Product code
HWJ
Distribution
The product was distributed to the following US states: Nevada, Colorado, Texas, Washington, Indiana, Georgia, Michigan, Florida, Pennsylvania, Illinois, Tennessee, Virginia, Missouri, South Dakota, California, Montana, Minnesota, Arizona

Product description

Brand Name: Meridian Angled Awl Fixed Product Name: Angled Awl Fixed Model/Catalog Number: RA2-003015 Product Description: Surgical Awl Component: No

Code information (models, lots, serial numbers)

Lot Code: UDI 10889981216368 Lot Numbers TT0190H, TT0191H

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page