Brivo MR355, with affected software versions: SV20.3… recallZ-2477-2024 · GE Medical Systems, LLC · initiated Jun 14, 2024
Class II Open, Classified All recalls posted in August 2024
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from GE Medical Systems, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.
FDA root cause category: Device Design
What the company is doing
GE Healthcare issue an URGENT MEDICAL DEVICE CORRECTION to its consignees via letter on 06/14/2024. The letter explained the problem with the device, safety issue, and requested the following: "Actions to be taken by Customer/User: You can continue to use your MR system by following the instructions below: To ensure that the displayed B1+RMS value on the Scan Prescription User Interface does not exceed the user prescribed limit specified in the Low SAR Mode User Interface: " If your software version includes the Optimized T2 FLAIR Sequence option, enable this option by setting the value to 1.00. This option is present on the Advanced Tab of the Scan Prescription User Interface. " If your software version does not include the Optimized T2 FLAIR Sequence option, adjust scan parameters such as TR , # Slices or Auto Refocus Flip Angle until the displayed B1+RMS value on the Scan Prescription User Interface does not exceed the user prescribed limit. Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions." GE HealthCare will correct all affected products at no cost to you. A GE HealthCare representative will contact you to arrange for the correction. If you have any questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.
Recall details
- Recall number
- Z-2477-2024
- Recall class
- Class II
- Status
- Open, Classified
- Initiated by the company
- Jun 14, 2024
- Posted by the FDA
- Aug 2, 2024
- Recalling firm
- GE Medical Systems, LLC
- Quantity
- 375 units
- Product code
- LNH
- FDA submissions
- K103330, K120778, K123417
- Distribution
- Worldwide distribution.
Product description
Brivo MR355, with affected software versions: SV20.3, SV23.3, MR30.0, MR30.1; Nuclear Magnetic Resonance Imaging System
Code information (models, lots, serial numbers)
Show the full code information (4k characters)
System ID: MSF067-1-6AB94D05, DZ1570MR01, MRR9492, MRR10103, MRR10449, MRR9613, MRR9196, MRR9924, MRR9725, MRR10479, MRR9358, MRR9372MM3, MRR9496, 083027600051013, MRR8258, MRR10217, MRR8634, MRR9104, MRR10184, MRR9949, MRR8550, MRR9286, MRR9231, MRR9153, MRR8685, MRR10207, MRR8755, MRR9962, MRR8669, MRR8972, MRR9920, MRR10124, MRR10693, MRR10081, MRR9131, MRR8647, MRR9484, MRR8609, MRR10162, MR1714SX5, MRR8984, MRR8602, MRR10063, MRR9968, MRR9037, MRR9980, MRR9811, MRR10213, MRR10237, BG4608MR01, MRR9350, MRR9031, 082427110037, 082427110039, 082427220030, 082427220032, 082427220029, 082427220028, 082427190073, 082427210085, 082427150052, 082427150046, 082427030077, 082427160108, 082427170015, 082427170018, 082427040126, 082427200069, 082427010062, 082427010058, 082427210126, 082427200055, 082427290012, 082427160099, 082427230062, 082427090205, 082427270043, 082427310049, 082427160125, 082427160120, 082427090185, 082427110057, 082427160096, 082427230076, 082427100157, 082427150054, 082427150055, 082427160109, 082427180025, 082427180026, 082427240033, 082427240027, 082427240032, 082427210086, 082427070092, 082427070077, 082427090158, 082427090199, 082427090154, 082427090216, 082427090156, 082427090191, 082427090167, 082427090172, 082427090152, 082427090197, 082427090177, 082427090243, 082427210067, 082427210088, 082427210095, 082427190063, 082427190056, 082427190054, 082427190052, 082427190062, 082427190060, 082427190059, 082427190057, 082427190058, 082427200059, 082427200054, 082427200051, 082427200061, 082427200085, 082427200079, 082427200067, 082427200070, 082427020048, 082427020043, 082427020046, 082427030068, 082427200096, 082427090155, 082427100167, 082427100163, 082427100155, 082427100162, 082427310067, 082427230072, 082427230071, 082427230054, 082427230066, 082427230060, 082427270035, 082427060037, 082427020040, 082427230070, 082427110038, 082427100151, 082427120127, 082427200060, 082427200044, 082427200063, 082427200064, 082427070076, 082427070091, 082427090184, 082427090183, 082427090186, 082427170017, 082427210103, 082427210079, 082427210092, 082427190051, 082427190068, 082427150049, 082427240040, 082427220025, 082427310055, 082427310051, 082427010059, MRR8316, MRR10123, MRR9787, MRR8587, MRR8724, MRR9448, MRR8333, MRR9530, 2943672BRIVO, MRR8727, MRR9190, MRR8270, 805609MR01, GH5014MR01, MR5407GT, MRR8813, 083027600033412, 083027280005313, 083027280035313, 083027000002713, 083027240003113, 083027240009512, 083027240046912, 083027100004713, 083027100020013, 083027820034810, 083027860004512, 083027861104811, 083027868443311, 083027860028112, 083027200051414, 083027200494913, 083027208469611, 083027206189511, 083027201003119, 083027200037012, 083027207329111, 083027200044913, 083027205952123, 083027060039211, 083027044291311, 083027800002313, 083027800405712, 083027805544513, 083027849928710, 083027601012912, 083027607456310, 0835270033, 0835270043, 0835270030, 0835270023, 0835270047, A5163111, A364241704, MR332876HM2, MRR8312, PC5014, PC5022, PC5059, PC5079, PC5010, PC5017, PC5047, PC5073, PC5035, PC5045, PC5056, PC5060, PC5151, PC5072, PC5100, PC5033, PC5069, PC5001, PC5140, PC5015, PC5008, PC5023, PC5038, PC5025, PC5071, PC5046, PC5021, PC5150, PC5052, PC5039, PC5007, PC5051, PC5078, PC5049, PC5064, PC5057, PC5019, PC5044, PC5016, PC5076, PC5066, PC5036, PC5031, PC5011, PC5048, PC5018, PC5043, PC5054, PC5055, PC5003, PC5050, PC5040, PC5090, PC5067, PC5024, PC5063, PC5013, PC5026, PC5053, PC5068, PC5028, PC5061, PC5030, PC5037, PC5027, PC5006, PC5042, 0850270356, 0850270349, 0850270429, 0850270378, 0850270364, 0850270383, 0850270352, LV4058MR01, MRR8614, MA1085MR01, MA1421MR01, MA1014MR02, NG1065MR01, NG1127MR01, NG1097MR01, PL2893MR01, PL2997MR01, PL1809MR01, PL2717MR01, RO1221MR01, RU7464MR01, RU3028MR01, RU5248MR01, RU6147MR01, RU1203MR01, RU1778MR01, RU7816MR01, RU3240MR01, RU7539MR01, RU2652MR02, RU3908MR01, RU8225MR01, RU1997MR01, RU3759MR01, RU2607MR01, RU2272MR01, RU7426MR01, RU8300MR01, RU6261MR01, RU3609MR01, RU7865MR01, RU8407MR01, RU7312MR01, RU4306MR01, RU3794MR01, RU7827MR01, RU3469MR01, RU3776MR01, RU3951MR01, RU8151MR01, RF2883MR01, A5940817, ZA2315MR01, ZA2471MR01, UPW01801, PPS68403, MPD34400, NPD03800, 0856270060, 0856270059, 0856270058, 0856270056, 0856270057, 0856270051, 0856270069, 06138MRS01, 09015MRS01, 13007MRS01, 34340MRS01, 34334MRS01, 46014PRO06, 46014MRS01, 46016MRS01, 55021MRS01, 07027MRS01, 48005MRS01, UZ1020MR01, 0859270016, 0859270017, 0859270018.
Other devices under the same FDA submission
5 devices cleared under K103330, K120778, K123417 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- NA EXPRESS COIL - POSTERIOR ARRAYGe Healthcare(Tianjin) Company Limited · System, nuclear magnetic resonance spectroscopic1 recall0 reports
- NA EXPRESS COIL - HEAD NECK ARRAYGe Healthcare(Tianjin) Company Limited · System, nuclear magnetic resonance spectroscopic1 recall0 reports
- NA 1.5T EXPRESS COIL - 16E HEAD NECK ARRAYGe Healthcare(Tianjin) Company Limited · System, nuclear magnetic resonance imaging1 recall0 reports
- NA EXPRESS COIL - ANTERIOR ARRAYGe Healthcare(Tianjin) Company Limited · System, nuclear magnetic resonance spectroscopic1 recall0 reports
- NA 1.5T EXPRESS COIL - 9E ANTERIOR ARRAYGe Healthcare(Tianjin) Company Limited · System, nuclear magnetic resonance imaging1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 94874).
- Z-2459-2024 · Open, ClassifiedSIGNA Architect, with affected software versions: DV26.0 to DV26.3, DV27.0 to DV27.3, DV28.0 to DV28.5, DV29.0, DV29.1, DV29.2, MR30.0, MR30.1; Nuclear Magnetic Resonance Imaging …
- Z-2460-2024 · Open, ClassifiedSIGNA Pioneer, with affected software versions: PX25.0 to PX25.5, PX26.0, PX26.1, PX28.0 to PX28.4, PX29.0, PX29.1, MR30.0, MR30.1; Nuclear Magnetic Resonance Imaging System
- Z-2461-2024 · Open, ClassifiedSIGNA Hero, with affected software versions: PX29.1, MR30.0, MR30.1; Nuclear Magnetic Resonance Imaging System
- Z-2462-2024 · Open, ClassifiedSIGNA Premier, with affected software versions: RX27.0 to RX27.3, RX28.0, RX29.0 to RX29.2, MR30.0, MR30.1; Nuclear Magnetic Resonance Imaging System
- Z-2463-2024 · Open, Classified3.0T SIGNA HDxt Family, with affected software versions: RX27.0 to RX27.3, RX28.0, RX29.0 to RX29.2, MR30.0, MR30.1HD16.0_V03, HD23.0_V03; Nuclear Magnetic Resonance Imaging Syste…
- Z-2464-2024 · Open, ClassifiedDiscovery MR750 3.0T, with affected software versions: DV24.0, DV25.0, DV25.1, DV26.0 to DV26.4, DV29.1, MR30.0, MR30.1; Nuclear Magnetic Resonance Imaging System
- Z-2465-2024 · Open, ClassifiedDiscovery MR750w 3.0T, with affected software versions: DV24.0, DV25.0, DV25.1, DV26.0 to DV26.5, DV29.1, MR30.0, MR30.1; Nuclear Magnetic Resonance Imaging System
- Z-2466-2024 · Open, ClassifiedSIGNA PET/MR, with affected software versions: MP24.0, MP26.0, MP26.1, MR30.0, MR30.1; Tomographic imager combining emission computed tomography with nuclear magnetic resonance
- Z-2467-2024 · Open, ClassifiedDiscovery MR450 1.5T, with affected software versions: DV24.0, DV25.0, DV25.1, DV26.0; Nuclear Magnetic Resonance Imaging System
- Z-2468-2024 · Open, ClassifiedOptima MR450w 1.5T, with affected software versions: DV24.0, DV25.0, DV25.1, DV26.0, DV29.1, MR30.0, MR30.1; Nuclear Magnetic Resonance Imaging System
- Z-2469-2024 · Open, ClassifiedSIGNA Artist, with affected software versions: DV26.0, DV27.1, DV28.1, DV29.1, MR30.0, MR30.1; Nuclear Magnetic Resonance Imaging System
- Z-2470-2024 · Open, ClassifiedSIGNA Voyager, with affected software versions: PX26.0 to PX26.6, VX28.0, VX29.1, VX29.2, MR30.0, MR30.1; Nuclear Magnetic Resonance Imaging System
- Z-2471-2024 · Open, Classified1.5T SIGNA HDxt Family, with affected software versions: HD16.0_V03 to HD16.4, HD23.0_V03, HD28.0, HD29.1, MR30.0, MR30.1; Nuclear Magnetic Resonance Imaging System
- Z-2472-2024 · Open, ClassifiedSIGNA Creator, with affected software versions: HD16.0_V03 to HD16.4, HD23.0_V03, HD28.0, HD29.1, MR30.0, MR30.1; Nuclear Magnetic Resonance Imaging System
- Z-2473-2024 · Open, ClassifiedSIGNA Explorer, with affected software versions: SV25.5, SV25.6, SV29.2, MR30.0, MR30.1; Nuclear Magnetic Resonance Imaging System
- Z-2475-2024 · Open, ClassifiedSIGNA MR355, with affected software versions: SV25.5, SV25.6; Nuclear Magnetic Resonance Imaging System
- Z-2476-2024 · Open, ClassifiedSIGNA MR360, with affected software versions: SV25.5, SV25.6; Nuclear Magnetic Resonance Imaging System
- Z-2478-2024 · Open, ClassifiedOptima MR360, with affected software versions: SV20.3, SV23.3, MR30.0, MR30.1; Nuclear Magnetic Resonance Imaging System
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page