Skip to content

Perfusor recallZ-2434-2024 · B Braun Medical Inc · initiated Jun 14, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from B Braun Medical Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Perfusor Space Syringe Pump's instructions for use previously identified Monoject syringes as an option for use. However, changes to manufacturing to Cardinal Health Monoject syringes may result in issues with recognition, compatibility, and pump performance, such as overdose, underdose, delay in therapy, delay in occlusion alarms, and delay in feeding. Cardinal Health Monoject syringes are not compatible with the indicated pumps.

FDA root cause category: Device Design

What the company is doing

On June 14, 2024, the firm notified customers via URGENT MEDICAL DEVICE FIELD CORRECTION NOTIFICATION letters. Customers were advised to not use the recalled syringes in Perfusor Space Syringe Pumps. All users should be informed of the correction. Customers were provided with an attachment to post in care areas as a reminder to not use the recalled syringes with the specified pumps.

Recall details

Recall number
Z-2434-2024
Status
Open, Classified
Initiated by the company
Jun 14, 2024
Posted by the FDA
Jul 25, 2024
Recalling firm
B Braun Medical Inc
Quantity
9943
Product code
FRN
FDA submissions
K172831, K191910
Distribution
Domestic distribution nationwide. International distribution to Canada.

Product description

Brand Name: Perfusor Space Product Name: Perfusor Space Syringe Pump, Wireless Model/Catalog Number: 8713031U Software Version: 8713031U Product Description: Perfusor Space Syringe Pump, Infusion Pump, 1 each Component: N/A

Code information (models, lots, serial numbers)

Lot Code: Model No 8713031U; UDI-DI 4046964797958; All Serial Numbers

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 94890).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page