IntelliVue Patient Monitor MX550 with L.x and M.x recallZ-2511-2024 · Philips North America Llc · initiated Jun 28, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Philips North America Llc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Philips lntelliVue Patient Monitors MX400/450/500/550 Shipped or Upgraded with Incorrect Software Options. Changes were made to reflect some options becoming standard capabilities for software-version N.x. In the process, the entitlements of software options MOS, M06, and M20 were removed for software versions L.x and M.x. As a result of these changes Patient Monitors manufactured with or updated to the latest versions of software L.x or M.x will not offer the capabilities> Mmonitors with software version K.x may experience this issue too, due to software version K being out of support and the devices being provided with L.x entitlements offered by software options MOS, M06, and M20. Potential for incorrect or delayed treatment for the patient,
FDA root cause category: Software design
What the company is doing
Philips North America issued URGENT Medical Device Correction on 6/25/24 to End-Users. Philips Market Organizations will distribute the letters outside of the U.S. Letter states reason for recall, health risk and action to take: Use only default measurement capabilities, as indicated in Table 1 above, for monitoring purposes. Pass this notice to all those who need to be aware within your organization or to any organization where affected devices have been potentially transferred. A Philips representative will contact you to schedule a visit from a Philips Field Service Engineer who will reload the current device software to enable missing options (MOS, M06, M20). If you need any further information, please contact your local Philips representative or call 1-800-722-9377.
Recall details
- Recall number
- Z-2511-2024
- Status
- Open, Classified
- Initiated by the company
- Jun 28, 2024
- Posted by the FDA
- Aug 5, 2024
- Recalling firm
- Philips North America Llc
- Quantity
- 9 units US
- Product code
- MHX
- FDA submissions
- K130849
- Distribution
- Nationwide Foreign: Australia Austria Bangladesh Brazil Canada Chile China Colombia Denmark France Germany Hong Kong India Indonesia Italy Japan Netherlands Portugal Singapore South Africa South Korea Spain Sri Lanka Sweden Switzerland Taiwan Thailand United Kingdom Vietnam
Product description
IntelliVue Patient Monitor MX550 with L.x and M.x; K.x provided with L.x entitlements Model Number: 866066. Indicated for use by health care professionals for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adults, pediatrics, and neonates.
Code information (models, lots, serial numbers)
UDI: 00884838038783
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- IntelliVue MX550 Patient Monitor IntelliVue Patient Monitor MX550Philips Medizin Systeme Böblingen GmbH · MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)5 recalls12 deaths256 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 94919).
- Z-2508-2024 · Open, ClassifiedIntelliVue Patient Monitor MX400 with L.x and M.x; K.x provided with L.x entitlements Model Number: 866060. Indicated for use by health care professionals for monitoring and recor…
- Z-2509-2024 · Open, ClassifiedIntelliVue Patient Monitor MX450 with L.x and M.x; K.x provided with L.x entitlements Model Number: 866062. Indicated for use by health care professionals for monitoring and reco…
- Z-2510-2024 · Open, ClassifiedIntelliVue Patient Monitor MX500 with L.x and M.x; K.x provided with L.x entitlements Model Number: 866064. Indicated for use by health care professionals for monitoring and recor…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page