Skip to content

Preat Implant Prosthetics recallZ-2687-2024 · Preat Corp · initiated Jun 25, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Preat Corp or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Due to a manufacturing issue, the product has a out of specification rotation of the digital analog used to visually inspect dental abutments and restorations.

FDA root cause category: Process control

What the company is doing

On 07/22/2024, the firm sent an "URGENT: MEDICAL DEVICE RECALL UPDATE" Letter via certified mail informing customers that Digital Analogs were manufactured out of specification for the rotational feature which may result in the dental restoration/implant may be slightly different than the position of the actual implant in the patient's mouth. Customers are instructed to: Product in the scope of this recall should be discarded or returned to PREAT CORPORATION. For questions contact Quality Regulatory Affairs at 224-622-7191 or email [email protected]

Recall details

Recall number
Z-2687-2024
Status
Open, Classified
Initiated by the company
Jun 25, 2024
Posted by the FDA
Aug 19, 2024
Recalling firm
Preat Corp
Quantity
185 devices
Product code
NDP
Distribution
U.S. Nationwide distribution in the states of AK, AZ, FL, GA, IL, LA, MD, MI, MN, NY, OH, OK, OR, PA, RI, TX, UT, and WA

Product description

Multiple Digital Analog Products labeled as: NobelBiocare Tri-Lobe-compatible 6.0mm Digital Analog, REF 9000703 Straumann Tissue-Level-compatible WN Digital Analog, REF 9001303 3i Certain-compatible 5.0mm Digital Analog, REF 9001903 3i Certain-compatible 6.0mm Digital Analog, REF 9002003 Astra-compatible 3.0mm Digital Analog, REF 9002103 Astra-compatible Aqua 3.5/4.0mm Digital Analog, REF 9002203 Astra-compatible Lilac 4.5/5.0mm Digital Analog, REF 9002203 BioHorizons-compatible 3.0mm Digital Analog, REF 9005203 Legacy-compatible 3.0mm Digital Analog, REF 9005303 Digital Analog - BioHorizons - 3.5mm, Digital Analog - BioHorizons - 4.5mnm, REF 9005403 Legacy-compatible 3.5mm Digital Analog, REF 9005503 Legacy-compatible 5.7mm Digital Analog, REF 9008503 Astra EV 4.8mm Milled Titanium Abutment REF 9006767 Digital analogs are used for visual inspection of abutments and restoration in the dental laboratory.

Code information (models, lots, serial numbers)

Model Numbers: 9000703, 9001303, 9001903, 9002003, 9002103, 9002203, 9002203, 9005203, 9005203, 9005303, 9005403, 9005503, 9008503, 9006767 UDI-DI codes: 00842092178690, 00842092181898, 00842092181959, 00842092178737, 00842092178775, 00842092178812, 00842092178812, 00842092178898, 00842092178898, 00842092178997, 00842092178928, 00842092178959, 00842092182130 Lot Numbers: 250219, 248890, 251432, 250602, 250418, 250213, 251459, 249071, 250335, 250223, 250336, 249510, 250215.

Devices named in this recall

11 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page