CryoValve SG Pulmonary Human Heart Valve, SGPV00… recallZ-2531-2024 · Artivion, Inc · initiated May 28, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Artivion, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Grafts were found to have been approved, released, shipped and implanted with an invalid test results.

FDA root cause category: Under Investigation by firm

What the company is doing

The recalling firm issued letters to their Field Representatives on 5/24/2024 or 5/28/2024 for distribution to the consignees via email or hand delivery. The first date consignees were notified was on 5/28/2024 by their Field Representative. The letter provided the details on the tissue the consignee received, the date it was implanted, and the implanting surgeon. The issue was described and it was pointed out there was no indication the impacted tissues were contaminated and unlikely to pose a health risk to the patient. The letter discussed the testing conducted on each allograft and that the identified allograft was reported to have an invalid Post-Processing Sterility test result or a Pre-Processing Culture test result (depending on the consignee letter), which represents a technical failure in the test, not an indication of a positive result. A response form was enclosed for return to acknowledge receipt of the notification.

Recall details

Recall number
Z-2531-2024
Status
Open, Classified
Initiated by the company
May 28, 2024
Posted by the FDA
Aug 8, 2024
Recalling firm
Artivion, Inc
Quantity
2 valves
Product code
OHA
FDA submissions
K092021
Distribution
US Nationwide distribution in the states of CA, DC, FL, and NY.

Product description

CryoValve SG Pulmonary Human Heart Valve, SGPV00, Cryopreserved Decellularized Heart Valve Allograft Prepared with SynerGraft (SG) Technology.

Code information (models, lots, serial numbers)

Serial #11673489, exp. 27Jan2026; and serial #12657716, exp. 15Aug2028, UDI-DI: 00877234000447

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 94909).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page