Why it was recalled

Potential for defective test cartridges which may result in false positive results for the Staphylococcus lugdunensis and Enterococcus faecium targets.

FDA root cause category: Under Investigation by firm

What the company is doing

Diasorin sent a recall notification on 06/20/2024, via email. Consignees were instructed to examine inventory and quarantine any affected units, arrange for replacement and confirm disposal of any of affected units on hand. Customers were asked to perform a retrospective review of results for VERIGENE Gram-Positive Blood Culture Nucleic Acid Test (BC-GP) to look for any false positives, distribute the notification to any relevant personnel, notify consignees if further distributed and complete and return the "Acknowledgement & Receipt Form."

Recall details

Recall number
Z-2533-2024
Status
Open, Classified
Initiated by the company
Jun 20, 2024
Posted by the FDA
Aug 8, 2024
Recalling firm
Luminex Corporation
Quantity
1,100 kits
Product code
PAM
FDA submissions
K122514
Distribution
US Nationwide distribution including in the states of AZ, CA, CO, DE, FL, GA, IA, IL, KY, MD, MI, MN, MO, MS, NE, NV, OH, OK, SC, TN, TX, VA, VT, WA, WV.

Product description

Verigene Gram Positive Blood Culture (BC-GP) Nucleic Acid Test

Code information (models, lots, serial numbers)

GTIN Number: 00840487101575; Catalog Number: 20-006-018; Lot Number: 031824018A; Expiration Date: 09/16/2024