Nanoplasty System recallZ-2718-2024 · Treace Medical Concepts, Inc. · initiated Jun 2, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Treace Medical Concepts, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Potential issue causing a higher frequency of interference with locking screws and/or inserter rods than anticipated, resulting in screws stripping and/or not fully seating in the plates and/or interface issues between inserter rods and plates.

FDA root cause category: Under Investigation by firm

What the company is doing

Distributors were contacted via phone or in person to hold product starting on 06/02/2024. Recall notifications were disseminated on 06/13/2024 via email containing the recall letter, acknowledgement, and shipping label for return. On 07/29/2024, Treace Medical contacted surgeons via letter. Surgeons were instructed to monitor patients for adverse events and to complete and return the acknowledgement form.

Recall details

Recall number
Z-2718-2024
Status
Open, Classified
Initiated by the company
Jun 2, 2024
Posted by the FDA
Aug 23, 2024
Recalling firm
Treace Medical Concepts, Inc.
Quantity
40 kits containing 40 plates
Product code
HRS
FDA submissions
K240173
Distribution
US Nationwide distribution in the states of CO, FL, MD, MI, NC, SC, TX, VA.

Product description

Nanoplasty 3D Bunion Correction System Implant, Model Number: SK57. System contains the 01-00066 1st Metatarsal Plate as a component.

Code information (models, lots, serial numbers)

UDI: (01)00810111222464(17)290424(10)300542404

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page