Ventana anti-CD10 (SP67) Rabbit Monoclonal Primary… recallZ-2644-2024 · Ventana Medical Systems, Inc. · initiated Jun 27, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Ventana Medical Systems, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Their is a potential for high background and off-target staining when using the monoclonal primary antibody which can result in false positive results.

FDA root cause category: Unknown/Undetermined by firm

What the company is doing

On July 5, 2024, Ventana Medical Systems, Inc. a subsidiary of Roche Diagnostics issued a "Urgent Medical Device Correction" via UPS to affected consignees. Ventana asked consignees to take the following actions: 1. Discontinue use and discard any remaining affected product in your inventory according to your local waste guidelines. 2. Please consult with the physician or pathologist at your facility to determine specific clinical implications for your patients. 3. Review all positive results generated with the affected lots. Follow your standard laboratory operating procedures to investigate any suspected false positive results generated with the affected lots. 4. If you performed your initial validation using an affected lot, you should revalidate. 5. Refer to the Product Replacement or Credit section on page three of this UMDC. To receive replacement product OR credit for the unexpired product currently in your inventory, use the following link to access the form: https://go.roche.com/TP-02194. 6. Complete all sections of the enclosed Recall Response form (TP-02195); fax or email it according to the instructions on the form even if you are not requesting replacement product or credit. " Share this notification with all device users within your facility and network to ensure they are aware of this recall. This recall notice should also be shared with any organization where the potentially affected devices have been transferred. " File this Urgent Medical Device Correction (UMDC) for future reference.

Recall details

Recall number
Z-2644-2024
Status
Open, Classified
Initiated by the company
Jun 27, 2024
Posted by the FDA
Aug 16, 2024
Recalling firm
Ventana Medical Systems, Inc.
Quantity
22,839 kits
Product code
NJT
Distribution
Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV and the countries of Austria, Belgium, Croatia, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Greece, Hungary, Ireland, Italy, Latvia, Lithuania, Netherlands, Norway, Poland, Romania, Slovakia, Spain, Sweden, Switzerland, Albania, Argentina, Armenia, Australia, Azerbaijan, Bahrain, Belarus, Brazil, Cameroon, Chile, China, Colombia, Costa Rica, Cuba, Dominican Republic, Ecuador, Egypt, El Salvador, Georgia, Ghana, Guatemala, Honduras, Hong Kong, India, Iran, Iraq, Israel, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Lebanon, Malaysia, Mali, Mauritania, Mauritius, Mexico, Moldova, Mongolia, Morocco, Myanmar, Nepal, New Zealand, Nicaragua, Nigeria, North Macedonia, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Russian Fed., Rwanda, Saudi Arabia, Serbia, Singapore, South Africa, South Korea, Sri Lanka, Taiwan, Tanzania, Thailand, T¿rkiye, Uganda, United Arab Emirates, United Kingdom, Uruguay, and Vietnam.

Product description

Ventana anti-CD10 (SP67) Rabbit Monoclonal Primary Antibody, REF: 790-4506, IVD

Code information (models, lots, serial numbers)

UDI: 04015630972579/J04613, J11853, J17541, J25047, J30286, K00982, K06239, K09880, K14266, K19784, K26461, and M00669