Baxter Spectrum Wireless Battery Modules A/B/G/N, product… recallZ-2581-2024 · Baxter Healthcare Corporation · initiated Jul 17, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Baxter Healthcare Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The Spectrum WBM may fail to auto-document infusion status information back to the hospital's Electronic Medical Record (EMR) system. This software issue only impacts customers where the Spectrum V8 or Spectrum IQ infusion system is integrated with the hospital s EMR system.
FDA root cause category: Under Investigation by firm
What the company is doing
Baxter issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 07/16/2024 via USPS first class mail. The notice explained the issue, hazard involved, and requested the following: Actions to be Taken by Customers 1. Customers may continue to use Spectrum V8 and Spectrum IQ infusion pumps. If clinicians observe a lack of auto-documentation within the flow sheet or medication administration record, they should verify medication was given, and document in the EMR as if manually programming the pump. The pump will need to be programmed manually rather than through auto-programming for subsequent infusions. 2. A Baxter representative will contact your facility to determine the correction plan and schedule the software upgrade. The representative will work with you to determine a list of affected serial numbers at your facility. Please note you will be receiving this software upgrade from Baxter at no charge. 3. If you received this communication directly from Baxter, please acknowledge receipt of this notification by responding on our customer portal at https://BaxterFieldActionCustomerPortal.onprocess.com, even if you do not have any remaining inventory. Log in to the portal using the account number listed in the enclosed reply form instruction sheet. Acknowledging receipt of this notification will prevent you from receiving repeat notices. If you do not complete the acknowledgement, you will receive a phone call from OnProcess Technology on behalf of Baxter to confirm your receipt of this notification. 4. If you purchased this product from a distributor or wholesaler, please note that responding on the Baxter customer portal is not applicable. If a response is requested by your distributor or wholesaler, please respond to the supplier according to their instructions. 5. If you distributed this product to other facilities or departments within your institution, please forward a copy of this communication to them. 6. If you are a dealer, wholesaler, d
Recall details
- Recall number
- Z-2581-2024
- Status
- Open, Classified
- Initiated by the company
- Jul 17, 2024
- Posted by the FDA
- Aug 16, 2024
- Recalling firm
- Baxter Healthcare Corporation
- Quantity
- 17,377 units
- Product code
- FRN
- FDA submissions
- K230022
- Distribution
- US, Nationwide
Product description
Baxter Spectrum Wireless Battery Modules A/B/G/N, product code 35223, used with either of the following: 35700BAX2 (Spectrum V8 infusion pump) or 3570009 (Spectrum IQ infusion pump) Baxter SIGMA Spectrum V8 Infusion Pump
Code information (models, lots, serial numbers)
Pump GTIN 00085412498683; All serial numbers with radio software versions 22D28, 22D29, and 22D30
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- SIGMA Spectrum Infusion Pump 35700BAX2Baxter Healthcare Corporation · Infusion safety management software9 recalls12 deaths11,585 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 94988).
- Z-2582-2024 · Open, ClassifiedSpectrum Wireless Battery Modules A/B/G/N, product code 36010, used with either of the following 35700BAX2 (Spectrum V8 infusion pump) and 3570009 (Spectrum IQ infusion pump)
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page