FilmArray Gastrointestinal (GI) Panel recallZ-2514-2024 · BioFire Diagnostics, LLC · initiated May 16, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from BioFire Diagnostics, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Due to increase of false positive Vibrio/Vibrio Cholerae results when using a certain transport medium.
FDA root cause category: Under Investigation by firm
What the company is doing
On 07/11/2024, the firm emailed an "URGENT FIELD SAFETY NOTICE" to customers informing them the firm has identified an increased risk of false positive Vibrio/Vibrio Cholerae results when the BIOFIRE GI Panel is used with certain lots of Remel" Cary-Blair transport medium (see Appendix A - Affected Remel" Cary-Blair Lots). Customers are instructed to: -If the BIOFIRE GI Panel is used in conjunction with certain lots of Remel" Cary-Blair (Lot # 743930, 769240, 712258, 782425, 732357, and 742049), positive results for Vibrio/Vibrio Cholerae should be confirmed by another method prior to reporting the test results. -Distribute this information to all appropriate personnel in your laboratory, retain a copy for your files, post this letter in or near the laboratory, and forward this information to all parties that may use this product, including others to whom you may have transferred our product. -Complete the Acknowledgement Form in Attachment A and return it to [email protected], so that bioMrieux may acknowledge your receipt of this notification. It is important that you return the acknowledgment form to bioMrieux even if you determine this urgent field safety notice does not impact your facility. For assistance or any questions, contact the bioMrieux Technical Support Center by email at [email protected] or (800) 682-2666. The anticipated turnaround time for email inquiries is 24 48 hours during normal business hours.
Recall details
- Recall number
- Z-2514-2024
- Status
- Open, Classified
- Initiated by the company
- May 16, 2024
- Posted by the FDA
- Aug 6, 2024
- Recalling firm
- BioFire Diagnostics, LLC
- Quantity
- 67,613 kits (or 2,028,309 pouches). Each pouch is used for each individual test.
- Product code
- PCH
- FDA submissions
- K140407, K160459, K230404, K143005
- Distribution
- U.S.: Nationwide - AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, Virgin Island, VT, WA, WI, WV, and WY. O.U.S.: Worldwide - AR, AU, AT, BE, BR, CA, CL, HK, CN, CO, CZ, DE, FI, FR, GR, HU, IN, IT, JP, MX, NL, NO, PH, PL, PT, SG, ZA, KR, ES, SE, CH, TW, TH, TR, GB, AL, DZ, AO, AM, BH, BY, BO, BA, BW, BG, CR, HR, CY, CD, DO, AN, EC, SV, EE, FJ, GE, GU, GT, GN, HT, HN, IS, ID, IQ, IE, IL, CI, CI, JO, KE, KW, LV, LB, LT, MO, MY, ML, MR, MD, MN, MA, MM, NP, NZ, NZ, NI, OM, PK, PA, PG, PY, PE, QA, RO, SA, RS, SI, TN, TR, UY, AE, VE, VN, and ZW.
Product description
BIOFIRE FILMARRAY Gastrointestinal (GI) Panel REF RFIT-ASY-0116 (30 test kit) BIOFIRE FILMARRAY Gastrointestinal (GI) Panel REF RFIT-ASY-0104 (6 kit test) The BIOFIRE FILMARRAY Gastrointestinal (GI) Panel is a qualitative multiplexed nucleic acid-based in vitro diagnostic test intended for use with BIOFIRE FILMARRAY Systems. The BIOFIRE GI Panel is capable of the simultaneous detection and identification of nucleic acids from multiple bacteria, viruses, and parasites directly from stool samples in Cary Blair transport media obtained from individuals with signs and/or symptoms of gastrointestinal infection.
Code information (models, lots, serial numbers)
Part Number: RFIT-ASY-0116 (30 test kit) UDI-DI code: 00815381020109 0081538102011: All unexpired products are potentially impacted if used in conjunction with the Remel Cary Blair transport medium Part Number: RFIT-ASY-0104 (6 test kit) UDI-DI code: 0081538102011 All unexpired products are potentially impacted if used in conjunction with the Remel Cary Blair transport medium Remel Cary-Blair Transport Medium lot numbers: 743930 769240 712258 782425 732357 742049
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- FilmArray® Gastrointestinal (GI) Panel RFIT-ASY-0116Biofire Diagnostics, LLC · Gastrointestinal Pathogen Panel Multiplex Nucleic Acid-Based Assay System6 recalls2 deaths24 reports
Product lines
Other devices under the same FDA submission
1 device cleared under K140407, K160459, K230404, K143005 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- FilmArray® Gastrointestinal (GI) Panel RFIT-ASY-0104Biofire Diagnostics, LLC · Gastrointestinal Pathogen Panel Multiplex Nucleic Acid-Based Assay System1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page