Skip to content

Product Name: Salivary Testosterone ELISA recallZ-2622-2024 · DRG International, Inc. · initiated Jul 5, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from DRG International, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Invalid Controls. The Low Control is out of range; too high which causes invalid runs that would result in repeating the runs and delays for reporting results to patients.

FDA root cause category: Nonconforming Material/Component

What the company is doing

On July 8, 2024 Urgent Field Safety Notification letters were sent to customers. Action to be taken by Distributors and End-users: Please check your inventory for the respective lots 66K014 and 66K123. and stop using these kits immediately and promptly quarantine them. If you faced this issue of failed controls, kindly report with daya, and request a replacement for the kits used that were affected. Also, all unused kits still in your stock of the affected lot(s) will also be replaced. Immediately reply to email: [email protected] to confirm this notice and your compliance. Please indicate in the email the quantity of kits used which failed and quantity of stock remaining. We sincerely apologize for the inconvenience this may have caused you and/or your customer(s).

Recall details

Recall number
Z-2622-2024
Status
Open, Classified
Initiated by the company
Jul 5, 2024
Posted by the FDA
Aug 15, 2024
Recalling firm
DRG International, Inc.
Quantity
257 units
Product code
CDZ
FDA submissions
K052649
Distribution
US Nationwide distribution in the states of CA, IL, MN, PA, UT, VA.

Product description

Product Name: Salivary Testosterone ELISA Model/Catalog Number: SLV-3013 Product Description: An enzyme immunoassay for the quantitative in vitro diagnostic measurement of free active testosterone in saliva.

Code information (models, lots, serial numbers)

Model/Catalog Number: SLV-3013; UDI-DI: 04048474030136; Lot Nymbers: 66K014 and 66K123.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.