FilmArray Torch Module recallZ-3123-2024 · BioFire Diagnostics, LLC · initiated Aug 7, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from BioFire Diagnostics, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Some automated in vitro diagnostic modules and instruments manufactured or serviced between May 2022 and May 2024 were calibrated with internal equipment that may lead modules and instruments to report higher melting temperatures, which may lead to erroneous results including false positive Salmonella, Giardia lamblia, false positive/negative Norovirus, and false negative MRSA.

FDA root cause category: Under Investigation by firm

What the company is doing

On 8/7/24 Biomerieux sent recall notices to customers who were asked to do the following: 1) Utilize http://biofiredx.com/fsca5788-1/ to determine which of your instruments are impacted. 2) If your instrument is on the impacted list, it needs to be returned to the firm's Service Center for recalibration; your local contact will reach out to you to coordinate instrument return. 3) Continue to use all your instruments in accordance with product labeling (instructions for use) while waiting for service. 4) For BIOFIRE GI Panel Users, you may continue using the product as directed in the package insert, but if a positive Salmonella or Giardia lamblia result is inconsistent with clinical presentation, consider confirming the positive using another method or rerun the sample on an unaffected instrument. In patients suspected of Norovirus infection, please consider confirming Norovirus results via another method prior to reporting those results or rerun the sample on an unaffected instrument. 5) For BCID2 users, you may continue using the product as directed in the package insert; however, in patients with a Staphylococcus aureus detected result on the BCID2 panel AND suspected of a MRSA infection, please confirm all negative MRSA results via another method prior to reporting or rerun the sample on an unaffected instrument. 6) For RP2.1, RP2.1plus, ME, PN, PNplus and JI users, you may continue using the product as directed in the package insert; No confirmation of results are required. 7) If you have further distributed impacted instrument(s), identify any recipients and notify them at once by sending them a copy of the recall notice and acknowledgement form. 8) Complete all fields of the accompanying Acknowledgment of Receipt Form via email and return to [email protected] or the form can be filled out electronically by following URL: https://www.biofiredx.com/recall5788-1/ Customers with questions can contact Technical Support at biofiresupport@biomer

Recall details

Recall number
Z-3123-2024
Status
Open, Classified
Initiated by the company
Aug 7, 2024
Posted by the FDA
Sep 11, 2024
Recalling firm
BioFire Diagnostics, LLC
Quantity
3,439
Product code
OCC
FDA submissions
K160068
Distribution
Worldwide - US Nationwide distribution including in the states of NC, TX, OH, CA, KS, AL, WA, NY, PA, IL, WI, KY, LA, NE, FL, UT, IA, SC, VA, AZ, CO, WV, MO, TN, MN, DC, DE, OK, WY, IN, NJ, NM, ID, OR, MS, SD, MI, MT, GA, HI, NV, MD, MA, AR, NH, PR, ND, VT, CT, AK, ME, RI and the countries of Korea, Canada, Singapore distributors, Canada, Canada, China, China, Singapore, Canada, China, Mexico, Japan, Chile, Chile, Singapore, Mexico, Hong King, China, Chile, Singapore, Philippines, Korea, Chile, Korea, Australia, Greece.

Product description

FILMARRAY TORCH MODULE BOX, HTFA-ASY-0003, HTFA-ASY-0003-W, HTFA-ASY-0003U

Code information (models, lots, serial numbers)

Show the full code information (35k characters)

HTFA-ASY-0003, UDI: 00815381020277, Serial Numbers: KTM01040, KTM03252, KTM05330, KTM05411, KTM05602, KTM05734, KTM05800, KTM05810, KTM05887, KTM05997, KTM06000, KTM06003, KTM06005, KTM06038, KTM06057, KTM06063, KTM06139, KTM06639, KTM06756, KTM06870, KTM06902, KTM07116, KTM13705, KTM13710, KTM13815, KTM14196, KTM14329, KTM14357, KTM14918, KTM15242, KTM15798, KTM15801, KTM15819, KTM17194, KTM17531, KTM17570, KTM17576, KTM17696, KTM17812, KTM18083, KTM18448, KTM18453, KTM18464, KTM18559, KTM19173, KTM19183, KTM19386, KTM20255, KTM20260, KTM20273, KTM20280, KTM20296, KTM20299, KTM20301, KTM20308, KTM20314, KTM20318, KTM20321, KTM20324, KTM20325, KTM20330, KTM20354, KTM20398, KTM20403, KTM20502, KTM13752, KTM18382, KTM19197, KTM19634, KTM19610, KTM20269, KTM21459, KTM00143, KTM00173, KTM00221, KTM00230, KTM00237, KTM01401, KTM01437, KTM01439, KTM01446, KTM00247, KTM07097, KTM07224, KTM07422, KTM09682, KTM10535, KTM10550, KTM12833, KTM13438, KTM14845, KTM18045, KTM15398, KTM01947, KTM04128, KTM05500, KTM05501, KTM06798, KTM06929, KTM08608, KTM08780, KTM08830, KTM08977, KTM10092, KTM10527, KTM11388, KTM11583, KTM11617, KTM11624, KTM11687, KTM12214, KTM12456, KTM12622, KTM12637, KTM13777, KTM13784, KTM14421, KTM14536, KTM15336, KTM15984, KTM16048, KTM16148, KTM16426, KTM16708, KTM16729, KTM17030, KTM17237, KTM17427, KTM20259, KTM20262, KTM20272, KTM20281, KTM20287, KTM20291, KTM20295, KTM20306, KTM20311, KTM20313, KTM20317, KTM20326, KTM20339, KTM20340, KTM20357, KTM17378, KTM20242, KTM03150, KTM20239, KTM20240, KTM20245, KTM20305, KTM02088, KTM20352, KTM18296, KTM04024, KTM04291, KTM04336, KTM04339, KTM15727, KTM01193, KTM03483, KTM03828, KTM03835, KTM03837, KTM04108, KTM05011, KTM05574, KTM05641, KTM05686, KTM06842, KTM06847, KTM06882, KTM07019, KTM07039, KTM07088, KTM07093, KTM07094, KTM07292, KTM07331, KTM07734, KTM07976, KTM07994, KTM08495, KTM08496, KTM10488, KTM13627, KTM13862, KTM16633, KTM16676, KTM16721, KTM17960, KTM17970, KTM18062, KTM19036, KTM19142, KTM19304, KTM20140, KTM20303. HTFA-ASY-0003-W, UDI: 00815381029089, Serial Numbers: KTM00005, KTM00013, KTM00033, KTM00039, KTM00040, KTM00041, KTM00044, KTM00046, KTM00050, KTM00053, KTM00054, KTM00056, KTM00059, KTM00060, KTM00064, KTM00074, KTM00076, KTM00077, KTM00080, KTM00081, KTM00088, KTM00090, KTM00105, KTM00112, KTM00119, KTM00136, KTM00138, KTM00140, KTM00152, KTM00156, KTM00175, KTM00178, KTM00183, KTM00190, KTM00194, KTM00202, KTM00203, KTM00219, KTM00227, KTM00232, KTM00234, KTM00236, KTM00238, KTM00241, KTM00249, KTM00250,KTM00251,KTM00253, KTM00262, KTM00263,KTM00265, KTM00275, KTM00291, KTM00293, KTM00295, KTM00298, KTM00299,KTM00303, KTM00307, KTM00308, KTM00312, KTM00318,KTM00322, KTM00325, KTM00341, KTM00345, KTM00350, KTM00352, KTM00354, KTM00361, KTM00364, KTM00382, KTM00383, KTM00384, KTM00385, KTM00395, KTM00401,KTM00403, KTM00408, KTM00417, KTM00419, KTM00422, KTM00432, KTM00438, KTM00441, KTM00448, KTM00458,KTM00462, KTM00480, KTM00483, KTM00487, KTM00491, KTM00493, KTM00510, KTM00514, KTM00523, KTM00529, KTM00533, KTM00535, KTM00555, KTM00556, KTM00560, KTM00578, KTM00580, KTM00581, KTM00583, KTM00596, KTM00597, KTM00600, KTM00611, KTM00617, KTM00618, KTM00621, KTM00627, KTM00642, KTM00658, KTM00661, KTM00680, KTM00683, KTM00684, KTM00697, KTM00707, KTM00711, KTM00717, KTM00723, KTM00733, KTM00734, KTM00735, KTM00747, KTM00753, KTM00755, KTM00756, KTM00761, KTM00762, KTM00766, KTM00771, KTM00785, KTM00788, KTM00789, KTM00804, KTM00817, KTM00818, KTM00819, KTM00821, KTM00826, KTM00834, KTM00838, KTM00840, KTM00843, KTM00845, KTM00853, KTM00854, KTM00855, KTM00856, KTM00858, KTM00859, KTM00861, KTM00864, KTM00876, KTM00879, KTM00881, KTM00884, KTM00901, KTM00911, KTM00919, KTM00935, KTM00944, KTM00946, KTM00947, KTM00948, KTM00968, KTM00974, KTM00980, KTM00987, KTM00989, KTM00994, KTM00996, KTM00999, KTM009C5, KTM01004, KTM01007, KTM01009, KTM01017, KTM01019, KTM01024, KTM01035, KTM01038, KTM01042, KTM01043, KTM01045, KTM01048, KTM01052, KTM01054.KTM01058, KTM01061, KTM01064, KTM01071, KTM01077, KTM01083, KTM01085, KTM01090, KTM01105, KTM01110, KTM01113, KTM01129, KTM01135, KTM01153, KTM01158, KTM01164, KTM01166, KTM01171, KTM01175, KTM01181, KTM01191, KTM01197, KTM01203, KTM01228, KTM01234, KTM01242, KTM01263, KTM01265, KTM01273, KTM01276, KTM01278, KTM01284, KTM01292, KTM01297, KTM01302, KTM01315, KTM01316, KTM01325, KTM01327, KTM01349, KTM01355, KTM01361, KTM01367, KTM01372, KTM01373, KTM01379, KTM01380, KTM01386, KTM01390, KTM01394, KTM01395, KTM01399, KTM01421, KTM01443, KTM01445, KTM01447, KTM01449, KTM01452, KTM01470, KTM01480, KTM01494, KTM01508, KTM01516, KTM01518, KTM01519, KTM01533, KTM01535, KTM01536, KTM01541, KTM01545, KTM01561, KTM01566, KTM01569, KTM01571, KTM01581, KTM01590, KTM01599, KTM01612, KTM01613, KTM01618, KTM01620, KTM01625, KTM01629, KTM01631, KTM01649, KTM01650, KTM01652, KTM01674, KTM01683, KTM01698, KTM01699, KTM01705, KTM01711, KTM01716, KTM01718, KTM01723, KTM01726, KTM01730, KTM01736, KTM01738, KTM01742, KTM01743, KTM01744, KTM01745, KTM01760, KTM01767, KTM01774, KTM01775, KTM01776, KTM01778, KTM01783, KTM01787, KTM01789, KTM01790, KTM01792, KTM01795, KTM01799, KTM01802, KTM01803, KTM01811, KTM01818, KTM01834, KTM01836, KTM01838, KTM01846, KTM01858, KTM01863, KTM01867, KTM01877, KTM01878, KTM01882, KTM01883, KTM01902, KTM01903, KTM01912, KTM01925, KTM01944, KTM01954, KTM01962, KTM01969, KTM01973, KTM01982, KTM01989, KTM01990, KTM01994, KTM01995, KTM02012, KTM02013, KTM02014, KTM02015, KTM02028, KTM02047, KTM02059, KTM02061, KTM02063, KTM02064, KTM02068, KTM02075, KTM02082, KTM02083, KTM02092, KTM02094, KTM02099, KTM02110, KTM02123, KTM02125, KTM02128, KTM02133, KTM02134, KTM02135, KTM02139, KTM02142, KTM02143, KTM02150, KTM02152, KTM02156, KTM02159, KTM02162, KTM02163, KTM02174, KTM02184, KTM02186, KTM02188, KTM02209, KTM02210, KTM02212, KTM02214, KTM02216, KTM02220, KTM02222, KTM02224, KTM02225, KTM02235, KTM02239, KTM02242, KTM02246, KTM02247, KTM02249, KTM02270, KTM02271, KTM02280, KTM02286, KTM02297, KTM02298, KTM02300, KTM02321, KTM02324, KTM02329, KTM02334, KTM02336, KTM02342, KTM02343, KTM02344, KTM02351, KTM02353, KTM02357, KTM02361, KTM02365, KTM02366, KTM02367, KTM02369, KTM02372, KTM02373, KTM02374, KTM02375, KTM02376, KTM02383, KTM02387, KTM02390, KTM02391, KTM02393, KTM02394, KTM02402, KTM02404, KTM02405, KTM02406, KTM02407, KTM02413, KTM02417, KTM02418, KTM02423, KTM02427, KTM02437, KTM02449, KTM02452, KTM02462, KTM02477, KTM02481, KTM02482, KTM02484, KTM02487, KTM02497, KTM02502, KTM02518, KTM02531, KTM02540, KTM02546, KTM02548, KTM02552, KTM02554, KTM02562, KTM02573, KTM02578, KTM02581, KTM02582, KTM02585, KTM02599, KTM02602, KTM02604, KTM02627, KTM02629, KTM02632, KTM02633, KTM02645, KTM02647, KTM02650, KTM02655, KTM02657, KTM02658, KTM02666, KTM02670, KTM02673, KTM02677, KTM02681, KTM02685, KTM02695, KTM02702, KTM02704, KTM02713, KTM02717, KTM02720, KTM02722, KTM02733, KTM02735, KTM02736, KTM02741, KTM02747, KTM02750, KTM02753, KTM02757, KTM02782, KTM02798, KTM02800, KTM02807, KTM02824, KTM02825, KTM02848, KTM02849, KTM02851, KTM02860, KTM02868, KTM02879, KTM02882, KTM02890, KTM02907, KTM02915, KTM02923, KTM02928, KTM02932, KTM02934, KTM02935, KTM02942, KTM02948, KTM02960, KTM02964, KTM02979, KTM02981, KTM02984, KTM02988, KTM02999, KTM03004, KTM03005, KTM03008, KTM03021, KTM03022, KTM03037, KTM03043, KTM03057, KTM03062, KTM03086, KTM03093, KTM03102, KTM03105, KTM03106, KTM03117, KTM03130, KTM03138, KTM03139, KTM03160, KTM03172, KTM03190, KTM03211, KTM03213, KTM03216, KTM03219, KTM03221, KTM03225, KTM03226, KTM03232, KTM03233, KTM03258, KTM03267, KTM03270, KTM03271, KTM03274,KTM03276, KTM03278, KTM03281, KTM03282, KTM03285, KTM03286, KTM03289, KTM03292, KTM03309, KTM03316, KTM03322, KTM03325, KTM03326, KTM03327, KTM03338, KTM03341, KTM03347, KTM03349, KTM03351, KTM03357, KTM03358, KTM03361, KTM03362, KTM03389, KTM03397, KTM03402, KTM03406, KTM03410, KTM03413, KTM03415, KTM03418, KTM03424, KTM03425, KTM03429, KTM03430, KTM03434, KTM03435, KTM03436, KTM03447, KTM03453, KTM03457, KTM03459, KTM03461, KTM03470, KTM03471, KTM03476, KTM03480, KTM03481, KTM03484, KTM03491, KTM03504, KTM03505, KTM03510, KTM03520, KTM03521, KTM03522, KTM03527, KTM03547, KTM03550, KTM03553, KTM03562, KTM03565, KTM03571, KTM03577, KTM03580, KTM03582, KTM03589, KTM03591, KTM03603, KTM03615, KTM03619, KTM03626, KTM03627, KTM03638, KTM03639, KTM03640, KTM03646, KTM03660, KTM03661, KTM03662, KTM03664, KTM03669, KTM03673, KTM03677, KTM03680, KTM03683, KTM03686, KTM03694, KTM03701, KTM03703, KTM03708, KTM03711, KTM03715, KTM03722, KTM03723, KTM03724, KTM03729, KTM03731, KTM03734, KTM03736, KTM03738, KTM03748, KTM03753, KTM03755, KTM03760, KTM03761, KTM03771, KTM03775, KTM03776, KTM03784, KTM03788, KTM03790, KTM03803, KTM03811, KTM03814, KTM03817, KTM03818, KTM03823, KTM03832, KTM03839, KTM03843, KTM03849, KTM03850, KTM03858, KTM03860, KTM03868, KTM03873, KTM03876, KTM03885, KTM03887, KTM03889, KTM03892, KTM03898, KTM03910, KTM03912, KTM03914, KTM03916, KTM03937, KTM03943, KTM03946, KTM03953, KTM03958, KTM03963, KTM03976, KTM03990, KTM03994, KTM03997, KTM03998, KTM03999, KTM04005, KTM04006, KTM04008, KTM04011, KTM04014, KTM04017, KTM04018, KTM04028, KTM04032, KTM04037, KTM04041, KTM04050, KTM04056, KTM04059, KTM04064, KTM04066, KTM04067, KTM04077, KTM04079, KTM04081, KTM04082, KTM04091, KTM04094, KTM04099, KTM04102, KTM04110, KTM04115, KTM04118, KTM04121, KTM04127, KTM04130, KTM04131, KTM04137, KTM04141, KTM04154, KTM04164, KTM04166, KTM04169, KTM04177, KTM04180, KTM04185, KTM04201, KTM04205, KTM04207, KTM04210, KTM04215, KTM04226, KTM04227, KTM04234, KTM04242, KTM04243, KTM04254, KTM04256, KTM04265, KTM04268, KTM04272, KTM04275, KTM04277, KTM04282, KTM04286, KTM04295, KTM04300, KTM04303, KTM04304, KTM04321, KTM04322, KTM04331, KTM04333, KTM04335, KTM04338, KTM04341, KTM04345, KTM04360, KTM04365, KTM04369, KTM04370, KTM04376, KTM04378, KTM04395, KTM04401, KTM04403, KTM04404, KTM04405, KTM04409, KTM04419, KTM04422, KTM04435, KTM04451, KTM04455, KTM04458, KTM04464, KTM04469, KTM04472, KTM04484, KTM04485, KTM04486, KTM04503, KTM04506, KTM04510, KTM04517, KTM04528, KTM04537, KTM04546, KTM04547, KTM04551, KTM04555, KTM04562, KTM04566, KTM04571, KTM04577, KTM04582, KTM04586, KTM04590, KTM04593, KTM04612, KTM04616, KTM04620, KTM04627, KTM04632, KTM04638, KTM04641, KTM04642, KTM04667, KTM04670, KTM04672, KTM04721, KTM04722, KTM04739, KTM04742, KTM04754, KTM04765, KTM04783, KTM04785, KTM04801, KTM04803, KTM04810, KTM04813, KTM04816, KTM04827, KTM04834, KTM04835, KTM04846, KTM04847, KTM04849, KTM04850, KTM04862, KTM04865, KTM04881, KTM04882, KTM04888, KTM04933, KTM04934, KTM04942, KTM04950, KTM04953, KTM04958, KTM04964, KTM04966, KTM04977, KTM0497A, KTM04981, KTM04982, KTM04988, KTM04989, KTM04991, KTM04994, KTM05002, KTM05004, KTM05005, KTM05007, KTM05008, KTM05019, KTM05021, KTM05034, KTM05038, KTM05039, KTM05040, KTM05048, KTM05052, KTM05058, KTM05062, KTM05070, KTM05077, KTM05081, KTM05088, KTM05090, KTM05097, KTM05098, KTM05107, KTM05112, KTM05113, KTM05119, KTM05124, KTM05126, KTM05131, KTM05132, KTM05138, KTM05140, KTM05143, KTM05144, KTM05155, KTM05186, KTM05194, KTM05195, KTM05214, KTM05219, KTM05222, KTM05225, KTM05228, KTM05229, KTM05235, KTM05240, KTM05244, KTM05255, KTM05260, KTM05265, KTM05267, KTM05270, KTM05281, KTM05289, KTM05290, KTM05303, KTM05304, KTM05307, KTM05311, KTM05316, KTM05319, KTM05326, KTM05327, KTM05328, KTM05338, KTM05352, KTM05357, KTM05369, KTM05371, KTM05373, KTM05375, KTM05385, KTM05389, KTM05395, KTM05396, KTM05421, KTM05424, KTM05435, KTM05442, KTM05451, KTM05453, KTM05454, KTM05467, KTM05468, KTM05470, KTM05493, KTM05494, KTM05504, KTM05507, KTM05517, KTM05519, KTM05521, KTM05522, KTM05526, KTM05529, KTM05535, KTM05536, KTM05541, KTM05549, KTM05552, KTM05554, KTM05555, KTM05561, KTM05567, KTM05571, KTM05576, KTM05579, KTM05582, KTM05584, KTM05597, KTM05604, KTM05623, KTM05628, KTM05630, KTM05635, KTM05638, KTM05676, KTM05677, KTM05683, KTM05687, KTM05696, KTM05698, KTM05710, KTM05714, KTM05717, KTM05728, KTM05737, KTM05741, KTM05742, KTM05745, KTM05751, KTM05764, KTM05769, KTM05772, KTM05777, KTM05797, KTM05805, KTM05811, KTM05813, KTM05814, KTM05815, KTM05817, KTM05826, KTM05832, KTM05836, KTM05839, KTM05840, KTM05844, KTM05851, KTM05852, KTM05858, KTM05865, KTM05868, KTM05873, KTM05879, KTM05881, KTM05882, KTM05883, KTM05892, KTM05894, KTM05898, KTM05902, KTM05906, KTM05907, KTM05908, KTM05918, KTM05921, KTM05923, KTM05924, KTM05926, KTM05932, KTM05933, KTM05934, KTM05940, KTM05949, KTM05954, KTM05957, KTM05963, KTM05968, KTM05971, KTM05974, KTM05977, KTM05978, KTM05980, KTM05985, KTM05990, KTM05991, KTM05993, KTM05995, KTM06008, KTM06009, KTM06018, KTM06022, KTM06026, KTM06032, KTM06039, KTM06040, KTM06043, KTM06047, KTM06051, KTM06061, KTM06073, KTM06076, KTM06080, KTM06096, KTM06102, KTM06104, KTM06106, KTM06120, KTM06149, KTM06152, KTM06156, KTM06161, KTM06163, KTM06164, KTM06169, KTM06185, KTM06190, KTM06191, KTM06199, KTM06204, KTM06206, KTM06207, KTM06208, KTM06211, KTM06219, KTM06220, KTM06225, KTM06228, KTM06233, KTM06236, KTM06240, KTM06245, KTM06248, KTM06252, KTM06255, KTM06257, KTM06258, KTM06261, KTM06264, KTM06273, KTM06280, KTM06284, KTM06285, KTM06292, KTM06295, KTM06296, KTM06302, KTM06311, KTM06325, KTM06328, KTM06335, KTM06339, KTM06343, KTM06346, KTM06347, KTM06354, KTM06362, KTM06365, KTM06369, KTM06376, KTM06377, KTM06384, KTM06385, KTM06386, KTM06389, KTM06393, KTM06394, KTM06398, KTM06406, KTM06408, KTM06418, KTM06427, KTM06429, KTM06434, KTM06435, KTM06447, KTM06454, KTM06455, KTM06460, KTM06475, KTM06481, KTM06485, KTM06488, KTM06489, KTM06512, KTM06520, KTM06523, KTM06527, KTM06532, KTM06533, KTM06534, KTM06549, KTM06550, KTM06554, KTM06559, KTM06560, KTM06561, KTM06563, KTM06564, KTM06578, KTM06585, KTM06591, KTM06604, KTM06617, KTM06627, KTM06628, KTM06634, KTM06643, KTM06644, KTM06654, KTM06669, KTM06670, KTM06675, KTM06681, KTM06682, KTM06685, KTM06688, KTM06691, KTM06692, KTM06696, KTM06699, KTM06701, KTM06713, KTM06720, KTM06723, KTM06733, KTM06735, KTM06769, KTM06770, KTM06780, KTM06787, KTM06789, KTM06796, KTM06797, KTM06803, KTM06810, KTM06814, KTM06815, KTM06825, KTM06831, KTM06840, KTM06848, KTM06849, KTM06852, KTM06859, KTM06860, KTM06861, KTM06866, KTM06874, KTM06877, KTM06884, KTM06886, KTM06891, KTM06893, KTM06909, KTM06921, KTM06923, KTM06938, KTM06953, KTM06954, KTM06959, KTM06960, KTM06973, KTM06974, KTM06980, KTM06987, KTM06992, KTM06996, KTM06997, KTM06BE4, KTM07001, KTM07006, KTM07010, KTM07013, KTM07021, KTM07023, KTM07026, KTM07029, KTM07047, KTM07056, KTM07057, KTM07059, KTM07061, KTM07062, KTM07063, KTM07066, KTM07071, KTM07076, KTM07109, KTM07113, KTM07122, KTM07130, KTM07133, KTM07139, KTM07147, KTM07153, KTM07163, KTM07167, KTM07171, KTM07173, KTM07174, KTM07195, KTM07200, KTM07203, KTM07211, KTM07212, KTM07217, KTM07227, KTM07235, KTM07243, KTM07247, KTM07248, KTM07252, KTM07262, KTM07275, KTM07290, KTM07296, KTM07305, KTM07312, KTM07335, KTM07339, KTM07341, KTM07346, KTM07359, KTM07366, KTM07376, KTM07402, KTM07411, KTM07412, KTM07415, KTM07421, KTM07423, KTM07425, KTM07427, KTM07429, KTM07430, KTM07432, KTM07435, KTM07448, KTM07456, KTM07463, KTM07470, KTM07471, KTM07473, KTM07477, KTM07483, KTM07485, KTM07486 ,KTM07497, KTM07498, KTM07505, KTM07509, KTM07523, KTM07526, KTM07529, KTM07542, KTM07545, KTM07546, KTM07547, KTM07572, KTM07588, KTM07594, KTM07613, KTM07635, KTM07639, KTM07646, KTM07647, KTM07650, KTM07651, KTM07654, KTM07658, KTM07664, KTM07669, KTM07670, KTM07716, KTM07717, KTM07718, KTM07724, KTM07739, KTM07755,KTM07762, TM07771, TM07772, KTM07792, KTM07796, KTM07802, KTM07815, KTM07840, TM07847, KTM07864, KTM07870, KTM07875, KTM07879, KTM07890, KTM07902, KTM07926, KTM07929, KTM07945, KTM07991, KTM08015, KTM08064, TM08098, KTM08130, KTM08139, KTM08140, KTM08144, KTM08146, KTM08147, KTM08155, KTM08165, KTM08173, KTM08180, KTM08188, TM08190, KTM08200, KTM08209, TM08218, KTM08236, KTM08239, KTM08242, KTM08252, KTM08269, KTM08272, KTM08288, KTM08291, KTM08303,KTM08305, KTM08308, KTM08316, KTM08337, KTM08346, KTM08362, KTM08373, KTM08411, KTM08415, KTM08438, KTM08462, KTM08478, KTM08479, KTM08503, KTM08506, KTM08510,KTM08527, KTM08556,KTM08565,KTM08566, KTM08572, KTM08581, KTM08583, KTM08588, KTM08600, KTM08602, KTM08603, KTM08614, KTM08623, KTM08628, KTM08635, KTM08642, KTM08657, KTM08661, KTM08664, KTM08666, KTM08669, KTM08677, KTM08678, TM08681, KTM08684, KTM08686, KTM08692, KTM08693, KTM08695, KTM08702, KTM08703, KTM08710, KTM08712, KTM08715, KTM08720, KTM08722, KTM08743, KTM08750, KTM08754, KTM08756, KTM08768, KTM08769, KTM08775, KTM08792, KTM08810, KTM08814, KTM08815, KTM08832, KTM08834, KTM08837, KTM08840, KTM08849, KTM08852, KTM08855, KTM08857, KTM08871, KTM08881, KTM08896, KTM08913, KTM08917, KTM08922, KTM08924, KTM08925, KTM08933, KTM08940, KTM08941, KTM08943, KTM08956, KTM08964, KTM08978, KTM08979, KTM08980, KTM08991, KTM08996, KTM08998, KTM09003, KTM09004, KTM09009, KTM09010, KTM09014, KTM09019, KTM09021, KTM09027, KTM09028, KTM09030, KTM09033, KTM09053, KTM09056, KTM09058, KTM09068, KTM09070, KTM09078, KTM09104, KTM09105, KTM09110, KTM09112, KTM09116, KTM09117, KTM09120, KTM09137, KTM09138, KTM09139, KTM09147, KTM09154, KTM09162, KTM09164, KTM09166, KTM09177, KTM09178, KTM09181, KTM09193, KTM09208, KTM09212, KTM09217, KTM09220, KTM09223, KTM09251, KTM09264, KTM09269, KTM09274, KTM09280, KTM09292, KTM09294, KTM09311, KTM09316, KTM09323, KTM09334, KTM09350, KTM09361, KTM09370, KTM09371, KTM09382, KTM09386, KTM09391, KTM09405, KTM09407, KTM09416, KTM09422, KTM09425, KTM09431, KTM09441, KTM09449, KTM09450, KTM09452, KTM09470, KTM09480, KTM09492, KTM09501, KTM09503, KTM09506, KTM09509, KTM09511, KTM09520, KTM09527, KTM09530, KTM09536, KTM09542, KTM09547, KTM09556, KTM09571, KTM09578, KTM09591, KTM09603, KTM09604, KTM09610, KTM09611, KTM09633, KTM09642, KTM09648, KTM09649, KTM09657, KTM09662, KTM09671, KTM09674, KTM09687, KTM09690, KTM09704, KTM09705, KTM09717, KTM09719, KTM09724, KTM09725, KTM09737, KTM09742, KTM09752, KTM09762, KTM09763, KTM09777, KTM09779, KTM09782, KTM09784, KTM09785, KTM09793, KTM09802, KTM09803, KTM09811, KTM09813, KTM09814, KTM09821, KTM09823, KTM09826, KTM09832, KTM09834, KTM09835, KTM09839, KTM09844, KTM09846, KTM09848, KTM09854, KTM09865, KTM09872, KTM09882, KTM09892, KTM09897, KTM09901, KTM09911, KTM09913, KTM09933, KTM09942, KTM09948, KTM09951, KTM09958, KTM09965, KTM09982, KTM09988, KTM09989, KTM0BB79, KTM0CF31, KTM0F1D0, KTM10009, KTM10021, KTM10041, KTM10097, KTM10102, KTM10128, KTM10131, KTM10134, KTM10137, KTM10139, KTM10144, KTM10152, KTM10158, KTM10161, KTM10168, KTM10173, KTM10176, KTM10178, KTM10186, KTM10191, KTM10192, KTM10193, KTM10198, KTM101C3, KTM10203, KTM10230, KTM10234, KTM10246, KTM10255, KTM10265, KTM10268, KTM10275, KTM10285, KTM10293, KTM10294, KTM10297, KTM10298, KTM10301, KTM10305, KTM10309, KTM10311, KTM10315, KTM10335, KTM10340, KTM10343, KTM10351, KTM10362, KTM10375, KTM10379, KTM10386, KTM10389, KTM10390, KTM10395, KTM10420, KTM10422, KTM10434, KTM10438, KTM10439, KTM10444, KTM10450, KTM10456, KTM10471, KTM10478, KTM10485, KTM10501, KTM10526, KTM10556, KTM10579, KTM10586, KTM10588, KTM10603, KTM10611, KTM10652, KTM10665, KTM10674, KTM10729, KTM10731, KTM10737, KTM10745, KTM10752, KTM10754, KTM10761, KTM10765, KTM10770, KTM10777, KTM10782, KTM10784, KTM10803, KTM10822, KTM10828, KTM10835, KTM10872, KTM10877, KTM10885, KTM10890, KTM10900, KTM10902, KTM10906, KTM10928, KTM10945, KTM10946, KTM10956, KTM10983, KTM10985, KTM10987, KTM10997, KTM10998, KTM10AEB, KTM11027, KTM11047, KTM11048, KTM11060, KTM11062, KTM11083, KTM11098, KTM11112, KTM11120, KTM11129, KTM11138, KTM11141, KTM11147, KTM11166, KTM11172, KTM11182, KTM11184, KTM11193, KTM11194, KTM11204, KTM11207, KTM11209, KTM11213, KTM11221, KTM11222, KTM11224, KTM11230, KTM11237, KTM11249, KTM11256, KTM11257, KTM11258, KTM11264, KTM11268, KTM11280, KTM11286, KTM11291, KTM11299, KTM112BC, KTM11305,KTM11307, KTM11311, KTM11330, KTM11351, KTM11358, KTM11359, KTM11361, KTM11363, KTM11372, KTM11397, KTM11398, KTM11401, KTM11415, KTM11418, KTM11431, KTM11433, KTM11466, KTM11477, KTM11482, KTM11491, KTM11511, KTM11516, KTM11549, KTM11561, KTM11564, KTM11597, KTM11627, KTM11631, KTM11661, KTM11674, KTM11676, KTM11707, KTM11715, KTM11742, KTM11820, KTM11827, KTM11871, KTM11875, KTM11879, KTM11883, KTM11884, KTM11891, KTM11893, KTM11913, KTM11919, KTM11922 ,KTM11965, KTM11966, KTM12014, KTM12017, KTM12018, KTM12033, KTM12057, KTM12070, KTM12120, KTM12121, KTM12158, KTM12161, KTM12191, KTM12197, KTM12213, KTM12215, KTM12224, KTM12225, KTM12229, KTM12240, KTM12243, KTM12245, KTM12256, KTM12258, KTM12262, KTM12269, KTM12271, KTM12279, KTM12280, KTM12298, KTM12302, KTM12303, KTM12305, KTM12307, KTM12308, KTM12352, KTM12355, KTM12356, KTM12361, KTM12363, KTM12364, KTM12365, KTM12367, KTM12396, KTM12405, KTM12437, KTM12444, KTM12455, KTM12473, KTM12474, KTM12478, KTM12484, KTM12486, KTM12488, KTM12489, KTM12497, KTM12498, KTM12506, KTM12512, KTM12524, KTM12534, KTM12547, KTM12557, KTM12561, KTM12574, KTM12576, KTM12585, KTM12587, KTM12597, KTM12607, KTM12640, KTM12645, KTM12648, KTM12650, KTM12654, KTM12660, KTM12666, KTM12670, KTM12678, KTM12679, KTM12684, KTM12692, KTM12697, KTM12698, KTM12701, KTM12707, KTM12709, KTM12710, KTM12711, KTM12715, KTM12716, KTM12718, KTM12719, KTM12722, KTM12724, KTM12733, KTM12741, KTM12755, KTM12757, KTM12761, KTM12777, KTM12778, KTM12781, KTM12782, KTM12803, KTM12822, KTM12844, KTM12847, KTM12865, KTM12878, KTM12884, KTM12891, KTM12893, KTM12903, KTM12939, KTM12946, KTM12950, KTM12952, KTM12961, KTM12976, KTM12999, KTM13007, KTM13023, KTM13024, KTM13035, KTM13042, KTM13051, KTM13054, KTM13059, KTM13060, KTM13062, KTM13063, KTM13067, KTM13076, KTM13079, KTM13089, KTM13101, KTM13104, KTM13115, KTM13116, KTM13157, KTM13159, KTM13170, KTM13174, KTM13224, KTM13238, KTM13242, KTM13253, KTM13258, KTM13270, KTM13271, KTM13287, KTM13304, KTM13309, KTM13312, KTM13314, KTM13316, KTM13321, KTM13326, KTM13336, KTM13343, KTM13348, KTM13353, KTM13357, KTM13361, KTM13365, KTM13367, KTM13372, KTM13378, KTM13382, KTM13385, KTM13390, KTM13394, KTM13395, KTM13407, KTM13413, KTM13420, KTM13427, KTM13433, KTM13434, KTM13437, KTM13440, KTM13441, KTM13442, KTM1345B, KTM13462, KTM13467, KTM13481, KTM13487, KTM13492, KTM13494, KTM13496, KTM13500, KTM13513, KTM13515, KTM13525, KTM13526, KTM13528, KTM13541, KTM13543, KTM13559, KTM13598, KTM13613, KTM13621, KTM13636, KTM13637, KTM13641, KTM13645, KTM13649, KTM13663, KTM13665, KTM13671, KTM13689, KTM13692, KTM13694, KTM13712, KTM13716, KTM13717, KTM13724, KTM13731, KTM13732, KTM13755, KTM13760, KTM13761, KTM13766, KTM13769, KTM13786, KTM13789, KTM13795, KTM13796, KTM13804, KTM13839, KTM13848, KTM13857, KTM13864, KTM13874, KTM13915, KTM13920, KTM13936, KTM13939, KTM13945, KTM13946, KTM13953, KTM13958, KTM13959, KTM14016, KTM14022, KTM14036, KTM14085, KTM14086, KTM14105, KTM14108, KTM14118, KTM14120, KTM14124, KTM14128, KTM14130, KTM14134, KTM14137, KTM14153, KTM14161, KTM14163, KTM14164, KTM14167, KTM14192, KTM14218, KTM14224, KTM14226, KTM14230, KTM14232, KTM14241, KTM14245, KTM14263, KTM14267, KTM14269, KTM14291, KTM14292, KTM14313, KTM14315, KTM14318, KTM14321, KTM14328, KTM14330, KTM14331, KTM14345, KTM14347, KTM14359, KTM14375, KTM14387, KTM14394, KTM14398, KTM14413, KTM14427, KTM14447, KTM14453, KTM14460, KTM14485, KTM14488, KTM14490, KTM14515, KTM14520, KTM14521, KTM14522, KTM14524, KTM14529, KTM14541, KTM14544, KTM14549, KTM14551, KTM14563, KTM14572, KTM14597, KTM14622, KTM14623, KTM14640, KTM14641, KTM14642, KTM14651, KTM14654, KTM14662, KTM14664, KTM14665, KTM14672, KTM14683, KTM14695, KTM14696, KTM14706, KTM14709, KTM14710, KTM14742, KTM14770, KTM14790, KTM14793, KTM14794, KTM14799, KTM14801, KTM14805, KTM14811, KTM14812, KTM14818, KTM14823, KTM14831, KTM14832, KTM14835, KTM14839, KTM14846, KTM14849, KTM14850, KTM14852, KTM14861, KTM14873, KTM14888, KTM14892, KTM14905, KTM14915, KTM14922, KTM14923, KTM14934, KTM14941, KTM14963, KTM14985, KTM14C56, KTM15001, KTM15015, KTM15020, KTM15021, KTM15026, KTM15033, KTM15042, KTM15043, KTM15050, KTM15066, KTM15068, KTM15079, KTM15089, KTM15092, KTM15094, KTM15101, KTM15105, KTM15127, KTM15130, KTM15131, KTM15133, KTM15140, KTM15148, KTM15150, KTM15151, KTM15153, KTM15159, KTM15160, KTM15185, KTM15189, KTM15196, KTM15225, KTM15232, KTM15236, KTM15257, KTM15270, KTM15274, KTM15280, KTM15303, KTM15306, KTM15310, KTM15311, KTM15313, KTM15329, KTM15357, KTM15365, KTM15373, KTM15382, KTM15384, KTM15390, KTM15403, KTM15415, KTM15422, KTM15429, KTM15431, KTM15450, KTM15463, KTM15466, KTM15470, KTM15474, KTM15481, KTM15486, KTM15488, KTM15501, KTM15522, KTM15535, KTM15539, KTM15544, KTM15568, KTM15583, KTM15587, KTM15592, KTM15595, KTM15623, KTM15643, KTM15673, KTM15674, KTM15692, KTM15694, KTM15702, KTM15704, KTM15721, KTM15722, KTM15730, KTM15734, KTM15742, KTM15764, KTM15777, KTM15796, KTM15797, KTM15814, KTM15834, KTM15856, KTM15883, KTM15885, KTM15892, KTM158B0, KTM15902, KTM15904, KTM15916, KTM15929, KTM15942, KTM15952, KTM15958, KTM15963, KTM15965, KTM15985, KTM15990, KTM15995, KTM16005, KTM16011, KTM16052, KTM16062, KTM16066, KTM16068, KTM16072, KTM16075, KTM16090, KTM16106, KTM16133, KTM16138, KTM16179, KTM16191, KTM16194, KTM16231, KTM16242, KTM16247, KTM16272, KTM16274, KTM16305, KTM16312, KTM16327, KTM16333, KTM16349, KTM16383, KTM16395, KTM16398, KTM16413, KTM16418, KTM16420, KTM16429, KTM16437, KTM16439, KTM16442, KTM16493, KTM16501, KTM16559, KTM16597, KTM16598, KTM16630, KTM16653, KTM16732, KTM16735, KTM16736, KTM16741, KTM16749, KTM16750, KTM16814, KTM16829, KTM16843, KTM16844, KTM16847, KTM16868, KTM16869, KTM16873, KTM16875, KTM16879, KTM16892, KTM16900, KTM16904, KTM16906, KTM16907, KTM16910, KTM16912, KTM16916, KTM16917, KTM16926, KTM16937, KTM16949, KTM16950, KTM16969, KTM16975, KTM16984, KTM16987, KTM16999, KTM17001, KTM17007, KTM17012, KTM17021, KTM17025, KTM17027, KTM17042, KTM17043, KTM17046, KTM17047, KTM17056, KTM17073, KTM17074, KTM17076, KTM17113, KTM17132, KTM17150, KTM17158, KTM17180, KTM17219, KTM17235, KTM17266, KTM17289, KTM17300, KTM17302, KTM17328, KTM17362, KTM17371, KTM17399, KTM17475, KTM17476, KTM17477, KTM17500, KTM17502, KTM17566, KTM17652, KTM17776, KTM17782, KTM17790, KTM17838, KTM17946, KTM17968, KTM18017, KTM18146, KTM18155, KTM18171, KTM18173, KTM18210, KTM18291, KTM18323, KTM18337, KTM18371, KTM18415, KTM18460, KTM18462, KTM18465, KTM18484, KTM18486, KTM18542, KTM18563, KTM18625, KTM18629, KTM18652, KTM18660, KTM18879, KTM18964, KTM18975, KTM19020, KTM19026, KTM19034, KTM19076, KTM19102, KTM19189, KTM19191, KTM19214, KTM19280, KTM19282, KTM19301, KTM19384, KTM19438, KTM19444, KTM19447, KTM19457, KTM19499, KTM19519, KTM19533, KTM19538, KTM19581, KTM19585, KTM19603, KTM19606, KTM19656, KTM19660, KTM19661, KTM19679, KTM19710, KTM19714, KTM19750, KTM19758, KTM19870, KTM19874, KTM19933, KTM19956, KTM19978, KTM19987, KTM1B8FE, KTM1C3BD, KTM1C968, KTM1FABB, KTM20040, KTM20057, KTM20062, KTM20068, KTM20069, KTM20073, KTM20116, KTM20163, KTM20389, KTM20420, KTM20798, KTM22B24, KTM23B5B, KTM24612, KTM27C68, KTM284B8, KTM2872F, KTM2880D, KTM29AF7, KTM2A249, KTM2AC7E, KTM2D8D1, KTM2EF17, KTM320BD, KTM32768, KTM354E6, KTM368A1, KTM36CB9, KTM3756F, KTM382B0, KTM3BBF6, KTM3C9C0, KTM423C7, KTM42A9A, KTM44316, KTM45F15, KTM48105, KTM4939C, KTM4AA32, KTM4F30D, KTM50EC9, KTM528E9, KTM5314C, KTM55DE1, KTM56F94, KTM5B735, KTM5C161, KTM5CC10, KTM5D285, KTM5E715, KTM5F3FF, KTM60116, KTM636F8, KTM65B91, KTM66A28, KTM6780F, KTM6B4EC, KTM742FB, KTM76652, KTM77764, KTM77E36, KTM789E3, KTM7D257, KTM7E552, KTM7F52E, KTM7F62E, KTM80C9F, KTM82ADA, KTM84F39, KTM87256, KTM874A4, KTM8867A, KTM8AE97, KTM8BAED, KTM8CFD8, KTM8D65F, KTM8E095, KTM8F233, KTM91761, KTM92055, KTM94744, KTM9587D, KTM95B96, KTM96A7A, KTM97121, KTM9A6C3, KTM9AEFB, KTM9D1B1, KTM9D3C6, KTM9D4F9, KTM9F27D, KTMA0409, KTMA127D, KTMA2DB2, KTMA31A8, KTMA6F74, KTMAA5CF, KTMACB7B, KTMACF2F, KTMADEED, KTMAE4C2, KTMB2315, KTMB2E69, KTMB521E, KTMB7392, KTMB74DB, KTMBDF6C, KTMBE233, KTMC04C7, KTMC1326, KTMC188D, KTMC291B, KTMC5252, KTMC52E3, KTMC6D35, KTMC7C49, KTMCA2D2, KTMCAD7F, KTMCC5DB, KTMCE54F, KTMCEFE8, KTMCFF36, KTMD1BDC, KTMD2492, KTMD32E9, KTMD3DAB, KTMDA530, KTME0F31, KTME1A4A, KTME2F8E, KTME71ED, KTMEAF5D, KTMEB024, KTMEB462, KTMEBCF0, KTMED8BF, KTMEDBF3, KTMEE4EC, KTMF0049, KTMF3755, KTMF3E30, KTMF3FCD, KTMF7648, KTMFD6F8, KTMFD72B, KTMFDF96, KTMFEF37, KTM03421, KTM14096, KTM21292, KTM21311, KTM491BB, KTM00222, KTM01015, KTM01691, KTM01784, KTM01879, KTM02234, KTM02714, KTM03111, KTM06ED8, KTM2FD90, KTMB1A5C, KTMD8636, KTMFEF6E, KTM04407, KTM04504, KTM05305, KTM06002, KTM01269, KTM06202, KTM06312, KTM08205, KTM08781, KTM08718, KTM08888, KTM08822, KTM09602, KTM10175, KTM10907, KTM11396, KTM11925, KTM12322, KTM13780, KTM14611, KTM15609, KTM16851, KTM17748, KTM09050, KTM04070, KTM00127, KTM00765, KTM01152, KTM01576, KTM1BB7A, KTM41AE8, KTM05227, KTM06072, KTM06568, KTM09020, KTM06153, KTM11313, KTM12341, KTM12505, KTM13802, KTM13858, KTM15167, KTM17035, KTM18673, KTM02970, KTM04844, KTM05528, KTM05538, KTM09052, KTM13214, KTM00297, KTM00391, KTM00424, KTM00673, KTM01346, KTM03506, KTM04481, KTM05037, KTM06187, KTM06745, KTM08409, KTM10786, KTM13506, KTM14740, KTM8EA20, KTM01831, KTM03367, KTM03378, KTM00444, KTM00676, KTM00688, KTM00706, KTM01097, KTM01104, KTM01216, KTM01244, KTM01247, KTM01304, KTM01321, KTM01610, KTM01662, KTM01675, KTM03479, KTM01891, KTM02105, KTM02305, KTM02332, KTM02432, KTM02467, KTM02802, KTM03184, KTM03186, KTM03240, KTM03321, KTM03331, KTM03391, KTM03422,KTM03466, KTM08221, KTM03492, KTM03610, KTM03611, KTM03623, KTM03629, KTM03630, KTM03642, KTM03678, KTM03787, KTM03867, KTM03903, KTM03967, KTM03974, KTM03983, KTM03985, KTM04001, KTM04036, KTM04100, KTM04105, KTM04113, KTM04181, KTM04183, KTM04192, KTM04197, KTM04342, KTM04388, KTM04427, KTM04449, KTM04466, KTM04487, KTM04493, KTM04495, KTM04501, KTM04518, KTM04523, KTM04531, KTM04536, KTM04543, KTM04544, KTM04553, KTM04565, KTM04633, KTM04652, KTM04666, KTM04675, KTM04745, KTM04797, KTM04868, KTM04909, KTM04914, KTM04923, KTM05179, KTM05272, KTM05278, KTM05336, KTM05414, KTM05437, KTM05455, KTM05477, KTM05480, KTM05481, KTM05484, KTM05512, KTM05515, KTM05523, KTM05530, KTM05570, KTM05760, KTM06308, KTM06465, KTM06505, KTM06545, KTM06693, KTM06742, KTM06773, KTM06808, KTM06946, KTM06999, KTM07141, KTM07244, KTM07294, KTM07300, KTM07311, KTM07351, KTM07447, KTM07515, KTM07528, KTM07543, KTM07553, KTM07596, KTM07660, KTM07673, KTM07748, KTM07855, KTM07866, KTM07916, KTM07923, KTM07937, KTM07948, KTM07966, KTM08007, KTM08034, KTM08046, KTM08063, KTM08110, KTM08113, KTM08116, KTM08176, KTM08181, KTM08182, KTM08222, KTM08226, KTM08237, KTM08271, KTM08275, KTM08293, KTM08295, KTM08405, KTM08580, KTM08617, KTM08654, KTM08687, KTM08726, KTM08765, KTM08931, KTM08963, KTM08968, KTM08976, KTM09085, KTM09314, KTM09324, KTM09473, KTM09504, KTM09524, KTM09617, KTM09627, KTM09695, KTM09852, KTM09918, KTM10001, KTM10039, KTM10208, KTM10279, KTM10358, KTM10384, KTM10448, KTM10503, KTM10511, KTM10515, KTM10541, KTM10596, KTM10689, KTM10725, KTM10727, KTM10814, KTM10856, KTM10888, KTM10909, KTM10911, KTM10941, KTM10950, KTM11099, KTM11137, KTM11151, KTM11163, KTM11185, KTM11205, KTM11233, KTM11245, KTM11253, KTM11324, KTM11366, KTM11377, KTM11385, KTM11412, KTM11420, KTM11425, KTM11430, KTM11456, KTM11497, KTM11537, KTM11548, KTM11559, KTM11796, KTM11818, KTM11839, KTM11946, KTM12290, KTM12581, KTM12812, KTM12828, KTM12872, KTM13310, KTM13454, KTM13661, KTM13775, KTM13950, KTM13972, KTM14104, KTM14111, KTM14227, KTM14234, KTM14303, KTM14542, KTM14621, KTM14707, KTM14868, KTM15100, KTM15216, KTM15340, KTM15362, KTM15401, KTM15433, KTM15769, KTM15787, KTM15809, KTM15822, KTM15923, KTM15966, KTM15974, KTM16019, KTM16028, KTM16029, KTM16103, KTM16128, KTM16139, KTM16188, KTM16205, KTM16387, KTM16407, KTM16691, KTM16698, KTM16988, KTM17080, KTM17112, KTM17123, KTM17125, KTM17147, KTM17224, KTM17233, KTM17250, KTM17327, KTM17394, KTM17491, KTM17554, KTM18543, KTM18991, KTM19139, KTM19163, KTM19164, KTM19174, KTM19178, KTM19181, KTM19193, KTM19204, KTM19215, KTM19403, KTM19964, KTM20219, KTM20246, KTM20247, KTM20248, KTM20250, KTM20251, KTM20254, KTM20256, KTM20261, KTM20264, KTM20268, KTM20270, KTM20275, KTM20283, KTM20286, KTM20294, KTM20297, KTM20300, KTM20307, KTM20309, KTM20343, KTM20346, KTM20351, KTM20359, KTM20366, KTM20370, KTM20374, KTM20386, KTM20413, KTM20469, TM02055, TM02107, TM02774, TM01107, TM00775, TM03107, TM02625, TM03135, TM03535, TM03536, KTM09665, KTM11602, KTM03049, KTM04450, KTM10243, KTM10573, KTM17621, KTM17661, KTM18013, KTM18050, KTM19216, KTM20124, KTM20274, KTM21449, KTM01602, KTM03947, KTM04078, KTM03158, KTM00430, KTM01901, KTM02538, KTM08AB1, KTMA2F03, KTM05236, KTM05441, KTM05514, KTM05842, KTM06046, KTM07757, KTM08198, KTM08309, KTM08487, KTM12569, KTM14005, KTM13578, KTM15409, KTM17930, KTM20319, KTM00949, KTM01053, KTM01121, KTM01207, KTM01615, KTM01617, KTM02018, KTM02051, KTM02371, KTM02528, KTM02580, KTM02630, KTM02737, KTM03122, KTM03133, KTM03333, KTM03456, KTM03569, KTM03579, KTM03651, KTM03659, KTM03834, KTM03842, KTM03926, KTM03929, KTM04044, KTM04086, KTM04424, KTM04476, KTM04505, KTM04529, KTM04541, KTM04545, KTM04583, KTM04706, KTM04725, KTM04730, KTM05308, KTM05830, KTM05942, KTM06517, KTM06738, KTM07319, KTM07342, KTM07396, KTM07605, KTM07905, KTM07940, KTM07943, KTM08069, KTM08099, KTM08115, KTM08124, KTM08125, KTM08137, KTM08175, KTM08273, KTM08299, KTM08318, KTM08375, KTM08493, KTM08597, KTM09069, KTM09655, KTM09711, KTM11065, KTM11755, KTM11881, KTM11987, KTM11991, KTM12171, KTM12759, KTM12804, KTM12817, KTM12886, KTM12901, KTM14221, KTM14252, KTM14261, KTM14275, KTM14308, KTM14395, KTM14537, KTM14752, KTM15008, KTM15084, KTM15095, KTM15249, KTM15268, KTM15287, KTM15468, KTM15585, KTM15670, KTM16063, KTM16120, KTM17038, KTM17472, KTM17573, KTM17942, KTM19228, KTM20409, KTM20492, KTM65F00, KTMF44C1, TM01521, TM02004, KTM12460, KTM15711, KTM17008, KTM20377, KTM20391, KTM03815, KTM07179, KTM07586, KTM08186, KTM05315, KTM06607, KTM15756, KTM00693, KTM03119, KTM03125, KTM03127, KTM12543, KTM03894, KTM08233, KTM05042, KTM03560, KTM02833, KTM07124, KTM11094, KTM16181, KTM18572, KTM20367, KTM04281, KTM01453, KTM01474, KTM01682, KTM02275, KTM02549, KTM02951, KTM03081, KTM04814, KTM08056, KTM12843, KTM15941, KTMB48F2 HTFA-ASY-0003U, UDI: 00815381020277, Serial Number: TM03611

Devices named in this recall

2 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

1 device cleared under K160068 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 95019).