iBur recallZ-2725-2024 · Stryker Corporation · initiated Jul 24, 2024
Completed
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Stryker Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Potential for the product to exhibit temperatures higher than specification where the bur shank meets the distal bushing. Temperatures higher than specification may lead to the potential for minor tissue/structure damage or tissue/structure damage from thermal injury which may need medical/surgical intervention.
FDA root cause category: Under Investigation by firm
What the company is doing
Stryker issued Urgent Medical Device Recall letter on 7/24/424 to Materials Manager, Risk Manager, OR Director via Fed'x. Letter states reason for recall, health risk, and action to take: 1. Immediately review your inventory to locate and quarantine any affected products at your facility. 2. Return the enclosed Business Reply Form (BRF), even if the affected product is no longer in inventory, to the email [email protected]. 3. Upon receipt of the completed BRF, Stryker will provide a shipping label to return recalled product on-hand. 4. A credit will be provided upon receipt of the recalled product. 5. Maintain awareness of this communication internally and inform Stryker if any of the subject devices have been distributed to other organizations. If so, provide contact details so Stryker can inform the recipients accordingly. If you have any questions about this matter, contact Stryker Instruments by phone (269) 568-1481 or email at [email protected].
Recall details
- Recall number
- Z-2725-2024
- Status
- Completed
- Initiated by the company
- Jul 24, 2024
- Posted by the FDA
- Aug 26, 2024
- Recalling firm
- Stryker Corporation
- Quantity
- 2507 units
- Product code
- HBE
- FDA submissions
- K210377
- Distribution
- Worldwide distribution - US Nationwide and the countries of Australia, Canada, France, Hong Kong, India, Italy, Japan, Japan, Netherlands, South Korea, Spain, Sweden, Taiwan, United Kingdom.
Product description
Stryker iBur 3.0mm Diamond Round, Distal Bend - intended to cut bone in the following manner: drilling, reaming, decorticating, shaping, dissecting, shaving, and smoothing for the following medical applications: Neuro; Spine; Ear, Nose, and Throat (ENT)/ Otorhinolaryngology; and Endoscopic applications Ref: 8431-012-030D
Code information (models, lots, serial numbers)
GTIN: 7613327501230 Lot Numbers: 22216017 22230017 22334017 23004017 23004027 23009017 23157017 23208017 23233017 23233027 24003017
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- iBur 8431-012-030DStryker Corporation · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)1 recall2 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 95040).
- Z-2722-2024 · CompletedStryker iBur 3.0mm Precision Round, Distal Bend- Intended to cut bone in the following manner: drilling, reaming, decorticating, shaping, dissecting, shaving, and smoothing for the…
- Z-2723-2024 · CompletedStryker iBur 4.0mm Precision Round, Distal Bend- intended to cut bone in the following manner: drilling, reaming, decorticating, shaping, dissecting, shaving, and smoothing for the…
- Z-2724-2024 · CompletedStryker iBur 2.0mm Diamond Round, Distal Bend- intended to cut bone in the following manner: drilling, reaming, decorticating, shaping, dissecting, shaving, and smoothing for the f…
- Z-2726-2024 · CompletedStryker iBur 4.0mm Diamond Round, Distal Bend- intended to cut bone in the following manner: drilling, reaming, decorticating, shaping, dissecting, shaving, and smoothing for the f…
- Z-2727-2024 · CompletedStryker iBur 3.0mm Coarse Diamond Round, Distal Bend - intended to cut bone in the following manner: drilling, reaming, decorticating, shaping, dissecting, shaving, and smoothing f…
- Z-2728-2024 · CompletedStryker iBur 4.0mm Coarse Diamond Round, Distal Bend - intended to cut bone in the following manner: drilling, reaming, decorticating, shaping, dissecting, shaving, and smoothing f…
- Z-2729-2024 · CompletedStryker iBur 5.0mm Coarse Diamond Round, Distal Bend intended to cut bone in the following manner: drilling, reaming, decorticating, shaping, dissecting, shaving, and smoothing f…
- Z-2730-2024 · CompletedStryker iBur 3.0mm Diamond Match Head, Distal Bend - intended to cut bone in the following manner: drilling, reaming, decorticating, shaping, dissecting, shaving, and smoothing for…
- Z-2731-2024 · CompletedStryker iBur 3.0mm Precision Match Head, Distal Bend- intended to cut bone in the following manner: drilling, reaming, decorticating, shaping, dissecting, shaving, and smoothing f…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page