Vivo 45 LS, pressure and volume ventilator capable of… recallZ-2690-2024 · Breas Medical, Inc. · initiated Jul 25, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Breas Medical, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
There is a potential for short term (<7 days) elevated levels of formaldehyde emitted from the device into the patient breathing airpath in specific conditions.
FDA root cause category: Under Investigation by firm
What the company is doing
On August 5, 2024, URGENT MEDICAL DEVICE CORRECTION letters were sent to consignees. Actions to be taken by Customer / User: 1. Since the elevated levels of formaldehyde were observed with the Vivo 45 LS operated at higher room temperatures, 40C (104F), as a matter of precaution, Breas Medical is lowering the maximum room air temperature for operation of the device from 40C (104F) to 30C (86F). 2. The following additional measure applies to new Vivo 45 LS devices distributed on or before July 24, 2024, that have not been in use for at least 14 days: Pre-run the device for at least 14 days in an ambient room temperature of at least 20C (68F). The pre-run can be performed in any ventilation treatment mode (for example: CPAP at 4 cmH2O). Make sure that the device s air outlet is not blocked, so that the air flows out from the device. Devices being pre-run should be in a dedicated, well-ventilated room. Personnel working with the devices should be in the room only intermittently, for time needed to connect and setup or disconnect devices. 3. Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions. 4. Retain this document as an addendum to the Instructions for Use and for your records. Please keep a copy of this document available with the Instructions for Use. 5. Complete and return the attached acknowledgement form to [email protected] and [email protected] (You only need to send one acknowledgement form.) Contact Information: If you have any questions or concerns regarding this notification, please contact Breas Medical Inc. service or your local Service Representative. Breas Medical Inc. service can be reached at 855-436-8724 ext. 110 between 8.00am and 5.00pm ET Mondays through Fridays, or by e-mail [email protected].
Recall details
- Recall number
- Z-2690-2024
- Status
- Open, Classified
- Initiated by the company
- Jul 25, 2024
- Posted by the FDA
- Sep 3, 2024
- Recalling firm
- Breas Medical, Inc.
- Quantity
- 8298
- Product code
- NOU
- FDA submissions
- K193586
- Distribution
- US Nationwide.
Product description
Vivo 45 LS, pressure and volume ventilator capable of delivering continuous or intermittent ventilatory support for patients who require invasive or non-invasive mechanical ventilation, Model/Catalog Number: 230000 (230016, 230216)
Code information (models, lots, serial numbers)
Model No 230000; UDI-DI 07321822300004; Lot Code: all lots up to 240530.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Vivo 45 LS USBreas Medical AB · Ventilator, Continuous, Facility Use2 recalls6 deaths13 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page