Skip to content

CARDIOVIT AT-180. Model Number: 0A.110000 recallZ-2734-2024 · Schiller, Ag · initiated Jul 24, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Schiller, Ag or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Potential for high-frequency signal artifacts is recorded during an ECG acquisition performed by CARDIOVIT AT-180 electrocardiographs.

FDA root cause category: Software design

What the company is doing

Single US consignee was notified via letter on 07/24/2024. The consignee was instructed to send the notification to all consignees if further distributed, respond to the notification and return the reply form, and updated the affected devices according to Service Instructions. At the user level, customer are requested to respond by sending the Customer Reply Form back to their authorized distributor as confirmation that the Field Safety Notice was read and understood.

Recall details

Recall number
Z-2734-2024
Status
Open, Classified
Initiated by the company
Jul 24, 2024
Posted by the FDA
Aug 27, 2024
Recalling firm
Schiller, Ag
Quantity
1084 untis
Product code
DPS
FDA submissions
K183425
Distribution
Worldwide - US distribution in the state of Florida and the countries of Austria, Belgium, Switzerland, Cyprus, Czechia, Germany, Spain, Finland, France, Great Britain, Ireland, Iceland, Italy, Luxembourg, Malta, The Netherlands, Norway, Portugal, Romania, Sweden, Slovenia, Australia, The United Arab Emirates, Bahrain, Canada, China, Israel, India, Jordan, Kuwait, Libya, Morocco, Malaysia, Qatar, Serbia, South Africa.

Product description

CARDIOVIT AT-180. Model Number: 0A.110000; Catalog number: 3.920570

Code information (models, lots, serial numbers)

UDI-DI: 07613365002775. All serial numbers, all software versions.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.