IBA Proton Therapy System - PROTEUS 235 ProteusPLUS and… recallZ-3089-2024 · Ion Beam Applications S.A. · initiated Jul 8, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Ion Beam Applications S.A. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Safety Parameters Verification Mechanisms can be deactivated in clinical runtime

FDA root cause category: Under Investigation by firm

What the company is doing

IBA issued AMENDED Urgent Medical Device Correction letter (IBA REFERENCE NCIPT-27531) on 7/10/24 provided by IBA directly to the customer either by hand or by email. Letter states reason for recall, health risk and action to take: Regarding Proteus 235 with a PTS-8, PTS-10, PTS-11 or PTS-12 versions: actions: Regarding the malfunction of Safety Parameters Verification Mechanism within BAPP . IBA carried out an analysis to determine if the impacted sites have some safety parameters checked by the BAPP in their clinical site configuration with values leading to violations. The result of this analysis showed that there are no sites with safety parameters checked by the BAPP out of tolerance. Regarding the possibility to disable a Safety Parameters Verification Mechanism (concerns PTS-8, PTS-10, PTS-11 or PTS-12 versions): IBA will carry out an analysis to determine if the impacted sites have specific configuration parameters that allow to deactivate the Safety Parameters Verification Mechanisms in the clinical configuration and if the case, remove them from the clinical site configuration. The result of the analysis will be provided to you on July 31, 2024, at the latest. Final solutions: Regarding the malfunction of Safety Parameters Verification Mechanism within BAPP (only concerns PTS-10, PTS-11 or PTS-12 versions): IBA will ensure that the system cannot be used in clinical with safety parameters checked by the BAPP out of tolerances. Regarding the possibility to disable a Safety Parameters Verification Mechanism (concerns PTS-8, PTS-10, PTS-11 or PTS-12 versions): IBA will ensure that the system displays a Red Banner informing the customer that the system must not be used for clinical activities when safety parameter Verification Mechanisms are deactivated to prevent the users from treating patients. The intermediate action will be implemented for your site by August 31, 2024, at the latest. The final solution will be implemented for your site by June 2026. Fi

Recall details

Recall number
Z-3089-2024
Status
Open, Classified
Initiated by the company
Jul 8, 2024
Posted by the FDA
Sep 10, 2024
Recalling firm
Ion Beam Applications S.A.
Quantity
35 units
Product code
LHN
FDA submissions
K163500
Distribution
Worldwide - US Nationwide distribution in the states of AK, NJ, PA, VA,OK,FL, KA, IL, LA, MI, TN , TX, VA and the countries of BE, CZ, DE, ES, FR, IN, IT, JP, NL, PL, RU, SE, SG, TW.

Product description

IBA Proton Therapy System - PROTEUS 235 ProteusPLUS and ProteusONE Software versions: PTS-8 versions, PTS-10 versions, PTS-11 versions, and PTS-12 versions

Code information (models, lots, serial numbers)

UDI-DI: 05404013801138 Serial Numbers: PAT.107 (DE), PAT.108 (US), PAT.109 (US), PAT.110 (US), PAT.111 (FR), PAT.112 (US), PAT.113 (US), PAT.114 (CZ), PAT.115 (IT), SAT.116 (US), SAT.117 (PL), SAT.118 (RU), SAT.119 (US), SAT.120 (DE), SAT.122 (SE), SAT.123 (US), SAT.125 (IN), SAT.126 (US), SAT.127 (TW), SAT.132 (NL), SAT.133 (US), SAT.136 (IN), SAT.140 (US), SBF.101 (FR), SBF.103 (JP), SBF.104 (JP), SBF.105 (US), SBF.107 (FR), SBF.112 (BE), SBF.113 (US), SBF.117 (ES), SBF.124 (IT), SBF.125 (SG), SBF.128 (US), SBF.135 (US).