GM60A. Digital Diagnostic Mobile X-Ray System recallZ-2715-2024 · NeuroLogica Corporation · initiated Jul 8, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from NeuroLogica Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Potential for the anti-fall system of the device arm to fail.

FDA root cause category: Device Design

What the company is doing

Consignees were emailed an "Urgent -- Medical Device Correction; GM60A Digital X-ray Imaging System" notification the week of 7/8/2024. The notification instructs consignees to share the recall notification with those in their organization that utilize the affected devices and forward the notification if devices have been transferred. Affected systems may continue to be used. Consignees are to check on-hand devices to confirm if they are affected by the field action and return the response form back indicating any devices that require correction by email to [email protected] within 10 business days. Samsung Electronics will send service engineers to affected consignees to correct devices free of charge. Service will be coordinated by Samsung Electronics and should be completed by October 2024. Consignees with any questions are to contact their local Samsung Electronics representative.

Recall details

Recall number
Z-2715-2024
Status
Open, Classified
Initiated by the company
Jul 8, 2024
Posted by the FDA
Aug 22, 2024
Recalling firm
NeuroLogica Corporation
Quantity
31 units
Product code
IZL
FDA submissions
K150097
Distribution
US Nationwide distribution in the states of AR, CT, GA, IN, KS, MA, MD, MO, MS, NC, ND, NJ, NV, OH, PA, VA, WV & DC.

Product description

GM60A. Digital Diagnostic Mobile X-Ray System.

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

UDI: 08806088511153/ 08806088088266; Serial No. G69812, G69821, G69824, G69986, G69987, G71422, G71425, G71431, G71434, G71795, G71798, G71802, G72052, G72053, G72264, G72279, G72296, G72301, G72310, G72313, G72319, G72320, G72481, G72484, G72595, G72601, G72853, G72875, G73179, G73184, G73542; Samsung Serial No. 50WFM9BG900003M, 50WFM9BG900004J, 50WFM9BG900005H, 50WFM9BGA00001T, 50WFM9BGA00002B, 50WFM9BH300001E, 50WFM9BH300002W, 50WFM9BH300003Z, 50WFM9BH300004N, 50WFM9BH400001X, 50WFM9BH400002W, 50WFM9BH400003A, 50WFM9BH500001H, 50WFM9BH500002Y, 50WFM9BH600001P, 50WFM9BH600002L, 50WFM9BH600003F, 50WFM9BH600004Y, 50WFM9BH600005W, 50WFM9BH600006N, 50WFM9BH600007K, 50WFM9BH600008J, 50WFM9BH700002B, 50WFM9BH700001P, 50WFM9BH700003M, 50WFM9BH700004E, 50WFM9BH800002E, 50WFM9BH800003D, 50WFM9BH900001X, 50WFM9BH900002D, 50WFM9BHB00001T.