ADVIA Chemistry Urinary/Cerebrospinal Fluid Protein (UCFP) recallZ-3064-2024 · Siemens Healthcare Diagnostics, Inc. · initiated Jul 29, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Siemens Healthcare Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The potential for biased quality control (QC) and patient results when using Atellica CH UCFP lot 130414 and ADVIA Chemistry UCFP lot 140 reagents.
FDA root cause category: Under Investigation by firm
What the company is doing
On July 29, 2024 URGENT MEDICAL DEVICE CORRECTION letters were sent to consignees. Impact to Results Erroneously depressed or elevated urine protein or cerebrospinal fluid patient results may occur. Siemens investigation revealed a positive bias of up to 52% (at 27.3 mg/dL (273 mg/L); up to 19% at 68.7 mg/dL (687 mg/L)). However, analyte recovery decreases rapidly after opening a reagent pack and a negative bias of up to -35% was observed at 24 hours (at 27 mg/dL (270 mg/L); up to -9% at 67 mg/dL (670 mg/L)). Results of this assay would be interpreted in conjunction with the patient s medical history, clinical presentation, and other findings. Customer Actions: 1. Please review this letter with your Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable. 2. Discontinue use of and discard the kit lot listed in the table above (Products Section). 3. Complete and return the Field Correction Effectiveness Check and indicate product replacement needs on the form attached to this letter within 30 days. 4. Please retain this letter with your laboratory records and forward this letter to those who may have received this product. We apologize for the inconvenience this situation may cause. If you have any questions, please contact your Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative.
Recall details
- Recall number
- Z-3064-2024
- Status
- Open, Classified
- Initiated by the company
- Jul 29, 2024
- Posted by the FDA
- Sep 6, 2024
- Recalling firm
- Siemens Healthcare Diagnostics, Inc.
- Quantity
- 143 units (17 US, 126 OUS)
- Product code
- CEK
- Distribution
- Worldwide - US Nationwide distribution in the states of CA, FL, GA, IA, IL, KY, LA, MS, NJ, NY, NV, PA, TX, VA, PR and the countries of Australia, Austria, Bahrain, Bangladesh, Belgium, Brazil, Canada, Chile, China, Columbia, Croatia, Cura¿ao, St Eus, Czech Republic, Egypt, Estonia, Finland, France, Georgia, Germany, Greece, Hong Kong, Hungary, India, India, Ireland, Italy, Jordan, Kenya, Korea, Kuwait, Latvia, Malaysia, Mexico, Netherlands, New Zealand, Norway, Oman, Pakistan, Paraguay, Peru, Philippines, Poland, Portugal, Romania, Saudi Arabia, Singapore, Slovakia, South Africa, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, U.A.E., United Kingdom, Uruguay, Vatikancity.
Product description
ADVIA Chemistry Urinary/Cerebrospinal Fluid Protein (UCFP); Siemens Material Number (SMN): 11319151
Code information (models, lots, serial numbers)
Siemens Material Number: 11319151; UDI-DI: 00630414279176; Lot number: 140
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- ADVIA Chemistry UCFP 11319151Siemens Healthcare Diagnostics Inc. · BIURET (COLORIMETRIC), TOTAL PROTEIN2 recalls10 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 95112).
- Z-3063-2024 · Open, ClassifiedAtellica CH Urinary/Cerebrospinal Fluid Protein (UCFP); Siemens Material Number (SMN): 11097543
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page