DR 800. Digital Radiography X-ray System recallZ-2738-2024 · Agfa N.V. · initiated Jul 29, 2024

Class II Open, Classified All recalls posted in August 2024

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Agfa N.V. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Potential for the front lever chain of the DR 800 table to fail.

FDA root cause category: Device Design

What the company is doing

Consignees were sent an URGENT FIELD SAFETY NOTICE by email, dated 7/29/24. In the notice, consignees are asked to check the correct tightening of the floor plate fixing bolts if a high and persistent noise is heard from the device during the tilting or elevation movement of the tabletop. If the noise persists, the device must be stopped immediately in a horizontal position at the maximum height. Agfa will coordinate with consignees to schedule an inspection of affected devices and to install hardware intended to reinforce equipment before 10/15/24. The recall notification is to be shared within consignee organizations and forwarded to those devices were transferred to. Consignees are asked to maintain awareness of this notice until devices are corrected. Completed response forms are to be provided via email to [email protected]. Consignees with any questions are to call 1-877-777-2432 and reference PRB2001106.

Recall details

Recall number
Z-2738-2024
Recall class
Class II
Status
Open, Classified
Initiated by the company
Jul 29, 2024
Posted by the FDA
Aug 27, 2024
Recalling firm
Agfa N.V.
Quantity
141 units
Product code
JAA
FDA submissions
K180589, K183275
Distribution
Worldwide - US Nationwide distribution in the states of CA, FL, IL, IN, KS, LA, MA, MO, NJ, NC, OH, OK, SC, VA, WV, WI; and the countries of AU, BE, BW, CA, CO, CZ, DE, ES, FR, GB, GR, IL, IN, IT, KZ, MX, PL, RS, UZ, ZA.

Product description

DR 800. Digital Radiography X-ray System.

Code information (models, lots, serial numbers)

Show the full code information (2k characters)

Type No. 6010/200; UDI-DI: 05414904251805; Device Serial No. 925, 945, 948, 949, 990, 1003, 1022, 1116, 1136, 1161, 1229, 1239, 1401, 1404, 1453, 1460, 1489, 1506, 1532, 1568, 1704, 1733, 1757, 1888, 1893, 1950, 1954, 1957, 2010, 2070, 2130, 2244, 2297, 2371, 2378, 2901, 2931, 3068, 3217, 1507, 643, 763, 903, 915, 930, 969, 984, 992, 1032, 1042, 1049, 1052, 1123, 1148, 1197, 1217, 1323, 1341, 1203, 1279, 1458, 1509, 1521, 1627, 1676, 1691, 1756, 1828, 1850, 1878, 1921, 1924, 1926, 1932, 1938, 1941, 1943, 1956, 2079, 2088, 2094, 2101, 2115, 2129, 2131, 2157, 2169, 2192, 2193, 2225, 2233, 2243, 2251, 2253, 2254, 2277, 2289, 2310, 2314, 2318, 2333, 2340, 2359, 2377, 2409, 2415, 2416, 2428, 2436, 2437, 2449, 2450, 2475, 2483, 2513, 2538, 2580, 2642, 2678, 2687, 2756, 2781, 2833, 2842, 2873, 2886, 2892, 2941, 2955, 2992, 3101, 3134, 3143, 3151, 3161, 1920, 3006, 3040; Table Serial No. 32497, 34673, 34695, 34713, 34717, 34976, 34990, 35350, 36241, 36267, 36462, 37150, 37177, 38592, 38595, 38624, 38648, 39729, 39763, 39773, 39796, 41160, 41179, 41192, 42084, 42088, 42139, 42138, 43208, 43239, 43250, 44203, 45270, 45274, 46108, 46130, 49894, 51360, 52275, 53884, 39755, 31908, 32486, 32977, 34257, 34353, 34355, 34678, 34957, 34970, 34979, 35365, 35371, 38591, 35756, 36246, 36442, 36883, 37137, 37869, 37909, 36889, 37221, 38630, 39739, 39757, 40634, 41132, 41147, 41191, 42066, 42071, 42076, 42098, 42101, 42102, 42113, 42115, 42117, 42118, 42140, 43259, 43265, 43269, 43288, 44180, 44188, 44192, 44212, 44233, 44253, 44254, 45219, 45227, 45235, 45259, 45249, 45264, 45260, 45269, 45282, 45283, 45285, 45299, 45304, 46076, 46113, 46145, 46148, 46156, 46166, 46167, 47188, 46175, 47186, 47204, 47207, 47227, 47241, 47273, 48349, 48361, 48376, 48413, 49808, 49830, 49834, 49869, 49882, 49887, 51364, 51381, 51422, 52303, 52322, 52327, 52333, 52337, 42097, 51428, 52255.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page