ADVIA Centaur CA 19-9 recallZ-3103-2024 · Siemens Healthcare Diagnostics, Inc. · initiated Jul 26, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Siemens Healthcare Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Positive bias for some samples around the CA 19-9 upper limit of normal of 35 U/m as compared to the previous lot. The sample bias observed was greatest in the Asia Pacific sample population. The observed bias is not proportional across the assay measuring interval and is less pronounced for sample results above a concentration of 110 U/mL.
FDA root cause category: Under Investigation by firm
What the company is doing
Siemens has issued an Urgent Medical Device Correction (UMDC) AIMC 24-14.A.US to US customers, and an Urgent Field Safety Notice (UFSN) AIMC 24-14.A.OUS to Outside the US (OUS) customers on 7/26/24. Letter states reason for recall, health risk and action to take: " Please review this letter with your Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable. " Perform the instructions provided below: o Discontinue use of and discard the kit lots listed in the table above (Products Section). o If you experience this issue, you may request no-charge replacement product from your local Siemens Healthineers or distributor office. Please review your inventory of these products to determine your laboratory s replacement needs and to provide information to Siemens Healthineers for reporting to the authorities. o Complete and return the Field Correction Effectiveness Check and indicate product replacement needs on the form attached to this letter within 30 days. o Please retain this letter with your laboratory records and forward this letter to those who may have received this product. Resolution CA 19-9 kit lots ending in 535 are not available for purchase or use. You may request no-charge replacement product from your local Siemens Healthineers or distributor office. If you have any questions, please contact your Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative.
Recall details
- Recall number
- Z-3103-2024
- Status
- Open, Classified
- Initiated by the company
- Jul 26, 2024
- Posted by the FDA
- Sep 11, 2024
- Recalling firm
- Siemens Healthcare Diagnostics, Inc.
- Product code
- NIG
- FDA submissions
- K031393
- Distribution
- Worldwide - US Nationwide distribution
Product description
ADVIA Centaur CA 19-9 (50 Test)-In vitro diagnostic use in the quantitative measurement of the CA 19-9 tumor-associated antigen in human serum using the Atellica CI and Atellica IM Analyzers SMN: 10491379 SMN: 11206239
Code information (models, lots, serial numbers)
UDI-DI: (01)00630414574257(10)55970535(17)20240914; (01)00630414574257(10)55971535(17)20240914 Kit Lot: 55970535, 55971535
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- ADVIA Centaur® CA 19-9™ 10491379Siemens Healthcare Diagnostics Inc. · SYSTEM, TEST, CARBOHYDRATE ANTIGEN (CA19-9), FOR MONITORING AND MANAGEMENT OF PANCREATIC CANCER1 recall37 reports
Product lines
Other devices under the same FDA submission
4 devices cleared under K031393 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- ADVIA Centaur® CA 19-9™ 10491244Siemens Healthcare Diagnostics Inc. · SYSTEM, TEST, CARBOHYDRATE ANTIGEN (CA19-9), FOR MONITORING AND MANAGEMENT OF PANCREATIC CANCER1 recall87 reports
- Atellica IM CA 19-9 10995489Siemens Healthcare Diagnostics Inc. · SYSTEM, TEST, CARBOHYDRATE ANTIGEN (CA19-9), FOR MONITORING AND MANAGEMENT OF PANCREATIC CANCER1 recall73 reports
- Atellica IM CA 19-9 10995490Siemens Healthcare Diagnostics Inc. · SYSTEM, TEST, CARBOHYDRATE ANTIGEN (CA19-9), FOR MONITORING AND MANAGEMENT OF PANCREATIC CANCER1 recall48 reports
- Atellica IM CA 19-9 11206239Siemens Healthcare Diagnostics Inc. · SYSTEM, TEST, CARBOHYDRATE ANTIGEN (CA19-9), FOR MONITORING AND MANAGEMENT OF PANCREATIC CANCER1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 95118).
- Z-3100-2024 · Open, ClassifiedAtellica IM CA 19-9 (50 Test)-In vitro diagnostic use in the quantitative measurement of the CA 19-9 tumor-associated antigen in human serum using the Atellica CI and Atellica IM A…
- Z-3101-2024 · Open, ClassifiedAtellica IM CA 19-9 (250 Test)- In vitro diagnostic use in the quantitative measurement of the CA 19-9 tumor-associated antigen in human serum using the Atellica CI and Atellica IM…
- Z-3102-2024 · Open, ClassifiedAtellica IM CA 19-9 (250 Test - REF)-In vitro diagnostic use in the quantitative measurement of the CA 19-9 tumor-associated antigen in human serum using the Atellica CI and Atelli…
- Z-3104-2024 · Open, ClassifiedADVIA Centaur CA 19-9 (250 Test)-In vitro diagnostic use in the quantitative measurement of the CA 19-9 tumor-associated antigen in human serum using the Atellica CI and Atellica I…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page