CONMED recallZ-3164-2024 · Unimax Medical Systems Inc · initiated Jun 25, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Unimax Medical Systems Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The tube may fall into the surgical site during the grasping process.

FDA root cause category: Device Design

What the company is doing

The recalling firm issued letters dated 6/25/2024 to two distributors and dated 6/27/2024 to one distributor. The letters were issued via email beginning 6/25/2024. The letter provided the reason for recall and the actions to be taken, which included: (1) Identify the affected devices at your facility and complete the Acknowledgement and Receipt Form; (2) Distribute the notification to relevant people/facilities to where the affected lots have been distributed. (3) Quarantine the affected devices and implement the destruction after receiving confirmation from UNIMAX. A list of catalog numbers and lot numbers was enclosed. If the consignee had the affected product, they were to list the catalog number, lot number, and quantity in their possession on the Acknowledgement and Receipt Form. They were to also indicate whether they have notified the relevant person/facility to where the affected lot had been distributed and indicate they have quarantined the affected devices for destruction. The form was to be emailed to the recalling firm.

Recall details

Recall number
Z-3164-2024
Status
Open, Classified
Initiated by the company
Jun 25, 2024
Posted by the FDA
Sep 17, 2024
Recalling firm
Unimax Medical Systems Inc
Quantity
81,241 endo retrieval pouches
Product code
GCJ
FDA submissions
K103510
Distribution
US Nationwide distribution in the states of GA, MO, and PA.

Product description

1. Tech Medical Services Inc. Detachable Endo Retrieval Pouch, REF TM2001, 3" x 6", sterile; 2. Tech Medical Services Inc. Detachable Endo Retrieval Pouch, REF TM2000, 5" x 7", sterile; 3. Tech Medical Services, Inc. Detachable Endo Retrieval Pouch, REF TM2002, 8" x 10", sterile; 4. UNIMAX Detachable endo pocket, REF FEP936116, 3" x 6", sterile; 5. UNIMAX Detachable endo pocket, REF FEP979000, 8" x 10", sterile; 6. ConMed Corporation Detachable endo pocket, REF SB936, 3" x 6", sterile; 7. ConMed Corporation Detachable endo pocket, REF SB957, 5" x 7", sterile; 8. ConMed Corporation Detachable endo pocket, REF SB979, 7.5" x 9", sterile; 9. ConMed Corporation Detachable endo pocket, REF SB979-CA, 7.5 x 9", sterile; 10. ConMed Corporation Detachable endo pocket, REF SB936-CA, 3" x 6", sterile;

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

1. REF TM2001 - Lot numbers 6252207033, 6252211139, and 6252304152, UDI-DI 00851695005104; 2. REF TM2000 - Lot numbers 6252207032, 6252211138, and 6252304151, UDI-DI 00851695005098; 3. REF TM2002 - Lot numbers 6252207034, 6252211140, and 6252304153, UDI-DI 00851695005111; 4. REF FEP936116 - Lot number 6252303126, UDI-DI 04714127863504; 5. REF FEP979000 - Lot number 6252207216, UDI-DI 04714127863573; 6. REF SB936 - Lot numbers 6252207165, 6252207184, 6252207186, 6252304017, 6252304033, 6252305036, 6252305087, UDI-DI 0471412786076; 7. REF SB957 - Lot numbers 6252207180, 6252211177, 6252211178, 6252303004, 6252303018, 6252304025, 6252304034, 6252304046, UDI-DI 04714127860770; 8. REF SB979 - Lot numbers 6252207022, 6252207182, 6252303007, 6252303017, 6252304026, 6252304035, 6252304047, UDI-DI 04714127860909; 9. REF SB979-CA - Lot numbers 6252212124, 6252207201, 6252209205, UDI-DI 04714127861395; 10. REF SB936-CA - Lot number 6252212017, UDI-DI 04714127861371.