Skip to content

CARDIOSAVE RESCUE INTRA-AORTIC BALLOON PUMP (IABP).… recallZ-3135-2024 · Datascope Corp. · initiated Aug 8, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Datascope Corp. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Firm has developed a software correction (Revision D.00) outside the US to mitigate product problems that pose a risk of hemodynamic instability and failure of transmission of clinical data in Cardiosave Hybrid and Cardiosave Rescue Intra-Aortic Balloon Pump devices.

FDA root cause category: Software design

What the company is doing

Consignees were mailed an URGENT FIELD SAFETY NOTICE dated 8/5/2024. This notification informs consignees of the deployment of previously installed software to mitigate previously identified issues in Cardiosave Hybrid and/or Cardiosave Rescue Intra-Aortic Balloon Pump devices. Consignees are asked to examine their inventory immediately to identify if they have any Cardiosave Hybrid and/or Rescue IABP devices, and to identify what software version is currently installed in devices. Devices with software revision D.00 or D.01 have been corrected. If consignee units have a software version lower than D.00 or D.01 (e.g.: B.XX or C.XX) the device still requires the software update and has not yet been corrected. If consignee units do not have the corrected software revision installed, they are to continue to follow the instructions as listed int he original recall notification until the affected device(s) has been updated. Consignees are to complete and return the provided Response Form to the provided email address. Datascope/Getinge service representatives will contact consignees to schedule installation of software, if needed. Consignees are to forward the recall notice to all device users in their facility and to any customers to whom the device was shipped.

Recall details

Recall number
Z-3135-2024
Status
Open, Classified
Initiated by the company
Aug 8, 2024
Posted by the FDA
Sep 20, 2024
Recalling firm
Datascope Corp.
Quantity
5,475 units
Product code
DSP
FDA submissions
K181122
Distribution
International distribution to the countries of UNITED ARAB EMIRATES, ARGENTINA , AUSTRIA, AUSTRALIA, BELGIUM, BULGARIA, BAHRAIN, BRAZIL, CANADA, SWITZERLAND, CHILE, CHINA, COLOMBIA, CZECH REPUBLIC, GERMANY, DENMARK, ECUADOR, EGYPT, SPAIN, FINLAND, FRANCE, UNITED KINGDOM, INDONESIA, IRELAND, INDIA, IRAQ, IRAN, ITALY, JORDAN, JAPAN, SOUTH KOREA, KUWAIT, LEBANON, LITHUANIA, LIBYA, MOROCCO, MYANMAR, MEXICO, MALAYSIA, NETHERLANDS, NORWAY, NEW ZEALAND, OMAN, PHILIPPINES, POLAND, PALESTINE, PORTUGAL, SERBIA, RUSSIAN FEDERATION, SAUDI ARABIA, SWEDEN, SINGAPORE, SLOVAKIA, THAILAND, TURKEY, TAIWAN, UNITED KINGDOM, VIETNAM, YEMEN.

Product description

CARDIOSAVE RESCUE INTRA-AORTIC BALLOON PUMP (IABP). Electromechanical system used to inflate and deflate intra-aortic balloons.

Code information (models, lots, serial numbers)

Product No. 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85; UDI: 10607567112312, 10607567108407, 10607567113449; Lot No. ALL LOTS.

Devices named in this recall

3 devices in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 95150).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page