Beckman Coulter IRISPEC CA/CB/CC Urine Chemistry Controls… recallZ-0093-2025 · Beckman Coulter, Inc. · initiated Jul 16, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Beckman Coulter, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Beckman Coulter has become aware of an increase in customer complaints for intermittent IRISpec CB negative glucose control failure, resulting in false positive control results when used in conjunction with iChemVELOCITY Urine Chemistry Strips (PNs 800-7204 and 800-7212) on the iChemVELOCITY Analyzer. This issue may delay the reporting of patient results, with a worst-case scenario of a delay in diagnosis or treatment of metabolic disorders, kidney function abnormalities, urinary tract infections, and liver function.

FDA root cause category: Under Investigation by firm

What the company is doing

Beckman Coulter issued an URGENT MEDICAL DEVICE RECALL notice to its consignees beginning on 07/15/2024 by letter delivered by either USPS first class mail or email. The notice explained the issue, potential impact, and provided updated QC ranges for Glucose in the Analyzer. The information is to be shared with all affected parties. For questions: please contact Customer Support Center or your local representative; From our website: http://www.beckmancoulter.com By phone: Call 800-526-7694 in the United States and Canada. o Outside the United States and Canada, contact your local Beckman Coulter representative.

Recall details

Recall number
Z-0093-2025
Status
Open, Classified
Initiated by the company
Jul 16, 2024
Posted by the FDA
Oct 16, 2024
Recalling firm
Beckman Coulter, Inc.
Quantity
5589 units
Product code
JJW
FDA submissions
K072640
Distribution
worldwide

Product description

Beckman Coulter IRISPEC CA/CB/CC Urine Chemistry Controls, Catalog Number REF 800-7211 and 800-7702, in vitro diagnostic device.

Code information (models, lots, serial numbers)

UDI/DI 10837461002611, Lot Numbers: 432-24 and higher

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.