CARDIOSAVE RESCUE INTRA-AORTIC BALLOON PUMP (IABP).… recallZ-3137-2024 · Datascope Corp. · initiated Aug 8, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Datascope Corp. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Firm has developed a software correction (Revision D.01) for outside the US to mitigate product problems that pose a risk of hemodynamic instability and failure of transmission of clinical data in Cardiosave Hybrid and Cardiosave Rescue Intra-Aortic Balloon Pump devices.
FDA root cause category: Software design
What the company is doing
Consignees were mailed an URGENT FIELD SAFETY NOTICE dated 8/5/2024. This notification informs consignees of the deployment of previously installed software to mitigate previously identified issues in Cardiosave Hybrid and/or Cardiosave Rescue Intra-Aortic Balloon Pump devices. Consignees are asked to examine their inventory immediately to identify if they have any Cardiosave Hybrid and/or Rescue IABP devices, and to identify what software version is currently installed in devices. Devices with software revision D.00 or D.01 have been corrected. If consignee units have a software version lower than D.00 or D.01 (e.g.: B.XX or C.XX) the device still requires the software update and has not yet been corrected. If consignee units do not have the corrected software revision installed, they are to continue
Recall details
- Recall number
- Z-3137-2024
- Status
- Open, Classified
- Initiated by the company
- Aug 8, 2024
- Posted by the FDA
- Sep 20, 2024
- Recalling firm
- Datascope Corp.
- Quantity
- 5,475 units
- Product code
- DSP
- FDA submissions
- K181122
- Distribution
- International distribution to the countries of UNITED ARAB EMIRATES, ARGENTINA , AUSTRIA, AUSTRALIA, BELGIUM, BULGARIA, BAHRAIN, BRAZIL, CANADA, SWITZERLAND, CHILE, CHINA, COLOMBIA, CZECH REPUBLIC, GERMANY, DENMARK, ECUADOR, EGYPT, SPAIN, FINLAND, FRANCE, UNITED KINGDOM, INDONESIA, IRELAND, INDIA, IRAQ, IRAN, ITALY, JORDAN, JAPAN, SOUTH KOREA, KUWAIT, LEBANON, LITHUANIA, LIBYA, MOROCCO, MYANMAR, MEXICO, MALAYSIA, NETHERLANDS, NORWAY, NEW ZEALAND, OMAN, PHILIPPINES, POLAND, PALESTINE, PORTUGAL, SERBIA, RUSSIAN FEDERATION, SAUDI ARABIA, SWEDEN, SINGAPORE, SLOVAKIA, THAILAND, TURKEY, TAIWAN, UNITED KINGDOM, VIETNAM, YEMEN.
Product description
CARDIOSAVE RESCUE INTRA-AORTIC BALLOON PUMP (IABP). Electromechanical system used to inflate and deflate intra-aortic balloons.
Code information (models, lots, serial numbers)
Product No. 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85; UDI: 10607567112312, 10607567108407, 10607567113449; Lot No. ALL LOTS.
Devices named in this recall
3 devices in the FDA device registry whose identifier appears in the recall's code information.
- CARDIOSAVE RESCUE 0998-00-0800-83Datascope Corp. · System, Balloon, Intra-Aortic And Control26 recalls167 reports
- CARDIOSAVE RESCUE, CHINESE 0998-00-0800-75Datascope Corp. · System, Balloon, Intra-Aortic And Control21 recalls12 reports
- CARDIOSAVE RESCUE, 3.1 EDITION 0998-00-0800-85Datascope Corp. · System, Balloon, Intra-Aortic And Control21 recalls1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 95168).
- Z-3136-2024 · Open, ClassifiedCARDIOSAVE HYBRID INTRA-AORTIC BALLOON PUMP (IABP). Electromechanical system used to inflate and deflate intra-aortic balloons.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page