Mazor X System recallZ-3156-2024 · Mazor Robotics Ltd · initiated Jul 17, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Mazor Robotics Ltd or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Software anomalies which can result in two separate issues: A lagging screen display during a navigated procedure; or the potential for a mismatch between the navigated screw and the screw information presented on the screen.
FDA root cause category: Device Design
What the company is doing
Medtronic/Mazor Robotics, Ltd. issued an URGENT: Medical Device Software Correction to its consignees on 07/17/2024 via 2-day tracked mailing. The notice explained the issue, potential risk, and provided a workaround until the software update is installed. Medtronic will schedule a time to perform a software update that will resolve these anomalies. Medtronic field service engineers will update all Mazor X systems from version 5.1.1 to version 5.1.2 or later. Questions: contact Medtronic Technical Services at 1-888-826-5603 or [email protected] .
Recall details
- Recall number
- Z-3156-2024
- Status
- Open, Classified
- Initiated by the company
- Jul 17, 2024
- Posted by the FDA
- Sep 16, 2024
- Recalling firm
- Mazor Robotics Ltd
- Quantity
- 94 devices
- Product code
- OLO
- FDA submissions
- K230064
- Distribution
- US Nationwide distribution in PR.
Product description
Mazor X. Model Number REF TPL0059; stereotaxic orthopedic instrument
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
UDI/DI 07290109180465, Lot Numbers: 400-38-122016, 400-44-022017, 400-67-062017, 400-69-062017; UDI/DI 07290109181158, Lot Numbers: 4001200218, 4001970419, 4001011117, 4001041117, 4001061117, 4001320518, 4001330618, 4001390818, 4001430918, 4001631218, 4001690119, 4001710119, 4001780219, 4001790219, 4001800219, 4001960419, 4001980419, 4002050419R, 4002170619, 4002230819, 4002240819, 4002310919, 4002511119, 4002531119, 4002541119, 4002591119, 4002641219, 4002661219, 4002850220, 4002860220R, 4002900220, 4003020320, 4003030320, 4003180320, 4003190320, 4003430920, 4006951223, 4006901223, 4006890923, 4006860923, 4006830923, 4006680723, 4006170323, 4006090223, 4006050223, 4006040223, 4005210922, 4005770123, 4005720123, 4005431122, 4005750123, 4006130323, 4004091121; UDI/DI 07290109184098, Lot Numbers: 4003840821, 4003860821, 4004151121, 4004231221, 4004290122, 4004310122; UDI/DI 07290115751376, Lot Numbers: 4004860522, 4005180822, 4004670322, 4004700322, 4004870522, 4005080722, 4005130822, 4005340922, 4005401122, 4005421122, 4005451122, 4005481122, 4005491122, 4005591222, 4005760123, 4005790123, 4005870123, 4005900223, 4005910223, 4005940223, 4005980223, 4006030223, 4006140323, 4006230423, 4006330523, 4006630723, 4006660723, 4006670723, 4006740823, 4006790823, 4006800823
Devices named in this recall
4 devices in the FDA device registry whose identifier appears in the recall's code information.
- Mazor X System TPL0059-09Mazor Robotics Ltd · Orthopedic Stereotaxic Instrument3 recalls1 death137 reports
- Mazor X System TPL0059-04Mazor Robotics Ltd · Orthopedic Stereotaxic Instrument3 recalls1 death131 reports
- Mazor X System TPL0059-05Mazor Robotics Ltd · Orthopedic Stereotaxic Instrument3 recalls65 reports
- Mazor X System TPL0059-07Mazor Robotics Ltd · Orthopedic Stereotaxic Instrument2 recalls11 reports
Product lines
Other devices under the same FDA submission
1 device cleared under K230064 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Mazor X System TPL0059-10Mazor Robotics Ltd · Orthopedic Stereotaxic Instrument1 recall1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page