Skip to content

Mazor X System recallZ-3156-2024 · Mazor Robotics Ltd · initiated Jul 17, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Mazor Robotics Ltd or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Software anomalies which can result in two separate issues: A lagging screen display during a navigated procedure; or the potential for a mismatch between the navigated screw and the screw information presented on the screen.

FDA root cause category: Device Design

What the company is doing

Medtronic/Mazor Robotics, Ltd. issued an URGENT: Medical Device Software Correction to its consignees on 07/17/2024 via 2-day tracked mailing. The notice explained the issue, potential risk, and provided a workaround until the software update is installed. Medtronic will schedule a time to perform a software update that will resolve these anomalies. Medtronic field service engineers will update all Mazor X systems from version 5.1.1 to version 5.1.2 or later. Questions: contact Medtronic Technical Services at 1-888-826-5603 or [email protected] .

Recall details

Recall number
Z-3156-2024
Status
Open, Classified
Initiated by the company
Jul 17, 2024
Posted by the FDA
Sep 16, 2024
Recalling firm
Mazor Robotics Ltd
Quantity
94 devices
Product code
OLO
FDA submissions
K230064
Distribution
US Nationwide distribution in PR.

Product description

Mazor X. Model Number REF TPL0059; stereotaxic orthopedic instrument

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

UDI/DI 07290109180465, Lot Numbers: 400-38-122016, 400-44-022017, 400-67-062017, 400-69-062017; UDI/DI 07290109181158, Lot Numbers: 4001200218, 4001970419, 4001011117, 4001041117, 4001061117, 4001320518, 4001330618, 4001390818, 4001430918, 4001631218, 4001690119, 4001710119, 4001780219, 4001790219, 4001800219, 4001960419, 4001980419, 4002050419R, 4002170619, 4002230819, 4002240819, 4002310919, 4002511119, 4002531119, 4002541119, 4002591119, 4002641219, 4002661219, 4002850220, 4002860220R, 4002900220, 4003020320, 4003030320, 4003180320, 4003190320, 4003430920, 4006951223, 4006901223, 4006890923, 4006860923, 4006830923, 4006680723, 4006170323, 4006090223, 4006050223, 4006040223, 4005210922, 4005770123, 4005720123, 4005431122, 4005750123, 4006130323, 4004091121; UDI/DI 07290109184098, Lot Numbers: 4003840821, 4003860821, 4004151121, 4004231221, 4004290122, 4004310122; UDI/DI 07290115751376, Lot Numbers: 4004860522, 4005180822, 4004670322, 4004700322, 4004870522, 4005080722, 4005130822, 4005340922, 4005401122, 4005421122, 4005451122, 4005481122, 4005491122, 4005591222, 4005760123, 4005790123, 4005870123, 4005900223, 4005910223, 4005940223, 4005980223, 4006030223, 4006140323, 4006230423, 4006330523, 4006630723, 4006660723, 4006670723, 4006740823, 4006790823, 4006800823

Devices named in this recall

4 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

1 device cleared under K230064 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page