EXACTAMIX recallZ-3083-2024 · Baxter Healthcare Corporation · initiated Aug 20, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Baxter Healthcare Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Baxter Healthcare Corporation has received increased customer reports of particulate matter in the Automated Compounding Device Inlets (disposable inlet), Product Codes H938173, H938174, H938175, H938176, used with the ExactaMix and ExactaMix Pro compounders. Particulate matter has been observed within the inlet primary packaging inlet components, including within the sterile fluid path tubing, before use.
FDA root cause category: Other
What the company is doing
Baxter issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 08/20/2024 via USPS first class mail. The notice explained the issue, hazard involved, and requested the following: "Actions to be Taken by Customers 1. Disseminate this information to anyone who may interact with the ExactaMix and ExactaMix Pro compounders and the products they produce (Pharmacy and Clinical Staff). 2. Pharmacy Staff: Inspect the inlets before use, including the inlet primary packaging, tubing, connectors, and spikes. Perform the inspection in accordance with the enclosed instructions. o If particulate matter is observed, do not use the inlet and contact Baxter Corporate Product Surveillance to report the complaint and to arrange for the safe return of the product for further investigation, see contact information below. Please have your Baxter 8-digit ship-to account number, product code, lot number, and quantity of product to be returned ready when contacting Baxter. The product code and lot number can be found on the individual product pouch and carton. o If no particulate matter is observed, the inlet can be used for compounding. Please ensure the inlet is primed before use according to the instructions provided in the Priming and Verifying section of the ExactaMix and ExactaMix Pro compounder Operator s Manual. 3. Pharmacy and Clinical Staff: After compounding, visually inspect the finished solution in the patient bag for precipitates and particulates per the Fulfilling the Order section in the ExactaMix and ExactaMix Pro compounder Operator s Manual. 4. Use a minimum of 1.2 micron in-line filter during product administration. The American Society for Parenteral and Enteral Nutrition (ASPEN) recommends using a 1.2 microns in-line filter for administration of total nutrient admixtures (TNAs), dextrose-amino acid admixtures, and lipid injectable emulsion. If you are already using in-line filtration per ASPEN recommendation, no additional action is necessary. 5. On
Recall details
- Recall number
- Z-3083-2024
- Status
- Open, Classified
- Initiated by the company
- Aug 20, 2024
- Posted by the FDA
- Sep 19, 2024
- Recalling firm
- Baxter Healthcare Corporation
- Quantity
- 1,548,782 units
- Product code
- LHI
- FDA submissions
- K002705
- Distribution
- Worldwide distribution.
Product description
Baxter EXACTAMIX Inlet, Vented, High Volume Inlet, REF H938174, for use in a pharmaceutical compounding device
Code information (models, lots, serial numbers)
UDI/DI 00085412475790, Lot Numbers: 803799 and lower
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- EXACTAMIX H938174Baxter Healthcare Corporation · SYSTEM/DEVICE, PHARMACY COMPOUNDING2 recalls238 reports
Product lines
Other devices under the same FDA submission
6 devices cleared under K002705 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- EXACTAMIX H938724Baxter Healthcare Corporation · Set, i.V. Fluid transfer1 recall513 reports
- EXACTAMIX H938175Baxter Healthcare Corporation · SYSTEM/DEVICE, PHARMACY COMPOUNDING3 recalls263 reports
- EXACTAMIX H938173Baxter Healthcare Corporation · Set, i.V. Fluid transfer2 recalls260 reports
- EXACTAMIX H938176Baxter Healthcare Corporation · Set, i.V. Fluid transfer2 recalls232 reports
- EXACTAMIX H938724EBaxter Healthcare Corporation · Set, i.V. Fluid transfer105 reports
- EXACTAMIX H938792Baxter Healthcare Corporation · Set, i.V. Fluid transfer12 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 95192).
- Z-3082-2024 · Open, ClassifiedBaxter EXACTAMIX Inlet, Non-Vented, High Volume Inlet, REF H938173, for use in a pharmaceutical compounding device
- Z-3084-2024 · Open, ClassifiedBaxter EXACTAMIX Inlet, Vented, Micro-Volume Inlet, REF H938175, for use in a pharmaceutical compounding device
- Z-3085-2024 · Open, ClassifiedBaxter EXACTAMIX Inlet, Vented, Syringe Inlet, REF H938176, for use in a pharmaceutical compounding device
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page