BIOLOX Option recallZ-3132-2024 · Zimmer GmbH · initiated Aug 8, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Zimmer GmbH or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Mislabeled: Incorrect sleeve in the packaging. The outer packaging is labeled as a standard neck sleeve, however, the product within the box is a -6mm neck sleeve, and vice versa. The neck of the sleeve can be identified by the device etching as either STD or -6
FDA root cause category: Under Investigation by firm
What the company is doing
Zimmer issued Urgent Medical Device Recall letter on 8/8/24 to Distributors, Risk Managers and Surgeons. Letter states reason for recall, health risk and action to take: Risk Manager Responsibilities: 1.Review this notification and ensure that affected personnel are aware of the contents. 2.If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3. If the product has been further distributed, provide your customers with this notice and ensure documentation. 4. Complete Attachment 1 Certificate of Acknowledgement and send to [email protected]. This form shall be returned even if you do not have affected products at your facility. 5. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facility s documentation. 6. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to [email protected]. Surgeon Responsibilities: 1. Review this notification for awareness of the contents. 2. There are no specific patient monitoring instructions related to this recall that are recommended beyond your existing follow-up schedule. 3. Complete Attachment 1 Certificate of Acknowledgement and send to [email protected]. 4. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facility s documentation. 5. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371
Recall details
- Recall number
- Z-3132-2024
- Status
- Open, Classified
- Initiated by the company
- Aug 8, 2024
- Posted by the FDA
- Sep 12, 2024
- Recalling firm
- Zimmer GmbH
- Quantity
- 69 units
- Product code
- LZO
- FDA submissions
- K082996
- Distribution
- Worldwide distribution - US Nationwide and the countries of Japan, Netherlands.
Product description
Biolox Option Taper Sleeve, Type 1 Taper, Standard Neck - Indicated for use in total hip replacement with cemented or non-cemented femoral and acetabular components. Item Number: 650-1066
Code information (models, lots, serial numbers)
UDI: (01)00887868271489(17)331130(10)3185263 Lot Number: 3185263
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- BIOLOX® Option 650-1066Biomet Uk Ltd · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED1 recall12 reports
Product lines
Other devices under the same FDA submission
18 devices cleared under K082996 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- BIOLOX DELTA HIP SYSTEM 650-1057Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented10 reports
- BIOLOX DELTA OPTION HIP SYSTEM 650-1055Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented7 reports
- BIOLOX OPTION HIP SYSTEM 650-1064Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented6 reports
- BIOLOX OPTION HIP SYSTEM 650-1068Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented4 reports
- BIOLOX® Delta Hip System 650-1057Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented4 reports
- BIOLOX® Delta Hip System 650-1058Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented4 reports
- BIOLOX OPTION HIP SYSTEM 650-1065Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented3 reports
- BIOLOX OPTION HIP SYSTEM 650-1066Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented2 reports
- BIOLOX DELTA HIP SYSTEM 650-1058Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented2 reports
- BIOLOX OPTION HIP SYSTEM 650-1067Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented2 reports
- BIOLOX® Delta Hip System 650-1056Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented2 reports
- BIOLOX® OPTION Head System 650-1065Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented2 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 95231).
- Z-3133-2024 · Open, ClassifiedBiolox Option Taper Sleeve, Type 1 Taper, -6mm Neck - Indicated for use in total hip replacement with cemented or non-cemented femoral and acetabular components. Item Number: 650-1…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page