Skip to content

Acceava Respiratory Strep A, CLIA Waived, Catalog Number… recallZ-3185-2024 · Cypress Medical Products LLC · initiated Aug 20, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Cypress Medical Products LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Due to a classification error in our inventory management system, these medical devices were shipped to customers that did not have the requisite license to purchase such devices.

FDA root cause category: Process control

What the company is doing

An URGENT PRODUCT RECALL notification letter dated 8/19/24 was sent to customers. Actions to be taken by the customer 1. Immediately examine your inventory and quarantine all product subject to this urgent product recall detailed in the affected products table above. 2. Please complete, sign, and return the enclosed response form (Attachment #1 Customer Response Form) as soon as possible, even if you do not have any product on hand. Return completed signed form to [email protected]. 3. Destroy any affected product and complete the attached customer response form (Attachment #1 Costumer Response Form). If you have any questions, you can call directly by phone at (800) 688-8840. Representatives are available Monday Friday, 8:00 am 5:00 pm. 4. If you have further distributed this product, please notify them immediately of this urgent product recall. You are encouraged to include a copy of this notification letter in your communication. If you have questions regarding this notification, please contact McKesson Medical- Surgical directly by phone at (800) 688-8840. Representatives are available Monday Friday, 8:00 am 5:00 pm.

Recall details

Recall number
Z-3185-2024
Status
Open, Classified
Initiated by the company
Aug 20, 2024
Posted by the FDA
Sep 19, 2024
Recalling firm
Cypress Medical Products LLC
Quantity
5 kits
Product code
GTY
FDA submissions
K010582
Distribution
US Nationwide distribution.

Product description

Acceava Respiratory Strep A, CLIA Waived, Catalog Number 4580295008

Code information (models, lots, serial numbers)

GTIN: 10815845020031; All Lots

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 95235).