Origio Sperm Wash Medium Pack of Ten 10ml bottles -… recallZ-3296-2024 · CooperSurgical, Inc. · initiated Feb 16, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from CooperSurgical, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Low concentration of sodium pyruvate in the wash could compromise gamete development
FDA root cause category: Under Investigation by firm
What the company is doing
CooperSurgical issued Urgent: Voluntary Medical Device Field Safety Notice on 2/16/24 via UPS Next Day Air via third party Sedgewick. Letter states reason for recall, health risk and action to take: 1) Inspect your inventory, quarantine any affected product, and immediately stop using the product. 2) Complete the appropriate version of the attached form (Customer Acknowledgement Form or Distributor Acknowledgement Form). Once completed, return the form to CooperSurgical as indicated at the top of the form. Note: Even if you do not have any affected product in your inventory ( or have not distributed any affected product to customers), please complete and return the enclosed form so that we may document receipt of this FSN. When the completed form is received by CooperSurgical, arrangements will be made for the return of any affected product at no additional cost to you. Credit will be applied to your account for product returned under this action. CooperSurgical is committed to high-quality, safe, and effective products. This corrective action has been initiated to ensure this failure mode does not reoccur and future potential patient harms can be avoided. If you have any questions, please feel free to reach us at +1 203-601-5200 ext. 3300 or [email protected].
Recall details
- Recall number
- Z-3296-2024
- Status
- Open, Classified
- Initiated by the company
- Feb 16, 2024
- Posted by the FDA
- Sep 26, 2024
- Recalling firm
- CooperSurgical, Inc.
- Quantity
- 980 units
- Product code
- MQL
- FDA submissions
- K153267
- Distribution
- Worldwide distribution - US Nationwide.
Product description
Origio Sperm Wash Medium Pack of Ten 10ml bottles - Intended for washing of spermatozoa Part Number: 84051010A
Code information (models, lots, serial numbers)
UDI: 20888937816536 Lot 230922-018006 Expiration date: 20 September 2024
Other devices under the same FDA submission
1 device cleared under K153267 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- ORIGIO Sperm Wash 84050060DOrigio A/S · Media, Reproductive1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 95246).
- Z-3297-2024 · Open, ClassifiedOrigio Sperm Wash Medium Pack of Five 60ml bottles - Intended for washing of spermatozoa. Part Number: 84055060A
- Z-3298-2024 · Open, ClassifiedOrigio Sperm Wash Medium Pack of Five 60ml bottles - Intended for washing of spermatozoa. Part Number: 84055060D
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page