Tornier Perform recallZ-3155-2024 · Tornier, Inc · initiated Aug 29, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Tornier, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

One lot of Perform Humeral Nucleus may exceed the specification for bacterial endotoxin load (max 20 EU/device).

FDA root cause category: Under Investigation by firm

What the company is doing

Stryker issued an URGENT: MEDICAL DEVICE RECALL notice to its consignees on 08/29/2024 via mail and email. The notice explained the issue, potential risk, and requested the following actions be taken: identify and isolate all affected product to prevent accidental use, and complete and return the response form. Upon receipt, Stryker will contact the consignee and arrange for the return of the product. Distributor were directed to notify their customers. Stryker issued an URGENT: MEDICAL DEVICE RECALL notice to its consignees on 08/29/2024 via mail and email. The notice explained the issue, potential risk, and requested the following actions be taken: identify and isolate all affected product to prevent accidental use, and complete and return the response form. Upon receipt, Stryker will contact the consignee and arrange for the return of the product. Distributors were directed to notify their customers. For questions or concerns, please contact [email protected].

Recall details

Recall number
Z-3155-2024
Status
Open, Classified
Initiated by the company
Aug 29, 2024
Posted by the FDA
Sep 16, 2024
Recalling firm
Tornier, Inc
Quantity
5 units
Product code
PKC
FDA submissions
K220418
Distribution
US Nationwide distribution in the states of IL, WI, AZ, NV, SD.

Product description

stryker TORNIER PERFORM HUMERAL SYSTEM - STEMLESS US Anatomic Nucleus, REF DWM003A; anatomic total arthroplasty of the shoulder

Code information (models, lots, serial numbers)

UDI/DI 00846832092192, Lot number 3268BA