ADVIA Centaur aTPO Assay, Material Number 10630886 (100… recallZ-3304-2024 · Siemens Healthcare Diagnostics, Inc. · initiated Aug 21, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Siemens Healthcare Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Positive bias in patient results at and below the cut-off of 60 U/mL with the lots listed in the table below when compared to alternate lots on the Atellica IM analyzer.
FDA root cause category: Under Investigation by firm
What the company is doing
An Urgent Medical Device Correction notification letter dated 8/21/24 was sent to customers. Customer Actions " Please review this letter with your Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable. " Perform the actions provided below: o Discontinue use of and discard the kit lots listed in the table above (Products Section). o You may request no-charge replacement product from your local Siemens Healthineers or distributor office. Please review your inventory of these products to determine your laboratory s replacement needs and to provide information to Siemens Healthineers for reporting to the authorities. o Complete and return the Field Correction Effectiveness Check and indicate product replacement needs on the form attached to this letter within 30 days. o Please retain this letter with your laboratory records and forward this letter to those who may have received this product. You may request no-charge replacement product from your local Siemens Healthineers or distributor office. Unaffected reagent lots are available for ordering. Please send a scanned copy of the completed form via email to: [email protected] Or to fax this completed form to the Customer Care Center at: (312) 275-7795 We apologize for the inconvenience this situation may cause. If you have any questions, please contact your Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative.
Recall details
- Recall number
- Z-3304-2024
- Status
- Open, Classified
- Initiated by the company
- Aug 21, 2024
- Posted by the FDA
- Sep 27, 2024
- Recalling firm
- Siemens Healthcare Diagnostics, Inc.
- Quantity
- 5622 kits
- Product code
- JZO
- FDA submissions
- K003291
- Distribution
- Worldwide distribution - US Nationwide.
Product description
ADVIA Centaur aTPO Assay, Material Number 10630886 (100 Test) and 10630887 (500 Test)
Code information (models, lots, serial numbers)
UDI-DI: Material Number 10630886 (100 Test) - 00630414587837; Material Number 10630887 (500 Test) - 00630414587844; Lot Number: 26053337 26054337 26051337 26052337
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- ADVIA Centaur® aTPO 10630886Siemens Healthcare Diagnostics Inc. · SYSTEM, TEST, THYROID AUTOANTIBODY1 recall0 reports
- ADVIA Centaur® aTPO +0 10630887Siemens Healthcare Diagnostics Inc. · SYSTEM, TEST, THYROID AUTOANTIBODY1 recall0 reports
Other devices under the same FDA submission
2 devices cleared under K003291 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Atellica IM aTPO 10995466Siemens Healthcare Diagnostics Inc. · SYSTEM, TEST, THYROID AUTOANTIBODY1 recall0 reports
- Atellica IM aTPO 10995467Siemens Healthcare Diagnostics Inc. · SYSTEM, TEST, THYROID AUTOANTIBODY1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 95259).
- Z-3305-2024 · Open, ClassifiedAtellica IM aTPO Assay, Material Number 10995466 (100 Test) and 10995467 (500 Test)
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page